Neuation iTHERM D150-1 Product User Manual

Neuation iTHERM D150-1 Product User Manual

Brand: Neuation Technologies

1. Introduction

This manual provides important safety information for this Dry Bath. It should be kept near the equipment for quick & easy reference. The Dry bath is equipped with a better display for safe operations. This dry bath has a temperature range from ambient +5°C to 120°C, making it useful for a wide range of applications. A probe is provided for monitoring accurate temperature of samples.

2. Intended Use

The dry bath is intended for use in specific applications such as Enzymatic / kinetic reactions, Denaturation of DNA/RNA and protein samples, Labelling of nucleic acids and proteins, quantification assays, Lysis reactions, Immunoprecipitation, etc.

Note: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repair, please contact the supplier.

3. Features

  • Dry heating block with temperature control range from ambient +5°C to 120°C.
  • Aluminum made blocks, resistant to chemical spillage & temperature shocks.
  • 1 heating block with a long timer range from 1min - 19hr:59mins & infinite mode.
  • Clear display shows critical parameters.
  • Can be used as a bead bath & water bath.
  • Various type tube capacity with convenient lifting gear for cleaning & disinfecting.
  • Rapid heating & constant temperature across the blocks throughout operation.
  • Built-in safe temperature protection.

4. Standard Accessories

  • Power adapter - Input voltage 100-240 VAC 50/60Hz (Attached to the unit) Output voltage 24V 4A.
  • Product user manual & warranty card.
  • Power cord, Dry block, Dry block lifter.

5. Technical Specifications

ParameterValue
Temperature RangeAmbient +5 to 120°C
Temperature Uniformity± 0.5°C
Temperature Stability± 0.5°C
Set-up Block materialAluminium alloy
Number of Blocks1 Nos
DisplayDigital
Timer1min - 19 hr:59mins & infinite
Dimension with Stack Lid190 x 87 x 145 mm
Input Voltage100-240 VAC, 50 / 60Hz
Power Consumption85 W

6. Safety Precautions

  • Never use the instrument in any manner not specified in this manual or by the manufacturer, as this might result in the laps of warranty.
  • Repairs must only be performed by an authorized service technician.
  • Only use recommended original spare parts for best result & product safety.
  • If liquids are spilled in blocks, the instrument must be cleaned carefully and properly before being used again.
  • Prior to operation, the tubes should be visually inspected for material damage.
  • Damaged tubes must not be used, as this can result in sample loss and the contamination of the product.
  • This product must be used for specified applications only. It must not be operated in a hazardous or flammable environment and must not be used to mix explosive or highly reactive substances.
  • Do not place potentially hazardous material within the clearance area/envelope.
  • Wear your personal protective equipment in accordance with the hazards category of the medium to be processed. There is a risk of liquid splashing.
  • Do not move the instrument while it is operating or connected to the main power supply.
  • Do not touch the Block surface when the temperature of the dry bath is over 50°C, as this could result in serious burns or injury. Pay attention to the residual heat after switching off.
  • Properly lift the device with both hands while moving or installing. Also, the device should only be moved from its position once it attains room temperature.
  • Ensure that only closed tubes are used in thermal reactions.
  • Do not lean on the equipment; it may damage the equipment or harm the operator.
  • Do not fill tubes near the device. Liquid spillage may harm the device.
  • In the event of contamination caused by aggressive agents, the device must be cleaned immediately using a natural cleaning liquid. If any damage is seen, contact the service technician.
  • Only process media which will not react dangerously to the extra energy produced by mixing.
  • Do not operate the appliance in an explosive atmosphere with hazardous substances or under water.

7. Installation

Gently remove the upper packaging and take out the dry bath by holding it from the bottom. When this equipment is used for the first time, ensure that all packaging accessories are removed from the product.

Location & Mounting

Place the device on a flat and levelled surface and ensure that all four feet stand on the surface firmly. Avoid installing on a slippery surface or a surface prone to vibration.

Ideal ambient temperature: 20°C ± 5°C; avoid placing the device in direct sunlight.

Clearance: Keep clearance of at least 30 cm from all sides to guarantee cooling efficiency.

Placement: Keep this away from heat or water to avoid sample temperature issues. Do not place the device in such a place that it becomes difficult to operate.

Lid Use: The lid can be used with 15 and 50ml blocks to increase the height of the dry bath, ensuring the closure of the lid to maintain temperature uniformity.

Block Handling: Device blocks can be removed/changed using a simple screwing device into the blocks for lifting/removing/changing the blocks. [Diagram Description: An illustration shows the process of using a screw mechanism to lift and remove blocks from the unit.]

8. Removing & Installing Blocks

Device blocks can be removed or changed by a simple screwing device into the blocks for lifting/removing/changing the blocks. [Diagram Description: An image shows the step-by-step process to remove blocks.] Follow the exact reverse process to insert the blocks back into the dry bath. [Diagram Description: An image shows the step-by-step procedure to replace the blocks back in the unit.]

9. User Interface & Display

Control Panel Operation

[Diagram Description: The control panel features buttons labeled PROBE, MODE, START, STOP, and adjustment buttons (+/-). It also displays Temperature (°C) and TIME. The display shows current settings and operational status.]

Button Functions:

  • START/STOP: Start/stop button for operation start and stop.
  • MODE: 1. Probe / plate mode selection (long press in standby condition). 2. Temp / time change (single press).
  • (+) / (-): Increase and decrease the value of TEMPERATURE/TIME and Prog command.

All LED will blink as per status of command.

10. Operation

Switch On the Device

After connecting the power cord, switch ON the main power supply. Make sure that tubes are placed properly in the wells.

10.1 Start/Stop Operation

  • Single press this button to "START" Operation (as per set Time, Temp & Mode).
  • With the start of operation, temperature will increase on display. Press this button again to "STOP".

10.2 Mode

  • When you select the "MODE" button, we can enable Time and Temperature Mode alternately. To change its value, press (+) and (-) buttons.
  • You can set time from 1 min to 19 hr:59 min or infinite mode.
  • The TIME value will increase by 1 min by single press (+).
  • After adjusting temp & time, the value will be saved after 5 blinks and then it will show your home screen alternatively.

10.2.1 Heating Function

  • You can set Temperature from Ambient + 5°C to 120°C.
  • The temperature value will increase by 0.1°C by single press (+) button & decrease by (-) button. The temperature remains in CELSIUS Unit.
  • When the unit starts, the display shows the default temperature value. By pressing (+) or (-) button, the temp Set value increases or decreases. The Set Value blinks for 5 times & saves the value. After that, you will see "Room temperature" on screen.
  • For fast increment and decrement of temperature setting, long press (+) or (-) button continuously. (After setting the desired value in the window, the value will blink for 5 times and then blink will stop, it means the value is saved.)

Note: Heating function works only when the set temperature is above room temperature.

10.2.2 Probe Selection

  • By long pressing the MODE Key (about 1.5 sec), Probe mode will be activated.
  • Probe LED will flash with the probe mode activation.
  • Without probe mode, the device will automatically work in plate mode. User can change Plate/Probe mode by long pressing MODE button only at standby operation.

10.2.3 Time Function

  • By pressing the MODE button, the time LED will flash. It will show the Time command activation, and we can change time parameters in this situation.
  • "19.59" (19hr:59min) value will display on the window at first time the machine is initiated.
  • To set Time value, press (+) and (-) for increment and decrement of value from 00:01 to 19:59 min. After setting the desired value in the window, the Value will blink for 5 times then stop, it means the value is saved.
  • The time value will increment in 1 minute by single press. Time unit will be in (hour. minute) as shown upon display.
  • In running condition, if time is changed, the timer will reset & count from the beginning. After 19hr:59 mins, there is infinite mode which is shown as " ][ ".

11. Calibration (Probe Mode)

  • Press (MODE) and (+) button together after the set temperature is reached in the display of the Unit.
  • Set reference temperature as per master instrument. The temperature display will blink 5 times and the value is saved.
  • Example: If set value is 100 and reference value is 101.2, then set 101.2 in calibration mode and calibration is done once the value is saved.
  • In case of reset of calibrated value, press [Mode] and (+) together and change the value of "set temperature".

Note: In Probe calibration, for perfect results, it is suggested to calibrate twice.

12. Cleaning & Maintenance

As the dry bath has no mechanical moving parts, it requires very little maintenance and cleaning. Cleaning should be done with mild detergent and water. Unplug the unit from its power source before any cleanup process. In case of any spillage, allow the device to cool before any cleanup or removal of blocks. The holes of the blocks should be regularly cleaned with a damp cloth to ensure the tube is well contacted with the wall to have good heat conduction. Use a slightly damp cloth for cleaning process. Wipe the exterior of the machine with a damp cloth. Cover the samples with the device lid over the block; this cover provides thermal insulation and reduces water condensation.

Note: User is solely responsible for decontamination of the unit in case of spillage of hazardous material.

13. Troubleshooting

Error CodeCauseEffectSolution
Er 01Break in safety circuitHeating offSwitch off device & allow to cool down. Otherwise, contact authorized service person.
Er 02Heating not progressHeating offSwitch off device. Otherwise, contact authorized service person.
Er 03NTC probe sensor failedHeating offSwitch off device. Only to be carried out by authorized service person.
Er 04Probe not in liquid mediumHeating offPut probe in liquid medium.
Er 05Heater NTC 100k failedHeating offSwitch off device. Only to be carried out by authorized service person.

14. Warranty Statement

This product is warranted to be free from defects in material and workmanship for a period of two (2) years from the date of purchase. Your product will be duly repaired upon prompt notification in compliance with the following conditions: This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in this instruction manual. This warranty does not cover damage caused by accident, neglect, misuse, improper service, natural forces, or other causes not arising from defects in original material or workmanship. This warranty does not cover any incidental or consequential damages, commercial loss, or any other damages from the use of this product. All items returned for service should be sent postage prepaid in the original packaging or other suitable carton, padded to avoid damage. This warranty is valid only if the warranty is registered with the supplier within 30 days from the date of purchase.

For your reference, make a note of the serial number, date of purchase and supplier here.

Serial No.: [_______] Purchase Date: [_______] Supplier: [_______]

15. Product Disposal

In case the product is to be disposed of, the relevant legal regulations are to be observed. Information on the disposal of electrical and electronic devices in the European Community. The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2012/19/EU on waste electrical and electronics equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business-to-business sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. They are marked with the following symbol to indicate this. [Symbol Description: A symbol of a crossed-out wheeled bin, indicating separate collection for electrical and electronic equipment.] As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.

Models: iTHERM D150-1, Single Block Dry Bath

PDF preview unavailable. Download the PDF instead.

iTherm-D150-1-Manual Corel PDF Engine Version 17.0.0.491 CorelDRAW X7

Related Documents

Preview Neuation iTherm CH25 Product User Manual: Features, Specifications, and Operation
Comprehensive user manual for the Neuation iTherm CH25 mini incubator. Learn about its features, technical specifications, installation, operation, safety precautions, cleaning, warranty, and disposal. Ideal for laboratory incubation experiments.
Preview Neuation iFuge M15K Centrifuge User Manual - Features, Operation, Safety
Detailed user manual for the Neuation iFuge M15K microcentrifuge. Learn about its features, technical specifications, safe operation, installation, troubleshooting, maintenance, and warranty.
Preview Neuation iSTIR HP320 Product User Manual
User manual for the Neuation iSTIR HP320 Hot Plate Magnetic Stirrer, detailing its features, technical specifications, safety instructions, operation, troubleshooting, and warranty information.
Preview Neuation iFuge D06 User Manual
Comprehensive user manual for the Neuation iFuge D06 centrifuge, detailing operation, safety, maintenance, and technical specifications for laboratory use.
Preview Neuation iTherm CH25 Mini Incubator User Manual
User manual for the Neuation iTherm CH25 mini incubator. Learn about its features, technical specifications, installation, operation, safety precautions, cleaning, warranty, and product disposal for precise temperature control in laboratory incubation experiments.
Preview Neuation iSHAK TS4 NXT Product User Manual
Comprehensive user manual for the Neuation iSHAK TS4 NXT Thermo Shaker, detailing features, specifications, operation, troubleshooting, and safety precautions for laboratory use.
Preview Neuation iFUGE FX2P Micro-Plate Centrifuge User Manual
Comprehensive user manual for the Neuation iFUGE FX2P, a microprocessor-controlled digital centrifuge designed for 96/384 well micro-plates. Covers features, specifications, installation, operation, safety precautions, and troubleshooting.