Fisher & Paykel AIRVO™ 2 User Manual

This manual is intended for healthcare professionals and applies to AIRVO 2 units with LOT numbers 130621 and above.

1. Overview

The AIRVO 2 is a humidifier with an integrated flow generator that delivers high-flow warmed and humidified respiratory gases to spontaneously breathing patients through various patient interfaces. It is designed for patients in hospitals and long-term care facilities.

Intended Use

The AIRVO 2 is for the treatment of spontaneously breathing patients who benefit from receiving high-flow warmed and humidified respiratory gases, including those with bypassed upper airways. The flow rate can range from 2 to 60 L/min, depending on the patient interface.

Warnings

2. Setting up AIRVO 2

Before You Begin

The AIRVO 2 should be fixed on a pole mounting tray (900PT405) below patient head height. Open the packaging of the tube & chamber kit.

Install Water Chamber

Remove blue port caps, fit the supplied adapter over the chamber's vertical ports, and attach the water supply tube. Fit the water chamber to the unit by aligning it with the blue chamber port and pushing firmly until it clicks.

Warnings: Do not start the unit without the water chamber. The water becomes hot during use; exercise caution when removing/emptying. Do not touch the heater plate, chamber, or base during use. Avoid tilting the machine to prevent water entering the unit. Empty the water chamber before transporting.

Cautions: For optimal therapy (MR290 only), do not use the auto-fill MR290 chamber if dropped or run dry with the "water out" alarm activated.

Connect Water Bag

Attach the sterile water bag to the hanging bracket above the unit. Insert the bag spike into the fitting at the bottom of the bag and open the vent cap. The chamber will automatically fill and maintain the water level. Ensure the chamber and/or water bag do not run out of water.

MR290 Flow setting vs usage time (2-litre sterile water bag):

L/min 2 5 10 15 20 25 30 35 40 45 50 55 60
hrs 379 152 76 51 38 30 25 22 19 17 15 14 13

3. Using AIRVO 2

Switch On Unit

Plug the power cord into the mains supply, ensuring the connector is secure. Ensure the unit is dry before plugging in. Switch on using the On/Off button.

Check Disinfection Status

The unit displays whether it is safe for use on a new patient based on its last disinfection.

Warm-up

The unit warms up, displaying current dew-point temperature, flow, and oxygen values. These values pulse until they reach target settings. This is the "Summary screen".

Junior Mode

For Optiflow Junior nasal cannulas (OPT316/OPT318), activate Junior Mode by holding the Mode button for 5 seconds. This limits target settings to 34°C and 2-25 L/min. The screen icons indicate Junior Mode. Hold the Mode button for 5 seconds to deactivate.

Configure Target Settings

Press the Mode button to view target settings (locked by default).

Target Dew-Point Temperature

Settings: 37°C (98.6°F), 34°C (93°F) [for compliance issues], 31°C (88°F) [for face masks only]. Settings may be limited based on configuration. The unit returns to 37°C after disinfection.

To change: Hold Up/Down buttons for 3 seconds to unlock. Use Up/Down to select, then press Mode to lock.

Transport Mode

Activate by holding the "Audio pause" button for 5 seconds. Enters a low-power, low-humidity mode for 20 minutes for patient transport. Deactivate by holding the "Audio pause" button for 5 seconds.

Target Flow

Settings: 10-60 L/min (increments of 1 L/min for 10-25 L/min, 5 L/min for 25-60 L/min). Settings may be limited. The unit remembers the last flow setting.

To change: Follow the same steps as for changing dew-point temperature.

Note: Low ambient temperatures may prevent reaching 37°C at high flow settings. Consider decreasing the target flow.

Connect Your Patient

Wait for the "Ready for use" symbol on the Summary screen. Connect the patient interface to the heated breathing tube. Monitor flow and oxygen values, adjusting oxygen as needed. The air will feel warm initially.

During Use

A screensaver activates after 1 minute of inactivity. If excess condensate forms in the heated breathing tube, drain it into the water chamber.

After Use

Switch off the unit by pressing the On/Off button.

4. Reprocessing

The AIRVO 2 must be cleaned and disinfected between patients per the Disinfection Kit Manual (900PT600). Follow aseptic techniques to minimize contamination. Dispose of consumables properly.

Schedule for Changing Accessories

Accessories must be changed frequently to prevent infection risk. Replace immediately if damaged or discolored, or within the specified periods:

Maximum period of use Part number and description
1 week (single-patient use) All patient interfaces: OPT316 (Infant), OPT318 (Pediatric), OPT842 (Small), OPT844 (Medium), OPT846 (Large), OPT870 (Tracheostomy), RT013 (Mask Interface Adapter - 22mm)
2 weeks (single-patient use) All tube & chamber kits: 900PT501, 900PT531
3 months or 1000 hours Air filter: 900PT913 (or more often if significantly discolored)

Filter Replacement

When prompted to change the filter:

  1. Remove the filter holder from the back of the unit and remove the old filter.
  2. Replace with a new filter.
  3. Reattach the filter holder, clipping the bottom in first, then rotating the top into place.
  4. Press the Mode button to proceed.

Servicing

This device contains no serviceable parts.

5. Technical Information

Symbol Definitions

Various symbols are used on the unit and in this manual, including: Caution, Hot Surfaces, Applied Part, IPX1 Drip Proof, Consult Accompanying Documents, Do Not Throw Away, Alternating Current, Class II Double Insulated, Power On/Off (Standby), CE 0123, 93/42/EEC Class IIa.

Product Specifications

Designed to conform to the requirements of:

IEC 60601-1, UL 60601-1, CSA C22.2/No. 601.1, AS 3200.1.0, EN 60601-1. Complies with IEC 60601-1-2 for electromagnetic compatibility. Ensure system compliance if connecting external equipment.

Operating Conditions

Storage and Transport Conditions

Store and transport in environmental conditions of -10°C to 60°C (14°F to 140°F), 10 to 95% RH, non-condensing.

Disposal Instructions

Unit Disposal: Do not discard with regular waste. Return to Fisher & Paykel Healthcare or dispose of according to local electronics disposal guidelines (WEEE directive in EU).

Consumables Disposal: Place interface, breathing tube, and chamber in a waste bag at the end of use. Dispose of according to hospital procedures for contaminated product.

6. Alarms

The AIRVO 2 has visual and auditory alarms for treatment interruptions. The alarm system processes sensor information and compares it to pre-programmed limits.

Alarm Signals

Alarm Conditions

Alarms are assessed as "Medium Priority". Simultaneous alarms display the highest priority.

Message Meaning Affects delivery of: Delays
Fault (E###) Internal fault, unit shut down. Restart. If persists, contact Fisher & Paykel Healthcare. Oxygen, humidity < 5 seconds
Check tube Heated breathing tube not detected. Check for damage and proper connection. Replace if necessary. Oxygen, humidity < 5 seconds
Check for leaks System leak detected. Check water chamber seating, tube connection, nasal interface fit, and filter fit. Oxygen, humidity < 5 seconds
Check for blockages System blockage detected. Check tube, interface, and air filter. Ensure Junior Mode is activated if using Junior cannula. Oxygen, humidity < 10 seconds
O₂ too low Measured oxygen level below limit. Check oxygen source connection and adjust level. Oxygen < 20 seconds
O₂ too high Measured oxygen level above limit. Adjust oxygen source level. Oxygen < 20 seconds
Cannot reach target flow Unit cannot reach target flow. Check tube/interface for blockage. Verify target flow is appropriate for the interface. Unit may suggest new settings. Oxygen 10 +/- 1 minutes
Check water Chamber empty. Replace chamber and water bag. Unit enters Warm-up Mode upon replacement. Ensure water is always present. Humidity Flows > 20 L/min: < 20 minutes; Flows ≤ 20 L/min: < 40 minutes
Cannot reach target temperature Unit cannot reach target temperature. May be due to high flow rate in low ambient conditions. Consider decreasing target flow. Humidity 30 +/- 3 minutes
Check operating conditions Unsuitable ambient conditions (below 10°C or above 30°C). Allow unit to run for 30 minutes or restart. Humidity 60 +/- 6 seconds
[Power out] Unit disconnected from mains power. No visual alarm. Auditory alarm sounds for 120 seconds. Oxygen, humidity < 5 seconds

Alarm Limits

Most alarm limits are pre-programmed. Some can be changed by authorized personnel.

Checking Alarm System Functionality

Remove the heated breathing tube. The "Check tube" alarm should activate. If not, contact Fisher & Paykel Healthcare.

Auditory Information Signals

7. Contact Information

For more information, contact your local Fisher & Paykel Healthcare representative.

Manufacturer: Fisher & Paykel Healthcare Ltd, 15 Maurice Paykel Place, East Tamaki, Auckland 2013, New Zealand. Tel: +64 9 574 0100, Fax: +64 9 574 0158, Email: info@fphcare.com, Web: www.fphcare.com

Contact details for Australia, Spain, Sweden, Switzerland, Taiwan, Turkey, UK, and USA/Canada are also provided.

Models: AIRVO 2, Water Spray Humidifier, AIRVO 2 Water Spray Humidifier, Spray Humidifier, Humidifier

File Info : application/pdf, 15 Pages, 1.28MB

PDF preview unavailable. Download the PDF instead.

myAIRVO-2-Humidifier-user-manual RAD PDF 3.19.2.2 - https://www.radpdf.com PDFescape Online - https://www escape.com

Related Documents

Preview SleepEh.ca F&P Eson & Eson 2 CPAP Mask Sizing Guide
Download the SleepEh.ca sizing guide for Fisher & Paykel Eson and Eson 2 CPAP masks. Learn how to properly size your mask for optimal comfort and performance with clear instructions and visual aids.
Preview Fisher & Paykel Series 11 Integrated Tall Double DishDrawer Dishwasher - Quick Reference Guide
Explore the features, specifications, and benefits of the Fisher & Paykel Series 11 Integrated Tall Double DishDrawer Dishwasher. Learn about its seamless kitchen integration, dual-drawer functionality, Wi-Fi connectivity, multiple wash programs, and efficient performance ratings.
Preview Fisher & Paykel 30-Inch Series 9 Contemporary Self-Cleaning Oven Quick Reference Guide
Explore the Fisher & Paykel 30-inch Series 9 Contemporary Self-Cleaning Oven. Features 4.1 cu ft capacity, 11 cooking functions, AeroTech™ technology, and Pyrolytic self-cleaning. Includes dimensions, specifications, and support information.
Preview Fisher & Paykel 48" Series 11 Professional Induction Range (RIV3-486) - Quick Reference
Comprehensive overview of the Fisher & Paykel 48" Series 11 Professional 6 Zone Induction Self-Cleaning Range (RIV3-486). Details features like SmartZones, generous capacity, guided cooking, and easy cleaning, alongside specifications, dimensions, and power requirements.
Preview Fisher & Paykel 30" Stainless Steel Kickstrip for Freestanding Range - Model 81813
Detailed specifications and dimensions for the Fisher & Paykel 30-inch stainless steel kickstrip, designed for Classic & Contemporary freestanding ranges compatible with OR30 models. Includes SKU and customer support information.
Preview Fisher & Paykel Professional Round Flush Handle Kit for Integrated Refrigerator Freezer, 60cm - Quick Reference Guide
Quick reference guide for the Fisher & Paykel Professional Round Flush Handle Kit for Integrated Refrigerator Freezer, 60cm. Provides product details, dimensions, SKU, compatibility information, and customer support contact details.
Preview Fisher & Paykel 24" Series 7 Contemporary Microwave Drawer Quick Reference Guide
Quick reference guide for the Fisher & Paykel 24-inch Series 7 Contemporary Microwave Drawer (OMD24SDB1), detailing features, benefits, specifications, and usage information.
Preview Fisher & Paykel 24" Series 9 Minimal Convection Speed Oven OM24NMTDB1 Quick Reference Guide
Quick reference guide for the Fisher & Paykel 24" Series 9 Minimal Compact Convection-Speed Oven (OM24NMTDB1). Features, specifications, dimensions, functions, performance, power requirements, and customer support.