Stryker Gynnie® OB/GYN Stretcher Operations Manual

Model: 1061

Document Revision: C.0

Publication Date: 2020/04

Symbols

? Operating instructions/Consult instructions for use
⚠️ General warning
Caution
[No pushing symbol] No pushing
[Oxygen bottle symbol with cross] Do not store the oxygen bottle
REF Catalogue number
SN Serial number
US Patents For US Patents see www.stryker.com/patents
CE CE mark
EC REP Authorized representative in the European Community
MD European medical device
[Manufacturer symbol] Manufacturer
[Safe working load symbol] Safe working load
[Lubricate symbol] Lubricate
[Type B applied part symbol] Type B applied part

Introduction

This manual assists with the operation or maintenance of your Stryker product. Read this manual before operating or maintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenance of this product.

CAUTION

This manual is a permanent part of the product and should remain with the product even if the product is sold.

Stryker continually seeks advancements in product design and quality. This manual contains the most current product information available at the time of printing. There may be minor discrepancies between your product and this manual. If you have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770.

Product description

The Stryker Model 1061 Gynnie® OB/GYN stretcher is a wheeled stretcher consisting of a platform mounted on a wheeled frame to transport patients in a horizontal position while providing a platform for the clinical treatment or examination of patients within the interior of a healthcare facility. The device has siderails and supports for fluid infusion devices.

Indications for use

The Gynnie OB/GYN stretcher is a wheeled device that provides a method of transporting patients within a healthcare facility by healthcare professionals or trained representatives of the healthcare facility. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications, such as short-term outpatient clinical evaluations, treatment, minor procedure and as a short-term outpatient recovery platform.

The Gynnie OB/GYN stretcher is for use in all establishments and may include use in, but is not limited to:

The Gynnie OB/GYN stretcher has a safe working load of 500 lb (225 kg).

The Gynnie OB/GYN stretcher is not for use for long-term inpatient treatment and recovery.

Clinical benefits

Patient transport and facilitate treatment

Expected service life

The Gynnie OB/GYN stretcher has a 10 year expected service life under normal use conditions and with appropriate periodic maintenance.

Disposal/recycle

Always follow the current local recommendations and/or regulations governing environmental protection and the risks associated with recycling or disposing of the equipment at the end of its useful life.

Contraindications

None known.

Specifications

500 lb 225 kg
Note - Safe working load indicates the sum of the patient, mattress, and accessory weight.
Length 81 in. 205.7 cm
Width Siderails up: 31.5 in. (80 cm) Siderails down: 28 in. (71 cm)
Height (to litter top) High: 35.75 in. (90.8 cm) Low: 22 in. (55.5 cm)
Litter positioning Backrest: 0° to 90° Knee Gatch: 0° to 90°
Trendelenburg/Reverse Trendelenburg Maximum angle: 87.5°
Patient surface 26 in. x 75.5 in. (66 cm x 191.5 cm)
Siderails 13 in. x 39 in. (34 cm x 99 cm)
Caster diameter 8 in. (20 cm)
Break away from wall point 68 in. (172.7 cm)

Stryker reserves the right to change specifications without notice.

Environmental conditions

Environmental conditions Operation Storage and transportation
Temperature 50 °F to 100 °F (10 °C to 38 °C) -4 °F to 140 °F (-20 °C to 60 °C)
Relative humidity 30% to 75% 10% to 95%
Atmospheric pressure 700 hPa to 1060 hPa 500 hPa to 1060 hPa

Product illustration

The Gynnie OB/GYN Stretcher features the following components:

[Illustration of the Stryker Gynnie OB/GYN Stretcher with labeled parts A through J]

Contact information

Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.

Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA

The user and/or the patient should report any serious product-related incident to both the manufacturer and the Competent authority of the European Member State where the user and/or patient is established.

To view your operations or maintenance manual online, see https://techweb.stryker.com/.

Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.

Serial number location

[Illustration showing the serial number location on the product, labeled 'A']

Figure 1 - Serial number location

Date of manufacture

The year of manufacture is the first four digits of the serial number.

Setup

Make sure that the product works before the product is put into service.

  1. Apply the brake. Push on the product to make sure that all four casters are locked.
  2. Release the brake. Push on the product to make sure that all four casters are unlocked.
  3. Raise and lower the litter with the hydraulic lift system.
  4. Raise the product to the highest position and put the product in the Trendelenburg position. Make sure that the head end lowers to the full down position.
  5. Raise the product to the highest position and put the product in the Reverse Trendelenburg position. Make sure that the foot end lowers to the full down position.
  6. Apply the fifth wheel and make sure that the fifth wheel guides and pivots the product.
  7. Make sure that the siderails raise, lower, and lock in place.
  8. Raise and lower the manual Fowler backrest (head section).
  9. Raise and lower the manual Gatch (foot section).
  10. Extend and retract each footrest and make sure that you can adjust the footrest position and lock the footrest.
  11. Raise and lower each push handle and make sure that the push handle locks.

Operation

Applying and releasing the brakes

WARNING - Always apply the brakes when a patient is getting on the product or off the product or when the product is not moving. Injury could result if the product moves while a patient is getting on the product or off the product.

To apply the brakes, push down on the brake (red) side of the brake/steer pedal. Push on the product to make sure that the brakes work.

To release the brakes, push down on the steer (green) side of the brake/steer pedal.

Raising or lowering the litter

WARNING

To raise the litter, press down on the pump pedal (A) until you achieve the desired height (Figure 2).

To lower the entire litter, press on the center of the uni-lower pedal (B).

To lower the head end of the litter, press down on the side of the uni-lower pedal (B) closest to the head end of the product.

To lower the foot end of the litter, press down on the side of the uni-lower pedal (B) closest to the foot end of the product.

The base may be equipped with optional variable descent controls. With variable descent controls, the farther you press down on the pedal, the faster the litter will lower.

[Illustration showing the uni-lower pedal (B) and pump pedal (A) for raising or lowering the litter]

Figure 2 - Raising or lowering the litter

Positioning the product in Trendelenburg

WARNING - Always remove any devices that may be in the way before you raise or lower the litter.

To position the product in the Trendelenburg position (head down), raise the litter to the highest height (Raising or lowering the litter (page 9)).

Raise the litter to the highest height for a greater Trendelenburg angle.

To lower the head end of the product, push down on the side of the uni-lower pedal (B) closest to the head end (Figure 2).

To lower the product from Trendelenburg position, push down on the center of the uni-lower pedal (B) until the litter is flat.

Positioning the product in Reverse Trendelenburg

WARNING - Always remove any devices that may be in the way before you raise or lower the litter.

To position the product in the Reverse Trendelenburg position (foot down), raise the litter to the highest height (Raising or lowering the litter (page 9)).

Raise the litter to the highest height for a greater Trendelenburg angle.

To lower the foot end of the product, push down on the side of the uni-lower pedal (B) closest to the foot end (Figure 2).

To lower the product from Reverse Trendelenburg position, push down on the center of the uni-lower pedal (B) until the litter is flat.

Transporting a patient with the retractable fifth wheel

WARNING

CAUTION - Do not use the hydraulics on the base to raise the product with a patient lift under the product.

To transport a patient with the retractable fifth wheel:

  1. Push down on the steer side of the brake/steer pedal to apply the fifth wheel.
  2. Put the pedal in the neutral position to move the product laterally. Move the product to the desired location.

Do not attempt to move the product laterally with the retractable fifth wheel applied.

  1. Apply the brakes to lock the product in place.

Always make sure that the brake is released before you move the product to avoid operator or patient injury.

Positioning or stowing the push handles (optional)

To position or stow the push handles:

  1. Pivot the handles up from the end of the product (Figure 3).
  2. Push down on the handles to lock them into position.
  3. Reverse steps to stow the handles.

Only use the push handles as push/pull devices unless otherwise specified to avoid product damage.

[Illustration showing the head end push handles in various positions]

Figure 3 – Positioning the head end push handles

Raising or lowering the siderails

WARNING

To raise the siderails, use two hands to grasp the siderail. Raise the siderail until the release latch clicks into place. Pull on the siderail to make sure that the siderail is locked.

To lower the siderails, pull up on the release latch. Guide the siderail to the lowest position.

Raising or lowering the Fowler backrest

Do not use siderails as restraint devices to keep the patient from exiting the product. The siderails keep the patient from rolling off the product. The operator must determine the degree of restraint necessary to make sure that the patient is safe.

WARNING

To raise the Fowler backrest, squeeze the red Fowler backrest release handle (A). Pull the Fowler backrest up to the desired position.

To lower the Fowler backrest, squeeze the red Fowler backrest release handle (A). Push the Fowler backrest down to the desired position.

[Illustration of the Fowler backrest with release handle labeled 'A']

Figure 4 - Fowler backrest

Positioning the foot supports

CAUTION

  1. Position the patient toward the foot end of the product.
  2. Grasp the red handle (B) and pull out the foot support (A).
  3. Lift the foot support until the foot support locks into the exam position.
  4. Lift up on the foot support and slide the foot support in or out to adjust the foot support position.
  5. Position the patient's feet in the foot supports.

[Illustration showing how to position the foot supports, with parts labeled 'A' and 'B']

Figure 5 – Positioning the foot supports

Storing the foot supports

CAUTION

To store the foot supports:

  1. Squeeze the red release handle.
  2. Lower the foot support.
  3. Slide the foot support arm under the litter.

Positioning the foot section

WARNING

To lower the foot section, lift up on the foot section. Squeeze the red release handle (A). Lower the foot section.

To raise the foot section, pull up on the foot section frame until the foot section latches. Make sure that the foot section is secure before you place any patient weight on the foot section.

Positioning the waste bin holder (optional)

CAUTION

To position the waste bin holder:

  1. Position the patient's feet in the foot supports (Positioning the foot supports (page 12)) and lower the foot section (Positioning the foot section (page 13)).
  2. Slide the waste bin holder (A) out from under the litter frame (Figure 6).
  3. Place the waste basin in the holder.
  4. Secure the top of the basin on the "wings” (B) of the waste bin holder.

To store the waste bin holder, remove and dispose of the waste basin. Slide the waste bin holder under the litter frame.

[Illustration showing how to position the waste bin holder, with parts labeled 'A' and 'B']

Figure 6 - Position the waste bin holder

Accessories and parts

These accessories may be available for use with your product. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770.

Name Number
Calf rest, free standing 1061-040-036
Calf support mounting brackets 1061-130-000
Caster assembly with hardware (no caster covers) 0715-259-100
Caster covers (four wheels) 1010-056-200
HAVASU™ IV pole, removable 0390-025-010
IV pole, three-stage, head end, left 1061-110-001
IV pole, two-stage, head end, left 1061-110-000
Hydraulic jack, constant descent 1231-070-210
Hydraulic oil, Mobil™ Aero™ HFA - 1 quart 2020-070-475
Hoop for drainage basin 1061-120-000
Mattress, Ultra Comfort, 4" x 26" 1061-426-200
Oxygen bottle holder assembly, upright 1020-130-000
Oxygen bottle retainer 1040-010-090
Restraint strap, ankle 0946-043-000
Restraint strap, body 0390-019-000
Restraint strap, chest 1010-058-000
Restraint strap, wrist 0946-044-000
Restraint strap, full package 1010-077-000
Serving tray 0785-045-700
Siderail pads 1010-052-000
Waste basin, disposable 1061-120-110

Attaching, removing, and positioning the free-standing calf rest

To remove the calf rest and secure the calf rest in place, remove the cap and lift the calf support and rod off of the bracket.

To attach the calf rest and secure the calf rest in place, tighten the cap.

To position the calf rest, adjust the knob and position the calf rest.

Attaching and positioning the removable IV pole

CAUTION

Always make sure that the IV pole is at a low height to pass through door openings when you transport a patient.

To attach and position the removable IV pole (Figure 7):

  1. Insert the IV pole into a socket at the head end or foot end of the product.
  2. Turn the knob (A) counterclockwise and pull up on the telescoping portion (B) until you reach the desired height.
  3. Turn the knob (A) clockwise to lock the telescoping portion in place.

[Illustration of a removable IV pole with parts labeled 'A' and 'B']

Figure 7 - Removable IV pole

Attaching the upright oxygen bottle holder

WARNING

The upright oxygen bottle holder supports an oxygen bottle in a vertical position.

To attach the upright oxygen bottle holder:

  1. Insert the support bar into any of the IV sockets.
  2. Insert the cotter pin through the hole in the support bar to secure the bottle holder to the product.

Do not use the upright oxygen bottle holder as a push/pull device. Product damage may occur.

Extending or stowing the serving tray holder/footboard

WARNING - Do not place objects that exceed 30 lb (14 kg) on the serving tray.

To fit the serving tray on the siderail, pull out on both sides of the serving tray and position the tray over the siderails.

To stow the serving tray:

  1. Remove the serving tray from the siderails.
  2. Push in the sides of the serving tray.
  3. Store the serving tray in the footboard.

Do not use the serving tray/footboard as a push/pull device. Product damage may occur.

Accessing the Fowler X-ray cassette holder

To access the Fowler X-ray cassette holder:

  1. Raise the Fowler backrest (Raising or lowering the Fowler backrest (page 12)).
  2. Grasp and squeeze the handles (A) until the locating pins (B) release from the mounting brackets (C).
  3. Lower the tray and insert the X-ray cassette into the Fowler X-ray cassette holder.
  4. Loosen the knob on the front of the tray and then slide the tray to the desired position. Tighten the knob.
  5. Reverse steps to secure the X-ray cassette in the mounting brackets (C).

To remove the tray from the Fowler, lift the bottom of the tray out of the mounting brackets (D).

[Illustration showing how to access the Fowler X-ray cassette holder, with parts labeled A, B, C, and D]

Figure 8 - Accessing the Fowler X-ray cassette holder

Locating the patient restraint strap tie-ins

WARNING

There are six patient restraint strap tie-in locations on the litter assembly for attaching patient restraint straps (Figure 9).

[Illustration showing the six patient restraint strap tie-in locations on the litter assembly]

Figure 9 - Restraint strap tie-in locations

Cleaning and disinfecting with wipes

For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770.

Stryker's preferred wipes (2060-000-001 6" x 10" or 2060-000-002 9" x 12") include the following active ingredients:

Non-active ingredient: Ethylene Glycol Monobutyl Ether < 3%

For safety information, read the product label.

To clean or disinfect the external product surface:

  1. To clean, wipe external surfaces with a fresh, clean wipe to remove all visible soils. Repeat as necessary until the product is clean.

Use as many wipes as necessary.

Complete step 1 before you disinfect.

  1. To disinfect, wipe external surfaces with a fresh, clean wipe until wet. Allow the external surface to remain wet for two minutes at room temperature.
  2. Allow the product to dry before you return it to service.

Cleaning

Cleaning the product

WARNING - Do not clean, service, or perform maintenance while the product is in use.

Recommended cleaning method:

  1. Follow the cleaning solution manufacturer's dilution recommendations.
  2. Hand wash all surfaces of the product with warm water and mild detergent.
  3. Avoid over-saturation and make sure that the product does not stay wet longer than the detergent manufacturer's guidelines for proper cleaning.
  4. Dry thoroughly. Do not replace the mattress on the product until the product is dry.
  5. Check functionality before you return the product to service.

Direct skin contact with visibly soiled, permeable material may increase the risk of infection.

Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left on the surface of the product that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.

Do not steam clean, power wash, hose off, or ultrasonically clean the product. Use of these methods of cleaning is not recommended and may void the product's warranty.

Clean the base hood.

Clean the bottom of the brake pads to prevent wax or floor remnant buildup.

Cleaning the mattress

WARNING

The life of the mattress can be affected by an increase in frequency of usage, which might include more frequent cleaning and disinfection.

Recommended cleaning method:

  1. Use a clean, soft cloth to wipe down the entire mattress with a mild soap and water solution to remove foreign material.
  2. Wipe down the mattress with a clean, dry cloth to remove any excess liquid or cleaning agents.
  3. Rinse and dry covers after cleaning.
  4. Disinfect as needed with a hospital grade disinfectant after cleaning has been completed (see Disinfecting the mattress).

Do not iron, dry-clean, or tumble dry the mattress, as this will cause malfunction and damage the product.

The mattress cover must be completely dry before you store, add linens, or place a patient on the mattress to prevent impairment of the product performance.

Avoid over-exposure to alcohol or hydrogen peroxide. The cover material will swell.

Do not allow liquid to seep into the zipper area and watershed cover barrier. Fluids allowed to come in contact with the zipper may leak into the mattress which could impair the product performance.

Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.

Remove iodine

  1. Make a solution of 1 to 2 tablespoons of sodium thiosulfate in a pint of warm water. Use the solution to wipe down the stained area.
  2. Clean the stain as soon as possible after the stain occurs.
  3. If stains are not immediately removed, allow solution to soak or stand on the mattress before you wipe the mattress.
  4. Rinse the mattresses which have been exposed to the solution with clear water before you return the mattresses to service.

Failure to follow these directions when you use these types of cleaners may void this product's warranty.

Special instructions

Condition Procedure
Velcro® Saturate with disinfectant, rinse with water, and allow the solution to evaporate.
Solids or stains Use neutral soaps and warm water. Do not use harsh cleansers, solvents, or abrasive cleaners.
Hard-to-clean spots Use standard household cleansers or vinyl cleansers and a soft bristle brush on troublesome spots or stains. Pre-soak dried-on soil.
Laundering Laundering is not recommended. Laundering may substantially decrease the useful life of the mattress.

Disinfecting

Disinfecting the product

WARNING

Recommended disinfectants:

Recommended disinfection method:

  1. Follow the disinfectant solution manufacturer's dilution recommendations.
  2. Hand wash all surfaces of the product with a disinfectant solution.
  3. Avoid over-saturation and make sure that the product does not stay wet longer than the chemical manufacturer's guidelines for proper disinfecting.
  4. Dry the product. Do not place the mattress on the product until the product is dry.
  5. Disinfect the Velcro® after every use. Saturate the Velcro® with disinfectant, rinse with water, and allow the disinfectant to evaporate (appropriate disinfectant is determined by the facility).
  6. Check functionality before you return the product to service.

Direct skin contact with visibly soiled, permeable material may increase the risk of infection.

Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.

Disinfecting the mattress

WARNING - Always disinfect the mattress between patients. Failure to do so could result in cross-contamination and infection.

Recommended disinfectants:

Recommended disinfection method:

  1. Make sure that the mattress is clean and dry before you apply disinfectants.
  2. Wipe down the mattress with a clean, dry cloth to remove any excess liquid or disinfectant.
  3. Rinse and dry covers after disinfection.

The mattress cover must be dry before you store or add linens. Failure to remove excess disinfectant could cause degradation of the cover material.

Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.

Frequent or prolonged exposure to higher concentrations of disinfectant solutions may prematurely age the cover fabric.

The use of accelerated hydrogen peroxides or quaternaries that contain glycol ethers may damage the cover.

Preventive maintenance

Remove product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.

Clean and disinfect the exterior of the mattress before inspection, if applicable.

Inspect the following items:

Product serial number:

Completed by:

Date:

PDF preview unavailable. Download the PDF instead.

1061-109-001C PDFlib PDI 9.0.6 (C /Win64) PTC Arbortext Publishing Engine

Related Documents

Preview Stryker Gynnie OB/GYN Stretcher Maintenance Manual
This maintenance manual provides detailed instructions and procedures for the operation, service, and preventive maintenance of the Stryker Gynnie OB/GYN Stretcher (Model 1061). It is intended for healthcare professionals and trained personnel.
Preview Stryker Prime Series Stretcher: Mobile, Ergonomic, Adjustable Patient Transport
Discover the Stryker Prime Series Stretcher, designed for enhanced patient comfort, reduced caregiver strain, and improved clinical efficiency. Features include motorized drive, advanced mobility, and ergonomic design for patient transport and care.
Preview Stryker ST1 Stretcher: Basic Patient Transport Specifications
Detailed specifications and features of the Stryker ST1 stretcher, a basic patient transport solution designed for fundamental department needs, including mobility, siderails, and optional accessories.
Preview Stryker Model 1004 ASC Stretcher Operations Manual
Comprehensive operations manual for the Stryker Model 1004 ASC Stretcher, covering setup, operation, maintenance, cleaning, and warranty information.
Preview Stryker 747/748 Transport Stretcher Maintenance Manual
This maintenance manual provides essential guidance for the operation, cleaning, preventative maintenance, and servicing of the Stryker Model 747/748 Transport Stretcher, ensuring safe and effective use in healthcare environments.
Preview Stryker Prime Connect Smart Transport Stretcher
Discover the Stryker Prime Connect, an intelligent and connected transport stretcher designed for reliable patient care. Features include visual and audible bed exit alarms, enhanced mobility options like Big Wheel and Zoom-motorized drive, and patient comfort controls.
Preview Stryker Prime Series Stretchers: Mobile, Smart, Efficient
Detailed specifications and features of Stryker's Prime Series Stretchers, highlighting patient comfort, caregiver efficiency, and advanced mobility options.
Preview Stryker Prime Series Stretcher Operations Manual
Comprehensive operations manual for the Stryker Prime Series Stretcher with Zoom Motorized Drive, detailing setup, operation, safety, cleaning, and maintenance procedures for healthcare professionals.