OtoClear® Powered by Aquabot® Ear Irrigation Kit
Instructions for Use
Product Offering
#7270 - OtoClear/Aquabot, 20 tips, 1 Ear Basin
#7275 - OtoClear/Aquabot, 5 tips, 1 Ear Basin
Bionix - Innovative Solutions for Everyday Problems
Introduction
Thank you for purchasing the OtoClear®, powered by Aquabot®. This kit provides a simple, effective solution for patient safety, comfort, and compacted cerumen removal. The OtoClear ear irrigation tips direct water to the ear canal wall at a 30-degree angle, avoiding direct impact on the tympanic membrane.
This manual covers assembly, cleaning, and tip placement. Bionix is not liable for uses outside of the intended purpose.
Getting Started: Unpacking and Verifying Contents
Inspect the package for any shipping damage. If damage is found, notify your local distributor. Ensure all components of your OtoClear®, powered by Aquabot® kit are present:
- Aquabot® Bottle
- Adapter wand
- Either 20 OtoClear Tips or 5 OtoClear Tips
- Ear basin
Contact your distributor if any item is missing.
Intended Use
The OtoClear®, powered by Aquabot® is intended for Ear Irrigation by a medical professional.
Before You Begin
Examine the ear canal and tympanic membrane with an otoscope. Note the type and location of any cerumen. Hard or impacted wax may require a cerumenolytic agent.
WARNING: DO NOT perform irrigation if the tympanic membrane is perforated or if tympanostomy tubes are present. Discontinue irrigation immediately if bleeding, irritation, or other trauma to the ear canal or tympanic membrane occurs.
Aquabot® Parts ID
1. Pump Handle
2. Trigger
3. Spring
4. Pinch Tube
5. Lid
6. Lid Nozzle
7. Nozzle Adapter
8. Soft Nozzle
9. Nozzle Cap
10. Lid O-ring
11. Draw Tube Adapter
12. Draw Tube O-ring
13. Piston Rod
14. Pump Barrel O-ring
15. Pump Barrel
16. Draw Tube
17. Umbrella Valve
18. Piston O-ring
19. Weighted Screen
Aquabot® Warnings
- Do not over-pressurize the unit.
- Do not use liquids with temperatures above 120°F / 49°C.
- CHOKING HAZARD: Unit contains small parts. Not suitable for children under 3 years.
- Only use the lid with Aquabot bases.
- Do not use the lid on any metal or glass bases.
FAILURE TO READ AND FOLLOW THE ABOVE WARNINGS AND OPERATING INSTRUCTIONS COULD LEAD TO SERIOUS BODILY INJURY.
Aquabot® Spray Modes
The Aquabot has two operating modes: Stealth and Pulse.
- Stealth: Ideal for drinking, misting, and conserving pressure.
- Pulse: Ideal for cleaning and conserving water.
a. Expose the adapter hole to use in Pulse mode.
b. Cover the adapter hole to use in Stealth mode.
OtoClear®, powered by Aquabot® Use Instructions
- Clean the base by filling it half full with soapy water. Twist on the lid, shake to clean, and spray soapy water through the unit. Repeat with fresh water.
- Fill the base to the 32 oz. mark on the 1000mL bottle or the 19 oz. mark on the 650mL bottle.
- Attach the Aquabot to the top of the bottle and close firmly by rotating clockwise.
- Grasp the OtoClear Tip and twist onto the luer lock fitting of the adapter until it stops.
- Pump no more than 6 times. This will pressurize the unit to approximately 35 psi. Note the force required on the last pump to use as a guide for future charges to avoid over-pressurization.
- To release water, firmly press the colored circle on the back of the trigger.
- To depressurize the unit safely, hold it upside down with the weighted screen above the water line. Press the trigger to spray out compressed air.
WARNING: Failure to remove the OtoClear Tip risks cross-contamination between patients. The tip is designed for single-use.
Aquabot® Maintenance
- Clean with soap and water as needed. The base is dishwasher safe for the top rack only. The lid is NOT dishwasher safe.
- Do not put the unit in the freezer.
- Store out of direct sunlight.
- Do not store pressurized for more than 36 hours.
- Spray air through the unit to dry the silicone tube before storing for long periods.
- Lock the trigger in the ON position when storing for long periods to prevent the Pinch Tube from kinking.
- Spare parts are available at lunatecgear.com.
- Instructional videos are available on product pages at lunatecgear.com.
- For assistance, contact support@lunatecgear.com.
High Level Disinfection of OtoClear®, powered by Aquabot®
High level disinfection is recommended after any use involving contact with blood, infectious material, or non-intact skin.
- Use one of the following high level disinfecting (HLD) solutions compatible with OtoClear®, powered by Aquabot®:
- Ortho-Phthalaldehyde Solution (0.55% maximum)
- Accelerated Hydrogen Peroxide Solution (2% maximum)
- Glutaraldehyde Solution (3.0% maximum)
- Prepare the HLD solution according to the manufacturer's instructions.
- Remove the spray head from the bottle.
- Drain all water from the bottle and spray head.
- Fill the OtoClear®, powered by Aquabot® with the HLD solution and reattach the spray head.
- Pump the OtoClear®, powered by Aquabot® until HLD solution discharges.
- Remove the spray head and completely fill the bottle with HLD solution.
- Reattach the spray head.
- Allow the HLD solution to soak in the OtoClear®, powered by Aquabot® as per the HLD manufacturer's instructions.
- After the recommended soak, remove the spray head and empty the HLD solution from the bottle.
- Rinse the outer surfaces of the bottle and spray head thoroughly with water.
- Fill the bottle with clean rinse water and reattach the spray head.
- Pump until half of the bottle is discharged.
- Repeat steps 11-13 until the bottle has been rinsed three times.
- Remove the spray head from the bottle.
- Drain all water from the bottle.
- Pump the spray head until all water is discharged.
- Dry the outer surfaces of the bottle and spray head, leaving them separated for complete drying.
- To high level disinfect the ear basin, repeat steps 1 & 2 and follow the manufacturer's instructions.
Aquabot® Troubleshooting Chart
PROBLEM | REASON | SOLUTION |
---|---|---|
Unit does not build or hold pressure | Lid O-ring is missing | Replacement parts available at lunatecgear.com |
Pump Barrel is too loose | Tighten Pump Barrel | |
Something else | Email support@lunatecgear.com or call 858.653.0401 | |
Builds pressure but does not spray | Pinch Tube is kinked | Remove trigger and unkink Pinch Tube |
Something else | Email support@lunatecgear.com or call 858.653.0401 | |
Water is getting inside of pump barrel | Umbrella Valve is not sealing | Replacement parts available at lunatecgear.com |
Pump Barrel is too loose | Tighten Pump Barrel | |
Water is leaking from under trigger | Pinch Tube is damaged | Replacement parts available at lunatecgear.com |
Something else | Email support@lunatecgear.com or call 858.653.0401 | |
Water is leaking from under trigger | Soft Nozzle is missing | Replacement parts available at lunatecgear.com |
Wear and tear on Soft Nozzle | Replacement parts available at lunatecgear.com | |
Something else | Email support@lunatecgear.com or call 858.653.0401 | |
Mist is not fine | Wear and tear | Try sanding 0.5mm off the open side of the Nozzle Cap. If that doesn't work, order a new Nozzle Cap. |
Operating mode is in Pulse | Switch to Stealth mode | |
Something else | Email support@lunatecgear.com or call 858.653.0401 | |
Spray is too noisy | Operating mode is in Pulse | Switch to Stealth mode |