
User Guide for Mammotome models including: MEP10 Elite Biopsy Device, MEP10, Elite Biopsy Device, Biopsy Device, Device
Mammotome Elite Tetherless Vacuum-Assisted Biopsy System | Mammotome Product Support
File Info : application/pdf, 3 Pages, 1.53MB
DocumentDocumentQuick Reference Guide Charging the Holster After cleaning and disinfecting, place the holster into the charging base. The holster will emit a two-tone beep to indicate correct placement into the base. Holster On Charger SOLID = Battery is fully charged. FLASH = Battery is charging. FLASH = Battery is charging. Holster must remain on charging base. Holster Not On Charger SOLID = Sufficient charge available to complete one or more procedure(s). FLASH = Sufficient charge available to complete one procedure. SOLID = Insufficient charge. Return holster to charging base. When charging, set the holster on top of this guide for quick, easy reference. Troubleshooting If the fault indicator light is illuminated: · Ensure the probe is outside the patient. · Disassemble the probe from the holster and replace it with a new probe. · If the indicator light does not turn off, contact Mammotome Technical Support. If the battery indicator light is not illuminated after probe loading: · Check the battery indicator to ensure there is sufficient charge. · Ensure the alignment and locking tabs are securely in place. · Ensure it is the initial use of the probe. · Replace the probe with a new probe. · If the indicator light does not turn on, contact Mammotome Technical Support. Please read the full Instructions For Use before using this product. 1 Loading the Probe Step 1 A. B. Grasp the probe body and align the alignment tab with the holster notch (A). Push the probe base into the holster while maintaining the alignment (B). Step 2 Push the probe body firmly into the holster until a click is heard, indicating the locking tabs have engaged the holster (C). C. Note: Automatic initialization will begin. If the holster is left undisturbed for 10 minutes after loading a probe, the holster will enter a sleep mode to conserve battery charge. During sleep mode, the biopsy and battery indicator lights will turn off and the buttons will be inactive. Once the holster is moved, the device will wake from sleep mode. Any time the holster is motionless for one hour with or without a probe attached, the holster will go into a deep sleep mode. Place the holster into the charging base without the probe and remove from the charging base to turn the holster back on. 2 Priming the Probe Interrupting the Biopsy Cycle Interrupt a biopsy cycle (as described below) leaving the aperture A. exposed (A). Use a sterile syringe to insert 0.5cc of saline into the needle aperture (approximately 7-10 drops). Press the active biopsy button to advance the cutter. Push the active biopsy button once to initiate the biopsy cycle. To interrupt the biopsy cycle, push the active biopsy button again when the cutter is retracting from position 1 to 3 (B) or when the cutter is advancing from position 3 to 2. The cutter will retract to position 3, B. and the biopsy indicator light will flash orange. Complete the cycle by pushing the active biopsy button. 3 Biopsy Button Selection Ensure the probe is attached to the holster. Press and hold the two inactive biopsy buttons simultaneously for three seconds. There will be an audible indicator and all three indicator lights will A. B. flash green (A). Press and release the desired active biopsy button. The active biopsy button is stored in the device for future use (B). 4 Tissue Sampling Procedure Introduce the probe through the incision and position the probe tip in the target area with the tip past the lesion. Imaging guidance should be A. used as appropriate. Push the active biopsy button on the holster once to acquire a tissue sample. The biopsy cycle proceeds: a vacuum is created, the inner cutter of the probe is retracted, tissue is drawn into the aperture, and the cutter rotates and translates forward, cutting tissue (A). When the cutter reaches the distal end of the aperture, the vacuum will transport the sample to the specimen collection cup. B. C. Needle Rotation (optional) Pull the needle rotation knob out (B). Then rotate to desired clock position and release (C). 5 Removing the Probe from the Coaxial Cannula Disconnect the integrated coaxial cannula and remove the probe by pressing the coaxial release button, leaving the cannula in place. 6 Post Sampling Instructions Step 1 A. B. Rotate the specimen collection cap counterclockwise 1/4 turn (A). Step 2 Remove the cap and tissue basket together (B). Step 3 C. Press the tissue basket release button to remove the tissue basket from the specimen collection cap for ease of transport to pathology (C). Note: Sterile forceps may be used to aid in removing the samples from the tissue basket, if needed. Removing the Probe from the Holster While pressing down on the probe's locking tabs, pull the probe out of the holster cavity (A). Dispose of the probe in accordance with the institution's A. B. policy for infectious waste disposal (B). Product Codes Code Description Qty MEH1 Mammotome® Elite Holster (includes charger base and power cord) 1 MEP10 10G Mammotome® Elite Ultrasound Probe 5 MEP13 13G Mammotome® Elite Ultrasound Probe 5 MEI10 10G Mammotome® Elite Introducer Stylet 5 MEI13 13G Mammotome® Elite Introducer Stylet 5 Technical Specifications Attribute Needle Length (w/o coaxial cannula) Needle Length (w/ coaxial cannula) Needle Diameter Aperture Length Dead Space Tip to Center MEP10 136mm 122mm 3.5mm 19.1mm 10.0mm 19.6mm MEP13 136mm 122mm 2.6mm 18.4mm 10.4mm 19.6mm Technical Support For technical support contact us via phone at 1-877-926-2666 (option 3) or by email at TechnicalSupport@mammotome.com. Order Placement Contact your Customer Service Representative at 1-877-926-2666 between the hours of 8:00am and 5:00pm EST. Shipping details and an order confirmation number will be provided to you upon completion of your order. Send your order via fax to your Customer Service Representative at 1-888-260-6362. Send your order via email to your Customer Service Representative at us.customerservice@mammotome.com. For complete product details, see Instructions for Use. mammotome.com Product may not be approved or available in your region. Please check with your local Mammotome representative. © 2022 Devicor Medical Products, Inc. All rights reserved. Devicor, Mammotome and Mammotome Elite are registered trademarks of Devicor Medical Products, Inc., in the USA and optionally in other countries. Other logos, product and/or company names might be trademarks of their respective owners. MDM# 12-0041 Rev 9/22