Triplett RHDL30/RHDL40 PDF Dataloggers User Manual

Display

  • [REC]: flashes every second when in logging mode. For Program Start Delay: [REC] is displayed but does not flash during standby status.
  • [T1]: is Air Temperature (T2 is for Probe – N/A on these models)
  • [H/L]: High and Low Set Point Limit exceeded Indicator
  • °C/°F: Temperature units of measure
  • 0000.8: Measured Temperature, Humidity or Pressure. Toggle Rate is every 5 seconds.
  • [MXMN]: MAX/MIN Readings
  • [Low battery indicator]

Faceplate (Buttons and LED Indicator)

  • [MAX/MIN]: Press to view MAX or MIN Values of all logged data.
  • LED Indicator: LED will flash GREEN every 10 seconds when logging is started and RED if any Low or High alarms have been tripped during a log session
  • [START/STOP]: Once batteries are installed. Pressing this button will POWER ON and OFF the logger. After Programming, Pressing and Holding for 5 seconds will Start and Stop the logging. In logging mode a short press will not turn OFF the logger.
  • [MARK]: To mark an event manually during the measurement period, press the [MARK] button for 2 seconds until the measured data flash three times. This [MARK] function is usually used when there is a transition, from one location to another. At most 8 marks will be identified in the PDF report. Short press [MARK] button any time can also display the leftover battery life time as a quick reference. However, precise battery life time is still decided by operation condition & battery brand.

Operation

NOTES

  • Use a coin to twist battery compartment to the UNLOCK position and Install (2) CR2032 Batteries observing polarity.
  • Adobe Reader Software is required
  • Program and generate the reports at Ambient Room Temperature. For example, if removed from a freezer allow stabilization time to room temp before generating the reports.

Step 1: Configuring the data logger

  • It is possible to make configuration changes at any time before the logging function is started. Once the logger is started, configuration changes cannot be made unless stopping the logging first. If the logger is locked with a Password, the Password is required to make configuration changes.
  • Press [ON/OFF] to power on the logger
  • Connect the data logger to a PC via the [USB] port. The green LED is ON while building the connection with computer.
  • Windows file display window will appear.
  • If the correct file folder is not open, click on Folder to view files.
  • Open the file "[PDF Logger Configuration Tool.exe]".
  • The default language is English. The user may change it to one of several alternate languages. There are English, German, French, Italian, Spanish and Portuguese.
  • To review the instruction manual, you may click "Manual" to open the manual PDF file.
  • The user can check the calibration report for the logger by selecting the PDF file named "[Calibration Certificate]".
  • User programmable parameters are as follows:
  • [Sampling rate]: Select the sampling interval you need from 30 seconds to 2 hours.
  • [Start delay]: Select the start delay from 0 min to 24 hours. For example: If the delay is 5 minutes, and the sampling rate is 10 minutes, the real time to log the first temperature measurement is 5 minutes after the [START] button is pressed. All measurements after the first measurement will be at a 10 minute (or selected) interval.
  • [Unit of Measurement (UoM)]: Select the unit that will be displayed on the PDF report and LCD. The selections are Metric or Imperial. In Metric, it is Celsius for temperature and hpa for pressure. In Imperial, it is Fahrenheit for temperature and inHg for pressure.
  • [Password]: The Password function is default OFF. The user may enable it to prevent unauthorized reprogramming prior to pressing the [START] button. A Password may have up to 16 alpha or numeric characters.
  • [Company Name]: A user defined name, or descriptor, can be input under Company Name. It will be displayed on the PDF report as the Title, with a maximum of 20 characters.
  • [Alarm types]: Regardless of the Alarm Type, if the red LED is triggered, it won't stop even the reading return to normal range or logging is stopped. The way to stop red alarm LED is to plug logger into computer to generate report or power off the logger.
    • Single: an alarm is triggered immediately when the measured value exceeds the alarm threshold.
    • Cumulative: an alarm is not triggered when the measured value exceeds the alarm threshold, but only once the overall average value during alarm delay duration exceeds the alarm threshold.
    • Disable: No alarm function during the logging process
  • [Alarm delay]: The preset alarm delay interval for a single alarm type is always ZERO. The adjustable alarm delay interval for a cumulative alarm type can be 5 minutes to 2 hours.
  • [Alarm Limits]: Select the alarm threshold values. For example: if 2 - 8 deg C is selected, it means that an alarm condition will occur below 2 deg C, or above 8 deg C. The programmable alarm limit of each parameter is limited to one decimal. To summarize, to activate a alarm when 30 min average value of hazardous alarm delay period is higher than 8°C, please program the alarm delay as 30 mins, alarm type as cumulative and range as 2~8°C
  • [Time zone]: Before programming the logger the user must assure that the PC is set for the correct time zone. The logger will auto synchronize to the time zone of the PC, when Save is pressed. Time zone changes over the transit distance are not adjusted in the logged data.
  • The default value of above parameters are:
    • Sampling rate: 5 min
    • Alarm Type: Disable
    • Start delay: 0 min
    • Alarm delay: 0 min
    • Temp. Unit: °C
    • Alarm Limits: Temp. blank, RH: blank, Pressure: blank
    • Password: disable
    • Company Name: blank
    • Language: English
  • Once all the programming is done, press "Save" to confirm the setting and then you may close the setup window and remove the logger from PC [USB] port.

Step 2: Start logging

  • Press "[START]" key for 5 seconds when you want to start the logging.
  • "[REC]" will appear and flash on LCD to indicate the logging is activated. Measured parameters update and interexchange on LCD every 10 seconds.
  • If Logger is programmed as Start delay, after pressing the start key, the "[REC]" will appear (not flashing) to show the logging is started and in standby status.
  • During the logging, the green LED will blink every 10 seconds if there is no alarm occurred. If any, it turns into flashing red LED.
  • If the red LED is triggered, it won't stop even the reading return to normal range or logging is stopped. The way to stop red alarm LED is to plug logger into computer to generate report or power off the logger.
  • To place a bookmark manually during the delivery transition, long press [MARK] key until measured reading flashes three times.
  • Press [MAX/MIN] key any time to review the maximum and minimum data stored in memory.
  • Press [MARK] key can review the approximate leftover battery life time.

Step 3: Download data

  • Press "[STOP]" key for 5 seconds to stop the logging.
  • Plug logger into PC [USB] port can also stop the logging.
  • Open the file "[PDF Logger Configuration Tool.exe]".
  • Choose function “Convert to PDF” or “Convert to Excel” to generate the report in preferred format. The default language is English. The user may change it to one of several alternate languages.
  • Select the preference location to save the generated report to.
  • The created excel report contains all data shown in PDF report except the graph.
  • NOTE: The generated "excel" file is a tab-delimited ASCII text file which can be easily read by many programs. However, when opening the file in Microsoft Excel, a warning message may be displayed because the file is named ".csv" but the contents are like a ".txt" file. It can be safely opened.

Technical Data

ParameterValue
ModelRHDL30: Temperature + Humidity logger
RHDL40: Temperature +Humidity +Barometric logger
Temp. & Ext.By NTC thermistor
Measurement-30.0~70.0°C (-22.0~158.0°F)
Temp. Resolution0.1°C (0.1°F)
Temp. Accuracy+/-0.5°C
Humidity0.1~99.9%rH
Humidity res.0.1%rH
Humidity accu.+/-3% at 25 °C(10-90%rH), others +/-5%
Pressure300~1100hpa; 8.9~32.5inHg
Resolution0.1hpa;0.1inHg
Accuracy+/-3(0~60°C), +/-5 (-20~0 °C), the rest is not specified.
Logging TypeMultiple use
Sampling pointsRHDL30: 24000 Temp. +24000 Humidity
RHDL40: 16000 Temp.+ 16000 Humidity+ 16000 Baro.
Battery Life3 months
Operating temp.-30~70°C (Logging status); room temp.(PC status)
Operating RH%Humidity < 80%
Storage temp.-40~85°C
Storage RH%Humidity <90%
Weight~90g
Battery2PC 3.0V CR2032
Sampling Interval30 seconds, 5, 10, 30, 60, 90, 120 minutes
Start delay0, 5, 30, 45, 60, 90,120 minutes, 24 hours
Alarm rangeTemp: -30.0~70.0°C
Humidity:1~99%rH
Baro:300~1100hpa
Alarm delay0, 5, 30, 45, 60, 90, 120 minutes
Alarm typeSingle, Cumulative, Disable
Operation keys3 Keys, Start/Stop & Mark & MX/Mn
LED indicatorREC, High /Low alarm
Protection classModel 2: IP65
DirectivesEN12830
Operating SystemWindows only

Error Codes

CodeWhy?Solution
In model 2 only. External probe is unpluggedCheck the contact between probe and logger
E02Measured value is below specified rangePut the logger in specified range
E03Measured value is above specified rangePut the logger in specified range
E04In humidity display of model 3 & 4 only, means temperature is in error modeSolve the error issue of temp. sensor
E11In humidity display of model 3 &4 only, means humidity calibration is failRe-calibrate the humidity
E31Microprocessor is failContact distributor for after service
E33In humidity display of model 3&4 only, means humidity measuring is failContact distributor for after service
Why red LED is flashing but recorded data is OKAlarm detection works every 10 seconds but data logging are possibly programmed as every 2 hours. In this condition, if alarm happens between two logging points, it might lead the recorded data is OK but red LED is flashing, especially if the alarm mode is set as "Single”. To program the alarm function as cumulative can reduce this situation

Warranty Information

Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty.

Copyright © 2020 Triplett

Models: RHDL40-NIST Temperature Humidity and Barometric Pressure Data Logger, RHDL40-NIST, Temperature Humidity and Barometric Pressure Data Logger, Barometric Pressure Data Logger, Pressure Data Logger, Data Logger

File Info : application/pdf, 7 Pages, 968.86KB

PDF preview unavailable. Download the PDF instead.

rhdl40-nist-manual

References

Adobe PDF library 16.07

Related Documents

Preview Triplett VDL48 DC Voltage Datalogger User Manual
Comprehensive user manual for the Triplett VDL48 DC Voltage Datalogger, detailing its features, operation, LCD indicators, specifications, firmware updates, and warranty information.
Preview Triplett RHT37 Dual Input Hygro-Thermometer Psychrometer User Manual
User manual for the Triplett RHT37 Dual Input Hygro-Thermometer Psychrometer, detailing its features, operation, specifications, maintenance, and warranty.
Preview TMDL20 Single Use USB Dry Ice Datalogger for Cold Chain Monitoring
The TMDL20 is a single-use USB dry ice datalogger designed for precise temperature monitoring in extreme cold conditions (-112°F/-80°C). It features a high-precision internal PT1000 sensor, stores up to 8,192 data points, and generates convenient PDF reports. With customizable settings, IP67 waterproof design, and compliance with EN12830:2018 and RTCA DO-160G, it is ideal for pharmaceutical, biotech, cold chain logistics, and food industries.
Preview Triplett RHT01 Humidity/Temperature Pen User Manual
User manual for the Triplett RHT01 Humidity/Temperature Pen, detailing its features, operation, battery replacement, specifications, and warranty information.
Preview Triplett GSM220 CO2 Monitor and Datalogger User Manual
Comprehensive user manual for the Triplett GSM220 CO2 Monitor and Datalogger, detailing its features, operation, setup, calibration, specifications, troubleshooting, and warranty information.
Preview Triplett RHT415 Hygro-Thermometer with Remote Probe User Manual
User manual for the Triplett RHT415 Hygro-Thermometer with Remote Probe, detailing its features, operation, specifications, and warranty. This device measures temperature and humidity with high precision.
Preview Triplett RHT02/RHT05 Hygro-Thermometer Pen User Manual
User manual for the Triplett RHT02 and RHT05 Hygro-Thermometer Pens, detailing features, operation, calibration, troubleshooting, and specifications for measuring humidity and temperature.
Preview Triplett RHT32 Digital Psychrometer User Manual
User manual for the Triplett RHT32 Digital Psychrometer, detailing its features, operation, calibration, maintenance, specifications, and warranty. This device measures humidity, dry bulb, dew point, wet bulb, and temperature, suitable for HVAC engineers.