Calibrating a Remote CXT2 Mounted at the Ceiling

Critical Environment Technologies™

Procedure to Perform ZERO and SPAN Calibrations

  1. Enter Calibration mode from any of the Data Display Screens by swiping the Down/Cal key and then swipe the Edit key to enter Cal Mode.
  2. Apply a clean Zero Gas using the Calibration Cup or be sure there is no background target gas in the monitored area. After the reading is stable, swipe the Edit key to set the Zero Calibration. To skip the Zero calibration, and go to the Span calibration, swipe the Next key. Once a message that the Zero calibration was completed successfully has been displayed, proceed to the next step.
  3. Apply the correct, as indicated, span gas. After the reading is stable, swipe the Edit key to set the Span Calibration. To skip the Span Calibration, swipe the Next key. When a message that the Span Calibration was completed successfully is displayed, the gas detector will exit back to the Data Display Screen.
  4. Remove the calibration gas. Once the Cal Purge Delay has expired, normal alarm and relay functionality will be restored.

WARNING: The SPAN gas used must match the Cal Span Value specified in the CXT. The Cal Span Value may be edited if it becomes necessary to apply a different SPAN gas concentration.

7. When tubing is clear of residual span gas, attach a filter fitting to the end that was attached to the gas cylinder to protect it from clogging or damage and coil or roll it up securely out of the way.

Diagram Descriptions:

A remote CXT2 sensor is installed on or near the ceiling with 6.35 mm (¼ in) tubing attached (Teflon is recommended).

A permanently installed length of calibration tubing is attached to the calibration adapter on the remote CXT2 sensor. The other end can be easily attached to a cylinder of gas during the calibration process.

The calibration adapter is screwed onto the end of the remote sensor.

The standoffs create space for the gas to pass over the sensor.

The calibration tubing is attached to the nozzle.

A remote sensor can be mounted up to 4,000 ft or 1,219 m away from the CXT2 transmitter.

Calibration is performed at the main CXT2 which is mounted at eye level.

*When finished calibration, detach the tubing from the gas cylinder. If allowing diffusion to clear the tubing, the CXT2 may not return to a zero reading for several minutes. If using zero air with a 0.5 LPM regulator, expect 5 seconds per 10 meter length before the CXT2 returns to a zero reading. When the reading reaches zero stop the air flow.

Models: CXT2, Gas Detectors, CXT2 Gas Detectors

File Info : application/pdf, 1 Pages, 515.41KB

PDF preview unavailable. Download the PDF instead.

Calibration-Remote-CXT2

References

Adobe PDF Library 16.0.7 Adobe InDesign 17.2 (Windows)

Related Documents

Preview Critical Environment Technologies Internal Silicone Heater Quick Reference Guide
Quick reference guide for the internal silicone heater used in Critical Environment Technologies (CET) transmitters, detailing compatibility, installation, and power requirements for cold environment applications.
Preview RDM Remote Display Module Operation Manual
This operation manual provides detailed instructions for the Critical Environment Technologies RDM Remote Display Module, covering installation, operation, configuration, maintenance, and troubleshooting.
Preview Critical Environment Technologies cGas-SD Sample Draw Gas Detector Datasheet
Datasheet for the Critical Environment Technologies cGas-SD Sample Draw Gas Detector system, detailing its features, technical specifications, product codes for various gas sensors (CO, H2S, NO, O2, SO2, NH3, Refrigerants), and accessories for monitoring hard-to-access locations.
Preview cGas-SD Sample Draw System Installation Manual - Critical Environment Technologies
Installation manual for the Critical Environment Technologies cGas-SD Sample Draw System. This document provides detailed instructions for installing, configuring, and operating the cGas-SD gas detection unit, including safety information, technical specifications, and troubleshooting guides.
Preview Critical Environment Technologies Remote Switch Datasheet
Datasheet for Critical Environment Technologies' Remote Switch (RSW-E and RSW-S), detailing features, specifications, applications, and wiring diagrams for manual equipment operation, emergency shutdowns, and ventilation control in gas detection systems.
Preview cGas-SD Sample Draw Guide - Critical Environment Technologies
A comprehensive guide for the cGas-SD Sample Draw System from Critical Environment Technologies, detailing its operation, maintenance, sensor replacement, and troubleshooting for accurate gas sampling.
Preview Critical Environment Technologies: DCC-MRI & DCC-SD Sample Draw System End of Sale Notification
Critical Environment Technologies (CET) announces the end of sale for its DCC-MRI and DCC-SD Sample Draw Systems. This notification details the reasons for discontinuation, the replacement product (cGas-SD), compatibility, and the end-of-service timeline.