CTS™ DynaCellect™ Magnetic Separation System

Catalog Number: A55867

Pub. No.: 100117621 Rev. A

Introduction and Safety Notes

Note: For safety and biohazard guidelines, see the “Safety” appendix in the CTS™ DynaCellect™ Magnetic Separation System User Guide (Pub. No. MAN0026480). Read the Safety Data Sheets (SDSs) and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.

Important Guidelines

Unpack and Set Up the Instrument

  1. Lift the cardboard box off the pallet.

    Image: A person lifting a large cardboard box off a pallet.

  2. Remove the two small cardboard boxes containing the bag hangers, the pole extensions and the pole adaptors.

    Image: Two small cardboard boxes being removed from larger packaging.

  3. Remove the top foam.

    Image: A piece of foam being lifted out of the main packaging.

  4. Unwrap packaging bag and remove the ethernet cable and power cable.

    Image: An ethernet cable and a power cable being unwrapped from protective packaging.

  5. Lift the instrument to bench or cart. The operation requires two people. Use the four holding recesses (front left side, back right side, and on both sides) to lift in a diagonal position.

    ⚠️ IMPORTANT! The instrument weighs 50 kg (109.7 lbs) and requires two people to lift.

    Image: Two individuals lifting the instrument using designated recesses, indicating the need for two people.

  6. Place the instrument in a protected location with at least 10 cm (4 inches) to the rear for proper operation of the exhaust fans. See "Important guidelines" on page 1.

    Image: The instrument placed on a surface with adequate clearance at the rear.

  7. Remove the rocker homing position foam.

    Note: It is acceptable for the rocker to freefall by gravity after removing the foam. Do not manually move the rocker once it has settled into its locked resting position.

    Image: Foam being removed from the instrument's rocker mechanism.

  8. Install the two pole adaptors to the two pole extensions, and install both on the pole base on the instrument.

    Diagram labels: ① Pole adaptor, ② Pole extension, ③ Pole base.

    Image: Close-up of pole adaptors being attached to pole extensions, which are then mounted onto the instrument's base.

  9. Place the bag hangers on the pole adaptors.

    Image: Bag hangers being placed onto the installed pole adaptors.

  10. Connect the power cable to the power inlet, and insert the plug into an electrical outlet.

  11. (Optional) Connect one end of an ethernet cable to the instrument ethernet port, and the other end to an ethernet port wall plug.

Contents

ComponentQuantity
CTS™ DynaCellect™ Magnetic Separation System (Cat. No. A55867)1
Region specific power cord (for US/Canada/Taiwan/Japan, Europe, or UK)1
Ethernet cable1
Pole extension2
Pole adaptor2
Bag hanger2
Quick Reference Guide1
Installation Guide1

Working Space

The CTS™ DynaCellect™ Magnetic Separation System is designed for benchtop use and requires a working space of approximately height 495 mm (19") with pole base and 434 mm (17") without pole base, width 749.5 mm (29.5"), depth 562 mm (22.12"). The system can also be placed on a cart.

Diagram illustrating the required working space dimensions for the CTS DynaCellect Magnetic Separation System, showing height, width, and depth.

Product Specifications

CategoryPropertyValue
ElectricalSupply voltage100–240 VAC ±10%, 50/60 Hz, 600 W
PhasesSingle
Maximum rated input current6 A
Fuses2 x 10 A
InstrumentRegulatory requirementsHEPA filter meets ISO 14644-14
Installation categoryII
Instrument typeBenchtop/cart unit
Instrument dimensions495 mm (19") with pole base/ 434 mm (17") without pole base (H) x 749.5 mm (29.5") (W) x 562 mm (22.12") (D)
Instrument weight49.75 kg (109.7 lbs) (50.6 kg/111.6 lbs with poles and bag hangers)
Sound levelOperating sound level73 dBA (measured 1 m from instrument)
Non-operating sound level63 dBA (measured 1 m from instrument)
Environmental rangesAmbient temp15°C to 30°C
Transport temperature-30°C to 60°C (extreme end temperatures at maximum 72 hours)
Storage temperature15°C to 30°C
Relative humidity15%–80% (non-condensing)
Operating limitsMagnetEngaged/Disengaged
Pinch valvesOpen/Closed
Pump speed10–150 mL/min
Rocker angle-30° to 30°
Rocker speed1-30 CPM (cycles per minute)
Pressure occlusion sensorMax 2 Bar. Only detect liquid pressure (not pressure build up by air). Following lines are not covered: CTS™ DynaCellect™ Cell Isolation Kit – Ports B and K; CTS™ DynaCellect™ Bead Removal Kit – Ports B, D and E
Height rocker lid (isolation)35 mm (1.4")
MagnetHeight rocker lid (bead removal)5 mm (0.2")
Overfilled detection1120 mL (static and rocking)
Magnet dimensions265 x 205 mm (10.4" x 8.07")
Magnetic strength≥ 6000 G
PoleMax height1070 mm (42.13") from the bottom of the rubber foot of the instrument to the tip of the bag hanger hook point
Max weight3 kg (6.6 lbs) per hook on bag hanger. Total 15 kg (33 lbs) per bag hanger
UseIndoor use only
Not intended for use in a wet environment
Intended for for use in Pollution degree 2 environment

Limited Product Warranty

Life Technologies Corporation and/or its affiliate(s) warrant their products as set forth in the Life Technologies' General Terms and Conditions of Sale at www.thermofisher.com/us/en/home/global/terms-and-conditions.html. If you have any questions, please contact Life Technologies at www.thermofisher.com/support.

Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256

For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.

Revision History

RevisionDateDescription
A17 October 2022New document for CTS™ DynaCellect™ Magnetic Separation System.

Disclaimers and Legal Information

The information in this guide is subject to change without notice.

DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.

Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all applicable Limited Use Label Licenses.

©2022 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.

PDF preview unavailable. Download the PDF instead.

100117621-CTSDynaCellectMagneticSeparationSystem-IG Antenna House PDF Output Library 7.2.1732

Related Documents

Preview Thermo Fisher Scientific Immuno-Oncology Product Resource Guide
A comprehensive guide to Thermo Fisher Scientific's tools, technologies, and solutions for immuno-oncology (I-O) research, covering checkpoint inhibition, CAR T cell therapy, and cancer vaccines.
Preview Recombinant Human Interferon-γ (IFN-γ) Product Information Sheet | Thermo Fisher Scientific
Detailed product information sheet for Recombinant Human Interferon-γ (IFN-γ) from Thermo Fisher Scientific, including specifications, biological activity, storage, references, and warranty. Catalog numbers PHC4031 and PHC4033.
Preview Lipofectamine 3000 Transfection Protocol for BT-474 Breast Cancer Cells
Detailed protocol for using Invitrogen Lipofectamine 3000 Transfection Reagent with BT-474 breast cancer cells, covering cell culture, passaging, seeding, transfection steps, and efficiency analysis.
Preview Thermo Fisher Connect: Remote Access for Thermal Cyclers - Quick Reference Guide
A quick reference guide on setting up and using Thermo Fisher Connect for remote access and control of SimpliAmp and ProFlex PCR systems. Covers requirements, account creation, firmware upgrades, instrument linking, and remote monitoring via desktop and mobile devices.
Preview Thermo Compound Discoverer User Guide - Software Version 3.0
Comprehensive user guide for Thermo Fisher Scientific's Compound Discoverer software (v3.0). Learn data processing, compound identification, workflow creation, and results analysis for mass spectrometry research.
Preview Thermo Fisher Scientific NEXT Magazine - March 2022
Explore the latest in life sciences research with Thermo Fisher Scientific's NEXT magazine (Issue 61, March 2022). This issue features an interview on SARS-CoV-2 antibodies, user voices on various research applications, and highlights of cutting-edge instruments and reagents, including cell sorters, gene editing tools, NGS solutions, and cell culture products.
Preview Thermo Fisher Scientific Bioprocessing Solutions Catalog
Explore Thermo Fisher Scientific's comprehensive bioprocessing solutions, covering everything from drug development to commercial production. This catalog details single-use technologies, production chemicals, and analytical methods designed to optimize biomanufacturing processes.
Preview Thermo Scientific Apreo 2 SEM: Advanced Field Emission Scanning Electron Microscope
Datasheet for the Thermo Scientific Apreo 2 SEM, a high-performance Field Emission Gun (FEG) Scanning Electron Microscope (SEM) featuring ColorSEM Technology for live elemental imaging, SmartAlign Technology for automated alignment, and FLASH Technology for advanced image tuning. Discover its versatile applications, detailed specifications, and advanced electron optics.