MerlinOnDemand™ capability with Merlin@home™ Transmitter Model EX1100

Quick Start Guide

Note: This transmitter with MerlinOnDemand™ capability retrieves full follow up data. It is configured for multiple patient use.

Send Patient Information

  1. Place the wand, green side towards the patient, over the implant.
  2. Quickly press and release the Start button. The Stars icon lights up.
  3. Quickly press and release the Start button again.
  4. Read: The Read icon flashes when the transmitter searches for the implant. It will stop flashing but remains lit while the transmitter reads the implant and the progress lights will move left to right.
  5. Send: When the Tower icon lights up, you can remove the wand. The transmitter dials out and sends information about the implant. This process may take several minutes. Depending on your connectivity this can be between 5-15 minutes. Do not use the phone line while the Tower icon is lit.
  6. Done: When the Stars icon lights up, the information is successfully sent. The Stars icon remains lit for one minute.
  7. Fax/email: You will receive a fax and/or email(s) based on your enrollment settings showing the patient's device information.

Troubleshooting

For Technical Support, contact your local St. Jude Medical representative.

Progress Lights

An error has occurred if all progress lights are lit or one or all are blinking. See below for error details.

Can't read information
  1. Press the Start button to stop the beeping.
  2. Reposition the wand over the implant.
  3. Press the Start button to continue.
  4. After 3 tries, check the Merlin.net manual to see if implant is supported.
Can't send information
  1. Press the Start button to stop the beeping.
  2. Check your wall jack and phone connector.
  3. Turn off any fax, computer, or TV that shares your phone line.
  4. Press the Start button to continue.
Unsupported implant detected
  1. Press the Start button to stop the beeping.
  2. Do not try to read the implant again. The transmitter cannot communicate with the implant.
Call Technical Support
  1. If all five progress lights are lit, call Technical Support.

For additional information, see the accompanying manual.

PDF preview unavailable. Download the PDF instead.

72d550f0-7e87-42b6-8d2c-455cfebb6fce Acrobat Distiller 10.1.1 (Windows)

Related Documents

Preview Merlin@home Transmitter User's Manual: EX1150, EX1150W
This user's manual provides comprehensive instructions for the St. Jude Medical Merlin@home Transmitter, models EX1150 and EX1150W. Learn about setup, operation, safety precautions, troubleshooting, and technical specifications for this remote monitoring system.
Preview Merlin Patient Care System Help Manual - St. Jude Medical
A comprehensive help manual for the St. Jude Medical Merlin Patient Care System (PCS), detailing its features, supported cardiac devices, operational procedures, error messages, and technical support information.
Preview Confirm Rx DM3500 Insertable Cardiac Monitor User's Manual
User's manual for the St. Jude Medical Confirm Rx DM3500 Insertable Cardiac Monitor, detailing its description, indications for use, contraindications, warnings, precautions, insertion, explantation, and technical specifications.
Preview Manuel des systèmes prêts pour l'IRM - St. Jude Medical
Ce manuel explique les procédures et précautions à suivre lors de l'examen IRM d'un patient porteur d'un système IRM sous conditions St. Jude Medical™. Il fournit des informations essentielles pour l'utilisation sécuritaire des systèmes St. Jude Medical™ dans un environnement IRM.
Preview MRI-Ready Systems Manual: St. Jude Medical MR Conditional Systems
Comprehensive guide for using St. Jude Medical MR Conditional Systems during MRI scans, detailing procedures, parameters, and safety precautions for cardiac physicians, clinicians, radiologists, and MRI technologists.
Preview Tendril STS Model 2088TC Pacing Leads User Manual
This user manual provides detailed information on the Tendril STS Model 2088TC pacing leads, including their description, intended use, contraindications, warnings, precautions, handling, implantation procedures, potential adverse effects, packaging, storage, sterilization, lead extraction, and technical specifications. It is designed for healthcare professionals involved in cardiac pacing.
Preview Confirm Rx™ Insertable Cardiac Monitor System Patient Manual
A patient manual for the St. Jude Medical Confirm Rx™ Insertable Cardiac Monitor System, including the DM3500 device and the myMerlin™ Mobile Application (Models APP1000, APP1001). This manual provides information on system operation, setup, precautions, and troubleshooting.
Preview St. Jude Medical Pulse Generator User's Manual
Comprehensive user's manual for St. Jude Medical pulse generators, including Assurity, Endurity, Zenex, Zenus, and Allure series. Details indications, contraindications, warnings, and technical specifications for cardiac rhythm management devices.