All-in-One Blood Glucose Meter
Quick Start Guide
Cassette
Integrated Lancing
Click here to watch video

Contents Parts summary Preparing the meter Setting the time and date Run a test Connecting to the App Clean the Meter E-4, E-5, E-9, E-12 Key to Error Messages

Parts summary

1. Sensor cover

7. Lancet

2. Display

8. Lancet ejector

3. Sample/settings button

4. Memory recall button

5. Penetration depth selector

6. Lancet holder

9. Test Cassette 10. Thumbwheel 11. Battery cover 12. Lancet cover 13. Penetration depth indicator 14. Sensor

Preparing the meter
1
Remove the protective film from the screen
Pull protective tab from the battery compartment
2 Slide the sensor cover to the left as
far as it will go. Meter will turn on.

3 Insert cassette, rounded end first
and press flat. Close sensor cover.
Press Flat
4 Slide lancet cover open

5 Insert lancet and twist off the cap
6 Close lancet cover and set desired
penetration depth

Setting the time & date Video 1
1 Hold down the S button to enter the
settings menu.
Meter may show `exp' (cassette expired) if the correct date is not set.
2 Press M to change the flashing
character. Press S to save.
Minutes Month Hours Day
YEAR

Connecting to the App (Optional)
1 Download the App to your mobile
device.
Video 4
BETACHEK Diabetes Management App Apple and Android
2 Hold down the S + M buttons until
the pairing symbol* appears.
Open the app on your mobile device and follow the instructions.
*or equivalent symbol

Run a test

Watch Video 2

1 Wash your hands in warm water

2 Dry your hands
thoroughly

3 Open the sensor cover and
wait for the wind-on symbol:
4 Turn the thumbwheel
until a test zone is in position and screen displays "OK"

5 Press depth selector firmly against
selected puncture site, then press S*
Yes you can prick again * To repeat, close & open the sensor cover to re-cock
6
· Gently squeeze out a small round drop of blood.
· Keep the palm of your hand facing down.
· Use immediately. Do not allow the
drop to run, dry or spread.

7 Wait for the flashing blood drop
symbol and then apply blood
Aim for the red flashing light

You may press down on the blood drop to spread it. Do not touch the test zone
Cover the red illuminated area with blood or the meter will display E-7 or produce an inaccurate result.
Do not spread blood off the edge of the tape and onto the meter.
See Cleaning the Meter.

Do not smear a small drop across the test zone. A larger drop should be used.
8 Results in approx. 5 seconds

9 Turn the thumbwheel
to dispose of the used test, until "OK" is displayed.
10 Close the sensor cover. The meter
will turn off. NOTE: If you close the cover before winding the used test away E-3 will be displayed - open the cover and wind away.

Removing a cassette

Cleaning the Meter E-4, E-5, E-9, E-12
1 Remove the cassette: Push on
the curved grip + lift out from the top.
Clean the meter if you spread blood off the tape and onto the meter or if you get any of the above error messages
2 Clean sensor window with a damp
cotton tip or tissue in the direction shown.
Video 5

Key to Common Error Messages
Used test not wound into waste chamber. Open sensor cover and wind away the used test. Blood applied too early or Sensor window dirty. Wait for blood drop icon before applying blood or clean meter.
Sensor window dirty or too much direct sunlight. Clean the meter or move to shade. Cassette wound too far. Continue winding until meter says "OK". Insufficient blood. Use a larger drop or press down on the drop to spread it. Test zone error or Sensor window dirty. Clean the sensor window and repeat the test. Expired cassette. Cassette past expiry date or opened longer than 90 days or date set in meter is a future date - check year.
Go to the complete user manual

· The BETACHEK C50 meter and lancing device are intended for personal use only. Not for multi-patient use.
· The BETACHEK C50 is covered by a 5 Year Warranty from date of purchase. The warranty does not apply to accidental damage, normal wear and tear or if the meter is misused. To claim your warranty please contact: support@betachek.com
· This Quick Start Guide does not replace the user manual which contains important additional information.