User Manual for PPI models including: LabCon Multi-Purpose Temperature Controller, LabCon, Multi-Purpose Temperature Controller, Temperature Controller, Controller

LabCon Basic Version

abc

LabCon Basic Operator Manual

Temperature control system | Temperature Recorders India - PPI controller


File Info : application/pdf, 1 Pages, 430.90KB

PDF preview unavailable. Download the PDF instead.

Product-065-OM-2
LabCon
Multi-Purpose Temperature Controller

Operation Manual
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

OPERATOR PAGE PARAMETERS

Parameters

Settings (Default Value)

Time Start Command >> Time Abort Command >>

Yes No (Default : No)

Time Interval (H:M) >> Ctrl Set Value >>
Ctrl Lo Deviation >>

0.00 to 500.00 (HH:MM)
(Default : 0.10)
Setpoint LO limit to Setpoint HI limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 25.0)
For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to 99 (Default : 2.0)

Ctrl Hi Deviation >>

For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to 99
(Default : 2.0)

Change Password >>

1 to 100 (Default : 0)

SUPERVISORY > SENSOR INPUT

Parameters

Settings (Default Value)

Ctrl Zero Offset >>

(Resolutio-5n00t.o1°5C0for RTD / DC Linear & 1°C for Thermocouple)

(Default : 0.0)

SUPERVISORY > CONTROL

Parameters

Settings (Default Value)

Tune >>

Yes No (Default : No)

Setpoint LO Limit >> Setpoint HI Limit >>

Min Range for the Selected Input Type to Setpoint HI Limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 0.0)
Setpoint LO Limit to Max Range for the Selected
Input Type (Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 600.0)

Compressor Setpoint >>

0 to 100 (Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 45.0)

Compressor Hyst >>

0.1 to 99.9 (Default : 2.0)

Heat Ctrl Action >>

ON-OFF PID
(Default : PID)

Heat Hyst >>

0.1 to 99.9 (Default : 0.2)

Heat Only Control

Heat + Cool Control Zone : Single

Heat + Cool Control Zone : Dual

Proportional Band >> Proportional Band >> PrCozpoPrrtioopnaBl aBnadnd>>for

0.1 to 999.9

0.1 to 999.9

Cool Pre-dominant zone

(Default : 50.0)

(Default : 50.0)

0.1 to 999.9 (Default : 50.0)

Integral Time >> 0 to 3600 sec
(Default : 100 sec)

Integral Time >> 0 to 3600 sec
(Default : 100 sec)

CzInItnetgergarlaTlimTiemfeor>> Cool Pre-dominant zone
0 to 3600 sec (Default : 100 sec)

Derivative Time >>
0 to 600 sec (Default : 16 sec)

Derivative Time >>
0 to 600 sec (Default : 16 sec)

Cz Derivative Time >> Derivative Time for
Cool Pre-dominant zone
0 to 600 sec (Default : 16 sec)

Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)

Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)

PrHozpoPrrtoiopnaBl aBnadnd>>for Heat Pre-dominant zone
0.1 to 999.9 (Default : 50.0)

Overshoot Inhibit >>
Enable Disable

Overshoot Inhibit >>
Enable Disable

HzInIntetgergarlaTlimTiemfeor>> Heat Pre-dominant zone

(Default : Disable)

(Default : Disable)

0 to 3600 sec (Default : 100 sec)

Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)

Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)

Hz Derivative Time >> Derivative Time for
Heat Pre-dominant zone 0 to 600 sec
(Default : 16 sec)

Cycle Time >> 0.5 to 100.0 sec (Default : 10.0 sec)

Overshoot Inhibit >>

Enable
Disable (Default : Disable)

Cutoff Factor >> 1.0 to 2.0 sec
(Default : 1.2 sec)

SUPERVISORY > PASSWORD

Parameters

Settings (Default Value)

Change Password >>

1000 to 1999 (Default : 123)

SUPERVISORY > EXIT

Parameters

Settings (Default Value)

Exit Setup Mode >>

Yes No
(Default : No)

FACTORY > CONTROL SENSOR INPUT

Parameters

Settings (Default Value)

Input Type >>

Refer Table 1 (Default : RTD Pt100)

Signal LO >>

Input Type 0 to 20mA 4 to 20mA
0 to 5V 0 to 10V 1 to 5V

Settings 0.00 to Signal High 4.00 to Signal High 0.000 to Signal High 0.00 to Signal High 1.000 to Signal High

Default 0.00 4.00 0.000 0.00 1.000

Signal HI >>

Input Type 0 to 20mA 4 to 20mA
0 to 5V 0 to 10V 1 to 5V

Settings Signal Low to 20.00 Signal Low to 20.00 Signal Low to 5.000 Signal Low to 10.00 Signal Low to 5.000

Default 20.00 20.00 5.000 10.00 5.000

Range LO >> Range HI >>

-199.9 to RANGE HI (Default : 0.0)
RANGE LO to 999.9 (Default : 100.0)

FACTORY > ALARM PARAMETERS

Parameters

Settings (Default Value)

Hysteresis >>

0.1 to 99.9 (Default : 0.2)

Inhibit >>

Yes No
(Default : Yes)

FACTORY > HEAT COOL SELECT

Parameters

Settings (Default Value)

Control Strategy >>

Heat Only Cool Only Heat + Cool
(Default : Heat + Cool)

Control Strategy : Cool Only

Time Delay (sec) >>

0 to 1000 Sec (Default : 200 Sec)

Control Strategy : Heat + Cool

Compressor Strategy >>

CONT. OFF CONT. ON SP BASED
PV BASED (Default : CONT. ON)

CONT. ON

SP BASED

Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)

Boundary Set Value >> 0 to 100
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 45.0)

PV BASED
Time Delay (sec) >> 0 to 1000 Sec
(Default : 200 Sec)

Control Zones >> Single Dual
(Default : Single)

Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)

FACTORY > SOAK TIMER PARAMETERS

Parameters

Settings (Default Value)

Enable >>

No
Yes (Default : No)

Start Band >>

0 to 999.9 (Default : 0.5)

Parameters Holdback Strategy >>
Hold Band >> Heat Off >> Cool Off >> Power Recovery >>

Settings (Default Value)
None Up
Down Both (Default : None)
0.1 to 999.9 (Default : 0.5)
No Yes (Default : No)
No Yes (Default : No)
Abort Restart Continuous (Default : Restart)

FACTORY > DOOR OPEN

Parameters

Settings (Default Value)

Enable >>

Yes No (Default : No)

Switch Logic >> Door Alrm Dly (sec) >>

Close : Door Open Open : Door Open (Default : Close : Door Open)
0 to 1000 Sec (Default : 60 Sec)

FACTORY > MAINS FAILURE

Parameters

Settings (Default Value)

Enable >>

Yes No
(Default : No)

Switch Logic >>

Close : Mains Fail
Open : Mains Fail (Default : Close : Mains Fail)

FACTORY > PASSWORD

Parameters

Settings (Default Value)

Change Password >>

2000 to 2999 (Default : 321)

FACTORY > FACTORY DEFAULT

Parameters

Settings (Default Value)

Set To Default >>

Yes
No (Default : No)

Parameters Exit Setup Mode >>

FACTORY > EXIT Settings
(Default Value)
Yes No (Default : No)

TABLE 1

What it means

Range (Min. to Max.)

Resolution

Type J Thermocouple

0 to +960°C

Type K Thermocouple

-200 to +1376°C

Type T Thermocouple Type R Thermocouple Type S Thermocouple

-200 to +385°C 0 to +1770°C 0 to +1765°C

Fixed 1°C

Type B Thermocouple

0 to +1825°C

Type N Thermocouple

0 to +1300°C

Reserve

Reserved for customer specific Thermocouple type not listed above. The type shall be specified in accordance with the ordered (optional on request) Thermocouple type.

3-wire, RTD Pt100 0 to 20mA DC current

-199.9 to 600.0°C

Fixed 0.1°C

4 to 20mA DC current 0 to 5.0V DC voltage 0 to 10.0V DC voltage

-199.9 to 999.9 units

Fixed 0.1 unit

1 to 5.0V DC voltage

101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist. Palghar - 401 210. Sales : 8208199048 / 8208141446 Support : 07498799226 / 08767395333 E: sales@ppiindia.net, support@ppiindia.net
ELECTRICAL CONNECTIONS

SOAK START ACK
BATTERY

01 234
COMMON DOOR

RS485

Control Outputs

COMM HTR CMP ALM
+DEV- +PC- + - + - + -

9 10 11 12 13 14 15 16 17 18

LabCon (w/o map) Configuration PC Printer Recording

5678

Pt100

++mVA +

-

TC
Supply

18~32VDC +-
19 20 21

FRONT PANEL KEYS

Symbol

Key
Scroll

Function
Press to scroll through various Process Information Screens in Normal Operation Mode.

Alarm Acknowledge
DOWN
UP

Press to acknowledge and mute (if active) alarm output.
Press to decrease the parameter value. Pressing once decreases the value by one count; keeping pressed speeds up the change.
Press to increase the parameter value. Pressing once increases the value by one count; keeping pressed speeds up the change.

SET-UP

Press to enter or exit set-up mode.

ENTER

Press to store the set parameter value and to scroll to the next parameter.

PV ERROR INDICATIONS

Message
OPEN OVR UNR

Error Type
Sensor Open Over-range Under-range

Cause
Sensor (RTD Pt100) Broken / Open
Temperature above Max. Specified Range
Temperature below Min. Specified Range

Jan 2022



References