1290 Infinity LC/1260 Infinity LC Site Preparation Checklist

Introduction

Thank you for purchasing an Agilent instrument. To ensure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide and checklist outlining the supplies, consumables, space, and utility requirements for your equipment.

For additional information about Agilent solutions, please visit our website at www.chem.agilent.com.

Customer Responsibilities

Ensure your site meets the following prior to the installation date, using the checklist below:

If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services to receive important operational, maintenance, and safety information.

Important Customer Information

  1. If you have questions or problems regarding the 'Customer Responsibilities', please contact your local Agilent or partner support/service organization for assistance prior to delivery. Agilent and/or its partners reserve the right to reschedule the installation based on the readiness of your laboratory.
  2. Should your site not be ready for any reason, please contact Agilent as soon as possible to rearrange any purchased services.
  3. Other optional services, such as additional training, operational qualification (OQ), and consultation for user-specific applications, may be provided at the time of installation when ordered with the system, but should be contracted separately.

Module List

The following table lists the available modules for the 1290 Infinity LC and 1260 Infinity LC systems:

Module Instrument Description
G1170A1290 Infinity Valve Drive and Valve Heads
G1310B1260 Infinity Isocratic Pump
G1311B1260 Infinity Quatenary Pump
G1311C1260 Infinity Quaternary Pump VL
G1312B1260 Infinity Binary Pump
G1312C1260 Infinity Binary Pump VL
G1314B1260 Infinity Variable Wavelength Detector VL
G1314C1260 Infinity Variable Wavelength Detector VL+
G1314E1290 Infinity Variable Wavelength Detector
G1314F1260 Infinity Variable Wavelength Detector
G1315C1260 Infinity Diode Array Detector VL+
G1315D1260 Infinity Diode Array Detector VL
G1316A1260 Infinity Thermostatted Column Compartment
G1316C1290 Infinity Thermostatted Column Compartment
G1321B1260 Infinity Fluorescence Detector
G1321C1260 Infinity Fluorescence Detector
G1322A1260 Infinity Standard Degasser
G1329B1260 Infinity Standard Autosampler
G1330B1290 Infinity Thermostat
G1361A1260 Infinity Preparative Pump
G1362A1260 Infinity Refractive Index Detector
G1364B1260 Infinity Preperative-scale Fraction Collector
G1364C1260 Infinity Analytical-scale Fraction Collector
G1364D1260 Infinity Micro-scale Fraction Collector
G1365C1260 Infinity Multiple Wavelength Detector
G1365D1260 Infinity Multiple Wavelength Detector VL
G1367E1260 Infinity High Performance Autosampler
G1376A1260 Infinity Capillary Pump
G1377A1260 Infinity High Performance Micro Autosampler
G1379B1260 Infinity Micro Degasser
G1390B1200 Infinity Series Universal Interface Box
G1968FActive Splitter
G2226A1260 Infinity Nanoflow Pump
G2257A1260 Infinity Sample Capacity Extension
G2258A1260 Infinity Dual-Loop Autosampler
G2260A1260 Infinity Preparative Autosampler
G4204A1290 Infinity Quaternary Pump
G4208A1200 Infinity Series Instant Pilot
G4212A1290 Infinity Diode Array Detector
G4212B1260 Infinity Diode Array Detector
G4220A1290 Infinity Binary Pump
G4220B1290 Infinity Binary Pump
G4225A1260 Infinity High Performance Degasser Module
G4226A1290 Infinity Autosampler
G4227A1290 Infinity Flexible Cube
G4240A1260 Infinity Chip Cube MS Interface
G4260AAgilent 380-ELSD
G4260B1260 Infinity Evaporative Light Scattering Detector
G4261AAgilent 385-ELSD
G4261B1290 Infinity Evaporative Light Scattering Detector
G4277A1290 Infinity LC Injector HTS
G4278A1290 Infinity LC Injector HTC
G5611A1260 Infinity Bio-inert Quaternary Pump
G5664A1260 Infinity Bio-inert Analytical-scale Fraction Collector
G5667A1260 Infinity Bio-inert High-Performance Autosampler

Dimensions and Weight

Identify the laboratory bench space required for your system based on the table below. Pay special attention to the total height and total weight requirements for all system components you have ordered. Avoid bench space with overhanging shelves. Also, pay special attention to the total weight of the modules you have ordered to ensure your laboratory bench can support this weight.

Special Notes

The thermostatted version of all samplers and the fraction collector includes the G1330B thermostat module. The thermostat module must be placed directly under the sampler or the fraction collector to be thermostatted. It is recommended that the thermostat module is positioned as the bottom module of the stack, directly on the laboratory bench. Any stack containing a G1330B thermostat module needs at least 25 cm (10 inches) of space on either side to guarantee proper ventilation.

Instrument Description Weight (kg/lbs) Height (cm/in) Depth (cm/in) Width (cm/in)
G1322A, G1379B7.5 / 16.58 / 343.5 / 1734.5 / 13.5
G4225A5 / 118 / 3.143.5 / 1734.5 / 13.5
G1310B11 / 2518 / 743.5 / 1734.5 / 13.5
G1311B/C, G5611A14.5 / 3218 / 743.5 / 1734.5 / 13.5
G1312B/C15.5 / 3418 / 743.5 / 1734.5 / 13.5
G1376A, G2226A17 / 3918 / 743.5 / 1734.5 / 13.5
G1361A15 / 32.920 / 843.5 / 1734.5 / 13.5
G4220A/B21.8 / 4824 / 9.343.5 / 1734.5 / 13.5
G4204A15.2 / 33.420 / 843.5 / 1734.5 / 13.5
G2258A14 / 29.820 / 843.5 / 1734.5 / 13.5
G1329B, G2260A14.2 / 31.320 / 843.5 / 134.5 / 13.5
G1367E, G1377A, G4226A, G5667A15.5 / 34.220 / 843.5 / 1734.5 / 13.5
G4277A10 / 22.864.8 / 25.538.5 / 15.182.8 / 32.6
G4278A8 / 18.264.8 / 25.538.5 / 15.154.4 / 21.5
G1364B/C/D, G5664A17 / 3818 / 743.5 / 1734.5 / 13.5
G4240A14 / 3136 / 14.130 / 11.735 / 13.7
G1330B18.5 / 40.714.4 / 5.543.5 / 1734.5 / 13.5
G4227A8.2 / 18.114.4 / 5.543.5 / 1734.5 / 13.5
G1316A/C11.2 / 22.514 / 5.543.5 / 1741 / 16
G1314B/C/E/F11 / 2514 / 5.543.5 / 1734.5 / 13.5
G1315C/D, G1365C/D11.5 / 2614 / 5.543.5 / 1734.5 / 13.5
G1321B/C11.5 / 25.414 / 5.543.5 / 1734.5 / 13.5
G4212A/B11.5 / 2614 / 5.543.5 / 1734.5 / 13.5
G1362A17 / 3818 / 743.5 / 1734.5 / 13.5
G4260A/B11 / 24.241.5 / 16.345 / 17.720 / 7.9
G4261A/B13 / 28.641.5 / 16.345 / 17.720 / 7.9
G1390B0.9 / 216.5 / 6.55.5 / 2.213.5 / 5.3
G1968F2.1 / 4.614.1 / 5.619.1 / 7.57.8 / 3.1

Environmental Conditions

Operating your instrument within the recommended temperature ranges ensures optimum instrument performance and lifetime.

Special Notes

  1. Performance can be affected by sources of heat and cold, e.g., direct sunlight, heating/cooling from air conditioning outlets, drafts, and/or vibrations.
  2. The site's ambient temperature conditions must be stable for optimum performance.
  3. The following table summarizes some key physical specifications. For the complete set of physical specifications, please refer to the corresponding module manual.
Instrument Description Operating temp range °C (°F) Operating humidity range (%)
G1314B/C/E/F, G1315C/D, G1316A/C, G1322A, G1365C/D, G1362A, G4225A0-55 °C (32-131 °F), constant temperature.< 95%, non-condensing
G1321B/C0-40 °C (32-104 °F), constant temperature.< 95%, non-condensing
G4212A/B4-40 °C (39-104 °F), constant temperature.< 80%, non-condensing
G1379B0-45 °C (32-113 °F)< 95%, non-condensing
G1330B, G1361A, G1364B/C/D, G5664A, G2258A, G4212B4-40 °C (39-104 °F)< 95%, non-condensing
G4260A/B, G4261A/B10-35 °C (50 – 95 °F)< 95%, non-condensing
G4240A5-40 °C (41-104 °F)< 80%, non-condensing
G4277A, G4278A4-40 °C (39-104 °F)< 75%, non-condensing
G1968F0-40 °C (32-104 °F), constant temperature80% r.h. up to 31 °C, decreasing to 50% at 40 °C, non-condensing
All other modules4-55 °C (39-131 °F), constant temperature< 95% r.h. at 40 °C, non-condensing

Power Consumption

Special Notes:

  1. If a computer system is supplied with your instrument, be sure to account for those electrical outlets.
  2. The heat dissipation can be calculated from the active power, using the following equation: 1 W = 3.413 BTU/h.
Instrument Description Line Voltage & Frequency (V, Hz) Maximum Power Consumption (VA) Maximum Power Consumption (W)
G1379B, G1322A, G4225A100-240 V(AC), 50 or 60 Hz30 VA30 W
G1310B100-240 V(AC), 50 or 60 Hz180 VA55 W
G1311B/C, G5611A100-240 V(AC), 50 or 60 Hz180 VA110 W
G1312B/C100-240 V(AC), 50 or 60 Hz220 VA74 W
G1376A, G2226A100-240 V(AC), 50 or 60 Hz220 VA75 W
G1361A100-240 V(AC), 50 or 60 Hz250 VA210 W
G4220A/B100 – 240 V(AC), 50 or 60 Hz350 VA270 W
G4204A100 – 240 V(AC), 50 or 60 Hz220 VA180 W
G1329B, G2260A, G1367E, G1377A, G4226A, G5667A100-240 V(AC), 50 or 60 Hz300 VA200 W
G2258A100-240 V(AC), 50 or 60 Hz260 VA210 W
G4277A, G4278A100-240 V(AC), 50 – 60 HzN/A120 W
G1364D100-240 V(AC), 50 or 60 Hz200 VA180 W
G5664A100-240 V(AC), 50 or 60 Hz180 VA180 W
G4240A100 – 240 V(AC), 50 or 60 HzN/A80 W
G1330B100-240 V(AC), 50 or 60 Hz260 VA210 W
G4227A100-240 V(AC), 50 or 60 Hz80 VA45 W
G1316A/C100-240 V(AC), 50 or 60 Hz320 VA150 W
G1314B/C/E/F100-240 V(AC), 50 or 60 Hz220 VA85 W
G1315C/D, G1365C/D100 – 240 V(AC), 50 or 60 Hz160 VA160 W
G4212A/B100-240 V(AC), 50 or 60 Hz160 VA130 W
G1362A100-240 V(AC), 50 or 60 Hz160 VA65 W
G1321B/C100-240 V(AC), 50 or 60 Hz180 VA70 W
G4260A/B, G4261A/B100-120 V(AC) or 220 – 240 V, 50 or 60 HzN/A150 W
G1390B100-240 V(AC), 50 or 60 Hz140 VA65 W
G1968F100-240 V(AC), 50 or 60 Hz93-111 VAN/A

Required Operating Supplies by Customer

Special Notes:

For information on Agilent consumables, accessories, and laboratory operating supplies, please visit www.chem.agilent.com/en-US/Products/consumables/Pages/default.aspx.

Other/Special Requirements

G4260B, G4261B

Gas requirements

A supply of inert gas (typically nitrogen) is required to operate the detector. The gas supply needs to be free of oil, humidity, and particles, as such contaminations can create background noise in the chromatograms and may damage the built-in pressure sensor. In case of such noise, for example for newly installed gas lines, flush the gas lines for a sufficient time (might take days) and use additional filters of 0.5 µm or less. The typical gas pressure is 4 bar (60 psi) and must be set by an external pressure regulator. Pure gas is not required as the gas is only used as a carrier for the solid sample particles. The gas inlets of the detector have an outer diameter of 4 mm (0.157 inches). The lab installation must therefore allow the installation of tubing with 4 mm (0.157 inches) outer diameter. Gas consumption is typically 0.9 SLM to 3.25 SLM, depending on the detector settings.

Item description, (including dimensions etc) Vendor/Part Number (if applicable) Recommended quantity
G4260A/B, G4261A/B ELSD Gas Nitrogen (typical)N/AN/A

Solvent requirements

Customers should have available HPLC grade Acetonitrile and water with a dry residue below 1 ppm or MS grade solvents.

Exhaust venting and drain requirements

The exhaust from the detector must be directed into a fume hood or exhaust vent. If a vacuum is used, it should be moderate so as to avoid turbulence in the glass cell siphon. The potentially hazardous exhaust of evaporated solvent and sample must not be allowed to enter the laboratory atmosphere, and any appropriate accessory like solvent filters should be disposed of according to local environmental requirements.

If the extraction tube provided with the instrument is to be extended, it is recommended that the diameter of the extension is increased to at least 50 mm (2 in) diameter tubing so the extraction quality is not inhibited.

NOTE: Do not connect the exhaust vent directly to the detector. This might cause either positive pressure or negative back pressure, both of which will impact the quality of your measurement results.

The drain tube must be directed to a waste container supplied with the instrument. The user is responsible for decontamination or recycling of any residue, regarding local environmental requirements.

Further requirements

The 1290 Infinity ELSD (G4261B) can be controlled either via RS232 or via LAN. If the RS232 interface is used for control, the ELSD must be installed close to the control PC unless special data transmission systems are used. The length of the straight female/female RS 232 cable supplied with both detectors is 2.9 m.

G4240A 1260 Chip LC

Gas Requirements

For HPLC Chip Cube installation, high purity air is required. The high purity air is used to reduce background ions in the spray chamber and to enhance spray stability. More details regarding the use of the Background Reduction Kit can be found in the installation manual for that kit.

Please see Table 1 on page 12 for a summary of the air for Chip Cube requirements.

Table 1 Air for Chip Cube requirements

Purity Gas Pressure Flow
High pressure air source¹ for 6520A/B, 6530, 6538, 654099.99% pure or better and hydrocarbon free²5.5–6.9 bar² (80–120 psi)Up to 4 L/min (240 L/h)
High pressure air source¹ for 655099.99% pure or better and hydrocarbon free²5.5–6.9 bar² (80–120 psi)Up to 10 L/min (600 L/h)

¹ Air source must be able to deliver air at a constant pressure.

² Inlet gas pressure for air must be slightly higher than the pressure used for the nitrogen gas supply pressure.

Exhaust Venting

Please refer to the Site Preparation Guide of your Agilent mass spectrometer for detailed specifications of the spray chamber exhaust vent.

Interface

Please plan for 15 cm (6 inch) free space between the LC stack and the HPLC-Chip Cube Interface. It is required for lifting the Chip Cube off the mass spectrometer. The LC and the mass spectrometer should be installed on the same table. This allows to remove the Chip Cube quickly without disconnecting capillaries for routine maintenance tasks.

Module Stacking

Figure 1 Recommended single stack configuration for HPLC-Chip MS

Description: A diagram showing a single stack configuration of HPLC-Chip MS modules, including Solvent cabinet, Degasser, Nanoflow pump, Micro autosampler, Cooler, and Capillary pump.

Figure 2 Recommended 2-stack configuration for HPLC-Chip MS

Description: A diagram showing a two-stack configuration of HPLC-Chip MS modules, including Solvent cabinet, Degasser, Nanoflow pump, Micro autosampler, Cooler, and Capillary pump.

G4277A 1290 HTS, G4278A 1290 HTC

Module Stacking

The G4277A and G4278A have to be installed on free bench space next to the 1290 Infinity LC stack.

To avoid excessive delay volumes from the pump to the injection valve and from the injection valve to the LC columns:

Figure 3 Recommended stack configuration for 1290 Infinity LC Injectors HTS/HTC

Description: A diagram illustrating the recommended stack configuration for 1290 Infinity LC Injectors HTS/HTC, showing an Agilent LCMS system with a 1290 Infinity Series LC stack and the G4277A HTS injector.

Figure 4 Recommended 1-Stack Configuration for 1260 Infinity (without Thermostat, analytical scale)

Description: A diagram showing a single stack configuration for the 1260 Infinity system (analytical scale, without thermostat), including Solvent cabinet, Vacuum degasser, Pump, Autosampler or Micro Autosampler or Preparative Autosampler, Column compartment, and Detector. An Instant Pilot is also shown.

Figure 5 Recommended 2-Stack Configuration for 1260 Infinity (with Thermostat, analytical scale)

Description: A diagram showing a two-stack configuration for the 1260 Infinity system (analytical scale, with thermostat), including Instant Pilot, Detector, Column compartment, Solvent cabinet, Vacuum degasser, Pump, Autosampler, and Thermostat.

Figure 6 Recommended stack configuration for 1290 Infinity (front view)

Description: A diagram showing the front view of a recommended stack configuration for the 1290 Infinity system, including Instant Pilot, Solvent cabinet, Detector, Column compartment, Autosampler, and Pump.

Figure 7 Recommended two stack configuration for 1290 Infinity (front view)

Description: A diagram showing the front view of a recommended two-stack configuration for the 1290 Infinity system, including Instant Pilot, Detector, Column compartment, Solvent cabinet, Pump, Autosampler, and an optional Thermostat for the ALS.

Stack Configurations

Figure 8 Stack configuration for Agilent 1260 Infinity Purification Solution, combined system with UV- and MS-Detection

Description: A diagram illustrating a stack configuration for the Agilent 1260 Infinity Purification Solution, featuring a combined system with UV- and MS-Detection. It shows Solvent cabinet, Pump 1, FC 3, Pump 2, Valve, FC 2, Sampler, Splitter, DAD, Makeup and Analytical, UIB2, and MSD.

Figure 9 Stack configuration for Agilent 1260 Infinity Purification Solution with UV and MS detection

Description: A diagram illustrating a stack configuration for the Agilent 1260 Infinity Purification Solution with UV and MS detection. It shows Solvent cabinet, Pump 1, FC 3, Pump 2, Valve, FC 2, Sampler, Splitter, DAD, Makeup, UIB2, and MSD.

Figure 10 Stack configuration for Agilent 1260 Infinity Purification Solution with UV detections

Description: A diagram illustrating a stack configuration for the Agilent 1260 Infinity Purification Solution with UV detections. It shows Solvent cabinet, Pump 1, FC 3, Pump 2, Valve, FC 2, and Sampler, DAD.

NOTE: The Agilent 1260 Infinity Purification Solution requires special stack configurations. For details, refer to the Agilent 1260 Infinity Purification Solution - System User Guide and the Agilent 1260 Infinity Purification Solution - Installation Checklist.

Separate recommendations may be required for ELSD and Method Development Solution.

PDF preview unavailable. Download the PDF instead.

Agilent-1260 1290 Infinity Site Preparation Checklist Acrobat Distiller 10.1.10 (Windows)

Related Documents

Preview Agilent 1200 Infinity II LC Specification Compendium
Detailed technical specifications for Agilent 1200 Infinity II LC system modules, including pumps, injectors, detectors, and ovens. Covers physical and performance parameters for various Agilent LC components.
Preview Agilent LC and LC/MS: Your Essential Resource for Columns & Supplies
Explore Agilent's comprehensive range of LC and LC/MS systems, columns, and supplies designed to maximize performance and deliver quality results. Discover solutions for various analytical applications.
Preview Agilent InfinityLab LC Series Multisamplers: User Manual for G7167B, G7167A, G5668A
This user manual provides comprehensive guidance for Agilent's InfinityLab LC Series Multisamplers, including the 1290 Infinity II, 1260 Infinity II, and 1260 Infinity II Bio-inert models. It covers installation, operation, maintenance, troubleshooting, and specifications for these advanced liquid chromatography sample handling systems.
Preview Agilent LC and LC/MS Columns & Supplies: Your Essential Resource
Explore Agilent's comprehensive range of LC and LC/MS columns and supplies designed to maximize system performance and deliver quality results. Discover solutions for various applications, including small molecule separations, biomolecules, and more.
Preview Agilent 1260 Infinity Refractive Index Detector User Manual
Comprehensive user manual for the Agilent 1260 Infinity Refractive Index Detector (G1362A RID), covering installation, operation, maintenance, troubleshooting, and specifications for laboratory and analytical applications.
Preview Agilent 1260 Infinity II Binary LC System: Comprehensive User Manual & Quick Guide
This comprehensive manual and quick guide provides detailed information on the Agilent InfinityLab LC Series 1260 Infinity II Binary LC System, covering its introduction, configuration settings, quick start procedures, parts, and consumables. Learn about this trusted platform for routine HPLC and UHPLC analysis, designed for high operational efficiency, low carryover, and seamless integration into laboratory workflows.
Preview Agilent InfinityLab LC Series 1260 Infinity II Quaternary System Manual and Quick Guide
Explore the Agilent InfinityLab LC Series 1260 Infinity II Quaternary System with this comprehensive manual and quick guide. Learn about its advanced features, including high-pressure capabilities for HPLC and UHPLC, ultra-low carryover, dual-needle injection, and high sample capacity, designed for robust routine analysis and efficient method development in various laboratory settings.
Preview Optimization of Jet Stream ESI Parameters for Agilent 1260 Infinity SFC and 6230 TOF LC/MS
This technical overview details the optimization of Agilent Jet Stream ESI source parameters when coupling the Agilent 1260 Infinity Analytical SFC System with the Agilent 6230 TOF LC/MS, focusing on enhancing MS sensitivity through adjustments to drying gas, sheath gas, and capillary voltage.