MINIWARE DS211 Digital Oscilloscope User Manual V1.1
Product Overview
DS211 is a mini digital storage oscilloscope based ARM Cortex-M3 compatible with 32-bit platform, equipped with a 320*240 color screen and Micro USB interface which serves as PC connection and charging. Compact size, simple operation, and easy to use, it can meet the basic requirements for school experiment, appliance repair and electronic engineering.
General Safety Information
- Read carefully all the following safety precautions to avoid personal injury and prevent damage to the device or any products connected to it. In order to avoid possible dangers, be sure to use this product in accordance with the provisions in order to avoid fire or personal injury.
- Use proper power supply. Please use power supply specified for this product and certified for use of your country/region.
- Connect and disconnect properly. Do not connect or disconnect probe or test leads while they are connected to voltage source, disconnect the testing circuit before connecting or disconnecting the probe.
- Observe all the terminal ratings. To avoid fire or shock hazard, please do not measure signals at DC40V or above to avoid damage to the device. Consult the product manual for further rating information before making connections to the device.
- Do not operate in wet/damp conditions.
- Do not operate in a potentially inflammable/explosive environment.
- Please keep the surface of the product clean and dry.
Functions
Feature | Specification |
---|---|
Analog bandwidth | 0-200KHz |
Maximum sampling rate | 1MSa/s |
Maximum sample memory depth | 8K |
Horizontal sensitivity | 1uS/Div~2S/Div (1-2-5 sequence step) |
Vertical sensitivity | 20mV/Div~10V/Div(x1probe) 0.2V/Div~100V/Div(x10probe) |
Analog input impedance | >500KΩ |
Maximum input voltage | 40Vpp (x1probe) |
Coupling | AC/DC |
Synchronous mode | Auto, Normal, Single, Scan |
Trigger mode | Rising/Falling edge trigger Ascend/Descend Edge Trigger Mode |
Auto measurement | Signal sequence/cycle/duty ratio, peak voltage/virtual value/max value/min value/average |
Inbuilt signal Generator | 10Hz~1MHz (1-2-5 sequence step) |
Storage | 8MB Flash |
Parameters
Parameter | Value |
---|---|
Weight | 66g |
Screen | 2.8" full color TFT LCD |
Screen Resolution | 320×240 |
Battery | 500mAH |
Charging | Via Micro USB Cable |
PC Connection | Via Micro USB Cable |
Size | 106mm×55.5mm×11mm |
Operational Environment
Temperature | Operating Conditions | Non-Operating Conditions |
---|---|---|
Relative Humidity | High Temperature: 40℃~50℃ 0%-60%RH | High Temperature: 40℃~60℃ 5%-60%RH |
Low Temperature | 0℃~40℃ 10%-90%RH | 0℃~40℃ 5%-90%RH |
General Inspections
When you get a new DS211 Mini Oscilloscope, you are advised to inspect the product by the following steps.
- Inspect damages caused by shipping. If the packaging carton is seriously damaged, keep the package until the oscilloscope & accessories pass the electrical and the mechanical test.
- Inspect the product. Please contact the dealer if the following problems occur to DS211: 1) product appearance is damaged, 2) product doesn't work properly, 3) product does not pass performance test. If the damage to DS211 is resulted from shipping, please keep the package.
Functional Inspections
Make a quick inspection of functions to ensure the device is working soundly. Please perform following steps:
- Turn on power and access to the homepage of the oscilloscope.
- Place in the standard signal (e.g. square wave 1 KHz, Vpp=5V), Insert X1 probe's MCX end to "INP” channel, and the probe to "OUT" channel. Check if the measurement value and the standard value are equal, calibrate if different.
Battery Charging Instructions
When the battery voltage status turns to "?" or display brightness is relatively dim, please charge the battery in time; charging is workable in both power-on and off mode.
Firmware Upgrading
To upgrade the firmware of oscilloscope, please carry out the operation follow the steps below:
- Visit www.miniware.com.cn, download the firmware to your PC appropriate DS211.
- Hold “—” button on DS211 and switch on the Power to enter DFU firmware upgrading mode.
- Use USB data cord to connect DS211 to your PC, and a removable hard disk named: “DFU X.XX" will appear on your PC; copy the hex firmware to the root directory of that disk, after the extension of the firmware changes from "hex” to “rdy”, restart DS211, and then the upgrading process is completed.
Buttons and Interface
The device features ports for Signal Input (INP), Signal Output (OUT), and USB connection. The front panel includes buttons for Power, Stop, Up, Left, Right, Down, and Menu (M).
Button Functions
Button | Function |
---|---|
►► / ►► | Run/pause button |
+ | Save parameter |
- | Slide Up |
◄◄ | Slide Down |
►► | Set parameter (Reduce, Slide Left) |
◄◄ | Set parameter (Increase, Slide Right) |
M | On/Off Sub-menu |
M (long press) | Save BMP |
M (double press) | Auto Fit |
Measurement Instructions
1. Parameters Introduction
The main screen displays waveform data and parameters. Key areas are labeled A-L.
Main Screen Parameters:
- A: Y-axis Unit Amplitude
- B: Vertical Sensitivity (20mV-10V, 1-2-5 stepping)
- C: Coupling (AC/DC)
- D: Horizontal Sensitivity (1uS-2S, 1-2-5 stepping)
- E: Waveform Calculation (e.g., double waveform calculation)
- F: Trigger Mode (Rising/Falling edge)
- G: Trigger Accuracy
- H: Synchronous Mode (AUTO/NORM/SINGL/SCAN)
- I: Cursor Parameters (Vertical, Horizontal, Sub-item)
- J: Frequency
- K: Measurement values (Freq, Duty, Vrms, Vavg, Vp-p, Vmax, Vmin)
- L: Storage Depth/File Management
Yn Parameter Notes
Y Ranges: Ordinate unit amplitude
Coupling: Coupling method
ProbeAtt: Probe multiple choice X1/X10
Y Offset: Wave adjustment Y axis
CursorV1: Measurement cursor V1
CursorV2: Measurement cursor V2
Hide_Vn_: Show/hide measurement cursor
Xn Parameter Notes
TimeBase: Abscissa unit amplitude
ViewPosi: Move horizontally to check waveform
SmplDpth: Storage depth (1k ~ 8k)
CursorT1: Time measurement cursor T1
CursorT2: Time measurement cursor T2
Hide_Tn_: Show/hide measurement cursor
2. Options Introduction
Various options are available for detailed settings:
- Yn: Y-axis function setting
- Xn: X-axis function setting
- Tr: Trigger function setting
- Me: Measurement function setting
- Ex: Waveform calculation function setting
- Fn: Saving and loading function setting
- Sn: Waveform output parameter setting
- St: System function setting
3. Specific Parameter Introduction
Option Setting Method: Use "+" or "-" buttons to navigate options. Press "M" to open the menu. Use "+" or "-" to select parameters and change values with blinking cursor using "►►►►" or "◄◄" buttons. Long press "M" to save changes.
Example: Saving all parameters is prompted with "Save ALL ParamTab? Yes / No".
Tr Parameter Notes
SyncMode: Synchronous mode (AUTO/NORM/SINGL/SCAN)
TrigMode: Trigger mode (Rising/falling edge)
Auto Fit: Automatic adjustment (Icon "F" appears when open, double click "M" for auto fit)
Threshol: Horizontal Triggering Position Level
Sensitiv: Horizontal Triggering accuracy
Hide_Tri: Show/Hide Horizontal Triggering Position Level
Me Parameter Notes
Freq: Signal frequency
Duty: Duty ratio
Vrms: Root mean square voltage value
Vavg: Average value
Vp-p: Peak-to-peak voltage
Vmax: Maximum value
Vmin: Minimum value
Ex Parameter Notes
Ext Refn: Input/Data source selection
Ext Posi: Signal position
Ext Hide: Show/hide signal calculation. To hide waveform computation lines, select "Ext Hide", change value to "Hide" using navigation buttons.
Fn Parameter Notes
Save Bmp: Save BMP file (waveform figure) into flash disk
Save Dat: Save DAT file into flash disk
Save Buf: Save buffer file (sampling buffer data) into flash disk
Save Csv: Save CSV file (export sampling buffer data) into flash disk
Save Svg: Save SVG file (sampling buffer figure) into flash disk
Load Dat: Load DAT file
Load Buf: Load buffer file
Sn Parameter Notes
Out Type: Output signal type
Out Freq: Output signal frequency
Out Duty: Output signal duty cycle
St Parameter Notes
B-Light: Adjust backlight brightness (Use ◄ or ►►)
Auto Cal: Calibrate Zero (Press ►►►► for auto calibrate, save setting per prompt)
Restore: Restore Data (Press ►►►► for auto calibration, save setting per prompt)
Standby: Adjust standby time (Use ◄ or ►► to choose a standby time of 0-60 mins)
PowerOff: Adjust power off time (Use ◄ or ►► to choose a power-off time of 0-60 mins)
Note: When connected to PC via USB data cord, it will not activate auto power-off.