Sebastien Gallien1,2, Jing Wang2, Aaron S. Gajadhar3, Bhavin Patel4, Markus Kellmann5, Tabiwang N. Arrey5, Alexander Harder5, Romain Huguet5, Graeme McAlister3, Derek Bailey3, Shannon Eliuk3, Yue Xuan5, Andreas Huhmer3, Emily L. Chen2
A novel global plasma proteome quantification workflow has been developed. It relies on a new high-density targeted acquisition method, called 'SureQuant' method, implemented on next-generation Orbitrap mass spectrometers: Thermo Scientific™ Orbitrap Exploris™ 480 mass spectrometer and Thermo Scientific™ Orbitrap Eclipse™ Tribrid™ mass spectrometer. In addition, the workflow leverages a large set of stable isotopically labeled (SIL) peptides, used as internal standards, to drive the systematic screening of more than 500 plasma proteins (PQ500 kit from Biognosys).
The SureQuant PQ500 analyses of undepleted plasma samples were performed using Orbitrap Exploris 480 and Orbitrap Eclipse Tribrid mass spectrometers, coupled to Thermo Scientific™ UltiMate™ 3000 RSLC system or Thermo Scientific™ EASY-nLCTM 1200 system. An acquisition variant of the internal standard triggered targeted acquisition, dedicated to higher throughput analyses, was executed through an application programming interface (API) controlling a Thermo Scientific™ Q Exactive™ HF-X hybrid quadrupole-Orbitrap™ mass spectrometer coupled to a Evosep One LC system. PQ500 kit from Biognosys was spiked into the plasma samples to drive the targeted acquisition.
The SureQuant method has been adapted from the internal standard triggered PRM (IS-PRM) method to improve its usability and robustness, while retaining its analytical performance, and has been implemented in the native instrument control software of next-generation Orbitrap mass spectrometers. The application of the method to the analysis of plasma samples, leveraging spiked-in SIL peptides from PQ500 kit, allowed robust and precise quantification of around 560 endogenous peptides used as surrogate of 400 plasma proteins using a 70-min LC gradient. Therefore, the method combined data quality of targeted analyses with proteome coverage of state-of-the-art profiling experiments. In addition, the availability of pre-set PQ500 SureQuant methods embedded into the instrument control software enables a 'load and play' execution of such plasma proteomics experiments.
The success of plasma proteomics studies to deliver useful protein biomarkers has remained lower than initial expectations, prompting a redefinition of the plasma biomarker development pipeline. One proposed measure consisted in the implementation of a "rectangular" approach, relying exclusively on broad proteome profiling (through DDA or DIA analyses) across large cohorts for both the discovery and validation/verification stages. This document describes an alternative global quantification workflow, relying on a high-density targeted acquisition method implemented on next-generation Orbitrap mass spectrometers. This adaptation of the IS-PRM method leverages spiked-in stable isotopically labeled (SIL) peptides to drive the systematic screening of more than 500 plasma proteins per analysis across large sample sets.
A commercial human plasma sample pooled from several individuals purchased from BioreclamationIVT was processed using Thermo Scientific™ EasyPep™ Mini MS Sample Prep Kit.
A set of 804 SIL peptides from PQ500 kit (Biognosys PN#Ki-3019-96) was spiked at around 80 fmol (median value) into 1 µg of undepleted plasma sample. One µL of the sample was injected for LC-MS/MS analyses.
For SureQuant analyses, chromatographic separations were performed on UltiMate 3000 RSLC system equipped with C18 trap cartridge (5 µm, 0.3 x 5 mm) and analytical column (2 µm, 0.15 x 150 mm), and EASY-nLC 1200 system equipped with a C18 analytical column (2 µm, 0.15 x 150 mm), using a 70-min gradient separation. The LC systems were coupled to Orbitrap Exploris 480 and Orbitrap Eclipse Tribrid mass spectrometers. In SureQuant analyses, Full MS scans were acquired at a resolution of 120k with a maximum fill time of 50 ms, while ddMS2 of IS and endogenous peptides were acquired sequentially at a resolution of 7.5k or 60k and maximum fill times of 10 ms and 116 ms, respectively.
The analyses using the high throughput IS-PRM acquisition variant were performed on Evosep One LC system equipped with Evosep C18 analytical column (3 µm, 0.1 x 80 mm), using a 21-min gradient separation. The LC system was coupled to a Q Exactive HF-X controlled by an API. In the analyses performed with the 'Single Scan Msx IS-PRM' method (Figure 1, right panel), PRM scans in watch mode were acquired at a resolution of 7.5k and a maximum fill time of 10 ms. In the quant. mode, the pairs of IS and endogenous peptides were measured simultaneously by 'multiplex' acquisition of their fragments in a single PRM scan generated at a resolution of 120k and non-isochronous maximum fill times of 20 or 124 ms for the IS and endogenous peptides, respectively.
This figure illustrates two acquisition modes: Sequential acquisition and Multiplex acquisition, each followed by Data Analysis. Both modes show a workflow from Isolation to Accumulation to Detection, with HCD (Higher-energy collisional dissociation) applied. The diagrams depict peptide heavy/light (H/L) ratios and m/z values. The Multiplex acquisition mode includes "Non-Isochronous fill times" and "Intensity correction," indicating simultaneous measurement of IS and endogenous peptides, contrasting with the sequential collection of data points over full elution profiles.
The developed SureQuant acquisition method includes different scan events and filters, with parameters optimized for the highest data quality and sensitivity of triggering, but also to retain sufficient triggering specificity (Figure 2). During SureQuant analyses, a high resolution (120k) full scan MS (1) is acquired, using lock mass re-calibration, to monitor the predefined optimal precursor ions of the IS, based on the list of associated m/z values and intensity thresholds (defined at 1% of the expected MS1 intensity at the chromatographic peak apex) included in the 'Targeted Mass' filter (2). In case of the detection of a candidate precursor ion satisfying the filtering criteria (MS peak with an m/z value within 3-10 ppm tolerance and intensity exceeding the threshold), it is subjected to fast MS/MS acquisition (3) performed at low resolution and short fill times (typically 7.5k/10 ms), which is used as a second layer of confirmation of the detection of the IS. Several fragment ions, predefined as associated with the IS (6 optimal frag.) are included in the 'Targeted Mass Trigger' filter (4), and the detection of a subset of them (at least 5 frag., 10-20 ppm tolerance) confirms the actual elution of the IS. This two-step process interrogating MS1 and MS2 data to track IS elution is the 'watch' mode of the SureQuant method. The detection of an IS in watch mode triggers the 'quant' mode, and therefore MS/MS acquisition of the corresponding endogenous pept. (5) with parameters favoring data quality, i.e., high resolution and long fill times (typically 60k/116 ms). The process is repeated over the entire LC separation with a systematic cycle-to-cycle assessment of the elution of the IS.
The development of a SureQuant assay to be associated with a custom peptide panels requires the preliminary determination of the optimal precursor ion of each peptide target together with their intensity response and optimal associated fragment ions. This information is typically extracted from an initial directed DDA analysis of the corresponding IS peptide mixture spiked into a representative matrix. The target list of multiple theoretical precursor ions of each IS peptide included in such survey runs enables comprehensive data collection, and therefore the refinement of analytes definition for subsequent SureQuant analyses. The process is supported by Skyline software and facilitated by the generic data acquisition and processing templates provided. The transfer of the generated information can be achieved through copy/paste from Skyline reports to 'Targeted Mass' and 'Targeted Mass Trigger' tables of SureQuant method. The survey analysis of a SureQuant assay needs to be run one time as the onset of a project without further adjustment over time, unless a major change in the experimental design is applied (e.g., LC set up, spiked amount of SIL peptides). In the latter case, only the intensity responses of predefined IS precursor ions need to be re-adjusted though an additional survey run performed under the modified experimental conditions. Therefore, the ability of embedding pre-set methods associated with predefined kits of IS peptides expedites SureQuant assay development at the user sites. Such validated method templates have already been included in the instrument control software for the PQ500 related methods (Figure 3).
This diagram illustrates the SureQuant acquisition scheme. It begins with SIL peptides (internal standards) mixed with a sample, undergoing LC-MS Analysis leading to MS OT (Orbitrap Mass Spectrometry). The workflow proceeds through several steps: 1) Full scan MS acquisition, 2) Targeted Mass filter for optimal precursor ions, 3) Fast, low-resolution MS/MS acquisition (ddMS² OT HCD) for IS confirmation, 4) Targeted Mass Trigger for IS fragment ions, and 5) High-quality MS/MS acquisition (ddMS² OT HCD) for endogenous peptides. This process involves a "WATCH MODE" (MS1 Survey Scan, IS peptides trigger, Fast, Low Res MS2 of Trigger, Pseudo Spectral Matching) and a "QUANT MODE" (High Quality MS2 of Target).
This flowchart outlines the PQ500 SureQuant acquisition workflow. It starts with a PQ500 kit in plasma, leading to LC-MS Analysis and a Survey Run. The Survey Run Analysis uses Skyline PQ500 analysis template, generating a PQ500 Directed DDA (Data Dependent Acquisition) with inclusion list optimal IS precursor ions. This exports a Targeted Mass report, which provides intensity response of precursor ions. This information is then copied/pasted to update the Targeted Mass intensity threshold for a PQ500 kit Plasma sample. This updated sample then undergoes LC-MS Analysis for Target Quantification, utilizing Skyline PQ500 SureQuant and a PQ500 Skyline analysis template. The process highlights the "Ready-to-use template" and "Custom Templates" options.
The PQ500 SureQuant acquisition workflow has been applied to the analyses of 1 µg of undepleted plasma sample supplemented with PQ500 kit (median spiked-in amount of SIL peptides of 80 fmol) using a 70-min LC gradient on UltiMate 3000 RSLC system. The survey run revealed that 802 of 804 SIL peptides were compatible (intensity response and hydrophobicity) with the LC-MS setup used (Figure 4), and were retained for subsequent SureQuant analyses. The high overall intensity response of the SIL peptides (median MS1 intensity at chromatographic peak apex of 1e8) insured high acquisition efficiency (>90%) in SureQuant analyses, benefiting from the median MS1 triggering intensity threshold at 1e6. This translated into a good overall sampling rate over the LC separation, exceeding 8 data points collected over the elution profile of the vast majority of the peptides, which also remained acceptable in the most crowded time ranges, with 6 to 8 data points collected per peak (Figure 5). The 802 retained internal standards were systematically detected in the triplicated SureQuant analyses of plasma samples, and triggered high quality measurement of the corresponding endogenous peptides, translating into a broad coverage of the undepleted plasma proteome (Figure 6). Around 400 proteins were detected in each replicate (based on around 560 surrogate endogenous peptides), while 90% of the total sets were systematically quantified across triplicates.
This bar chart shows the Injected Amount of Internal Standard (IS) in fmol, ranging from 1.0 to 10,000.0 fmol, with a median amount of 80 fmol. The x-axis represents Peptide# from 1 to 802. A second bar chart shows the MS1 Intensity Apex of IS, ranging from 1.0E+06 to 1.0E+09, with a median intensity of 1E8. Both charts illustrate the distribution and consistency of the PQ500 kit peptides in the survey run.
This line graph displays Cycle Time in seconds (s) on the y-axis (0 to 4 s) against time in minutes (min) on the x-axis (0 to 90 min). The sampling rate is indicated as 6-8 data points/peak (20-s peak width) and ≥8 data points/peak (20-s peak width), showing the efficiency of data collection over the LC separation time.
This figure presents bar charts and pie charts illustrating peptide detection and quantification. The bar charts show the "Number of spectra including ≥ 5 ref. fragments" (0, 1-2, ≥ 3) for Internal Standards peptides and Endogenous peptides across three replicates (Rep. 1, Rep. 2, Rep. 3). For Internal Standards, 802 peptides were detected in each replicate. For Endogenous peptides, Rep. 1 detected 564 peptides / 397 proteins, Rep. 2 detected 560 peptides / 396 proteins, and Rep. 3 detected 550 peptides / 390 proteins. The pie charts summarize the total peptides and proteins detected: Rep 1 shows 526 peptides, Rep 2 shows 368 proteins, and Rep 3 shows 589 total peptides and 421 total proteins, indicating the overall detection and quantification performance.
In addition to delivering a broad plasma proteome coverage, rivalling that of state-of-the-art profiling methods, SureQuant PQ500 analyses still provide targeted quantification performance (Figure 7). The amounts of endogenous peptides were estimated based on the known amounts of IS injected and single-point quantification strategy. This revealed a quantification range between 4 amol and 15 pmol, spanning more than 6 orders of magnitude, which is in theory sufficient to cover the vast majority of current FDA-approved plasma biomarkers. The fragment ion signals of the lowest abundant quantified endogenous peptide (PFLVFIR, 4 amol/µg of plasma) have been extracted (Figure 7, lower panel), illustrating the acceptable data quality even obtained for this challenging analyte. The quantification precision obtained on the entire dataset was excellent, with 75% of the endogenous peptides quantified with a CV below 10% (median CV value of 6%). The experiments were reproduced at a second site by another operator, using a different LC set-up (EASY-nLC 1200 system operated in a one-column setup instead of UltiMate 3000 RSLC system operated in a two-column setup) (Figure 8). After initial survey analysis, enabling the adjustment of internal standards MS1 triggering intensity threshold, SureQuant PQ500 analyses of undepleted plasma sample were repeated. It turned out that the quantification of the endogenous peptides between sites / systems / operators was very consistent, as reflected by the quantitative values tightly aligned along the theoretical 'perfect' correlation line, and the absence of bias revealed by the equation of the regression. The acquisition scheme of SureQuant method, fully avoiding any dependency on time-scheduling, enables efficient method portability and strong robustness.
This figure presents two graphs. The top left graph is a scatter plot showing "Endo. Pep. Conc. [fmol/µg plasma]" on the y-axis (from 0.001 to 10000 Albu) against "Abundance Rank" on the x-axis (from 0 to 600), illustrating the concentration distribution of endogenous peptides. The top right graph is a bar chart showing "Number of endo. Peptide" on the y-axis (from 0 to 300) against Coefficient of Variation (CV) ranges (<5%, 5-10%, 10-20%, >20%), indicating that 75% of peptides had a CV below 10%. The bottom graph shows two chromatograms for PFLVFIR, comparing SIL (10fmol) and ENDO (calc: 4amol) over time in minutes (56.4 to 57 min), demonstrating the signal quality for a low-abundance analyte. The quantification range is 4 amol – 15 pmol (>6 orders of magnitude), with CV < 10% for more than 75% of the peptides (median CV: 5.7%).
This scatter plot compares "Site 2-Endo. Pep. Conc. [fmol/µg plasma]" on the y-axis (0.001 to 100000) against "Site 1-Endo. Pep. Conc. [fmol/µg plasma]" on the x-axis (0.001 to 100000). It shows a strong correlation with R² = 0.97 and a regression equation of y = 1.0009x + 0.0881, indicating high consistency between quantification results obtained at different sites/systems. A "Theoretical 'perfect' correlation" line is shown alongside the "Actual correlation" data points.
A variant of SureQuant acquisition, dedicated to higher throughput analyses, has been developed using API-based IS-PRM application, which is only compatible with QE MS platforms. This method has not been natively implemented in the instrument control software of orbitrap instruments and does not benefit from the revised acquisition strategy of SureQuant method, but relies on the acquisition scheme of the original IS-PRM method. More specifically, in this 'Single Scan Msx IS-PRM' acquisition variant, the detection of an IS peptide in watch mode triggers the concomitant measurement of the corresponding pair of IS and endogenous peptides in a single PRM scan in quant. mode close to their elution profile apex, following their 'multiplex' isolation (Figure 1). This contrasts with the regular sequential acquisition, in which data points are collected over the full elution profiles of the peptides. With this acquisition variant, quantification is performed based on the fragment ion intensities extracted from single high quality PRM scan, and not on the area under the curve of the corresponding extracted fragment traces. This quantification strategy alleviates sampling rate constraint.
The analyses of plasma sample supplemented with PQ500 kit were reproduced using the 'Single Scan Msx IS-PRM' acquisition method on QE HF-X MS coupled to Evosep One LC system operated at a 60 samples/day analytical throughput (21-min LC gradient). The sampling rate under these conditions remained significantly higher than the predefined (relaxed) limit of 4-5 data points collected per peptide chromatographic peak (Figure 9, upper panel). In comparison with SureQuant analyses performed with longer LC gradient, a substantial number of IS peptides (81) were not detected, corresponding to the most hydrophilic peptides, which are not compatible with the LC setup used (Figure 9, lower panel). The remaining 721 IS peptides were properly detected in the watch mode of the method and triggered multiplex PRM measurement in the quant. mode, enabling the detection of 421 endogenous peptides, surrogate of 301 plasma proteins. In Figure 10, the data generated by SureQuant and Single Scan Msx IS-PRM analyses are displayed for one endogenous peptide quantified by both methods. The quantification results were very consistent across the methods.
This figure presents two bar charts. The top chart shows "Cycle Time [s]" on the y-axis (0 to 4 s) against time in minutes (0 to 20 min), with a sampling rate of ≥ 4-5 data points/peak (12-s peak width). The bottom chart illustrates "Number of spectra including ≥ 4 ref. fragments" (0, 1, Non-detectable IS) against a "Hydrophobicity scale". It shows that 712 Internal Standards peptides and 421 Endogenous peptides (surrogate of 301 proteins) were detected, with 802 total peptides detected. This figure highlights the performance of the high-throughput Single Scan Msx IS-PRM method.
This figure displays two sets of chromatograms comparing endogenous peptide quantification using "Single-Scan Msx IS-PRM (QE HF-X)" and "SureQuant (Exploris 480)" methods for the peptide FQASVATPR. Each method shows two graphs: "WATCH MODE" and "QUANT MODE", with intensity on the y-axis (0.E+00 to 6.E+07) against time in minutes (4.45 to 4.85 for Single-Scan Msx IS-PRM, and 23 to 23.6 for SureQuant). The intensity ratio ENDO/SIL is 0.0051 for Single-Scan Msx IS-PRM and the area ratio ENDO/SIL is 0.0049 for SureQuant, demonstrating consistent quantification results between the two methods for this peptide.
![]() |
Global Plasma Proteome Quantification Using SureQuant Targeted Analyses This white paper details the SureQuant method, a novel high-density targeted acquisition workflow for global plasma proteome quantification. Implemented on Thermo Scientific Orbitrap mass spectrometers, it leverages stable isotopically labeled peptides as internal standards to systematically screen over 500 plasma proteins, offering robust, precise, and high-throughput quantification for biomarker discovery and validation. |
![]() |
High-Throughput Proteomics Quantification by Fast LC Separation and Advanced PRM Acquisition This application note details the development and application of high-throughput LC-PRM setups, utilizing fast LC separations and advanced PRM acquisition schemes for proteomics studies. It focuses on monitoring AKT/mTOR signaling pathway components, demonstrating increased analytical throughput and high sensitivity for endogenous peptide quantification in complex samples. |
![]() |
Thermo Xcalibur Data Acquisition and Processing User Guide Comprehensive user guide for Thermo Fisher Scientific's Xcalibur software, detailing data acquisition, processing, instrument setup, sequence creation, and batch reprocessing for scientific analysis. |
![]() |
Thermo Scientific Environmental Contaminant Analysis Compendium: Methods from Sample to Compliance Explore Thermo Fisher Scientific's comprehensive solutions for environmental contaminant analysis. This compendium details methods for organic, inorganic, trace elemental, and emerging contaminants in water, air, and soil using advanced chromatography and mass spectrometry techniques. |
![]() |
Thermo Compound Discoverer User Guide - Software Version 3.0 Comprehensive user guide for Thermo Fisher Scientific's Compound Discoverer software (v3.0). Learn data processing, compound identification, workflow creation, and results analysis for mass spectrometry research. |
![]() |
Fast and Accurate Disulfide Bridge Detection: A Proteomics Workflow This paper details a novel, rapid, and accurate method for disulfide bridge mapping in proteins using Microwave-Assisted Acid Hydrolysis (MAAH), Field Asymmetric Ion Mobility Spectrometry (FAIMS), and Electron Transfer Higher-energy Collision Dissociation (EThcD), integrated with the XlinkX/Proteome Discoverer software. |
![]() |
Thermo Fisher Scientific iCAP Q Operating Manual Comprehensive operating manual for the Thermo Fisher Scientific iCAP Q mass spectrometer, covering system description, setup, operation, troubleshooting, and maintenance. |
![]() |
FreeStyle 1.8 SP1 Installation Guide Instructions for installing the FreeStyle 1.8 SP1 application to view and analyze raw mass spectrometry data, including system requirements and upgrade procedures. |