Rebel Tools Digital Multimeter

Model: MIE-RB-55

Safety Instructions

  1. Before using the device, inspect the case for any mechanical damage. If the case has cracks or is missing any part, do not use the device.
  2. Before using the device, check the test leads for any insulation damage. If the test leads has insulation damage or cuts, do not use the device.
  3. Do not input voltages higher than said in the specification.
  4. During measurements, do not change the position of the rotary switch.
  5. Take extra precaution while measuring voltages above 60 V DC and 30 V AC.
  6. Before measurement position the rotary switch to correct position and connect the test leads to the proper terminals.
  7. Do not use the device in temperatures exceeding the range in specification, in high humidity, and near explosives and flammables.
  8. During measurement, hold the test leads above the finger guards.
  9. Before measurement disconnect the measured circuit from power and discharge all capacitors.
  10. If the display shows [low battery indicator], change the batteries. Otherwise the accuracy of measurements may be decreased.
  11. Before opening the battery case disconnect the test leads from measured circuit and from the device.
  12. Battery must be replaced with a new one and with the same specification (9 V 6LR061 battery).
  13. Do not tamper with devices internal circuits. This may lead to measurement decrease or damaging the device.
  14. The device is intended for indoors use only.
  15. Take out the batteries, if the device is not going to be used for a long time.

Product Description

  1. Display
  2. ON/OFF button
  3. Measurement hold button
  4. Rotary switch
  5. hFE terminal
  6. Connection terminals

Operation

AC and DC Voltage Measurement

  1. Connect the red test lead to the VΩ terminal.
  2. Connect the black test lead to the COM terminal.
  3. Set the rotary switch to proper voltage range.
  4. Connect the test leads to a measured circuit.
  5. Turn on the circuits power. The voltage and polarization will show on the display.
  6. *If the range is not known, set the range to highest and gradually lower it, until getting the accurate measurement.

AC and DC Current Measurement

  1. Connect the red test lead to the VΩmA terminal. If the measurement will be in 200 mA - 20 A range, connect the red test lead to the 20 A terminal.
  2. Connect the black test lead to the COM terminal.
  3. Set the rotary switch to proper current range.
  4. Open the measured circuit and connect the test leads in series.
  5. The result will appear on the display.
  6. Caution: while connected to the 20 A terminal, measurements can last only 15 seconds in a few seconds intervals.

Resistance Measurement

  1. Connect the red test lead to the VΩ terminal.
  2. Connect the black test lead to the COM terminal.
  3. Set the rotary switch to proper Ω range.
  4. If the resistance measurement will be conducted on the circuit, disconnect the circuit from power and discharge all capacitors.
  5. Connect the test leads to the measured circuit.
  6. The result will appear on the display.

Diode Measurement

  1. Connect the red test lead to the VΩ terminal.
  2. Connect the black test lead to the COM terminal.
  3. Set the rotary switch to position [diode symbol].
  4. Connect the red lead to anode of the diode, and black to the cathode.
  5. Voltage drop will be displayed. If the display shows the "1", it means that the polarization is reversed.

Continuity Test

  1. Connect the red test lead to the VΩmA terminal.
  2. Connect the black test lead to the COM terminal.
  3. Set the rotary switch to position [continuity test symbol].
  4. Connect the test leads to a measured circuit. If the resistance is lower than 30 Ω ± 20 Ω, the buzzer will buzz.

hFE Transistor Test

  1. Set the rotary switch to the hFE position.
  2. Determine the type of transistor (PNP or NPN) and connect the emitter, base and collector to proper sockets.
  3. Approximate value of hFE will be displayed.

Capacitance Measurement

  1. Connect the red test lead to the VΩmA terminal.
  2. Connect the black test lead to the COM terminal.
  3. Connect the test leads to measured capacitor.
  4. Caution: before measurement, disconnect the measured circuit and discharge all of the capacitors.

Cleaning and Maintenance

  • The meter will turn off automatically after 15 minutes of idle.
  • Clean the device with soft, slightly damp cloth, without abrasive agents.
  • To change the battery, unscrew the screw of the battery compartment. Take out the battery compartment cover and replace the battery (note the polarity). Close the battery compartment cover and screw in the screw.

Specifications

DC Voltage

RangeResolutionAccuracy
200 mV0.1 µV±(0.5% + 3)
2 V1 mV±(0.8% + 5)
20 V10 mV
200 V100 mV±(1% + 5)
1000 V1 V
  • Input impedance: 10 MΩ
  • Overload protection: 1000 V DC or 750 V AC
  • Max input voltage: 1000 V DC

AC Voltage

RangeResolutionAccuracy
200 mV0.1 µV±(0.5% + 3)
2 V1 mV±(0.8% + 5)
20 V10 mV
200 V100 mV±(1% + 5)
750 V1 V
  • Input impedance: 10 MΩ
  • Frequency range: 40 Hz ~ 400 Hz
  • Max input voltage: 1000 V DC

DC Current

RangeResolutionAccuracy
20 µA10 nA±(1.8% + 2)
200 µA100 nA
2000 µA1 µA±(2% + 2)
20 mA10 µA
200 mA100 µA±(2% + 10)
2 A1 mA
20 A10 mA
  • Overload protection:
  • mA: F0.5A/250V
  • A: F2A/250V
  • 20A: no fuse
  • Voltage drop: 200 mV

AC Current

RangeResolutionAccuracy
20 µA10 nA±(2% + 5)
200 µA100 nA
2000 µA1 µA±(2% + 3)
20 mA10 µA
200 mA100 µA±(2% + 5)
2 A1 mA
20 A10 mA±(2.5% + 10)
  • Overload protection:
  • mA: F0.5A/250V
  • A: F2A/250V
  • 20A: no fuse
  • Voltage drop: 200 mV
  • Frequency range: 40 Hz ~ 400 Hz

Resistance

RangeResolutionAccuracy
200 Ω0.1 Ω±(1% + 10)
2 KΩ1 Ω
20 KΩ10 Ω±(1% + 4)
200 KΩ100 Ω
2 MΩ1 KΩ±(1% + 10)
20 MΩ10 KΩ
200 MΩ100 KΩ±(5% + 10)
  • Open circuit voltage: about 0.5 V (200 MΩ range: 3 V)
  • Overload protection: 250 V DC/AC

Capacitance

RangeResolutionAccuracy
2 nF1 pF±(4% + 5)
20 nF10 pF
200 nF100 pF
2 uF1 nF
20 uF10 nF
200 uF100 nF
  • Overload protection: fuse F0.5A/250V

Continuity and Diode Measurement

RangeDescriptionRemarks
[continuity test symbol]The approximate forward voltage drop will be displayed. If the resistance is lower than 30 Ω ± 20 Ω, the buzzer will buzz.Open circuit voltage: 2.8 V
[diode symbol]Forward voltage drop will be displayed.Open circuit voltage: approx. 2.8 V

Overload protection: 250 V DC/AC rms

Correct Disposal of This Product (Waste Electrical & Electronic Equipment)

(Applicable in the European Union and other European countries with separate collection systems)

This marking shown on the product or its literature, indicates that it should not be disposed with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources. Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.

Made in China for LECHPOL ELECTRONICS Sp. z o.o. Sp.k., ul. Garwolińska 1, 08-400 Miętne.

Models: MIE-RB-55, Digital Universal Multimeter, MIE-RB-55 Digital Universal Multimeter

File Info : application/pdf, 17 Pages, 648.29KB

PDF preview unavailable. Download the PDF instead.

a38a4db7a2745ba3e815911b2fb2364d PDFlib PDI 9.2.0p5-i (Win32)

Related Documents

Preview Rebel Digital Multimeter MIE-RB-33B, C, D User Manual
User manual for the Rebel Digital Multimeter models MIE-RB-33B, MIE-RB-33C, and MIE-RB-33D. Includes safety instructions, product description, operation guide, specifications, and maintenance information.
Preview Rebel RB-1140 Endoscope Camera User Manual - Setup and Operation
Comprehensive user manual for the Rebel RB-1140 Endoscope Camera. Learn how to connect and operate the inspection camera with Android, Windows, and macOS devices. Includes safety instructions, operation guides, and technical specifications.
Preview Rebel RB-0701 4G LTE MiFi Router Owner's Manual
This owner's manual provides comprehensive guidance for the Rebel RB-0701 4G LTE MiFi Router, covering setup, operation, safety instructions, troubleshooting, and technical specifications for portable wireless internet connectivity.
Preview Rebel LIGHT LED Floodlight with Dusk and Motion Sensor - User Manual URZ3622
Comprehensive user manual for the Rebel LIGHT URZ3622 LED Floodlight. Features include dusk and motion sensors, energy efficiency, and technical specifications. Learn about installation, settings, troubleshooting, and proper disposal.
Preview Rebel KOM1060 Digital TV USB Tuner Owner's Manual
Comprehensive owner's manual for the Rebel KOM1060 Digital TV USB Tuner, covering safety instructions, setup, operation, features, and disposal guidelines.
Preview Rebel RB-0700 4G LTE Modem with Wi-Fi User Manual
User manual for the Rebel RB-0700 4G LTE Modem with Wi-Fi, covering safety instructions, device status LEDs, SIM card installation, wired and wireless connections, and network configuration. Includes multilingual support.
Preview Rebel Router 4G LTE RB-0702 - Specifications and Features
Detailed information about the Rebel Router 4G LTE RB-0702, including its construction, network capabilities, packaging data, wireless LAN functions, ports, antenna type, WAN connection, and mobile network support. Optimized for search engines.
Preview Rebel Power Inverter User Manual
Comprehensive user manual for Rebel Power Inverters, covering installation, operation, safety, troubleshooting, and specifications for models URZ3160-62, URZ3164-65, URZ3167-68.