User Guide for Honeywell models including: AQ3CO SENSOR, Carbon Monoxide CO Gas Sensor

AQ3CO SENSOR - prod-edam.honeywell.com

Sensor Part Number (without board): ABQ045-H00 Module Part Number (with board): QAB045-H00 PCBA Part Number: AQ3-BP01 FEATURES AND BENEFITS High resolution Low detection limit Custom-built low noise

AQ3CO SENSOR

Gas Sensor

The data provided in this document is based on the assumption that the sensor is used at 20°C, 50% rH, and 1013 mBar for three months from the date of ...


File Info : application/pdf, 4 Pages, 738.21KB

sps-ast-aq3-series-aq3co-datasheet-en
AQ3CO SENSOR
Carbon Monoxide (CO) Gas Sensor
The purpose of this document is to present the performance specification of the AQ3 Series AQ3CO Carbon Monoxide Gas Sensor.

DOCUMENT PURPOSE
This document should be used in conjunction with the AQ3CO Characterization Note and the Product Safety Datasheet (PSDS 15).

The data provided in this document is based on the assumption that the sensor is used at 20°C, 50% rH, and 1013 mBar for three months from the date of sensor manufacture. For guidance on sensor performance outside of these limits, please refer to the AQ3CO Characterization Note.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Characterization Note.

Sensor Part Number (without board): ABQ045-H00
Module Part Number (with board): QAB045-H00
PCBA Part Number: AQ3-BP01

FEATURES AND BENEFITS

High resolution

Low detection limit

Custom-built low noise board achieves high
accuracy under ppb level

Individual compensation for temperature and cross sensitivity

High correlation with control station

AQ3CO Carbon Monoxide Gas Sensor Technical Specifications

TECHNICAL SPECIFICATIONS

MEASUREMENT

Technology

4-electrode electrochemical

Measurement Range 0 ppm CO to 10 ppm CO

Maximum Overload

without board: 50 ppm CO with board: 30 ppm CO

Onboard Filter Sensitivity*

to remove SOx/NOx and H2S
without board: 240 nA/ppm ±80 nA/ppm with board: 90 mV/ppm ±35 mV/ppm

Response Time (T90) Resolution*

< 70 seconds
20 ppb when used with recommended circuitry

Baseline Offset*

without board: -200 nA ~ 360 nA with board: -80 mV ~ 140 mV

Repeatability*

< ±3% of signal

Linearity**

linear

Low Detection Limit* 40 ppb

ELECTRICAL

Recommended Load Resistor

68 

Sensor Bias Voltage no bias (without board)

Power Supply Required 5 Vdc (with board)

Power Consumption 370 mA @ 5 Vdc (with board)

MECHANICAL

Weight

without board: < 22 g with board: < 30 g

Outer Plastic Body Material

polycarbonate

Sealing Gasket Material

TPU

Contact Material

mild steel with gold flash-over nickel plate

Orientation Sensitivity none

ENVIRONMENTAL

Operating Temperature Range

-30°C to 50°C

Recommended Storage Temperature

0°C to 20°C in original sealed container

Operating Humidity Range

15% rH to 90% rH non-condensing

Operating Pressure Range

atmospheric ±10%

Typical Applications ambient environmental monitoring

LIFETIME

Product Dimensions

Dimensions mm.
All tolerances ±0,15 mm unless otherwise stated

Sensor without board
Ø 41,2 mm

17,8 mm

Counter Ø 3,0 mm

Ø 1 mm on 34.2 PCD

3,0 mm pin projection

Sensing Reference

Ø 42,5 mm

3 mounting holes equi-spaced on 34.4 PCD
Auxiliary
Sensor with board
Ø 41,2 mm

Non-connected pin

A SEE DETAIL
Ø 41 mm Ø 42,5 mm 0,6 mm

17,8 mm 25,3 mm
5,72 mm

13,8 mm

1,27 mm

60°

1 2 3 4 5 DETAIL A 1,27 mm

Storage Life

6 months in original sealed container

Long-Term Sensitivity Drift*

< 10% signal loss per annum

Expected Oper. Life two years in air

Warranty

one year

*Specifications are valid at 20°C, 50% RH, and 1013 mbar using AQ3 low noise board. Performance characteristics outline the performance of sensors supplied within the first 3 months. Output signal can drift below the lower limit over time. Please be aware that sensors' performance also reflected by circuit board design.

** Linear through the concentration range across the whole operational environment range.

15,4 mm

17,2 mm

B SEE DETAIL
4,8 mm

SAME PINS WITH DETAIL B

1,91 mm

3 ­ Ø 3 mm

DETAIL B: 2 COUPLE OF LOCATION PINS

Pin Definition
15,4 mm

1

Vin (4.8 V ~ 5.5 V)

2 GND

3 Aux

4 GND

5 Sensing

AQ3CO Carbon Monoxide Gas Sensor Datasheet | sps.honeywell.com/ast | 2

AQ3CO Carbon Monoxide Gas Sensor Technical Specifications

Filter Information
Removes acid gases such as SOx/NOx and H2S
Poisoning
Gas sensors are designed for operation in a wide range of environments and harsh conditions. However, it is important that exposure to high concentrations of solvent vapors is avoided, during storage, fitting into instruments, and operation.
When using sensors with printed circuit boards (PCBs), degreasing agents should be used before the sensor is fitted.
Do not glue directly on or near the sensor as the solvent may cause crazing of the plastic.

Cross Sensitivity Table
Whilst AQ3 Series gas sensors are designed to be highly specific to the gas they are intended to measure, they will still respond to some degree to various other gases. The table below is not exclusive and other gases not included in the table may still cause a sensor to react.
IMPORTANT NOTE: The cross sensitivity data shown below does not form part of the product specification and is supplied for guidance only. Values quoted are based on tests conducted on a small number of sensors and any batch may show significant variation. For the most accurate measurements, an instrument should be calibrated using the gas under investigation.

Gas Sulfur Dioxide (SO2) Nitrogen Dioxide (NO2) Ozone (O3) Isobutylene (C4H8)

Gas Concentration 5 ppm 5 ppm 1 ppm 1 ppm

Cross Interference 0%<x%<7% -3%<x%<5% -7%<x%<4% ~None

AQ3CO Carbon Monoxide Gas Sensor Datasheet | sps.honeywell.com/ast | 3

WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship during the applicable warranty period. Honeywell's standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgment or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items that Honeywell, in its sole discretion, finds defective. The foregoing is buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.
While Honeywell may provide application assistance personally, through our literature and the Honeywell web site, it is buyer's sole responsibility to determine the suitability of the product in the application.
Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this writing. However, Honeywell assumes no responsibility for its use.

m WARNING
MISUSE OF DOCUMENTATION
· The information presented in this product sheet is for reference only. Do not use this document as a product installation guide.
· Complete installation, operation, and maintenance information is provided in the instructions supplied with each product.
Failure to comply with these instructions could result in death or serious injury.
, SAFETY NOTE
This sensor is designed to be used in environmental applications. To ensure that the sensor and/or instrument in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas (bump check) before each use of the sensor and/or instrument. Failure to carry out such tests may jeopardize the safety of people and property.

FOR MORE INFORMATION

Honeywell Advanced Sensing Technologies services its customers through a worldwide network of sales offices and distributors. For application assistance, current specifications, pricing or the nearest Authorized Distributor, visit our website or call:

USA/Canada Latin America Europe Japan Singapore Greater China

+1 302 613 4491 +1 305 805 8188 +44 1344 238258 +81 (0) 3-6730-7152 +65 6355 2828 +86 4006396841

Honeywell Advanced Sensing Technologies 830 East Arapaho Road Richardson, TX 75081 sps.honeywell.com/ast

AQ3CO Datasheet ECN 5062 | 002722-1-EN | 1 | 05/21 © 2021 Honeywell International Inc.

