Rebel Tools Digital Multimeter
Model: MIE-RB-55
Safety Instructions
- Before using the device, inspect the case for any mechanical damage. If the case has cracks or is missing any part, do not use the device.
- Before using the device, check the test leads for any insulation damage. If the test leads has insulation damage or cuts, do not use the device.
- Do not input voltages higher than said in the specification.
- During measurements, do not change the position of the rotary switch.
- Take extra precaution while measuring voltages above 60 V DC and 30 V AC.
- Before measurement position the rotary switch to correct position and connect the test leads to the proper terminals.
- Do not use the device in temperatures exceeding the range in specification, in high humidity, and near explosives and flammables.
- During measurement, hold the test leads above the finger guards.
- Before measurement disconnect the measured circuit from power and discharge all capacitors.
- If the display shows [low battery indicator], change the batteries. Otherwise the accuracy of measurements may be decreased.
- Before opening the battery case disconnect the test leads from measured circuit and from the device.
- Battery must be replaced with a new one and with the same specification (9 V 6LR061 battery).
- Do not tamper with devices internal circuits. This may lead to measurement decrease or damaging the device.
- The device is intended for indoors use only.
- Take out the batteries, if the device is not going to be used for a long time.
Product Description
- Display
- ON/OFF button
- Measurement hold button
- Rotary switch
- hFE terminal
- Connection terminals
Operation
AC and DC Voltage Measurement
- Connect the red test lead to the VΩ terminal.
- Connect the black test lead to the COM terminal.
- Set the rotary switch to proper voltage range.
- Connect the test leads to a measured circuit.
- Turn on the circuits power. The voltage and polarization will show on the display.
- *If the range is not known, set the range to highest and gradually lower it, until getting the accurate measurement.
AC and DC Current Measurement
- Connect the red test lead to the VΩmA terminal. If the measurement will be in 200 mA - 20 A range, connect the red test lead to the 20 A terminal.
- Connect the black test lead to the COM terminal.
- Set the rotary switch to proper current range.
- Open the measured circuit and connect the test leads in series.
- The result will appear on the display.
- Caution: while connected to the 20 A terminal, measurements can last only 15 seconds in a few seconds intervals.
Resistance Measurement
- Connect the red test lead to the VΩ terminal.
- Connect the black test lead to the COM terminal.
- Set the rotary switch to proper Ω range.
- If the resistance measurement will be conducted on the circuit, disconnect the circuit from power and discharge all capacitors.
- Connect the test leads to the measured circuit.
- The result will appear on the display.
Diode Measurement
- Connect the red test lead to the VΩ terminal.
- Connect the black test lead to the COM terminal.
- Set the rotary switch to position [diode symbol].
- Connect the red lead to anode of the diode, and black to the cathode.
- Voltage drop will be displayed. If the display shows the "1", it means that the polarization is reversed.
Continuity Test
- Connect the red test lead to the VΩmA terminal.
- Connect the black test lead to the COM terminal.
- Set the rotary switch to position [continuity test symbol].
- Connect the test leads to a measured circuit. If the resistance is lower than 30 Ω ± 20 Ω, the buzzer will buzz.
hFE Transistor Test
- Set the rotary switch to the hFE position.
- Determine the type of transistor (PNP or NPN) and connect the emitter, base and collector to proper sockets.
- Approximate value of hFE will be displayed.
Capacitance Measurement
- Connect the red test lead to the VΩmA terminal.
- Connect the black test lead to the COM terminal.
- Connect the test leads to measured capacitor.
- Caution: before measurement, disconnect the measured circuit and discharge all of the capacitors.
Cleaning and Maintenance
- The meter will turn off automatically after 15 minutes of idle.
- Clean the device with soft, slightly damp cloth, without abrasive agents.
- To change the battery, unscrew the screw of the battery compartment. Take out the battery compartment cover and replace the battery (note the polarity). Close the battery compartment cover and screw in the screw.
Specifications
DC Voltage
Range | Resolution | Accuracy |
---|---|---|
200 mV | 0.1 µV | ±(0.5% + 3) |
2 V | 1 mV | ±(0.8% + 5) |
20 V | 10 mV | |
200 V | 100 mV | ±(1% + 5) |
1000 V | 1 V |
- Input impedance: 10 MΩ
- Overload protection: 1000 V DC or 750 V AC
- Max input voltage: 1000 V DC
AC Voltage
Range | Resolution | Accuracy |
---|---|---|
200 mV | 0.1 µV | ±(0.5% + 3) |
2 V | 1 mV | ±(0.8% + 5) |
20 V | 10 mV | |
200 V | 100 mV | ±(1% + 5) |
750 V | 1 V |
- Input impedance: 10 MΩ
- Frequency range: 40 Hz ~ 400 Hz
- Max input voltage: 1000 V DC
DC Current
Range | Resolution | Accuracy |
---|---|---|
20 µA | 10 nA | ±(1.8% + 2) |
200 µA | 100 nA | |
2000 µA | 1 µA | ±(2% + 2) |
20 mA | 10 µA | |
200 mA | 100 µA | ±(2% + 10) |
2 A | 1 mA | |
20 A | 10 mA |
- Overload protection:
- mA: F0.5A/250V
- A: F2A/250V
- 20A: no fuse
- Voltage drop: 200 mV
AC Current
Range | Resolution | Accuracy |
---|---|---|
20 µA | 10 nA | ±(2% + 5) |
200 µA | 100 nA | |
2000 µA | 1 µA | ±(2% + 3) |
20 mA | 10 µA | |
200 mA | 100 µA | ±(2% + 5) |
2 A | 1 mA | |
20 A | 10 mA | ±(2.5% + 10) |
- Overload protection:
- mA: F0.5A/250V
- A: F2A/250V
- 20A: no fuse
- Voltage drop: 200 mV
- Frequency range: 40 Hz ~ 400 Hz
Resistance
Range | Resolution | Accuracy |
---|---|---|
200 Ω | 0.1 Ω | ±(1% + 10) |
2 KΩ | 1 Ω | |
20 KΩ | 10 Ω | ±(1% + 4) |
200 KΩ | 100 Ω | |
2 MΩ | 1 KΩ | ±(1% + 10) |
20 MΩ | 10 KΩ | |
200 MΩ | 100 KΩ | ±(5% + 10) |
- Open circuit voltage: about 0.5 V (200 MΩ range: 3 V)
- Overload protection: 250 V DC/AC
Capacitance
Range | Resolution | Accuracy |
---|---|---|
2 nF | 1 pF | ±(4% + 5) |
20 nF | 10 pF | |
200 nF | 100 pF | |
2 uF | 1 nF | |
20 uF | 10 nF | |
200 uF | 100 nF |
- Overload protection: fuse F0.5A/250V
Continuity and Diode Measurement
Range | Description | Remarks |
---|---|---|
[continuity test symbol] | The approximate forward voltage drop will be displayed. If the resistance is lower than 30 Ω ± 20 Ω, the buzzer will buzz. | Open circuit voltage: 2.8 V |
[diode symbol] | Forward voltage drop will be displayed. | Open circuit voltage: approx. 2.8 V |
Overload protection: 250 V DC/AC rms
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
(Applicable in the European Union and other European countries with separate collection systems)
This marking shown on the product or its literature, indicates that it should not be disposed with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources. Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
Made in China for LECHPOL ELECTRONICS Sp. z o.o. Sp.k., ul. Garwolińska 1, 08-400 Miętne.