KAIWEETS HT200B Digital Clamp Meter User Manual

Introduction

The HT200B is a cost-effective AC clamp meter. It is compact, easy to use, and suitable for daily work.

Safety Information

Warnings: Read First

To avoid possible electric shock or personal injury, please obey the following instructions:

Product Overview

The KAIWEETS HT200B features:

The device is powered by 2 x AAA (1.5V) batteries.

Safety Explanation

Symbol Meaning Symbol Meaning
⚠️ Hazardous Voltage. Danger! Alternating Current (AC)
AC and DC Direct current (DC)
⚠️ Important Information ± Earth Ground
Fuse Low Battery Indication
Double Insulated
CE Conforms to requirements of European Free Trade
Do not dispose of this product in unsorted municipal waste.
CAT.III IEC Measurement Category III: CAT III equipment has protection against transients in equipment in fixed-equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.

Function Buttons

Operating Instructions

Insert and replace batteries

⚠️ Attention

Replace the batteries immediately when the low battery symbol [] appears in the display. Disconnect the meter from the power source. Remove the test leads from the measuring device, loosen the screws on the back, remove the lower half of the housing, replace the old batteries, and screw the cover back on.

Turn on/off the meter

The device is switched on when the rotary switch (3) is set to a measuring function. When the rotary switch is set to OFF, the measuring device is switched off.

Automatic shutdown

If no operation is performed for 15 minutes, the clamp meter beeps 3 times. Without further operation, the clamp meter beeps again and switches off automatically. Press any key to restore the clamp meter to operating status. The symbol [O] appears on the display. To deactivate the automatic switch-off, hold down the "FUNC." key for more than 2 seconds and switch on the clamp meter at the same time. The symbol [O] disappears. Restart after switching off the meter; the automatic switch-off can be restored.

Auto-ranging

The meter automatically selects the measuring range that is best suited for the measurement performed. The measuring unit and resolution are adjusted accordingly.

Connecting Test Leads

Do not test if leads are improperly seated, as results could cause intermittent display readings. To ensure proper connection, firmly press leads into the jack completely.

Incorrect connection example: Probe tip exposed beyond the insulated sleeve.

Correct connection example: Probe tip fully inserted into the jack.

Measurement Operation

AC current measurement

The current is measured via the current clamp (1). The sensors in the current clamp detect the magnetic field created by current-carrying conductors. You can take measurements on insulated and non-insulated conductors.

The test leads should be disconnected from the measuring device when measuring current.

  1. Estimate the current range (2 A, 20 A, or 200 A) and turn the knob to the desired range.
  2. Press the clamp release lever and open the clamp.
  3. Grasp the individual current conductor to be measured and close the current clamp again.

The conductor must be placed in the center of the clamp for accurate readings; otherwise, additional errors may occur.

The measured current is shown on the display. If the current is > 3 A, the orange display backlight illuminates.

Attention:

AC/DC voltage measurement

△ Voltage above 600 V cannot be measured! When measuring high voltage, pay special attention to safety to avoid electric shock or injury.

  1. Turn the knob to (DC Voltage) or (AC Voltage).
  2. Plug the black test lead into the COM socket and the red test lead into the INPUT socket.
  3. Bring the test probes into contact with the points to be measured.
  4. Read the result on the display. If the voltage is > 80 V, the orange display backlight illuminates.

Attention:

Resistance measurement

△ Do not apply a voltage above 10 V! Make sure that the power supply to the circuit is disconnected and all capacitors are discharged.

  1. Turn the rotary function switch to "" for resistance measurement. It displays "M" on the screen.
  2. Turn off the circuit under test.
  3. Plug the black test lead into the COM socket and the red test lead into the INPUT socket.
  4. Place the test probes in contact with the points at which the resistance must be measured.
  5. Read the result on the display.

Note: When measuring resistance on a circuit, the measured value can be influenced by other parallel circuits.

Overload protection: 250 V

Continuity test ▶※

Do not apply a voltage above 10 V! Make sure that the power supply to the circuit is disconnected and all capacitors are discharged.

  1. Turn the rotary switch to "" and press the "FUNC." key several times until ▶※ appears on the display. The continuity test is now active.
  2. Plug the red test lead into the INPUT socket and the black test lead into the COM socket.
  3. Connect the probes to the circuit or component under test.
  4. If the resistance is < 30 , an acoustic signal sounds continuously, and the orange backlight lights up.

Note: When measuring resistance on a circuit, the measured value can be influenced by other parallel circuits.

Open circuit voltage: approx. 2V

Overload protection: 250V

Diode Test

△ Do not apply a voltage above 10 V! Make sure that the power supply to the circuit is disconnected and all capacitors are discharged.

  1. Turn the rotary function switch to "". Press the "FUNC." key 2 times until appears on the display.
  2. Remove power from the circuit being tested.
  3. Connect the black test lead to the COM terminal and the red test lead to the INPUT terminal.
  4. Connect the black test lead to the cathode side and the red test lead to the anode side of the diode being tested.
  5. Read the forward bias voltage value on the LCD.
  6. If the polarity of the test leads is reversed relative to the diode polarity, the LCD reading shows "OL". This can be used for distinguishing the anode and cathode sides of a diode.

Test voltage approx. 2 V

Overload protection: 250 V

Capacitance measurement

  1. Turn the knob to .
  2. Insert the red probe into the INPUT socket and the black probe into the COM socket.
  3. Contact the probes to the capacitance.
  4. Read the measurement result on the screen.

Attention:

When measuring capacitance on the line, disconnect the power supply and discharge all high-voltage capacitors. Otherwise, the instrument may be damaged and may cause electric shocks.

Note: If the capacitance is > 100 uF, it will take a long time to measure correctly.

Overload protection: 250 V

Specification

Accuracy

Function Range Resolution Accuracy
DC Volts 200mV/2V/20V/200V/600V 0.1mV / 0.001V / 0.01V / 0.1V / 1V ±(0.5%+5)
AC Volts 2V/20V/200V/600V 0.001V / 0.01V / 0.1V / 1V ±(1.0%+5)
AC Amps 2A/20A/200A 0.001A / 0.01A / 0.1A ±(2.5%+8)
Resistance 200 / 2k / 20k / 200k / 2M / 20M 0.1 / 0.001k / 0.01k / 0.1k / 0.001M / 0.01M ±(1.0%+5)
Capacitance 2nF/20nF/200nF/2uF/20uF/200uF/ 2mF 0.001nF / 0.01nF / 0.1nF / 0.001uF / 0.01uF / 0.1uF / 0.001mF ±(4.0%+5)

Maintenance

Warning

Caution

Cleaning the Meter

Three Year Warranty

KAIWEETS will repair, without charge, any defects due to faulty materials or workmanship for three years from the date of purchase, provided that:

Defective products will be repaired or replaced, free of charge or at KAIWEETS' discretion, if sent together with proof of purchase to an authorized distributor.

For further detail of warranty coverage and warranty repair information, send an email to support@Kaiweets.com.

Connect with KAIWEETS on Facebook: @kaiweetstools

PDF preview unavailable. Download the PDF instead.

User Manual - English Adobe Illustrator 24.0 (Windows) Adobe PDF library 15.00

Related Documents

Preview KAIWEETS HT206D True-RMS Digital Clamp Meter User Manual
Comprehensive user manual for the KAIWEETS HT206D True-RMS Digital Clamp Meter. Learn about its features, safety information, operating instructions, and specifications for accurate electrical measurements.
Preview KAIWEETS HT208D True-RMS Digital Clamp Meter User Manual
Comprehensive user manual for the KAIWEETS HT208D True-RMS Digital Clamp Meter. Learn about its features, safety information, operating instructions, and specifications for measuring AC/DC current, voltage, frequency, resistance, capacitance, and temperature.
Preview KAIWEETS HT206A True-RMS Digital Clamp Meter User Manual
User manual for the KAIWEETS HT206A True-RMS Digital Clamp Meter, covering safety information, product overview, function buttons, operating instructions, and specifications. Available in English, German, French, Spanish, and Italian.
Preview KAIWEETS HT206B True-RMS Digital Clamp Meter User Manual
Comprehensive user manual for the KAIWEETS HT206B True-RMS Digital Clamp Meter, detailing its features, safety information, operating instructions, and specifications for various electrical measurements.
Preview KAIWEETS HT206D Digital Clamp Meter User Manual
User manual for the KAIWEETS HT206D True-RMS Digital Clamp Meter, covering safety information, product overview, function buttons, operating instructions, specifications, and maintenance.
Preview KAIWEETS KM100 Digital Multimeter User Manual
Comprehensive guide to the KAIWEETS KM100 Digital Multimeter, covering safety operations, meter diagram, functions, measurement procedures for DC/AC voltage, DC current, resistance, continuity, and diode testing, along with technical specifications and maintenance.
Preview Kaiweets HT118A Digital Multimeter User Manual
Comprehensive user manual for the Kaiweets HT118A digital multimeter, covering its features, functions, and safe operation for various electrical measurements.
Preview KAIWEETS HT113B Digital Multimeter User Manual
Comprehensive user manual for the KAIWEETS HT113B Digital Multimeter, detailing its features, specifications, and operating instructions for accurate electrical testing.