User Guide

Heathrow Scientific Digital Urea & Battery Acid Refractometer

PDF preview unavailable. Download the PDF instead.

1590
Model
MDX-602
DEF / Battery Refractometer
USER GUIDE

Included Parts
6 5 4
1 3
2 1. MDX-Series Digital Refractometer 2. Removable Rubber Grip 3. Removable Prism Cover 4. Disposable Pipet 5. AAA, 1.5V Battery (1ea.) 6. Hardshell Storage Case 7. Screwdriver (not shown)

Introduction

VEE GEE MDX-Series

The VEE GEE MDX-Series portable digital refractometers are a modern design utilizing

Instrument Components

advanced opto-electronic components to

provide years of fast, accurate liquid testing

1

6

with high reproducibility.

2

7

The MDX-Series models are designed to be 3

5

simple to use with little training necessary;

regardless though, it's highly recommended

CONT. ZERO °C/°F
4

that users read this User Guide in it's entirety

before using the instrument for the first

time.

1. Prism Surface

At any point if you have any questions 2. Sample Well

please contact our support team at:

3. LCD Display

4. Control Keys

5. Battery Compartment Cover

6. Removable Prism Cover

Warranty

7. Removable Rubber Grip

VEE GEE digital refractometers are warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. During this period the VEE GEE Service Center will, at their option and without charge, either repair or replace any part found to be defective in materials and workmanship.
All warranty work shall be performed by the VEE GEE Service Center. Contact your dealer or the VEE GEE Service Center for troubleshooting the issue and to receive a Return Authorization, if necessary, for the return of your instrument for repair. The party returning the product must prepay all postage, shipping, transportation, packaging, duties & taxes, and delivery costs to the VEE GEE Service Center.
This warranty is subject to the following limitations and will not apply if:
1) There is lack of proof of purchase date and place of purchase. The warranty is not assignable or transferable.
2) The damage is due to normal wear (including the prism), misuse, abuse, negligence, or any other cause not due to manufacturing of the product.
3) The serial number is altered or obliterated; or unauthorized repair or replacement of partts by any party other than the VEE GEE Service Center.
This warranty expressly excludes transportation damage and readjustment. In no case shall VEE GEE Scientific LLC be liable to the Buyer or any person for any special, indirect, incidental, or consequential damage whether claims are based in contract or otherwise with respect to or arising out of product furnished hereunder. For goods manufactured by any third party, VEE GEE Scientific's liability under warranty is limited to the terms of the warranty by the supplier of the goods.

Instrument Display

6

5

1 4
2
3
1. Battery Level Indicator 2. Reading Display =3. Percent or Permille Scale Indicator 4. Refractive Index Scale Indicator 5. Temperature Display (°C or °F) 6. Active Scale Indicator / Continuous Read Timer

3

4

Battery Level 100-80%

Display Indicator

80-50%

50-20%

<20% Expected battery life: 1000+ reads

Battery Compartment
=

· Unscrew the two screws to open the battery compartment.
· Place AAA, 1.5V (1ea.) battery into the battery slot. Pay careful attention to ensure proper battery polarity when inserting.
5

Control Keys
READ

SCALE

CONT. ZERO °C / °F
(2 SEC)
READ | Power On | CONT. · Press once and release to power on. · Press once and release while powered on to
take reading. · Press and hold for 2 seconds while powered
on to initiate continuous read loop function.
ZERO · Press and hold for 2 seconds then press once
again within 10 seconds to perform zero calibration procedure.
SCALE | °C / °F · Repeatedly press and release to toggle
through all available scales. · Press and hold for 2 seconds to toggle
between °C and °F temperature display.

6

Instrument Use
Power On and Calibration
One press of the power key will power on the instrument (figure 6). In order to maximize battery life the refractometer is designed to shut itself off after 1 minute of inactivity.

figure 6
A zero calibration should be performed daily to achieve optimal accuracy. · Clean and dry the sample well. Place a 4-5
drops of clean, distilled water in the sample well. · Close the prism cover to prevent stray light from affecting the calibration. · Press and hold the ZERO key for two seconds until "CAL" is displayed, then release (figure 7). · "CAL" will flash repeatedly for 10 seconds. Press ZERO once again while "CAL" flashes to perform the calibration. During this time it is crucial to not open the prism cover.
If ZERO is not pressed again then the calibration will not be performed. · The display will indicate "-END" when
finished calibrating (figure 8).

figure 7

figure 8

figure 9 7

· At this point the display will change back to the default screen showing the active scale and instrument temperature (figure 9). If the instrument does not read distilled water with the values shown below then clean the prism and recalibrate with fresh distilled water: Distilled Water 0.0% Brix ±0.2% 1.3330RI ±0.0003RI
Scale Selection
Repeatedly press and release the SCALE key to toggle through all available scales. The display will indicate which scale position is currently active- S01, S02, S03,... (figure 9).
You can find descriptions and ranges for the available scales on the bottom of the instrument, the storage case label, and this user guide.
Taking Measurements Single Readings: · Ensure that the sample well is clean and dry.
Place 4-5 drops of your liquid sample into the well and close the prism cover. · Press the READ key once and the sample reading will display on the screen (figure 10). Additional readings can be made by pressing the READ key again if desired. · Clean and dry the sample well when finished. The prism must be cleaned and dried before testing another sample.

figure 10 8

figure 11

Continuous Loop Readings: Pressing and holding the READ key for 2 seconds will start the continuous loop function. In this mode the refractometer will make 15 sequential reads spaced about 1 second apart, then display an average of all readings as the last reading. During this function the scale indicator will change to a counter and count down from 15 with each read (figure 11).
Troubleshooting
If unexpected results are experienced while testing your samples: · Ensure that you are closing the prism cover when
taking readings and during calibration. · Recalibrate with clean, fresh distilled water and be
sure that you are pressing the ZERO key once again while CAL is flashing on display. · Replace the battery with a new AAA 1.5V battery. · Fully clean the sample well and prism. · Ensure that you are working with highly homogenous samples and that it is free of large particulate matter; filter if necessary. · Keep in mind that as a sample sits in the sample well, suspended solids may settle to the bottom and affect the readings. · When testing samples that are of a different temperature than the instrument and/or environment it is best to wait a few seconds before pressing the READ key to allow the sample to acclimate properly.
Care and Cleaning
· It is important to clean the sample well completely when finished using and between samples to prevent solids from building up on the prism and causing erroneous readings.
· Use distilled water or diluted isopropanol with a soft cloth or tissue to clean the prism surface and sample well.
· Use a mild detergent to clean the instrument body; solvents are not recommended.
· When testing corrosive or acidic samples it is crucial to completely clean the sample well and prism between each reading and be careful to not get any of the sample on the plastic instrument housing.
9

Error Codes

figure 12

figure 13

figure 14

Error Code A01
A02
A03
HHH / LLL (sample) HHH / LLL (temperature)

Description
Calibration Temperature Error: O0-p4e0r°aCti.n(gfigteumrep1e2ra) ture must be
Calibration Sample Error: Use clean distilled water only for zero calibration.
Hardware Failure: Contact our Technical Support department.
Sample High/Low Value Error: Sabalme pralengaeb.o(vfeigourrebe1l3o)w measur-
Temperature High/Low Value Error: Eacncveirpotnambleenrtaanbgoev. e(fiogrubreel1o4w)

10

Model Catalog Number Scale 1 Range Scale 1 Resolution Scale 1 Accuracy Scale 2 Range Scale 2 Resolution Scale 2 Accuracy Scale 3 Range Scale 3 Resolution Scale 3 Accuracy

MDX-602 48602 0.0 - 51.0% DEF (Urea) 0.1% DEF ±0.2% DEF 1.3330 - 1.4200 nD 0.0001 nD ±0.0003 nD 1.000 - 1.500 s.g. 0.001 s.g. ±0.005 s.g.

Model ATC Range
IP Rating Wavelength Sample Cell Calibration Solution Min. Sample Volume Battery Dimensions Weight Warranty

MDX-602 0-40°C 32-104°F IP67: detecting unit IP65: body 589.3nm LED Stainless Steel Distilled Water 0.3mL 1 X AAA, 1.5V 121 x 58 x 25mm 90g 1-Yr Parts & Labor