OtoClear® Powered by Aquabot® Ear Irrigation Kit

Instructions for Use

Product Offering

#7270 - OtoClear/Aquabot, 20 tips, 1 Ear Basin

#7275 - OtoClear/Aquabot, 5 tips, 1 Ear Basin

Bionix - Innovative Solutions for Everyday Problems

Introduction

Thank you for purchasing the OtoClear®, powered by Aquabot®. This kit provides a simple, effective solution for patient safety, comfort, and compacted cerumen removal. The OtoClear ear irrigation tips direct water to the ear canal wall at a 30-degree angle, avoiding direct impact on the tympanic membrane.

This manual covers assembly, cleaning, and tip placement. Bionix is not liable for uses outside of the intended purpose.

Getting Started: Unpacking and Verifying Contents

Inspect the package for any shipping damage. If damage is found, notify your local distributor. Ensure all components of your OtoClear®, powered by Aquabot® kit are present:

Contact your distributor if any item is missing.

Intended Use

The OtoClear®, powered by Aquabot® is intended for Ear Irrigation by a medical professional.

Before You Begin

Examine the ear canal and tympanic membrane with an otoscope. Note the type and location of any cerumen. Hard or impacted wax may require a cerumenolytic agent.

WARNING: DO NOT perform irrigation if the tympanic membrane is perforated or if tympanostomy tubes are present. Discontinue irrigation immediately if bleeding, irritation, or other trauma to the ear canal or tympanic membrane occurs.

Aquabot® Parts ID

1. Pump Handle

2. Trigger

3. Spring

4. Pinch Tube

5. Lid

6. Lid Nozzle

7. Nozzle Adapter

8. Soft Nozzle

9. Nozzle Cap

10. Lid O-ring

11. Draw Tube Adapter

12. Draw Tube O-ring

13. Piston Rod

14. Pump Barrel O-ring

15. Pump Barrel

16. Draw Tube

17. Umbrella Valve

18. Piston O-ring

19. Weighted Screen

Aquabot® Warnings

FAILURE TO READ AND FOLLOW THE ABOVE WARNINGS AND OPERATING INSTRUCTIONS COULD LEAD TO SERIOUS BODILY INJURY.

Aquabot® Spray Modes

The Aquabot has two operating modes: Stealth and Pulse.

a. Expose the adapter hole to use in Pulse mode.

b. Cover the adapter hole to use in Stealth mode.

OtoClear®, powered by Aquabot® Use Instructions

  1. Clean the base by filling it half full with soapy water. Twist on the lid, shake to clean, and spray soapy water through the unit. Repeat with fresh water.
  2. Fill the base to the 32 oz. mark on the 1000mL bottle or the 19 oz. mark on the 650mL bottle.
  3. Attach the Aquabot to the top of the bottle and close firmly by rotating clockwise.
  4. Grasp the OtoClear Tip and twist onto the luer lock fitting of the adapter until it stops.
  5. Pump no more than 6 times. This will pressurize the unit to approximately 35 psi. Note the force required on the last pump to use as a guide for future charges to avoid over-pressurization.
  6. To release water, firmly press the colored circle on the back of the trigger.
  7. To depressurize the unit safely, hold it upside down with the weighted screen above the water line. Press the trigger to spray out compressed air.

WARNING: Failure to remove the OtoClear Tip risks cross-contamination between patients. The tip is designed for single-use.

Aquabot® Maintenance

High Level Disinfection of OtoClear®, powered by Aquabot®

High level disinfection is recommended after any use involving contact with blood, infectious material, or non-intact skin.

  1. Use one of the following high level disinfecting (HLD) solutions compatible with OtoClear®, powered by Aquabot®:
    • Ortho-Phthalaldehyde Solution (0.55% maximum)
    • Accelerated Hydrogen Peroxide Solution (2% maximum)
    • Glutaraldehyde Solution (3.0% maximum)
  2. Prepare the HLD solution according to the manufacturer's instructions.
  3. Remove the spray head from the bottle.
  4. Drain all water from the bottle and spray head.
  5. Fill the OtoClear®, powered by Aquabot® with the HLD solution and reattach the spray head.
  6. Pump the OtoClear®, powered by Aquabot® until HLD solution discharges.
  7. Remove the spray head and completely fill the bottle with HLD solution.
  8. Reattach the spray head.
  9. Allow the HLD solution to soak in the OtoClear®, powered by Aquabot® as per the HLD manufacturer's instructions.
  10. After the recommended soak, remove the spray head and empty the HLD solution from the bottle.
  11. Rinse the outer surfaces of the bottle and spray head thoroughly with water.
  12. Fill the bottle with clean rinse water and reattach the spray head.
  13. Pump until half of the bottle is discharged.
  14. Repeat steps 11-13 until the bottle has been rinsed three times.
  15. Remove the spray head from the bottle.
  16. Drain all water from the bottle.
  17. Pump the spray head until all water is discharged.
  18. Dry the outer surfaces of the bottle and spray head, leaving them separated for complete drying.
  19. To high level disinfect the ear basin, repeat steps 1 & 2 and follow the manufacturer's instructions.

Aquabot® Troubleshooting Chart

PROBLEMREASONSOLUTION
Unit does not build or hold pressureLid O-ring is missingReplacement parts available at lunatecgear.com
Pump Barrel is too looseTighten Pump Barrel
Something elseEmail support@lunatecgear.com or call 858.653.0401
Builds pressure but does not sprayPinch Tube is kinkedRemove trigger and unkink Pinch Tube
Something elseEmail support@lunatecgear.com or call 858.653.0401
Water is getting inside of pump barrelUmbrella Valve is not sealingReplacement parts available at lunatecgear.com
Pump Barrel is too looseTighten Pump Barrel
Water is leaking from under triggerPinch Tube is damagedReplacement parts available at lunatecgear.com
Something elseEmail support@lunatecgear.com or call 858.653.0401
Water is leaking from under triggerSoft Nozzle is missingReplacement parts available at lunatecgear.com
Wear and tear on Soft NozzleReplacement parts available at lunatecgear.com
Something elseEmail support@lunatecgear.com or call 858.653.0401
Mist is not fineWear and tearTry sanding 0.5mm off the open side of the Nozzle Cap. If that doesn't work, order a new Nozzle Cap.
Operating mode is in PulseSwitch to Stealth mode
Something elseEmail support@lunatecgear.com or call 858.653.0401
Spray is too noisyOperating mode is in PulseSwitch to Stealth mode
Models: 7270 Ear Irrigation Kit, 7270, Ear Irrigation Kit

File Info : application/pdf, 6 Pages, 602.15KB

PDF preview unavailable. Download the PDF instead.

rm95-7275 aquabot instructions 1 (1)

References

RAD PDF 3.19.2.2 - https://www.radpdf.com PDFescape Online - https://www escape.com