Owner's Manual for TIMOX models including: MF8050 Oxygen Timer Medical Fittings, MF8050, Oxygen Timer Medical Fittings, Timer Medical Fittings, Fittings

MEDICAL FITTINGS TIMOX MF8050 (01) PDF MANUAL


File Info : application/pdf, 2 Pages, 290.20KB

MEDICAL FITTINGS TIMOX MF8050 (01)
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Medical Fittings, Inc. warrants that the TIMOX Oxygen Timer ("the Product"), excluding the battery, will be free of defects in workmanship or material for the period of two (2) years from the date of shipment.
Should any failure to conform to this warranty appear within the applicable period, Medical Fittings, Inc. shall, upon written notiication thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Medical Fittings, Inc.'s instructions and standard industry practice, and that no modiications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representatives of Medical Fittings, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a inal, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED. Medical Fittings, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost proits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulillment of all liabilities of Medical Fittings, Inc. whether based on contract, negligence, strict tort or otherwise. Medical Fittings, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or speciications without notice.

Powered by TCPDF (www.tcpdf.org)

300 Held Drive·Northampton, PA Phone 800-331-2685·Toll Free Fax 800-352-1240
504599 rev4 2.5M 4/05 Made in USA

OPERATING INSTRUCTIONS
GENERAL The function of the TIMOX Oxygen Timer is to
display and record the amount of time oxygen has been lowing to a patient. The display is constantly visible, indicates when oxygen is lowing, and is easily reset.
OPERATION Place the TIMOX Oxygen Timer in service
by connecting its handtight inlet itting to the downstream side of a lowmeter or regulator. The timer generates a small amount of backpressure; therefore it should be used with a backpressure compensated lowmeter. Besides oxygen, the timer can be used with compressed air, helium, nitrogen, or carbon dioxide. DO NOT USE IN HAZARDOUS ATMOSPHERES.
READING HOURS The display is an LCD and is constantly visible.
The timer will start running when the low reaches .4 LPM ±.1 LPM. The decimal point blinks when the timer is running. The display is capable of showing a maximum elapsed time of 1999.9 hours.
RESETTING To reset the timer to 0 hours, simply insert a
pen or pencil tip into the hole marked "RESET" on the front of the timer to press the reset switch.
CARE The TIMOX Oxygen Timer is an electronic
device and must be handled with reasonable care. Do not submerge the unit in any liquid. Do not autoclave it. Clean the exterior with a mild detergent and a damp cloth.
Powered by TCPDF (www.tcpdf.org)

BATTERY REPLACEMENT 1. The battery should be replaced when the display becomes dim.
2. Remove the three small screws at the upper right and left and lower left outer corners on the back of the unit. (The screw at the lower right-hand corner does not need to be removed.)
3. Lift the backplate/pressure block assembly out of the housing. Be careful not to pull the wires loose from the PC board.
4. Remove the old battery by pulling it out of the battery holder.
5. Insert the new battery into the battery holder so it's + and ­ ends match the markings in the bottom of the battery holder. (If the battery is put in backwards, the LCD display will remain blank.) The replacement battery is a lithium type battery Sanyo #CR14250SE or equal. It can be purchased from Medical Fittings, Inc. (our part #8066) or from any electronics supply store such as Newark Electronics.
6. Replace the backplate/pressure block assembly; be careful not to pinch the wires between the PC board and the pressure block. Orient the backplate/pressure block assembly so the handtight connection is at the end away from the LCD display. Replace the three screws in the back and make sure the timer display is visible, it should read 000.0 hours.

