MICRONOR MANUAL FOTEMP1-OEM-MNT
FOTEMP1-OEM-MNT Manual
Doc No: 98-OPTC-09 Revision A3
MICRONOR SENSORS INC. 2085 Sperry Ave, Suite A-1 Ventura, CA 93003 USA +1-805-389-6600 sales@micronor.com www.micronor.com
For Support in Europe:
WEIDMANN TECHNOLOGIES DEUTSCHLAND GMBH Washingtonstrasse 16/16a D-01139 Dresden Germany +49-351-843-5990 info@optocon.de www.optocon.de
Notice of Proprietary Rights
The design concepts and engineering details embodied in this manual, which are the property of MICRONOR SENSORS INC. and WEIDMANN TECHNOLOGIES DEUTSCHLAND GMBH, are to be maintained in strict confidence; no element or detail of this manual is to be spuriously used, nor disclosed, without the express written permission of MICRONOR SENORS INC and WEIDMANN TECHNOLOGIES DEUTSCHLAND GMBH. All rights are reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission from MICRONOR SENSORS INC and WEIDMANN TECHNOLOGIES DEUTSCHLAND GMBH.
© COPYRIGHT 2016-2023, MICRONOR SENSORS INC. VENTURA, CALIFORNIA
UNITED STATES OF AMERICA

FOTEMP1-OEM-MNT Manual

Revision History

Revision A A1 A2 A3

Date 12/20/2016 10/17/2017 3/21/2018 7/11/2023

Notes Original Release Added FOTEMP1-OEM-MNT-4 (with 4-20mA output option) Corrected polarity of 4-20mA output connections Updated for Micronor Sensors Inc.

Page 2 of 17

FOTEMP1-OEM-MNT Manual
Table of Contents
Table of Contents ...................................................................................................................... 3 Table of Figures ......................................................................................................................... 3 1. Product Description ............................................................................................................ 4
1.1 FOTEMP1-OEM-MNT .................................................................................................. 4 1.2 Fields of Applications................................................................................................... 5 1.3 Features ....................................................................................................................... 5 1.4 Ordering Codes ........................................................................................................... 5 2. Specifications ...................................................................................................................... 6 2.1 OPTOCON SPI Protocol .............................................................................................. 8 2.2 Timing Diagrams.......................................................................................................... 9 2.3 Digital Input/Output Signal Requirements ................................................................. 10 2.4 Connections ............................................................................................................... 10 2.5 Application Specific Connections............................................................................... 11 3. Operation.......................................................................................................................... 13 3.1 Contents .................................................................................................................... 13 3.2 SPI Test Connections ................................................................................................. 14 3.3 Temperature Calibration Using FOTEMP-ASSISTANT Software................................. 15 3.4 FOTEMP-Assistant Software Installation..................................................................... 15 4. Warranty Information ........................................................................................................ 16 5. Mechanical Reference Drawing ......................................................................................... 17
Table of Figures
Figure 1. FOTEMP1-OEM-MNT PCB-Based Signal Conditioner/Controller. ................ 4 Figure 2: Various Views of Signal Conditioner..............................................................7 Figure 3: Timing Diagram of One Complete Read Access ........................................... 9 Figure 4: Oscilloscope Display Shows the "Data Gap" Between the Two Bytes .......... 9 Figure 5: FOTEMP1-OEM-MNT Interface Connections..............................................10 Figure 6: Location of DB9 RS232/SPI and J16 Power Connector ............................... 11 Figure 7: Front View and Pin-Outs of Female D89 Connector....................................11 Figure 8: Power Connections to J15 on FOTEMP-OEM-MNT....................................12 Figure 9: Connections to 4-20mA output on FOTEMP-OEM-MNT-4 ......................... 12 Figure 10: Key Contents of the FOTEMP Temperature System.................................. 13 Figure 11: Test Setup Shown With Micronor SPI Master Test Box.............................. 14
Page 3 of 17

1. Product Description
1.1 FOTEMP1-OEM-MNT

FOTEMP1-OEM-MNT Manual

Figure 1. FOTEMP1-OEM-MNT PCB-Based Signal Conditioner/Controller.
The FOTEMP-OEM-MNT is a small form factor PCB-sandwich module designed for low cost insertion of high precision GaAs-based fiber optic temperature sensor technology into OEM applications where common electrical temperature sensors and thermocouples cannot be used. This OEM product is compatible with all TS series fiber optic temperature probes. The sensors are entirely non-metallic and electrically passive, providing immunity to EMI, RFI, NMR, high voltage, microwaves and radiation The FOTEMP1-OEM-MNT PCB signal conditioner incorporates three digital interface; USB, RS232, and SPI. USB interface is Type B jack while RS232 and SPI share electrical interface connections via a DB9 connector.
Page 4 of 17

1.2 Fields of Applications
· EMI, RFI, and Microwave Environments · High Voltage Environments · Harsh and Hazardous Environments · Nuclear Environments · Aerospace Applications · Process Monitoring · Medical Applications (MR)

FOTEMP1-OEM-MNT Manual

1.3 Features
· Designed for OEM Integration · Very Small Size, Small Footprint · Modular Electronics Support Various Device Configurations · High Accuracy · Built in USB, RS232, and SPI Interface · Attractive OEM Volume Discounts

1.4 Ordering Codes

Standard configurations typically available from stock:

FOTEMP1-OEM-MNT FOTEMP1-OEM-MNT-4

1-Channel Controller, USB/RS232/SPI, No Analog Output 1-Channel Controller, USB/RS232/SPI, 4-20mA Output

Contact Micronor sales concerning other configurations.

Page 5 of 17

FOTEMP1-OEM-MNT Manual

2. Specifications

Table 1: FOTEMP-OEM-MNT Specifications

Parameter

Specification

Optical Interface

ST plug, Single input compatible with all TS Series fiber optic

temperature probes

Fiber

200/220, 0.22NA Step Index Multimode Fiber

Measuring Range

Standard Calibration: -40°C to +300°C

Extended Calibration: -200°C to +300°C

Internal Accuracy

+/- 0.2°C

Internal Sampling Rate 2 Hz

Internal Resolution

0.1°C

SPI Communication

SPI, Slave, Output Only, 16-bit packet (TI TMP123 format)

Protocol

SPI Clock

1 MHz max OR 10 MHz max with 1µs delay between bytes

Power Required

12 VDC, 350 mA

Electrical Interface

Power Connector (2-pin Molex): +V, GND

DB9 (Female): SPI (Ground, Clock, Serial Out, Chip Select);

RS232 (RxD and TxD)

USB via Type B receptacle

4-20mA (only available on FOTEMP1-OEM-MNT-4, 2pin Molex))

Operating Temperature Operating 0°C to +50°C, Storage -20°C to +70°C

Humidity

0-95%, Non-condensing

Ingress Protection

IP00

Dimensions

100mm L x 80mm W x 34mm H

Page 6 of 17

FOTEMP1-OEM-MNT Manual
Figure 2: Various Views of Signal Conditioner
Page 7 of 17

FOTEMP1-OEM-MNT Manual

2.1 OPTOCON SPI Protocol

The OPTOCON FOTEMP-OEM-MNT offers various interface options.

The SPI interface is specifically designed to be compatible with the Texas Instruments TMP123 temperature sensor chip.

The temperature read is started by pulling /SS low and clocking out a 16-bit word with the high byte read first. Only the high order 13 bits are relevant to the temperature reading. The 13 bits are standard two's complement for a signed integer. The most significant bit is the sign while the following 12 bits represents temperature in 1/16°C increments. The lowest three bits are not used. See Table 2 for binary output format of the temperature reading.

The following limitations apply: · Maximum read out clock speed is 1MHz, or 10MHz with a 1µs "inter byte spacing" clock delay. · The value of the three least significant bits are always 0.

Table 2: Temperature Conversion
Temperature (Degree C) 150 C 125 C 25 C 0.0625 C 0 C -0.625 C -25 C -55 C No Sensor Connected

Digital Code 0x4B00 0x3E80 0x0C80 0x0008 0x0000 0xFFF8 0xF380 0xE480 0x9600

Figures 3 and 4 illustrate the "data gap" or "inter byte spacing". The time between the first and second data byte must be 1µs and be guaranteed > 750ns.

Page 8 of 17

2.2 Timing Diagrams
MOSI MOSI
/SS SCK

FOTEMP1-OEM-MNT Manual

Temperature High Byte 0x12

Inter-byte spacing

Temperature Low Byte 0x34

Figure 3: Timing Diagram of One Complete Read Access

Figure 4: Oscilloscope Display Shows the "Data Gap" Between the Two Bytes
Page 9 of 17

FOTEMP1-OEM-MNT Manual

2.3 Digital Input/Output Signal Requirements

Table 3: Digital Input/Output Electrical Characteristics

Signal

Condition

Min

Input Logic Levels

VIH

VCC = 2.7 - 3.6V

VIL

Output Logic Levels

VOL (IOL=2mA)

VOH (IOH=-2mA) VCC = 2.7 - 3.6V

SPI Clock

Frequency

2 --
-2.4V
--

Max
-0.8V
0.4V --
1 MHz or 10 MHz with 1µs
Spacing

2.4 Connections
ST Connection for TS
Temperature Probes
4-20mA Output (FOTEMP1-OEM-
MNT-4 Only)

DSUB9 Connector provides RS232 and SPI Interface
12V Power Connection
USB Interface

Figure 5: FOTEMP1-OEM-MNT Interface Connections

Page 10 of 17

2.5 Application Specific Connections

FOTEMP1-OEM-MNT Manual

Figure 6: Location of DB9 RS232/SPI and J16 Power Connector
Function: SPI and RS232 Connector: DB9, Female with 9x Socket Contacts Mating Connector: DB9 Male Plug with 9x Pin Contacts

Figure 7: Front View and Pin-Outs of Female D89 Connector

IMPORTANT NOTE: The RS232 signals uses 12V signals. A special cable must be used that does not connect pins 6-9 as the normal RS232 12V handshake signals appearing on these pins may damage the SPI 5V signals.

Table 4: D89 Signal Connections

Pin

Signal

1

Not connected

2

RxD (RS232)

3

TxD (RS232)

4

Not connected

5

GND

6

MOSI (SPI)

7

/SS (SPI)

8

SCK (SPI)

9

MISO (SPI)

Notes
Data Out, RS232, DCE, 12V Data In, RS232, DCE, 12V
GND L2C, SPI, 5V L1C, SPI, 5V L4C, SPI, 5V L3C, SPI, 5V
Page 11 of 17

FOTEMP1-OEM-MNT Manual
Figure 8: Power Connections to J15 on FOTEMP-OEM-MNT
Function: Power, J15 Connector: 2C Pin Header, Mating Connector: 2C, Socket Plug
Figure 9: Connections to 4-20mA output on FOTEMP-OEM-MNT-4
Function: 4-20mA Analog Output, J_MA Connector: 2C Pin Header, Mating Connector: 2C, Socket Plug
Page 12 of 17

3. Operation
3.1 Contents
FOTEMP1-OEM-MNT Signal Conditioner

FOTEMP1-OEM-MNT Manual

TS Temperature Probe (Purchased Separately)

Figure 10: Key Contents of the FOTEMP Temperature System
The FOTEMP1-OEM-MNT system contains the following: · FOTEMP1-OEM-MNT Signal Conditioner · AC Power Supply (Upon request for engineering evaluation applications only) · CDROM with FOTEMP-ASSISTANT Software, USB Drivers and other support documentation (not shown)
Available Separately: · TS Series Temperature Probes

Page 13 of 17

FOTEMP1-OEM-MNT Manual
3.2 SPI Test Connections
The system was connected and tested as shown in Figure 10. The Micronor SPI Master Emulator displays the both the TMP123-type binary reading and temperature in °C. From the FOTEMP signal conditioner.
FOTEMP1-OEM-MNT Signal Conditioner
TS Temperature
Probe
Figure 11: Test Setup Shown With Micronor SPI Master Test Box
Steps 1. Connect TS2 temperature probe to ST optical interface 2. Connect 12 V Power to J15 pin header 3. Connect SPI interface using DB9 Male Connector With Pin Contacts
Page 14 of 17

FOTEMP1-OEM-MNT Manual
3.3 Temperature Calibration Using FOTEMP-ASSISTANT Software
The FOTEMP1 will need to be calibrated with the actual temperature sensor used. This is achieved using a PC running the supplied FOTEMP-ASSISTANT graphical display/data logging software and connecting to the FOTEMP1 signal conditioner either via RS232 or USB interface.
For more information regarding calibration using FOTEMP-Assistant please refer to the FOTEMP-ASSITANT2 Manual.
3.4 FOTEMP-Assistant Software Installation
The FOTEMP-ASSISTANT V2 software is provided on the supplied CDROM. It will work with either USB or RS232 Interface
For more information regarding the installation of FOTEMP-Assistant please refer to the FOTEMP-ASSITANT2 Manual.
For RS232 operation 1. Connect the FOTEMP1 signal conditioner to the PC using Specially Wired RS232 cable. This special cable has Pins 6-9 disconnected so that the 12V RS232 signals do not damage the 5V SPI signals wired to these DB9 pins. 2. Click on the Program File and then [PortSettings]. Be sure that COM1 is selected. 3. Select [OK] and then [Connect] to start the program. After a few seconds, the program should start and display temperature.
For USB operation 1. Connect the FOTEMP1 signal conditioner to the PC using supplied USB cable. If Windows tells you that the device has been properly installed, then continue to Step 3. 2. Disconnect the USB cable. Windows was not able to load the correct Virtual Comm Port (VCP) drivers for the FTDI serial interface chip. You will need to install VCP drivers (either 32-bit or 64-bit depending on your system) supplied on the CD. Then repeat Step 1. 3. Click on the Program File and then [PortSettings]. Select the COMx port that corresponds to VCP assigned to the FOTEMP unit. 4. Select [OK] and then [Connect] to start the program. After a few seconds, the program should start and display temperature. 5. If the program will not start, then contact Micronor for assistance.
Note: If developing custom communications software, contact Micronor for ASCII Protocol. Sample National Instruments LabVIEWTM Driver is also available.
Page 15 of 17

FOTEMP1-OEM-MNT Manual

4. Warranty Information

Warranty

MICRONOR SENSORS INC. warrants this product to be free from defects in material and workmanship for a period of 1 (one) year from date of shipment. During the warranty period we will, at our option, either repair or replace any product that proves to be defective.

To exercise this warranty, write or call MICRONOR SENORS INC. You will be given prompt assistance and return instructions. Send the instrument, transportation prepaid, to the indicated service facility. Repairs will be made and the instrument returned transportation prepaid. Repaired products are warranted for the balance of the original warranty period, or at least 90 days.

Limitations of Warranty This warranty does not apply to defects resulting from unauthorized modification or misuse of any product or part. This warranty also does not apply to Fiber Optic Connector interfaces, fuses or AC line cords. This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability of fitness for a particular use. MICRONOR SENSORS INC. shall not be liable for any indirect, special or consequent damages.

Contact Information:

Micronor Sensors Inc. 2085 Sperry Ave, Suite A-1 Ventura, CA 93003 USA

Phone +1-805-389-6600
Email sales@micronor,com URL www.micronor.com

For Europe:
Weidman Technologies Deutschland GmbH Washingtonstrasse 16/16a D-01139 Dresden GERMANY

Phone +49-351-843-5990 Fax +49-351-843-5991111 Email info@optocon.de URL www.optocon.de

Page 16 of 17

5. Mechanical Reference Drawing

FOTEMP1-OEM-MNT Manual

Page 17 of 17

