Installation Guide for ELEMENTAL MACHINES models including: ED1, EK-120i, EK-200i, EK-300i, EK-600i, EK-1200i, EK-2000i, EK-3000i, EK-6000i, EK-12Ki, EW-150i, EW-1500i, EW-12Ki, EK-120i Compact Balance, EK-120i, Compact Balance, Balance

Ian Harding

772-00037 re...

Setting up Element-D with AND EK-i and EW-i Series Balance : Elemental Machines Helpdesk


File Info : application/pdf, 4 Pages, 228.40KB

PDF preview unavailable. Download the PDF instead.

19114441443
Element-D Installation Guide
Model: ED1
For: AND EK-i and EW-i Series Balance

Applicable Asset Models
 EK-120i  EK-200i  EK-300i  EK-600i

 EK-1200i  EK-2000i  EK-3000i  EK-6000i

 EK-12Ki
 EW-150i
 EW-1500i  EW-12Ki

Package Contents
 Element-D  Power Supply  Connection Cable
Note: You may receive multiple types of cables if you are connecting Element-D's to different types of assets.

Safety Information
Review and abide by the safety information found in the Element-D and AND EK-i and EW-i Series Balance user manuals. Tidy connection cable to prevent tripping hazards.

Installation Instructions

Receiving the Shipment
1. If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.
2. When devices are shipped, they will be added to your Elemental Insights DashboardTM with default names. When you first log in you will see all your devices with a 'disconnected' status.
3. When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com.
4. Confirm the Element-D's in your package are labeled with the assets to which they are to connect. If the labels do not match the intended assets for connection, please contact help@elementalmachines.com.
Element-D Initial Power-up
5. Plug the Element-D in using the provided power supply. 6. Complete Wi-Fi Connection following instructions in the provided 771-00010 Wi-Fi Connection Guide -
Gateway GW2/3, Element-C/D. a. 771-00010 is included with your Element-D as a printed copy. Digital versions of this are available in the
Support section of the Elemental Insights DashboardTM.

772-00037 rev 01 Element-D, ED1 AND EK-i and EW-i Series

Balance Installation Guide

1/4

Copyright © 2023 Elemental Machines help@elementalmachines.com

Element-D Installation Guide
Model: ED1
For: AND EK-i and EW-i Series Balance
Attaching Connection Cable to Asset
7. Locate the data port on the balance (RS232 C terminal shown in Figure 1) to which the Element-D will connect.

Figure 1 Data port on asset
8. Plug the RS232 end of the Connection Cable (Figure 2) into the data port on the balance.

Figure 2 Connector cable RS232 end (left) and USB end (right)
Configuring Asset Menus
9. While balance is off, press and hold the RE-ZERO button and press the ON:OFF while still holding the RE-ZERO button.

10. The screen will read Func. Press PRINT and the screen will read PoFF followed by a number.

772-00037 rev 01 Element-D, ED1 AND EK-i and EW-i Series

Balance Installation Guide

2/4

Copyright © 2023 Elemental Machines help@elementalmachines.com

Element-D Installation Guide
Model: ED1
For: AND EK-i and EW-i Series Balance
11. Press SAMPLE multiple times until the screen reads bPS followed by a number.
12. Use the RE-ZERO button until the screen displays bPS 2. 13. Press SAMPLE until the screen displays btPr followed by a number. 14. Use the RE-ZERO button until the screen displays btPr 2. 15. Press PRINT to store the parameters. 16. Press the ON:OFF button to return to the weighing display.
Attaching Connection Cable to Element
17. Plug the other end of the Connection Cable into the data port on the Element-D (Show in Figure 3). Use any of the four USB ports on the Element-D.

Figure 3 Data ports on Element-D
Positioning the Element-D
18. Determine where the Element-D should be placed to allow for power access and a tidy connection cable. 19. Securely position the Element-D in the chosen location. The Element-D has magnets which make it easily
mountable to magnetic surfaces. a. Note: If applicable, an adhesive metal mounting plate available from EM may be used.

772-00037 rev 01 Element-D, ED1 AND EK-i and EW-i Series

Balance Installation Guide

3/4

Copyright © 2023 Elemental Machines help@elementalmachines.com

Element-D Installation Guide
Model: ED1
For: AND EK-i and EW-i Series Balance
Finalization
20. To send a weight measurement to the Elemental Insights Dashboard, press PRINT.
21. Confirm that the value appears on the dashboard. (The dashboard may only display two digits after the decimal point. However, transfer to any third party platforms will include the full value.).
22. If the dashboard does not display any value, check cable connection and menu settings. Contact help@elementalmachines.com if problem persists.
23. Contact any third-party vendor for next steps regarding communication to the external platform (if applicable).

772-00037 rev 01 Element-D, ED1 AND EK-i and EW-i Series

Balance Installation Guide

4/4

Copyright © 2023 Elemental Machines help@elementalmachines.com



References