· Measurement range -100 to +100°C (-148 to +212°F)
· T-type probe on 1m cable · Dual channel recording through two external thermocouple probes · Wirelessly stream and view data on the EasyLog Cloud, App
or on a PC
· View and analyse multiple sensors, including graphing of historic data · Configurable high and low alarms with indicator · Easy sensor set-up using free PC software or App

The EL-WiFi-DULT is designed to monitor vaccines in cryogenic dry ice storage.
Data is streamed wirelessly over any compatible WiFi network and can be viewed on a PC using free software or on the EasyLog Cloud or App.

To configure the sensor for use on a given wireless network, either connect it via USB to a PC running EasyLog WiFi software on the network, or configure wirelessly using the EasyLog Cloud app on a mobile phone with access to the network. The sensor can then be placed anywhere within range of the network. If the sensor temporarily loses connectivity with the network, it will log readings until it is able to communicate again with the PC application or EasyLog Cloud (max 30 days at 10 second sample interval).

The sensor is IEEE 802.11bgn* (2.4GHz) compliant, supports WEP, WPA/WPA2 encryption and enterprise networks (PEAP, TTLS, FAST).

EL-WiFi-DULT has a protection rating of IP40. The unit is freestanding, but can be attached to a wall or surface using the bracket provided. The unit can be clipped in and out of the bracket as required.

SPECIFICATIONS

ACCESSORIES

Minimum

Typical

Maximum

Unit

Battery life

>6

Months

Temperature measurement range

-100 (-148)

+100 (+212) °C (°F)

USB supply voltage (@500mA)

4.5

5.0

5.5

Vdc

Operating temperature range (logger)

-20 (-4)

+60 (+140) °C (°F)

Logging period (user configurable)

10 sec

10 min

12 hrs

Transmission period (user configurable)

1 min

1 hr

24 hrs

Temperature measurement resolution

0.1 (0.2)

°C (°F)

Temperature display resolution

0.1

Accuracy

±2.5 (±4.5)

°C (°F)

IP Rating

IP40
(Bung fully inserted, not permanently powered, thermocouple connector fitted and
fully inserted, device mounted vertically.)

Dimensions

82 x 70 x 36mm (3.22 x 2.75 x 1.41")**

Probe dimensions

300 x  1.5mm, 1m cable (11.8 x  0.05, 39.3" cable)

* MAC Address starting 98:8B:AD:2....... only **Excluding probe and mounting bracket

PSU-5VDC-USB-USA USB Mains Power Adapter for USA

PSU USB-UK

USB Mains Power Adapter for UK

PSU USB-EU

USB Mains Power Adapter for EU

EL-P-TC-T-ULT

Replacement T-type Thermocouple probe

INCLUDED IN THE BOX

EL-WIFI WALL BRACKET

Wall mounting bracket for EL-WiFi sensors

EL-P-TC-T-ULT

Thermocouple probe

CABLE USB A-MICRO B USB Type A to Micro B

CALIBRATION CERTIFICATES NOW AVAILABLE

Lascar now offers a Traceable Calibration Certificate Service on Temperature Data Loggers. Using reference equipment

which has been calibrated international standards. For

by a more

inUfKoArmS/aNtiIoSwTn wpalcewcarse.eldaistseeecdawlrawebwloer.lacattsocrraoyrnealinecdcstr.uocsnoinicmgs.c/aopdmpaa.traat-ulsogtrgaceerasble

to

national

or

Issue 1 11/2020 Page 1 of 2

EL-WiFi-DULT
Dual Channel Ultra Low Temperature Cryogenic Vaccine WiFi Data Logger
Easylog Software
Lascar's WiFi software is available as a free download from: www.lascarelectronics.com/software/easylog-wifi. Easy to install and use, EL-WiFi-WIN allows easy connection of sensors to a WiFi network. The user can select where data is stored - the PC or the Cloud.
When setting up your logger, make sure you select the T-Type probe.

Your data, anytime, anywhere

Battery Life and Power Supply
The battery can be recharged (unit must be between 0 - 40 °C) via a PC, a USB +5V wall adapter, or a portable USB battery pack using the USB lead provided. It can also be permanently powered by a USB wall adapter or USB battery pack. Readings may be affected while the internal battery is being charged. However, once charged, continued connection of the charger will have no effect.
Battery life is dependent on: transmission period, WiFi encryption method, WiFi encryption key rotation frequency (determined by the router/access point), signal strength between router/access point and WiFi device, presence volume and type of WiFi traffic from other devices, sample rate and operating temperature.