INSTRUCTION MANUAL
DIGITAL MULTIMETER
KEW1021R

LCD SELECT Key

HOLD / Back light Key REL / MAXMIN Key

Function Switch

Current Terminal COM Terminal Cap

Voltage Terminal Protect fingerguard

Protective figerguard: It is a part providing protection against electrical shock and ensuring the minimum required air and creepage distances. Cap: Test leads can be used under the CAT II and CAT III and CAT IV environments by attaching a protective cap as illustrated below. Use of our protective cap offers different lengths suitable for the test environments.
When the instrument and the test lead are combined and used together, whichever lower category either of them belongs to will be applied. This instrument is designed for CAT IV 300V/ CAT III 600V. Test leads
M-7066A with the supplied caps are designed for CAT IV600V/ CAT
III 1000V and without the caps are for CAT II 1000V.

1. Safety Warnings
This instrument has been designed, manufactured and tested according to IEC 61010: Safety requirements for electrical equipment, and delivered in the best condition after passing quality control tests. This instruction manual contains warnings and safety rules which have to be observed by the user to ensure safe operation of the instrument and to maintain it in safe condition. Therefore, read through these operating instructions before using the instrument.

WARNING
 Read through and understand the instructions contained in this manual before using the instrument.
 Keep the manual at hand to enable quick reference whenever necessary.
 The instrument is to be used only in its intended applications.  Understand and follow all the safety instructions contained in the
manual. Failure to follow the instructions may cause injury, instrument damage and/or damage to equipment under test. Kyoritsu is by no means liable for any damage resulting from the instrument in contradiction to this cautionary note.
The symbol indicated on the instrument means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.

DANGER is reserved for conditions and actions that are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause injury or instrument damage.

 Symbols listed below are used on this instrument. User must refer to the manual. Instrument with double or reinforced insulation.

AC

DC

Ground (Earth)

This instrument complies to WEEE Directive (2002/96/EC). Please contact your local distributor at disposal.

Measurement Category
O Circuits which are not directly connected to the mains power supply.
CAT II Primary electrical circuits of equipment connected to an AC electrical outlet by a power cord.
CAT III Primary electrical circuits of the equipment connected directly to the distribution panel, and feeders from the distribution panel to outlets.
CAT IV The circuit from the service drop to the service entrance, and to the power meter and primary over current protection device(distribution panel).
DANGER
 Never make measurements under the circumstances exceeding the designed measurement category and the rated voltage of the instrument.
 Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
 Never attempt to use the instrument if its surface or your hand is wet.  Do not exceed the maximum allowable input of any measuring
range.  Never open the Battery compartment cover during a measurement.  To avoid electrical shock by touching the equipment under test or its
surroundings, be sure to wear insulated protective gear.  Test leads to be used for voltage measurements shall be rated as
appropriate for Measurement Category III or IV according to IEC 61010-031 and shall have a voltage rating of 600V or higher.  Protective fingerguard on the test leads provide protection to keep your fingers and hands from touching an object under test. Keep your fingers and hands behind the protective fingerguard during measurement.
WARNING
 Never attempt to make measurement if any abnormal conditions, such as broken case and exposed metal parts are found on the instrument or test leads.
 Verify proper operation on a known source before use or take action as a result of the indication of the instrument.
 Firmly attach the caps to the test leads when performing measurements in CAT III or higher test environments. When KEW1021R and the test leads are combined and used together, whichever is lower category & voltage to earth either of them belong to is applied.
 Do not rotate the Function Switch if the instrument and the equipment under test are connected.
 Do not install substitute parts or make any modification to the

instrument. For repair or re-calibration, return the instrument to your
local KYORITSU distributor.
Stop using the test lead if the outer jacket is damaged and the inner
metal or color jacket is exposed.
CAUTION
 Use of this instrument is limited to domestic, commercial and light industry applications. Strong electromagnetic interference or strong
magnetic fields, generated by large currents, may cause malfunction
of the instrument.
 Firmly insert the test leads.  Do not pull or twist the test leads to prevent the risk of damage.  Power off the instrument after use. Remove the battery if the
instrument is to be stored and will not be in use for a long period.
 Do not expose the instrument to direct sunlight, high temperature and humidity or dewfall.
 Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
NOTE
 The LCD shows some digits at the Voltage or the Current range even while the test leads are open. And, it may show some digits instead of 0 even if the test leads are shorted. However, these phenomena don't affect measurement results.
 A resistance measurement takes time to settle the reading if there is a high resistance or capacitance components.

2. Specification

 Accuracy (Temperature: 23 ± 5°C, Humidity: 45 - 75%)

ACV / RMS

Range

Display Range

6V

0.000, 0.006 - 6.299V

60V

5.70 - 62.99V

600V 57.0 - 629.9V

Guaranteed accuracy : 0.01V-600V,

Input impedance : approx. 10M

(Auto Range) Accuracy (sine wave)
±1.0 %rdg±3dgt (40-500Hz)
less than 900V peak

Hz Frequency - ACV measurement

Range

Display Range

99.99Hz

10.00 - 99.99Hz

999.9Hz

95.0 - 999.9Hz

9.999kHz 0.950 - 9.999kHz

99.99kHz 95.0 - 99.99kHz

Guaranteed accuracy : 10Hz-99kHz

(Auto Range) Accuracy (sine wave)
±0.1%rdg±3dgt

% DUTY - ACV measurement

Range

Display Range

99.9 %

0.0 ­ 99.9 %

Guaranteed accuracy : 10%-90%

Accuracy (Square wave)
±1.0%rdg±3dgt (50/60Hz)

DCV

(Auto Range)

Range

Display Range

Accuracy

6.000V

0.000 - ±6.299V

60.00V

±5.70 - ±62.99V

±0.5 %rdg±3dgt

600.0V

±57.0 - ±629.9V

Guaranteed accuracy : 0V-±600V

Input impedance : approx. 11M(6V range) / 10M(60/600V range)

ACmV / RMS

Range

Display Range

600.0mV

0.0, 0.9 - 629.9mV

Guaranteed accuracy : 1.2mV-600mV, Input impedance : approx. 900k

Accuracy (sine wave) ±2.0 %rdg±3dgt (40-500Hz)
less than 900mV peak

AC Clamp Sensor / RMS

(Auto Range)

Range

Display Range

Accuracy (sine wave)

60.00A 200.0A

0.00, 0.09 - 62.99A 57.0 - 209.9A

±2.0 %rdg±3dgt + Sensor accuracy (40-500Hz)

Direct reading from 10mV / A output Clamp sensor

Guaranteed accuracy : 0.12A-200A, less than 300A peak

Input impedance : approx. 900k

DCmV

Range

Display Range

600.0mV

0.0 - ±629.9mV

Guaranteed accuracy : 0mV-±600mV,

Accuracy ±1.5 %rdg±3dgt
Input impedance : approx. 900k

DC Clamp Sensor

Range

Display Range

60.00A

0.00 - ±62.99A

(Auto Range) Accuracy ±1.5 %rdg±3dgt

200.0A

±57.0 - ±209.9A

+ Sensor accuracy

Direct reading from 10mV / A output Clamp sensor

Guaranteed accuracy : 0A-±200A, Input impedance : approx. 900k

Resistance

(Auto Range)

Range

Display Range

Accuracy

600.0

0.0 - 629.9

±0.5 %rdg±4dgt

6.000k

0.570 - 6.299k

60.00k 600.0k

5.70 - 62.99k 57.0 - 629.9k

±0.5 %rdg±2dgt

6.000M

0.570 - 6.299M

40.00M

5.70 - 41.99M

±1.5 %rdg±3dgt

Guaranteed accuracy : 0-40M, Open-loop Voltage : less than 3V

Measurement current : less than 1mA

Continuity

Range

Display Range

Accuracy

600.0

0.0 - 629.9

Buzzer threshold value : less than 90

Open-loop Voltage : less than 3V, Measurement current : less than 1mA

Diode

Range 2.000V

Display Range 0.000 - 2.099V

Accuracy ±5 %rdg±5dgt

Guaranteed accuracy : 0V-2V, Open-loop Voltage : less than 3V

Measurement current : approx.0.5mA (Vf=0.6V)

Capacitance

(Auto Range)

Range

Display Range

Accuracy

60.00nF 600.0nF

0.00 - 62.99nF 57.0 - 629.9nF

±2.0 %rdg±5dgt*

6.000µF

0.570 - 6.299µF

60.00µF 600.0µF

5.70 - 62.99µF 57.0 - 629.9µF

±5.0 %rdg±5dgt

1000µF

570 - 1049µF

* Accuracy after canceling floating capacitance using REL function.

Guaranteed accuracy : 0nF-1000µF

ACA / RMS

(Auto Range)

Range

Display Range

Accuracy (sine wave)

6.000A

0.000, 0.006 - 6.299A

±1.5 %rdg±3dgt

10.00A

5.70 - 10.49A

(40-500Hz)

Guaranteed accuracy : 0.01A - 10A, less than 15Apeak

Hz Frequency - ACA measurement

Range

Display Range

99.99Hz

10.00 - 99.99Hz

999.9Hz

95.0 - 999.9Hz

9.999kHz 0.950 - 9.999kHz

Guaranteed accuracy : 10Hz-9.9kHz

(Auto Range) Accuracy
±0.1 %rdg±3dgt

% DUTY - ACA measurement

Range

Display Range

Accuracy

99.9 %

0.0 ­ 99.9 %

±1.0%rdg±3dgt (50/60Hz)

Guaranteed accuracy : 10%-90% (Square wave)

DCA

(Auto Range)

Range 6.000A 10.00A

Display Range 0.000 - ±6.299A ±5.70 - ±10.49A

Accuracy (sine wave) ±1.5 %rdg±3dgt

Guaranteed accuracy : 0A-±10A

Measuring method :  modulation Over-range indication : OL Measurement cycle : 2.5 times per second
(1000µF range of Capacitance function 0.05 times per second)
Crest factor : less than 3 (45-65Hz) For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt
(Applicable functions : ACV, ACmV, AC clamp sensor, ACA)
Applicable standards : IEC 61010-1 /61010-2-033 : CAT IV 300V / CAT III 600V
Pollution degree 2, Indoor use, Altitude up to 2000m
IEC61010-31 (Test leads Model 7066A)
IEC 61326 (EMC) , EN 50581 (RoHS)
Withstand voltage : AC5160Vrms 5sec between circuit and enclosure IP rating : IP40 (IEC60529) Insulation resistance : 100M or more /1000V between enclosure
and electrical circuit
Operating temperature and humidity range : 0 to 40°C, 80%RH or less (no condensation)
Storage Temperature and humidity range : -20 to 60°C, 80%RH or less (no condensation)
Power source : DC3V R03/LR03 (AAA) × 2 Current consumption : 3mA or less Battery life : Approx. 200 hours (ACV, continuous, no load, with R03) Dimension, Weight : 155(L)×75(W)×40(D)mm,
approx. 250g (including batteries and Wing-type holder)
Accessories : Test leads (M-7066A), soft case(M-9097) Instruction manual, 10A/600V Fuse (M-8919, included)

Battery R03 (AAA) 2pcs Flat-type holder, Wing-type holder Option : Magnet hanger strap (M-9189)
Test leads with alligator clips (M-7234) AC Clamp Sensor (KEW8161), AC/DC Clamp Sensor (KEW8115)
3. Other Functions
 REL Function Press the REL key to enable this function and store the measured value to display the differences between the stored value and the values measured in further tests. The measurement range will be fixed when the REL function is enabled, and the measuring range will be between the initial value and the full scale value. Press the REL Key again to release the stored value.
"" appears and "AUTO" disappears when REL Key is pressed.

One press: REL ON Another press: REL OFF
*To activate REL function, MAX/MIN function should be disabled.  Max / Min value display function
This function is to display the measured max and min values on the LCD during a measurement. Press the MAX/MIN key 1 sec or longer to start recoding of max and min values. Then the LCD shows the latest max value. After that, the min and present values can be toggled and checked by pressing the MAX/MIN key (less than 1 sec). To disable this function, press the MAX/MIN key 1 sec or longer.

Max value

Min value

Present value

Less than 1 sec

Less than 1 sec

Less than 1 sec

* To activate MAX/MIN function, REL function should be disabled.

 Data hold function

Press the HOLD key (less than 1 sec). The LCD shows "H" mark and

the reading will be held (Data hold mode). The max/ min values are

not updated in MAX/MIN mode. Press the HOLD key

again (less than 1 sec) to release the display.  Backlight function

Press the Backlight key 1 sec or longer to turn

on the backlight. Press the Backlight key another 1 sec or longer to

turn it off. The light automatically turns off in 1 min.

 Low battery indication

The LCD shows "

" mark when the batteries fall below the

normal operating voltage.

Replace the batteries with new ones when this mark appears.

 Sleep Function Automatically powers off the instrument in about 15 min after the last switch operation. To exit from the Sleep mode, rotate the Function switch or press any key. To disable the Sleep function, press the HOLD/Backlight key and power on the instrument. Confirm that the LCD shows "P.OFF" about 1 sec.
How to store test leads Attach the Wing-type holder to the back of the instrument to store the test leads. 1. Attach the Wing-type holder to the back of the instrument. 2. Fit the barrier on the test lead into the groove between the instrument and the Wing-type holder.

Flat-type Wing-type

1. holder

holder

2.

3.

3. Then it will be easier to see the displayed readings during a measurement.
 Other convenient functions

Magnet hanger strap (option)

Tilt stand
4. ACV / DCV / ACmV / DCmV Measurement
DANGER
 Before starting a measurement, always check and confirm the Function switch is in the appropriate measurement position and the test leads are connected to the proper input terminals.
 Never make measurement on a circuit in which voltage over 600V exists.
 Keep your fingers and hands behind the protective fingerguard during measurement.
4.1 ACV / DCV/ ACmV / DCmV measurement
(1) Set the Function switch to ACV, DCV, ACmV or DCmV position. For frequency or DUTY measurement, set the switch to ACV and press the SELECT key.

Default : ACV

DUTY

Frequency

Press the SELECT key to toggle functions

(2) Connect the test leads to the Voltage and COM terminals.

ACV / ACmV

DCV / DCmV

Black

Red

4.2 Clamp sensor (option) measurement
(1) Set the Function switch to ACmV or DCmV position and press the SELECT key. LCD shows "SENSOR".

Default :ACmV Default :DCmV

Sensor mode (ACA)
Sensor mode (DCA)

Press the SELECT key to toggle functions

(2) Connect the clamp sensor to the Voltage and COM terminals.

AC clamp sensor

DC clamp sensor

 The sensor mode corresponds to direct reading with 10mV/A output clamp sensor.
5. Resistance / Diode / Continuity / Capacitance Measurement
WARNING
Never use the instrument on an energized circuit. Discharge the capacitor before starting a capacitance measurement.
(1) Set the Function switch to the resistance position. For continuity check or diode/ capacitance measurement, press the SELECT key.

Default:Resistance

Capacitance

Diode

Continuity

Press the SELECT key to toggle functions

(2) Connect the test leads to the Voltage and COM terminals.

Resistance or Capacitance

Continuity

Diode

Black

Red



NOTE
 LCD shows "OL" when the test leads are open. (except for capacitance measurement)
 The LCD shows "OL" if the test lead connection is reversed for diode measurement.
 Measurement time on 600F/1000F range is a bit long. (20sec max.)

6. ACA / DCA Measurement
WARNING
 The value for the maximum input current on ACA and DCA ranges is 10A (protected by fuse). Do not apply current that exceeds 10A.
 If measuring a current of 6A or higher, the duration of measurement should be within 2 min. After that, a rest period of 10 min is required.
(1) Set the Function switch to ACA or DCA position. For frequency/ DUTY measurement, press the SELECT key while the Function switch is in ACA position.

Default : ACA

DUTY

Frequency

Press the SELECT key to toggle functions (2) Connect the test leads to the Current and COM terminals.

Black

Red

NOTE  If the connection is reversed, the "  " mark will be displayed on the
LCD. (DCV measurement).
 Press the REL key to adjust the reading on the DC clamp sensor to "0".

Red

Black

NOTE  If the connection is reversed, the "  " mark will be displayed on the
LCD. (DCA measurement).
 If the fuse is blown, please refer to 7. Fuse Replacement and replace it with a new one.

7. Battery / Fuse Replacement

WARNING

 Replace the battery when the "

" mark- low battery voltage

warning- appears on the LCD.

Otherwise, precise measurement cannot be made. If the battery

is completely exhausted, the LCD goes blank without showing

the "

" mark.

 Disconnect the test leads from the object under test and power

off the instrument before opening the Battery compartment

cover for battery or fuse replacement.

 The fuse gets hot after current measurement and may cause

burn injury. If replacing the batteries or fuse after current

measurement, a rest period of 10 min before you do.

 Use the designated fuse only.

 Do not try to replace the battery or fuse if the surface of the

instrument is wet.

CAUTION
 Do not mix old and new batteries.  Install a batteries in correct polarity as indicated in the
Battery Compartment.

(1) Set the Function Switch to "OFF" position. (2) Untighten the screw on the back of the instrument. (3) Remove the Battery compartment cover and replace the
batteries or fuse. (4) Attach the cover to the instrument and then secure the cover
by tightening the screw.

Battery

Screw

Fuse (10A/600V, size : 6.3 × 32mm, Fast Acting type)