User Manual for PPI models including: LabCon Recording PC Software, LabCon, Recording PC Software, PC Software, Software

LabCon PC

abc

LabCon PC Operator Manual

Temperature control system | Temperature Recorders India - PPI controller


File Info : application/pdf, 1 Pages, 431.55KB

PDF preview unavailable. Download the PDF instead.

Product-065-OM
LabCon (PC)
Recording + PC Software

OPERATOR PAGE PARAMETERS

Parameters

Settings (Default Value)

Time Start Command >> Time Abort Command >> Time Interval (H:M) >> Ctrl Set Value >>
Ctrl Lo Deviation >> Ctrl Hi Deviation >> Change Password >>

Yes No (Default : No)
0.00 to 500.00 (HH:MM)
(Default : 0.10)
Setpoint LO limit to Setpoint HI limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple) (Default : 25.0)
For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to 99 (Default : 2.0)
For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to 99 (Default : 2.0)
1 to 100 (Default : 0)

SUPERVISORY > SENSOR INPUT

Parameters

Settings (Default Value)

Ctrl Zero Offset >>

-50 to 50 (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple) (Default : 0.0)

SUPERVISORY > RECORDING

Parameters

Settings (Default Value)

Rec. Interval (min) >>

0 to 250 Min (Default : 5 Min)

Alarm Toggle Record >>

Enable Disable (Default : Enable)

Del PC Records >>

Yes No (Default : No)

Del Display Record >>

Yes No
(Default : No)

GSM Machine ID >> (refer note)

1 to 128 (Default : 1)

RTC Time >> RTC Date >>

0 to 23 for Hour 0 to 59 for Min (Default : NA)
1 to 31 (Default : NA)

RTC Month >>

1 to 12 (Default : NA)

RTC Year >> Device Id >>

2000 to 2099 (Default : NA)
1 to 127 (Default : 1)

SUPERVISORY > CONTROL

Parameters

Settings (Default Value)

Tune >>

Yes No (Default : No)

Setpoint LO Limit >>

Min Range for the Selected Input Type to Setpoint HI Limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 0.0)

Setpoint HI Limit >>

Setpoint LO Limit to Max Range for the Selected
Input Type (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple)
(Default : 600.0)

Compressor Setpoint >> Compressor Hyst >>

0 to 100 (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple) (Default : 45.0)
0.1 to 99.9 (Default : 2.0)

Heat Ctrl Action >>

ON-OFF PID
(Default : PID)

Heat Hyst >>

0.1 to 99.9 (Default : 0.2)

Heat Only Control

Heat + Cool Control Zone : Single

Heat + Cool Control Zone : Dual

Proportional Band >> Proportional Band >> PrCozpoPrrtioopnaBl aBnadnd>>for

0.1 to 999.9

0.1 to 999.9

Cool Pre-dominant zone

(Default : 50.0)

(Default : 50.0)

0.1 to 999.9

(Default : 50.0)

Integral Time >> 0 to 3600 sec
(Default : 100 sec)

Integral Time >> 0 to 3600 sec
(Default : 100 sec)

CzInItnetgergarlaTlimTiemfeor>> Cool Pre-dominant zone
0 to 3600 sec (Default : 100 sec)

Derivative Time >>
0 to 600 sec (Default : 16 sec)

Derivative Time >>
0 to 600 sec (Default : 16 sec)

Cz Derivative Time >> Derivative Time for
Cool Pre-dominant zone
0 to 600 sec (Default : 16 sec)

Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)

Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)

PrHozpoPrrtoiopnaBl aBnadnd>>for Heat Pre-dominant zone
0.1 to 999.9 (Default : 50.0)

Overshoot Inhibit >>
Enable Disable

Overshoot Inhibit >>
Enable Disable

HzInIntetgergarlaTlimTiemfeor>> Heat Pre-dominant zone

(Default : Disable)

(Default : Disable)

0 to 3600 sec

(Default : 100 sec)

Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)

Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)

Hz Derivative Time >> Derivative Time for
Heat Pre-dominant zone 0 to 600 sec
(Default : 16 sec)
Cycle Time >> 0.5 to 100.0 sec (Default : 10.0 sec)
Overshoot Inhibit >> Enable Disable
(Default : Disable)

Cutoff Factor >> 1.0 to 2.0 sec
(Default : 1.2 sec)

SUPERVISORY > PASSWORD

Parameters

Settings (Default Value)

Change Password >>

1000 to 1999 (Default : 123)

SUPERVISORY > EXIT

Parameters

Settings (Default Value)

Exit Setup Mode >>

Yes No
(Default : No)

FACTORY > CONTROL SENSOR INPUT

Parameters

Settings (Default Value)

Input Type >>

Refer Table 1 (Default : RTD Pt100)

Signal LO >>

Input Type 0 to 20mA 4 to 20mA
0 to 5V 0 to 10V 1 to 5V

Settings 0.00 to Signal High 4.00 to Signal High 0.000 to Signal High 0.00 to Signal High 1.000 to Signal High

Default 0.00 4.00 0.000 0.00 1.000

Signal HI >>

Input Type 0 to 20mA 4 to 20mA
0 to 5V 0 to 10V 1 to 5V

Settings Signal Low to 20.00 Signal Low to 20.00 Signal Low to 5.000 Signal Low to 10.00 Signal Low to 5.000

Default 20.00 20.00 5.000 10.00 5.000

Range LO >> Range HI >>

-199.9 to RANGE HI (Default : 0.0)
RANGE LO to 999.9 (Default : 100.0)

FACTORY > ALARM PARAMETERS

Parameters

Settings (Default Value)

Hysteresis >>

0.1 to 99.9 (Default : 0.2)

Inhibit >>

Yes No
(Default : Yes)

Operation Manual
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist. Palghar - 401 210. Sales : 8208199048 / 8208141446 Support : 07498799226 / 08767395333 E: sales@ppiindia.net, support@ppiindia.net

FACTORY > HEAT COOL SELECT

Parameters Control Strategy >>

Settings (Default Value)
Heat Only Cool Only Heat + Cool (Default : Heat + Cool)

Control Strategy : Cool Only

Time Delay (sec) >>

0 to 1000 Sec (Default : 200 Sec)

Control Strategy : Heat + Cool

Compressor Strategy >>

CONT. OFF CONT. ON SP BASED
PV BASED (Default : CONT. ON)

CONT. ON Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)

SP BASED Boundary Set Value >>
0 to 100 (Resolution 0.1°C for RTD / DC Linear & 1°C
for Thermocouple) (Default : 45.0)

PV BASED Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)

Control Zones >> Single Dual
(Default : Single)

Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)

FACTORY > SOAK TIMER PARAMETERS

Parameters

Settings (Default Value)

Enable >>

No Yes (Default : No)

Start Band >>

0 to 999.9 (Default : 0.5)

Holdback Strategy >>

None Up
Down Both
(Default : None)

Hold Band >> Heat Off >>

0.1 to 999.9
(Default : 0.5)
No Yes (Default : No)

Cool Off >>

No
Yes (Default : No)

Power Recovery >>

Abort Restart Continuous (Default : Restart)

FACTORY > DOOR OPEN

Parameters

Settings (Default Value)

Enable >>

Yes No (Default : No)

Switch Logic >> Door Alrm Dly (sec) >>

Close : Door Open Open : Door Open (Default : Close : Door Open)
0 to 1000 Sec (Default : 60 Sec)

FACTORY > MAINS FAILURE

Parameters

Settings (Default Value)

Enable >>

Yes No
(Default : No)

Switch Logic >>

Close : Mains Fail
Open : Mains Fail (Default : Close : Mains Fail)

FACTORY > SMS ALERT

Parameters

Settings (Default Value)

Enable >> (refer note)

Yes No (Default : No)

Reset GSM Module >> (refer note)

Yes No
(Default : No)

Note : These parameters are available only if the controller supplied with GSM module version.

FACTORY > PASSWORD

Parameters

Settings (Default Value)

Change Password >>

2000 to 2999 (Default : 321)

FACTORY > FACTORY DEFAULT

Parameters

Settings (Default Value)

Set To Default >>

Yes No
(Default : No)

FACTORY > EXIT

Parameters

Settings (Default Value)

Exit Setup Mode >>

Yes No
(Default : No)

PV ERROR INDICATIONS

Message
OPEN OVR UNR

Error Type

Cause

Sensor Open Sensor (RTD Pt100) Broken / Open

Over-range Temperature above Max. Specified Range

Under-range Temperature below Min. Specified Range

What it means Type J Thermocouple Type K Thermocouple Type T Thermocouple Type R Thermocouple Type S Thermocouple Type B Thermocouple Type N Thermocouple

TABLE 1
Range (Min. to Max.) 0 to +960°C
-200 to +1376°C -200 to +385°C 0 to +1770°C 0 to +1765°C 0 to +1825°C 0 to +1300°C

Resolution
Fixed 1°C

Reserve

Reserved for customer specific Thermocouple type not listed above. The type shall be specified in accordance with the ordered (optional on request) Thermocouple type.

3-wire, RTD Pt100
0 to 20mA DC current 4 to 20mA DC current 0 to 5.0V DC voltage 0 to 10.0V DC voltage 1 to 5.0V DC voltage

-199.9 to 600.0°C -199.9 to 999.9 units

Fixed 0.1°C Fixed 0.1 unit

FRONT PANEL KEYS

Symbol

Key

Scroll
Alarm Acknowledge

DOWN

UP

Function
Press to scroll through various Process Information Screens in Normal Operation Mode.
Press to acknowledge and mute (if active) alarm output.
Press to decrease the parameter value. Pressing once decreases the value by one count; keeping pressed speeds up the change.
Press to increase the parameter value. Pressing once increases the value by one count; keeping pressed speeds up the change.

SET-UP ENTER

Press to enter or exit set-up mode.
Press to store the set parameter value and to scroll to the next parameter.

ELECTRICAL CONNECTIONS

SOAK START ACK
BATTERY

01 234
COMMON DOOR

RS485

Control Outputs

COMM HTR CMP ALM
+DEV- +PC- + - + - + -

9 10 11 12 13 14 15 16 17 18

LabCon (w/o map) Configuration PC Printer Recording

5678

Pt100

++mVA +

-

TC
Supply

18~32VDC
+-
19 20 21

Jan 2022



References

Corel PDF Engine Version 17.6.0.1021 CorelDRAW X7