Agilent MassHunter Workstation Requirements Guide

Notices

Document Identification

Doc No D0026036
DE50653432
October 2022 Revision A.00

Copyright

Agilent Technologies, Inc. 2022

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051

Software Revision

This guide is valid for MassHunter Acquisition for LC/TQ 12.0, MassHunter Acquisition for LC/TOF and LC/Q-TOF 11.0, MassHunter Qualitative Analysis 10.0, and MassHunter Quantitative Analysis 12.0.

Software Manufacturing

Manufactured for Agilent Technologies
5301 Stevens Creek Blvd
Santa Clara, CA 95051

Trademark Acknowledgment

PCIe® is a US registered trademark and/or service mark of PCI-SIG.

Warranty

The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend

U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).

Safety Notices

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

In This Book

This document details the minimum computer hardware, software, and network requirements as well as the minimum instrument firmware revisions required to run an Agilent MassHunter Workstation or Agilent MassHunter Networked Workstation. Also included are the supported instrument types and Windows operating system configurations.

Table 1 Terms and abbreviations used in this document

TermDescription
Content ManagementDatabase to manage your analytical data. The database is provided as a component of OpenLab Server. Always used in Client/Server systems, optional for Workstations.
Control PanelControl Panel for OpenLab Shared Services
Microsoft Control PanelPart of the Microsoft Windows operating system

1 Hardware and Software Requirements

This chapter contains the hardware and software requirements for the different components of a MassHunter system.

Depending on the type of installation, you may need different hardware components. The following descriptions show the required components for each scenario.

Description of Figure 1. MassHunter Workstation: A diagram showing a MassHunter Workstation setup. It depicts a computer connected to a network (Customer LAN) and to analytical instruments (LC/MS system) via a switch. All required components are installed on the workstation.

Description of Figure 2. Networked Workstation system: A diagram illustrating a networked workstation system. It shows multiple networked workstations, each connected to analytical instruments, and all workstations connected to a central OpenLab Server or OpenLab ECM XT server via a network. The system includes both Networked Workstations and an OpenLab Server or OpenLab ECM XT server or ECM 3.x server.

Software

General Software Requirements

ComponentDetails
.NET framework
  • NET 3.5.1 must be enabled on systems running on Windows 8.1 or Windows 10 or Windows 11 and
  • NET 4.7.2 or above (if needed, it will be installed automatically by the MassHunter Installer)
Web browser
  • Google Chrome 40 or higher
  • Edge
Anti-virus software*
  • Microsoft Windows Defender

* The listed anti-virus software has been tested to be compatible with the MassHunter software described in this document. While other third-party AV solutions may also be compatible, they have not been tested, and compatibility cannot be guaranteed.

Windows Compatibility

Only these Windows operating systems are supported:

Computer Hardware

Disk Space

Disk space requirements depend based on the number and type of instruments, archival frequency, and the method settings chosen for Acquisition. Agilent recommends providing enough disk space for one year of lab operation, in addition to the operating system and MassHunter Workstation requirements.

The MassHunter Workstation is available either with storage in the local file system (MassHunter Workstation) or on a remote OpenLab Server or OpenLab ECM XT server with built-in Content Management database (MassHunter Networked Workstation).

Note that on Networked Workstations, data is only temporarily stored in a secured location on the local computer until it is transferred to the OpenLab Server/ECM XT server.

PC Recommendation

Table 2 provides the recommended hardware configuration for MassHunter computers with a single LC/TOF or LC/Q-TOF instrument.

Table 1. Tested and recommended hardware configuration for Workstations and Networked Workstations for TOF/Q-TOF
ItemFor all LC/Q-TOF except 6546For 6546 only
DescriptionStandard MassHunter-ready ComputerHigh Capacity MassHunter-ready Computer
Processor speed (CPU)Intel Xeon W-2123, 4 core, 3.6 GHzIntel Xeon W-2235, 6 core, 3.8 GHz
Physical memory (RAM)32 GB64 GB
Hard disk1 TB M.2 NVMe SSD - Primary (C:\) Boot. 4 TB x 2 RAID1 (4TB) - Data (D:\)1 TB M.2 NVMe SSD - Primary (C:\) Boot. 6 TB × 4 RAID10 (12 TB) - Data (D:\)
Graphic Resolution1920 x 10801920 x 1080
USB port*1 USB port required for installation1 USB port required for installation
LAN card - House
LAN card - instrument+
Integrated Intel I217LM PCIe GbE Controller
Integrated Intel I217LM PCIe GbE Controller
1 Integrated Intel I217LM PCIe GbE Controller
1 Intel Ethernet 210-T1 PCIe

* If a USB port is not available, the installation media can be copied over the network or downloaded from https://agilent.subscribenet.com.

+ A second LAN interface is required to isolate the instrument's data traffic from the local area network.

Existing MassHunter Workstations with the Agilent bundled Z4 G4 PC are supported with MassHunter Workstation 12.0 running in Workstation configuration only.

Table 2. Minimum hardware configuration for Workstations
ItemFor All TQ systems
DescriptionHewlett-Packard Z4 G4 Minitower
Processor speed (CPU)Intel Xeon W-2123 (3.6 GHz, 8.25 MB cache, 4 cores)
Physical memory (RAM)16GB (2x8GB) DDR4 2666 DIMM ECC Registered Memory
Hard disk2 x 500GB 7200 RPM SATA 6G Hard Drive (RAID 1)
Graphic Resolution1920 x 1080
USB port*1 USB port required for installation
LAN card+2 x Integrated Intel I219 and I210 PCIe GbE

* If a USB port is not available, the installation media can be copied over the network or downloaded from https://agilent.subscribenet.com.

† A second LAN interface is required to isolate the instrument's data traffic from the local area network.

Windows 10 and Windows 11 Configuration

When setting up a MassHunter computer for the first time, some actions and settings are required. The MassHunter installer configures Windows 10 with required and performance settings.

These instructions show Windows 10. The changes apply to Windows 11 as well.

Required Actions

These steps must be taken for a new Windows installation, before the software is installed.

  1. In Control Panel > System, activate Windows with Microsoft.
    Do this step only if the computer is customer-supplied. Computers from Agilent are already activated.
  2. In Windows Settings > Update and Security, click Check for updates to check for updates and apply all patches.
    Make sure all updates are downloaded and installed before you proceed.
    Make sure that no reboot is pending.

CAUTION

Windows Update must not have a pending installation during installation of MassHunter Workstation.

Required Settings

Windows Control Panel Options

Power Options

Description of Power Options Screenshot: Screenshot of Windows Power Options settings, showing 'High performance' plan selected, 'Turn off the display' set to 'Never', 'Put the computer to sleep' set to 'Never', and 'Turn off hard disk after' set to '0 minutes (never)' in advanced settings.

Administrative Tools

In Local Security Policy > Local Policies > Security Options:

Date and Time

Network and Sharing Center

In Change adapter settings > Local Area Connection > Properties > Configure > Power Management:

Description of Network Adapter Properties Screenshot: Screenshot of network adapter properties, specifically the 'Power Management' tab, with checkboxes 'Allow the computer to turn off this device to save power' and 'Allow this device to wake the computer' visible. The instruction is to clear the 'Allow the computer to turn off this device to save power' checkbox.

Programs and Features

In Turn Windows features on or off:

This option requires an Internet connection.

If this procedure does not work as expected, or the computer has no Internet access, install .NET 3.5 from the Windows installation media (see details for Windows 10 in https://support.microsoft.com/en-us/kb/2734782). If you do not have installation media, create them as described https://www.microsoft.com/en-us/software-download/windows10.

This setting makes sure that the net.tcp components are properly initialized, enable Non-HTTP activation

The system must be rebooted after Windows features are turned on.

Windows Settings

System > Tablet Mode

Description of Tablet Mode Screenshot: Screenshot of Windows Settings for 'Tablet mode' showing 'Never use tablet mode' selected.

Apps > Offline Maps

Description of Offline Maps Screenshot: Screenshot of Windows Settings for 'Offline maps' settings showing 'Metered connections' and 'Map updates' turned off.

Required Windows Explorer options

This Windows Explorer option is required.

View > Navigation pane

Performance Options

Control Panel

These Control Panel options enhance your system performance.

In Advanced > Performance > Visual Effects:

In Advanced > Performance > Advanced > Virtual Memory:

In Advanced > Performance > Advanced > Data Execution Prevention:

Windows Settings

Personalization > Colors

2 Network Requirements

This chapter describes the network requirements that must be met in order to support the environmental computing needs of a MassHunter system.

Introduction

MassHunter systems rely on network infrastructure in order to support the communication between Networked Workstations and the OpenLab Server/ECM XT server. This communication is based on standard TCP/IP protocols. In order to provide optimum performance and uptime, the network must meet design criteria for available bandwidth, IP address assignment, name resolution and appropriate isolation of the lab subnet from the corporate network.

Refer to the Agilent OpenLab Server and OpenLab ECM XT Hardware and Software Requirements Guide (p/n D0007295) for more details about networking requirements.

LAN Connectivity

All MassHunter Networked Workstations include at least two 1 GB Ethernet network interfaces. One is reserved for PC-to-Instrument communication. The other is used to communicate with the OpenLab Server / ECM XT / ECM 3.x.

Note these LAN Connectivity requirements:

¹ Network Interface Card (NIC) teaming is also known as Load Balancing and Failover (LBFO)

LAN Power Management

Avoid data capture or transfer interruptions in your data acquisition system by making LAN communication cards available for instrument and system component communications.

Windows may be set to turn instruments/components off to save power while sleeping or hibernating. To change this setting:

  1. In the Microsoft Control Panel, open the Network and Sharing Center.
  2. Select Change adapter settings. Right-click Local Area Connection, and then click Properties > Configure.
  3. Select the Power Management tab.
  4. Clear the Allow the computer to turn off this device to save power check box.

Depending on the model of network adapter, the name of this option can also be Energy efficient Ethernet, Power saving Ethernet mode, or a similar name.

CAUTION

While applying Windows Updates, LAN Power Management might become reactivated. Be sure to check the LAN Power Management settings after all Windows Updates.

Firewall Settings

If you are using a third party firewall or anti-virus software on the network where MassHunter is installed, open these firewall ports to allow communication between the system components of MassHunter. These apply to workstations as well as to Client/Server systems as component communications rely on these communication channels:

Table 3. Firewall ports
PortsComment, required for
6570SubscribeNet: active retrieval and release of product licenses
8084Agilent OpenLab Licensing support
8085-8089Alternative to port 8084 if that port is in use by another page or process
8090Hosts the viewing page of current license grants and consumptions found in the OpenLab Control Panel administration interface
27000-27009Communication of license availability
30101Store and retrieve instrument data
53971Activity log messages display notification

The MassHunter Workstation installer will automatically open required ports on an enabled Windows firewall during installation.

You will need to disable Windows Defender Firewall when you update the firmware.

Domain Requirements

Domains support the flow of information and user access rights across machines in the network. This means that all machines and instruments within the MassHunter Networked Workstation system must reside within the same domain or have the appropriate cross domain trusts to allow name based communications between all components in the system. In the case of a workstation installation, domains are only relevant if you are using a Windows domain-based authentication model. In this case the workstation or client must always be able to communicate with domain components in order to function as expected.

Installing MassHunter Workstation will apply network exceptions to the Windows firewall under the domain profile to result in a functional system. The components necessary to support MassHunter on a domain are:

The domain name server (DNS) must be able to resolve the IPv4 address of all instrument controllers and instruments. Any unresolved instrument controller or instrument will disrupt the functionality of MassHunter resulting in errors or delays. IPv6 is not supported and must be deactivated.

MassHunter Workstation components must not be installed on the same machine as the domain controller.

The domain components above host a variety of services and settings that must be configured appropriately to allow communication across machines. The following services and settings will need to be configured to fit your domain. Your internal IT group is responsible for proper configuration of any custom domain solutions. These include settings for:

Environments with Proxy Servers

The OpenLab server must be accessible via http or https in the network. If you use proxy servers, verify that they can be accessed. If required, adjust the proxy settings.

Network Isolation

MassHunter Networked Workstations must be isolated from network environments that experience frequent failures due to faulty switching, viruses, or worms. If network isolation is not possible, the machines should be reconfigured and disconnected from the problematic network until these issues can be resolved. On an isolated network, name resolution services must be hosted by a separate machine to enable proper communications between system components by name.

An isolated network is completely physically isolated, so that no LAN switch connections on the network are shared with the corporate network infrastructure. Description of Figure 3. Sample client/server topology: Network Isolation: A diagram depicting a physically isolated client/server network topology. It shows multiple client workstations, each connected to analytical instruments, and all clients connected to a central server, with no shared LAN switch connections to a corporate network.

Specific Requirements for Compliant Systems

If you intend to use your system in a compliant environment, ensure the following settings related to time synchronization:

3 Instrument Connections

This chapter provides information on the instruments supported by the current revision of OpenLab CDS and the required respective instrument drivers and firmware revisions.

RC.NET Drivers and OpenLab CDS

MassHunter can control instruments and modules that use RC.NET based driver software only.

Agilent and other vendors may release RC.net drivers independent of the MassHunter releases. Agilent recommends always using the most recent firmware revisions which include latest firmware features and improvements. Agilent driver software is forward compatible with respect to firmware, i.e. the firmware can be updated without the need of updating the driver or CDS.

More information on instrument drivers and firmware is available in the respective RC.net driver release notes.

The following Agilent instrument driver software revisions are part of the MassHunter Acquisition software package, and are installed by default with the software:

Table 4. Agilent driver packages shipped and installed with MassHunter 12.0
RC.net Instrument DriverDriver Software Revision
Agilent LC3.5
CTC PAL XTB.01.08
Agilent TQ3.0
Agilent TOF/Q-TOFNot an RC.net driver

Additional RC.NET instrument drivers are supported with MassHunter 12.0 and will need to be installed separately.

Third Party Drivers

MassHunter Acquisition for 6400 Series Triple Quadrupole LC/MS or MassHunter Acquisition for Ultivo LC/TQ does not support third party instruments.

Agilent LC

Most Agilent LC Modules can be controlled with the current version of MassHunter Workstation. LC driver 3.5 has been tested with this revision and is installed by default with the software.

Recommended Firmware

Agilent recommends using current LC firmware sets with your MassHunter Workstation. The latest LC firmware sets are 7.xx. See Table 5.

Firmware Set 7.xx can be downloaded from http://www.agilent.com/en-us/firmwareDownload?whid=99818.

LC and CE firmware revisions are grouped into sets for each module or system. Firmware sets include just the latest firmware of each module.

Modules combined into one LC instrument always need to have firmware from the same set. Firmware of one set is fully compatible with your CDS that supports this firmware set. However, if one module uses a firmware revision lower that the specified Minimum Firmware Revision some functionality might not be supported.

Do not mix firmware revisions from one set with older or newer sets. Firmware is not tested across set borders!

Table 5. Recommended firmware
DeviceRecommended Firmware
Agilent 1100 Series, 1200 Series and 1200 InfinityB.07.30 or later
Agilent 1200 Series, 1200 Infinity and 1120 Compact LCB.07.30 or later
Agilent 1200 Infinity Hosted ModulesC.07.20 or later
Agilent 1260/1290 Infinity II ModulesD.07.33 or later

You need to upgrade all existing modules to the latest version only if

For more information on downloading the current LC firmware, please refer to the Agilent website at http://www.agilent.com/en-us/firmwareDownload?whid=69761 (LC Firmware News and Downloads) and http://www.agilent.com/en-us/firmwareDownload?whid=83974 (Firmware Update Tools & Procedures).

The tables that follow list the supported LC modules with the minimum required firmware. Please note that using some driver features like valve-thermostat clusters and new temperature control modes require firmware updates beyond the minimum required firmware for related modules. For more information please refer to the respective release notes.

Table 6. 1120 Compact LC / 1220 Infinity LC
1120 Compact LC / 1220 Infinity LCVersion
G4286A Compact LCB.07.30 [005]
G4286B Infinity LCB.07.30 [005]
G4286C Infinity LCB.07.30 [005]
G4287A Compact LCB.07.30 [005]
G4287B Infinity LCB.07.30 [005]
G4288A Compact LCB.07.30 [005]
G4288B Infinity LCB.07.30 [005]
G4288C Infinity LC VLB.07.30 [005]
G4289A Compact LCB.07.30 [005]
G4289B Infinity LCB.07.30 [005]
G4289C Infinity LC VLB.07.30 [005]
G4290A Compact LCB.07.30 [005]
G4290B Infinity LCB.07.30 [005]
G4290C Infinity LC VLB.07.30 [005]
G4291A Compact LCB.07.30 [005]
G4291B Infinity LCB.07.30 [005]
G4291C Infinity LC VLB.07.30 [005]
G4292A Compact LCB.07.30 [005]
G4292B Infinity LCB.07.30 [005]
G4292C Infinity LC VLB.07.30 [005]
G4293A Compact LCB.07.30 [005]
G4293B Infinity LCB.07.30 [005]
G4293C Infinity LC VLB.07.30 [005]
G4294B Infinity LC VLB.07.30 [005]
G4294C Infinity LC VLB.07.30 [005]
Table 7. 1260 Infinity LC
1260 Infinity LCVersion
G1156A 6-Position/7-Ports ValveA.07.01 [001]
G1157A 2-Position/10-Ports ValveA.07.01 [001]
G1158A 2-Position/6-Ports Valve (1100/1200)A.07.01 [001]
G1158B 2-Position/6-Ports ValveA.07.01 [001]
G1159A 6-Position Column Selector ValveA.07.01 [001]
G1160A 12-Position Selector ValveA.07.01 [001]
G1162A Micro Valve 2-Position/6-PortsA.07.01 [001]
G1163A Micro Valve 2-Position/10-PortsA.07.01 [001]
G1310A Isocratic Pump (1100/1200)A.07.01 [001]
G1310B Isocratic PumpA.07.01 [001]
G1311A Quaternary Pump (1100/1200)A.07.01 [001]
G1311B Quaternary PumpA.07.01 [001]
G1311C Quaternary Pump VLA.07.01 [001]
G1312A Binary Pump (1100/1200)A.07.01 [001]
G1312B Binary PumpA.07.01 [001]
G1312C Binary Pump VLA.07.01 [001]
G4302A SFC-Binary PumpA.07.01 [001]
G1313A Autosampler (1100)A.07.02 [005]
G1314A Variable Wavelength Detector (1100/1200)A.07.02 [001]
G1314B Variable Wavelength Detector VLA.07.02 [001]
G1314C Variable Wavelength Detector VL+A.07.02 [001]
G1314D Variable Wavelength Detector (1200)B.07.33 [003]
G1314F Variable Wavelength DetectorB.07.33 [003]
G1315A Diode Array Detector (1100/1200)A.07.02 [001]
G1315B Diode Array Detector (1100/1200)A.07.02 [001]
G1315C Diode Array Detector VL+B.07.33 [003]
G1315D Diode Array Detector VLB.07.33 [003]
G1316A Thermostatted Column CompartmentA.07.02 [001]
G1316B Thermostatted Column Compartment SL (1200)A.07.02 [001]
G1321A Fluorescence Detector (1100/1200)A.07.02 [001]
G1321B Fluorescence Detector SPECTRAA.07.02 [001]
G1321C Fluorescence DetectorA.07.02 [001]
G1329A Standard Autosampler (1100/1200)A.07.02 [005]
G1329B Standard AutosamplerA.07.02 [005]
G1365A Multiple Wavelength Detector (1100/1200)A.07.02 [001]
G1365B Multiple Wavelength Detector (1100/1200)A.07.02 [001]
G1365C Multiple Wavelength DetectorB.07.33 [003]
G1365D Multiple Wavelength Detector VLB.07.33 [003]
G1367A Well Plate Sampler (1100)A.07.02 [005]
G1367B High Performance Autosampler (1200)A.07.02 [005]
G1367C High Performance Autosampler SL (1200)A.07.02 [005]
G1367D High Performance Autosampler SL+ (1200)A.07.02 [005]
G1367E High Performance AutosamplerA.07.02 [005]
G1369C LAN Interface CardB.07.33 [003]
G1376A Capillary PumpA.07.01 [001]
G1377A Micro Well Plate SamplerA.07.02 [005]
Table 8. 1290 Infinity LC
1290 Infinity LCVersion
G1170A Universal Valve Drive*C.07.30 [001]
G1314E Variable Wavelength DetectorB.07.33 [003]
G1316C Thermostatted Column Compartment+A.07.02 [001]
G4204A Quaternary PumpB.07.33 [003]
G4220A Binary PumpB.07.33 [003]
G4212A Diode Array DetectorB.07.33 [003]
G4226A AutosamplerA.07.02 [005]
G4227A Flexible CubeC.07.30 [001]

* A host may be required. For details see latest driver release note.

+ Cluster with up to three G1316C with integrated 8pos/9port valves (products G4230A/B). Minimum two G1316C TCCs, the third TCC can be a G1316A, B or C.

Table 9. 1260/1290 Infinity II LC
1260/1290 Infinity II LCVersion
G4782A Binary Pump SFCD.07.33 [003]
G4767A Multisampler SFCD.07.33 [003]
G5654A Bio-inert Quaternary PumpD.07.33 [003]
G5668A Bio-inert MultisamplerD.07.33 [003]
G7104A Quaternary PumpB.07.33 [003]
G7104C Flexible PumpB.07.33 [003]
G7110B Isocratic PumpD.07.33 [003]
G7111A Quaternary Pump VLD.07.33 [003]
G7111B Quaternary PumpD.07.33 [003]
G7112B Binary PumpD.07.33 [003]
G7114A Variable Wavelength DetectorD.07.33 [003]
G7114B Variable Wavelength DetectorD.07.33 [003]
G7115A Diode Array DetectorD.07.33 [003]
G7116A Multicolumn ThermostatC.07.30 [001]
G7116B Multicolumn ThermostatC.07.30 [001]
G7117A Diode Array Detector FSD.07.33 [003]
G7117B Diode Array DetectorD.07.33 [003]
G7117C Diode Array Detector HSD.07.33 [003]
G7120A Binary PumpB.07.33 [003]
G7121A Fluorescence DetectorD.07.33 [003]
G7121B Fluorescence Detector SPECTRAD.07.33 [003]
G7129A VialsamplerD.07.33 [003]
G7129B VialsamplerD.07.33 [003]
G7129C VialsamplerD.07.33 [003]
G7132A Bio Binary PumpD.07.33 [003]
G7137A Bio MultisamplerD.07.33 [003]
G7165A Multiple Wavelength DetectorD.07.33 [003]
G7167A MultisamplerD.07.33 [003]
G7167B MultisamplerD.07.33 [003]
Table 10. Capillary Electrophoresis (CE)
ModuleVersion
G7100A Capillary Electrophoresis (CE)B.07.30 [001]
Table 11. Other modules
Product NumberModule NameMinimum Firmware Revision
G1330A1100 Series Thermostatn/a
G4277A1290 Infinity LC Injector HTS4.3.0
G1330B1290 Infinity Thermostatn/a
G4270-CTCHTC PAL Auto samplerAgilent 2 6.8 or 4.1.0
G4271-CTCHTS PAL Auto samplerAgilent 2 6.8 or 4.1.5
G7130AIntegrated Column Compartment ICCD.06.76
VTC Valve Thermostat ClusterCombinations of G7116B, G1170A and G1316C (valve or column hosts) and G1316A/B and G7130ASee LC Driver Release Notes: Valve-Thermostat Cluster
HDR-DAD Cluster2x G4212A, 2x G4212B, 2x G7117A or 2x G7117B, or G7117A and 1x G7117BB.06.57

Agilent LC/MS

Recommended Firmware

Always use the most recent Firmware or Embedded Software installation package that comes with the driver package.

Triple Quad LC/MS System

These Agilent LC/MS instruments can be controlled with MassHunter Workstation.

Table 12. 6400 Series Triple Quad LC/MS System
Product numberDescriptionSupported sourceAutotune sourceChecktune source
G6420A6420 LC/TQESIESIESI
G6460C6460C LC/TQAJS, ESIAJS, ESIAJS, ESI
G6470A6470 LC/TQAJS, ESI, APCI, MMI, APPI, ChipCubeAJS, ESI, APCI, MMIAJS, ESI, APCI (6470 A/B only), MMI
G6470B6470 LC/TQAJS, ESI, APCI, MMI, APPI, ChipCubeAJS, ESI, APCI, MMIAJS, ESI, APCI (6470 A/B only), MMI
G6475A6475 LC/TQAJS, ESI, APCI, MMI, Nano ESIAJS, ESI, APCI, MMIAJS, ESI, APCI, MMI
G6495A6495A LC/TQAJS, ESI, APCI, MMI, APPI, ChipCubeAJS, ESI, APCI, MMIAJS, ESI, APCI (6495C only), MMI
G6495B6495B LC/TQAJS, ESI, APCI, MMI, APPI, ChipCubeAJS, ESI, APCI, MMIAJS, ESI, APCI (6495C only), MMI
G6495C6495C LC/TQAJS, ESI, APCI, MMI, APPI, ChipCubeAJS, ESI, APCI, MMIAJS, ESI, APCI (6495C only), MMI
Table 13. 6400 Series Triple Quad LC/MS System
Product numberDescriptionSupported sourceAutotune sourceChecktune source
G6465AUltivo LC/TQAJS, ESI, APCI, MMIAJS, ESI, APCI, MMIAJS, ESI, APCI, MMI
G6465BUltivo LC/TQAJS, ESI, APCI, MMIAJS, ESI, APCI, MMIAJS, ESI, APCI, MMI

LC/TOF and LC/Q-TOF System

These Agilent LC/TOF and LC/Q-TOF instruments can be controlled with MassHunter Workstation.

Table 14. Agilent LC/TOF Systems
Product numberDescriptionNote
G6224A6224A Accurate-Mass TOF LC/MSNo support for AJS or Dual AJS Source
G6230A6230A Accurate-Mass TOF LC/MS
G6230B6230B Accurate-Mass TOF LC/MS
Table 15. Agilent LC/Q-TOF Systems
Product numberDescriptionNote
G6520B6520B Accurate-Mass Q-TOF LC/MSNo support for AJS or Dual AJS Source
G6530A6530A Accurate-Mass Q-TOF LC/MS
G6530B6530B Accurate-Mass Q-TOF LC/MS
G6530C6530C Accurate-Mass Q-TOF LC/MS
G6538A6538 Ultra High Definition (UHD) Accurate-Mass Q-TOF LC/MSNo support for AJS or Dual AJS Source
G6540A6540 Ultra High Definition (UHD) Accurate-Mass Q-TOF LC/MS
G6540B6540 Ultra High Definition (UHD) Accurate-Mass Q-TOF LC/MS
G6545A6545A LC/Q-TOF
G6545B6545B LC/Q-TOF
G6546A6546 LC/Q-TOF
G6549A6545XT AdvanceBio LC/Q-TOF
G6550A6550 iFunnel Q-TOF LC/MS System
G6550B6550B iFunnel Q-TOF LC/MS System
G6560A6560 Ion Mobility LC/Q-TOF
G6560B6560 Ion Mobility LC/Q-TOF

Drivers for non-Agilent Instruments

If any drivers are available for non-Agilent instruments, they will be available from SubscribeNet

Incompatible Instruments and Modules

The following tables summarize selected instruments or modules that can not be controlled with the current revision of MassHunter Workstation.

Incompatible LC and LC/MS Modules

Table 16. Incompatible LC and LC/MS Modules
Product NumberModule Name
G1361A1260 Infinity Preparative Pump
G1364A/B/C/DFraction Collector / Fraction Cluster
G1364EFraction Collector (Prep)
G1364FFraction Collector (Analytical)
G1389AMicro Sampler
G2258A1260 Infinity Dual-Loop Autosampler
G2260A1260 Infinity Preparative Autosampler (High flow)
G4218A, G7102A, G4260A, G4260B, G4261A, G4261BEvaporative Light Scattering Detector (ELSD)
G5664A1260 Infinity Bio-inert fraction collector AS
G5664BFraction Collector (BIO)
G7157A1260 Infinity II Preparative Autosampler
G7158BPreparative Open Bed Sampler/Collector
G7159BPreparative Open Bed Fraction Collector
G7161A1260 Infinity II Prep Binary Pump
G7161BBinary Prep Pump
G7162ARefractive Index Detector
G7162BRefractive Index Detector Micro
G7166AValve-based Fraction Collector
G7170BMS Flow Modulator

Incompatible CE Instruments

MassHunter Workstation cannot be used to control CE Instruments.

4 Software Compatibility

This chapter contains information on compatibility with other Agilent or Non-Agilent Software.

Software

The MassHunter Workstation 12.0 software includes MassHunter Acquisition for LC/TQ 12.0, MassHunter Quantitative Analysis 12.0, and MassHunter Qualitative Analysis 10.0.

These programs are compatible with MassHunter Workstation 12.0:

¹ In a MOBILion MOBIE system, MOBIE EyeOn Software is used to acquire data in HRIM mode. MassHunter Data Acquisition is used to acquire data in Q-TOF mode.

Libraries and Databases

The following libraries and databases are compatible with MassHunter Workstation.

PDF preview unavailable. Download the PDF instead.

MHRequirements EN , Inc Acrobat Distiller 22.0 (Windows)

Related Documents

Preview Agilent MassHunter Workstation Software Quantitative Analysis Familiarization Guide
A guide to help users learn the Agilent MassHunter Workstation Software Quantitative Analysis program through step-by-step exercises. Covers setting up batches, methods, target compounds, quantitation, and reviewing results for MRM and Q-TOF data.
Preview Agilent MassHunter GC/MS Hardware User Manuals and Tools
A comprehensive list of user manuals, quick start guides, and troubleshooting resources for Agilent's GC/MS systems, including the 5977B GC/MSD, 5975/77 Series MSD for OpenLAB, 7000D/7010B Series Triple Quad MS, and 7200 Accurate-Mass Q-TOF GC/MS systems, all designed to work with MassHunter software.
Preview Agilent InfinityLab LC Supplies Catalog
Explore the comprehensive Agilent InfinityLab LC Supplies catalog, featuring a wide range of instruments, columns, and accessories designed to optimize your liquid chromatography workflows for maximum efficiency and performance.
Preview Quantitative Screening of Multiresidue Veterinary Drugs in Milk and Egg Using Agilent 6495C Triple Quadrupole LC/MS
This application note details a workflow for the quantitative screening of 210 veterinary drug residues in milk and egg matrices using the Agilent Comprehensive Veterinary Drug dMRM Solution and the Agilent 6495C Triple Quadrupole LC/MS system. It covers sample preparation, chromatographic separation, MS detection, and data processing, demonstrating high sensitivity, accuracy, and repeatability.
Preview Agilent InfinityLab LC Series Refractive Index Detectors User Manual
User manual for the Agilent InfinityLab LC Series Refractive Index Detectors, including the 1260 Infinity III (G7162A) and 1290 Infinity III (G7162B) models. This guide covers installation, operation, maintenance, and troubleshooting.
Preview Agilent 1260 Infinity II Binary LC System: Manual and Quick Guide
This manual and quick guide provides comprehensive information on the Agilent 1260 Infinity II Binary LC System, covering product description, features, system components, configuration settings, quick start procedures, parts, consumables, and safety information.
Preview Agilent MassHunter Quantitative Data Analysis: A Comprehensive Guide
Explore the capabilities of Agilent MassHunter Quantitative Data Analysis software, covering batch table navigation, compound information setup, calibration curves, and global settings. This guide provides insights into method editor views and compound analysis.
Preview Agilent OpenLab CDS Configuration Guide: Integration with OpenLab ECM
This guide from Agilent Technologies provides essential guidance for installing and configuring the OpenLab Shared Services Server. It details the integration of OpenLab CDS 2.6 with OpenLab ECM 3.x environments.