WP4C MEASUREMENT USING LABROS SOILVIEW SOFTWARE
This application note describes how a WP4C measurement is processed using LABROS SoilView Software. For more details about the WP4C, please refer to the WP4C User Manual. For more information about taking a WP4C sample after HYPROP measurement, please refer to the How to Create a Full Moisture Release Curve Using the WP4C and HYPROP application note.
WP4C MEASUREMENT USING LABROS SOILVIEW SOFTWARE
Figure 1 Workflow Overview
WP4C MEASUREMENT PROCESS
1. Connect the WP4C via USB and turn it on. 2. Open LABROS SoilView Software and press the Show Devices button. 3. The connected device appears in the device tree on the left of the software screen and a measurement line in the
manager is created. There are two possible ways to create a WP4C measurement file. The procedures for these options are described in the Adding WP4C Data Points to a *.bhdx File and Adding WP4C Data Points to an Excel File sections.
METER Group, Inc. USA 2365 NE Hopkins Court, Pullman, WA 99163 T +1.509.332.2756 F +1.509.332.5158 E info@metergroup.com W metergroup.com

SETTING MENU

WP4C MEASURMENT USING LABROS SOILVIEW SOFTWARE

Figure 2 Settings Window Button
Use the drop-down menu where a default profile for each measurement mode may be selected. Load this as a default WP4C measurement profile by selecting the checkbox at the bottom of the Settings window.
To create an owner profile use the four buttons described in Figure 3. As a measurement parameter, the Measurement mode is set using the drop-down menu. Using Continuous mode, the count of WP4C data points may be selected as well as the tolerance. If the count of measurement cycle is e.g. set to three and the tolerance is set to 0.2 pF, the software verifies if the last three recorded data points are within the tolerance of +/- 0.2 pF. Select the Autostop checkbox to stop the measurements automatically.

Figure 3 Create Template
2

WP4C MEASURMENT USING LABROS SOILVIEW SOFTWARE
ADDING WP4C DATA POINTS TO A *.BHDX FILE
1. Select the button with the ellipsis as depicted in Figure 4.
Figure 4 Open *.bhdx File 2. Select the *.bhdx file for the WP4C data points that need to be added. 3. Select one of the default profiles depending on the required measurement mode, or create an owner profile in the
Settings menu (Figure 3). For more information about measurement modes, refer to the WP4C User Manual. 4. Select Start. 5. A new measurement tab opens with a message asking to close the WP4C measurement chamber, turn the knob
to Read mode and press Done. 6. The status of the measurement and the next steps to perfom is displayed in the upper right corner of the
software screen, see Figure 5.
Figure 5 Measurement Status 7. The single point measurements are displayed in the upper right table. 8. When using Fast or Precise mode only one data point is recorded and the measurement stops automatically.
When using Continuous mode the software records as many data points as defined in the profile Count of WP4C Data Points. The software will verify if the datapoints are within the defined tolerance; if so stop the measurement by pressing the button Stop Data Recording. The software stops automatically if Autostop is selected. The average is calculated and displayed in the lower table on the right (Figure 6). If the values are not within the defined tolerance the measurements continue. 9. Once measuring has stopped the soil sample must be removed from the measurement chamber and weighed immediately. The wet weight of the sample as well as the tare weight of the cup may be entered directly into the lower table on the right (Figure 6).
3

WP4C MEASURMENT USING LABROS SOILVIEW SOFTWARE
Figure 6 Result Table 10. The next sample may then be measured by inserting a new sample, pressing Start Data Recording, and following
step 5--9. When all samples referring to the *.bhdx file are measured, select the Stop Measurement and Save File button, see Figure 7. All values in the lower table are entered into the *.bhdx file.
NOTE: The Stop Measurement and Save File must be selected in order for the file to be finished and saved correctly.
Figure 7 Stop Measurement and Save 11. The measurement can be removed from the manager by selecting Remove from Manager.
4

WP4C MEASURMENT USING LABROS SOILVIEW SOFTWARE
ADDING WP4C DATA POINTS TO AN EXCEL FILE
1. Enter a sample name in the manager line (Figure 8).
Figure 8 Enter Sample Name and Filename 2. Follow step 3--10 in the Adding WP4C Data Points to a *.bhdx File section. 3. When all samples have been measured select the Stop Measurement and Save button. This button must be
selected to finish and save the file. All values in the lower table are entered into the Excel file. 4. The measurement may be removed from the manager by selecting Remove from Manager.
WP4C EVALUATION USING LABROS SOILVIEW-ANALYSIS SOFTWARE
To evaluate both file options, the LABROS SoilView-Analysis Software is used. For more detailed information about the LABROS SoilView-Analysis Software please refer to the LABROS SoilView-Analysis for HYPROP User Manual.
EVALUATING A *.BHDX FILE
The recorded WP4C data points are displayed in the Evaluation register located in the Add WP4 Data Points table. The gross dry mass of the sample, the gross wet mass, and the tare mass of the sample cup must be entered.
EVALUATING EXCEL FILE
To add the recorded WP4C data points saved in an Excel file to a *.bhdx file in LABROS SoilView-Analysis, the respective *.bhdx file must be opened. The data points are loaded by selecting Add Points From File and clicking the right hand mouse button on the Add WP4 Data Points table located in the Evaluation register (Figure 9). The gross dry mass of the sample, the gross wet mass, and the tare mass of the sample cup must be entered.
NOTE: Do not change the format of the created Excel file.
Figure 9 Add Points From File