Instructions for SCHREIBER models including: SCHREIBER, SG-600, Oscillating Plastersaw

GA-SG-600-E

A. Zepf

Instructions of use for Ed. 2.2 05/2018 Oscillating ...

Instructions of use for Oscillating Plastersaw SG-600 SCI Gmb nterer Damm 1 D Fridingen Tel: 4434 Fa: 4434 http://wwwschreibergmbhde infoschreibergmbhde 3 GA-SG-600-E

Content - Gima

Content. Validity. 1. Certificate of compliance. 1. Purpose of use. 2. Inspection. 2. First use. 2. Starting the machine. 2. Adjustment of oscillating speed ...


File Info : application/pdf, 4 Pages, 218.69KB

PDF preview unavailable. Download the PDF instead.

M59100EN
GA-SG-600-E Ed. 2.2 05/2018 IFU Code 1080

Instructions of use for Oscillating Plastersaw
SG-600

Content
Validity ______________________________________________________________________________________ 1 Certificate of compliance _______________________________________________________________________ 1 Purpose of use________________________________________________________________________________ 2 Inspection____________________________________________________________________________________ 2 First use _____________________________________________________________________________________ 2 Starting the machine ___________________________________________________________________________ 2 Adjustment of oscillating speed _________________________________________________________________ 2 Changing saw blade ___________________________________________________________________________ 2 Saw blades ___________________________________________________________________________________ 2 Safety Instructions ____________________________________________________________________________ 2 Scope _______________________________________________________________________________________ 2 Opening of bandages __________________________________________________________________________ 2 Special instructions ___________________________________________________________________________ 3 Additional Tips ________________________________________________________________________________ 3 Troubleshooting_______________________________________________________________________________ 3 Care_________________________________________________________________________________________ 3 Maintenance__________________________________________________________________________________ 3 Guaranty_____________________________________________________________________________________ 3 Technical data ________________________________________________________________________________ 4 Delivered parts _______________________________________________________________________________ 4 Spare parts/Accessoires _______________________________________________________________________ 4

Validity These instructions are valid for all SG-600 Plaster Saws and accessoires, sold by SCHREIBER Instruments.
Certificate of compliance Manufacturer´s name and address: SCHREIBER GmbH Chirurgische Instrumente Unterer Damm 15 D-78567 Fridingen

Product name: SG-600-01 and 02, oscillating plaster saw and accesoires (saw blades) in accordance to the valid catalogue and price-list.

The Schreiber GmbH confirms herewith in sole responsibility, that the above mentioned products comply with the requirements according to annex I of the council directive 93/42/EEC and EN 606011-2. These products are permitted to carry the CE-mark. We have established a certified quality management system according to DIN/EN/ISO 13485

Signature

Armin Zepf Quality Manager

SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-996949-0, Fax: +49-7463-996949-9, http://www.schreibergmbh.de, info@schreibergmbh.de

1

Instructions of use for Oscillating Plastersaw
SG-600

GA-SG-600-E Ed. 2.2 05/2018 IFU Code 1080

With purchasing this SCHREIBER device you have decided to use a high quality product. Proper function and safety are only guaranteed for a long time, if the following points are strictly observed.
Purpose of use Cutting of plaster and synthetic bandages. The safety rules of MDD (medical device directive) and the national laws for safe devices (Arbeitsmittel-Gerätesicherheitsgesetz) are valid.
Inspection SCHREIBER oscillating plaster saws must be inspected for completeness , damages or faults immediately after transportation. Claims for compensation are only possible, if shipper or dealer are informed immediately.

ning. The best oscillating frequency is

depending on the material and condi-

tion of the bandage.

Lowest frequency:

(1)

Highest frequency: (8)

Changing saw blade -unplug cable to prevent unintentional turning on. Use protection gloves to prevent injuries caused by the saw blade. 1.Lay down the saw on a stable and even surface (table). 2.Place one open end wrench (1) behind the saw blade on the shaft, the other (2) in front of the blade on the locking screw (3). 3. Retain the wrench (1) and use it for fixing the saw at the same moment. 4. Untighten locking screw (3) with the other wrench (2).

First use SCHREIBER oscillating plaster saws must not be used, before these instructions are worked through.

· For use and maintenance the valid (national) laws for safety and maintenance must be observed. · make sure, that the voltage on the machine is similar to your socket. · Check cables and extensions regularly. ·donotscreworrivetexternallabels on the body of the machine. This may cause electrical danger.

5.Remove the locking screw (3) and the fixation washer (4). 6. Take care to loose no parts.

Starting the machine ·Makesure,thattheplugisconnected to the socket properly. · Plaster saw is equipped with an electronic speed control button. The optimal speed must be adjusted, depending on the sawblade and use. Speed can be adjusted by turning the button (2) (pre-selection= 4). ·Turn switch  (1) forwards to "ON"position for turning on and backwards for turning off.
Adjustment of oscillating speed Adjust speed, when the motor is run-

7. Remove saw blade (5) 8. Clean thread and bearing surface 9. Insert new saw blade and center on the shaft. IMPORTANT: The pin on the axis must fit into the notches in the saw blade (6)perfectly (on both sides). 10.Mount the fixation washer (4). Notches of the washer must show towards the saw and must fit to the pins, too. 11. Screw down locking screw by hand and tighten with wrench (6-7 Nm). Make sure, that the shaft is not turned during untightening and tightening

the screw. Turning the excentric axis inside the machine may damage the transmission.
Saw blades Attention: blunt saw blades must be changed immediately. Hot saw blades may burn the patients skin and overload the machine. Available saw blades are shown on tha last page. Only use original SCRHEIBER blades.
Safety Instructions ·Onlyuseoriginalspareparts. ·Improperusemaycauseoverheating and damage of the machine. · Never bring the machine in contact with water or other solutions. ·Maschineisnotexplosionprotected. Keep away from flammable liquid or gas. · It is recommended to wear safety equipment (protection glasses, gloves, ear protectors)
Scope Generally it is not recommended to misuse the saw for medical or other purposes, that means the machine must only be used as saw for plaster or sysnthetic bandages. It is not suitable for surgical operations. The patient must not have direct contact to the sawblade.
Opening of bandages ATTENTION: Unpadded bandages must be opened with special care, because sticked skin might be burned! Choose a suitable saw blade first. Place the saw with the fixed saw blade on the bandage with slightly pressure. The thumb can serve as guide and depth gauge and prevents abruptly penetration into the bandage. As long as the saw blade is cutting, there is a resistance, that decreases, when the saw blade goes through the bandage. Then the saw must be lifted without beeing taken out of the groove and mooved about 15 mm into cutting direction. Repeat these steps, until the complete bandage is opened by several linear cuts.
Special instructions ·Bandageswithmetalinsertsshould not be opened with this saw.

2

SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-996949-0, Fax: +49-7463-996949-9, http://www.schreibergmbh.de, info@schreibergmbh.de

GA-SG-600-E Ed. 2.2 05/2018 IFU Code 1080

Instructions of use for Oscillating Plastersaw
SG-600

·Avoidpermanentfrictiononthesoft furnishing (heat).
Additional Tips
The various bandage materials have very different attributes. Please try to find out the best settings during performance. Usually synthetic bandages need lower speed, to avoid melting of the material. The cutting of unpadded bandages should be avoided as far as possible(see"openingofbandages"). If a close bandage is necessary, it is useful to integrate a gaze or cotton stripe on the patients skin and mark the position on the bandage with a marker. Later the bandage can be opened along the line. After surgical operations, the position of the wound should also be marked on the bandage, so it is possible to open a window for wound inspection, extraction of wires or incisions. The machine is equipped with a temperature switch, that avoids overloads. If the safety switch has turned the machine automatically off, the saw can be resaterted after a short cooling period of a few minutes. Blunt saw blades are overloading the motor. To reach a balanced abrasion, the saw blade can be turned to 90°.

and duration.
Maintenance Electrical devices must only be checked,repaired and maintenanced by experts. Alway unplug the machine, when it is not used. Blunt saw blades must be removed, to avoid overheating of the machine. Saw must be stored under dry conditions. Device must not be disassembled. There are no components inside, that can be repaired by the user.
Guaranty We give 15 months guaranty on this machine and accessoires on material and manufacturing faults. Improper use, normal wear, repair by third parties and modifications void guaranty and CE-marking. Please return the saw with an invoice copy to the manufacturer. Lables mounted by the manufacturer must not be removed.

Troubleshooting REPAIR OF THE MACHINE MUST BE MADE BY EXPERTS ONLY! If the saw does not work: · Let the motor cool down for a few minutes and retry to turn on. ·Checkthecablefordamages ·Checkthefuse. If the saw is too loud: Turn off and send to the manufacturer for inspection.

Care The machine does not need special service, but it should be cleaned with a moist duster regularly. The saw must not be immersed into liquids or rinsed with water. Take care, that no plaster debris is disturbing the oscillating shaft or the on/off switch. Keep the airing grids free of plaster debris and make sure that the cool air can circulate properly. The saw should be cleaned and checked all 6 months, depending on use

SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-996949-0, Fax: +49-7463-996949-9, http://www.schreibergmbh.de, info@schreibergmbh.de

3

Instructions of use for Oscillating Plastersaw
SG-600

GA-SG-600-E Ed. 2.2 05/2018 IFU Code 1080

Technical data
Ref.-N° Voltage Power
Strikes Noise, A- typical Weight
Conformity
Protection Class IP Code

SG-600-01

SG-600-02

220-240V/50-60Hz

110-120V/50-60Hz

250Watt

0.8A

1.6A

12000-21000 /min

68 - 82 dB(A)

1.3 Kg (1.7 kg with cable)
CE, 92/43 EWG, 62/5/EWG, EN 60101-1, IEC 60601-1, EN 60601-1,2, IEC 606011-2
II

IP2X

Delivered parts
SG-600-01 (02) SG-200-50H SG-200-65H SG-600-15 GA-SG-600-01D

1 Saw in Cardboard box 1 sawblade round for synthetics and plaster, Ø 50 mm 1 sawblade round for synthetics and plaster, Ø 65 mm 1 Set open end wrenchs 1 instructions of use

Spare parts/Accessoires

SG-600-10

Plastic case with insert

SG-200-50H 1 sawblade round for synthetics and plaster, Ø 50 mm

SG-200-65H 1 sawblade round for synthetics and plaster, Ø 65 mm

SG-600-15

1 Set open end wrenchs

SG-600-20

1 locking screw

SG-600-21

1 fixation washer

4

SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-996949-0, Fax: +49-7463-996949-9, http://www.schreibergmbh.de, info@schreibergmbh.de

