Neuation iFUGE L1000 Product User Manual

This document provides comprehensive instructions for the Neuation iFUGE L1000 centrifuge.

1. Introduction

This centrifuge is equipped with a maintenance-free motor drive, a large digital display & simple interface for efficient operation for daily lab usage. The programmable centrifuge can deliver up to 4200 RPM and can accommodate different types of rotors. It features various programmable modes to save time & adds convenience.

2. Intended Use

Centrifuge is used in laboratories to separate particles from liquid suspension according to their density. The different biological substances that are usually separated by centrifugation are cells, mammalian cells, organelles, etc. It is majorly used in laboratories.

Note: Before using the centrifuge, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repairs please contact the supplier.

3. Features

4. Standard Accessories

5. Technical Specifications

Parameter Value
Motor Type Brushless DC Motor
Max capacity 1000 ml (4 x 250 ml)
Speed Setting Variable 500 - 4200 rpm (Step of 100 rpm)
Speed Accuracy ± 100 rpm
Run Time 1 min to 99 mins & infinite mode
Min. Acceleration Time 60 seconds
Min. Deceleration Time 60 seconds
Noise Level <65 dB
Ambient Temperature 5-40°C
Permissible Relative Moisture <80%
Size (L x B x H) 475 x 585 x 360 mm
Weight 24.7 Kgs (without rotor)
Input Power 230 VAC ± 10 VAC
Power Consumption 400 W

6. Safety Precautions

7. Installation

The Laboratory centrifuge is supplied in packaging box. Open the box, remove the packaging and gently place the centrifuge out of the box. Before 1st time usage, open the centrifuge & ensure to remove all packaging from the rotor chamber. Please keep all packaging in safe storage for at least 2 years for warranty purpose.

Location & Mounting

Place the centrifuge on a flat, solid and leveled surface and ensure that all the four feet of this centrifuge stand on the surface firmly. Avoid installing on slippery or surface prone to vibration.

8. User Interface & Display

The centrifuge features a control panel with various buttons and a display for setting parameters and monitoring operation. The following table describes the interface elements:

Item Button Name Function
1 Start Single press, motor will start as per selected rpm and time if lid closed. Start button works only if lid closed. Motor start or stop, indicates with LED (12)
2 Stop/open Single press stop/open button, will stop motor as per defined DCC time if motor is running. Second press stop/open button, will open the lid if motor stopped. Also use as lid (11) open function if centrifuge motor is not running.
3 RPM/RCF Long Press, RPM/RCF will toggle display to RPM to RCF. RPM, RCF mode also indicate via LED on panel.
4 Time Single press, Time will be display like 99.00. Timer can be set from 1 to 99 min & infinite (][)
5 Increment / Decrements Short press of {+}Plus and {-}Minus buttons will increase and decrease speed, time, ACC, DCC values normally. Long press of these keys would change the parameters faster
6 ACC (Acceleration) First Single press, Acceleration will display on time segment like ACC1. We can change acceleration using {+} and {-} buttons. (1 to 9).
7 DCC (Deceleration) Second Single press, Deceleration will display on time segment like DCC1. We can change Deceleration using {+} and {-} buttons. (1 to 9).
8 Rotor To select among multiple rotors listed on page 6 & 7
9 PRG Press PRG button to select the program from 1-99. The different parameters can be set in each program for quick selection.
10 Quick Spin Press hold for immediate centrifugation up to last set RPM
11 LID (Indicates lid status)
12 LED (Indicates motor start/stop status)

9. Operating the Centrifuge

Switch On the Centrifuge

After connecting the power cord, switch ON the main power supply & then switch on the power switch located on the back side & then right side of the instrument. Make sure to check the rotor fitment before use. Centrifuge will not operate with open lid.

Note: Maintain a gap of at least 3 seconds between switch OFF and switch ON. DO NOT switch OFF and ON instantly.

Speed Setting

Press RPM/RCF button to set speed from 500 to 4200 RPM. Single press to set speed, long press to toggle between RPM/RCF. The parameter can be set when the speed screen blinks, by using “+” button for increment & "-" button for decrement. The values will automatically saves after screen stops blinking.

Timer Setting

Press the timer button to set timer. Time can be set by "+" for increment & "-" for decrement. The timer will save automatically once the blinking stops. The timer can be set in minutes only while seconds are for countdown timer while running operation.

Acc/Dcc

This button will set the parameters of acceleration or deceleration of the rotor ramp. Press acc/dcc button once to set the acceleration from 1-9 by using "+/-" button for increment/decrement respectively. Similarly press once again the acc/dcc button to set deceleration from 1-9 by using "+/-" button for increment/decrement respectively. The values of acceleration & deceleration timings are listed below.

- dcc L 480 Seconds Acc 5 dcc 5 120 Seconds
dcc 0 300 Seconds Acc 6 dcc 6 105 Seconds
Acc 1 dcc 1 180 Seconds Acc 7 dcc 7 90 Seconds
Acc 2 dcc 2 165 Seconds Acc 8 dcc 8 75 Seconds
Acc 3 dcc 3 150 Seconds Acc 9 dcc 9 60 Seconds
Acc 4 dcc 4 135 Seconds

PRG

Press PRG button to select the program from 1-99. The different parameters can be set in each program for quick selection. To set parameters in each program follow the parameters (Speed, time & Acc/Dcc) as described above. To select from programs press using +/- button for increment or decrement. All the parameters in the program saves automatically once the blinking stops. The saved programs also can be edited by following the instruction of setting Speed, time & Acc/Dcc as described above.

Quick Spin

Press & hold the quick spin button for immediate centrifugation of the samples up to previous run RPM/RCF. Releasing the hold will stop of the rotor gradually.

Pedit

Press PEdit button to enter all the different parameters - Speed, Timer, acceleration & deceleration automatically one after the other. These parameters will be asked by the instrument sequentially to be entered if required to be changed (if not edited the display value will set). These parameters will save automatically.

Rotor

Press rotor button to select rotor from the listed below. The rotor numbers can be changed by using "+/-" for increment or decrement of the rotor number.

Swing Out Rotors

Rotor Rotor No. Max. Volume Max. RPM Max. RCF
L1000-156 Rotor 21 4 x 100 ml with lid 4200 2766
L1000-142 Rotor 22 4 x 100 ml 4200 2884
L1000-141 Rotor 23 4 x 50 ml 4200 2884
L1000-139 Rotor 24 16 x 15 ml 4200 2884
L1000-140 Rotor 25 32 x 6 ml 4200 2766
L1000-124 Rotor 26 2 x Microplates 4200 2074
L1000-161 Rotor 27 4 x 250 ml 4200 3155

Fixed Angle Rotors

Rotor Rotor No. Max. Volume Max. RPM Max. RCF
L1000-146 Rotor 31 4 x 100 ml 4200 2272
L1000-147 Rotor 32 4 x 50 ml 4200 2331
L1000-148 Rotor 33 6 x 50 ml 4200 2410
L1000-150 Rotor 34 16 x 15 ml 4200 2410
L1000-149 Rotor 35 24 x 15 ml 4200 2489

Start & Stop

The centrifugation can be started by pressing the start button once the desired operation parameters are set. The operation can be stopped by pressing the stop button. To open the lid, press the stop/open key after the rotor comes to stop.

10. Maintenance and Cleaning

Note: Liquid should not come into contact with the motor.

11. Troubleshooting

This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed on the display screen and the operator can determine the malfunction with the warning code below.

ERROR PROBLEM SOLUTION
No display No main power connection. Power check & proper plug-in of mains cable at both ends.
Power failure Check the mains fuse of the lab.
Improper connection. Connect adaptor properly.
[Lock symbol] Lid not closed correctly. Close lid correctly.
Error with lid closing and opening mechanism. Connect power cord properly.
Err 55 Rotor not loaded symmetrically. Load rotor symmetrically & restart centrifuge.
Rotor is still spinning. Wait for the rotor to come to a stop.
Centrifuge lid cannot be opened Power failure Emergency lid release after rotor stops
Centrifuge shakes during acceleration & exceptional running noise Rotor not loaded symmetrically. Load rotor symmetrically & restart operation
Either a broken tube, damage to the rotor or motor is cause for run noise. Replace broken tube. For damaged rotor/motor contact service representative.
Display error Rotor damaged. Remove & change rotor
Loose connection of display. Contact service representative.
Err 1 Latch damaged, Latch jammed Contact service representative.
Err 52 Motor stuck or incorrect operating voltage Turn OFF the centrifuge, Check rotor fitment or apply correct 230VAC ± 10VAC operating voltage
Power tripping Cable not fit properly. Remove cable and connect properly.
Last run memory not displayed Turing ON centrifuge immediately after turning it OFF. Maintain 3 seconds gap Between switch OFF and switching ON again.
System gets hang Electronics error. Switch off centrifuge and then switch it ON again. If the error still shows, contact service representative.

IMPORTANT NOTE:

12. Warranty Statement

This product is warranted to be free from defects in material and workmanship for a period of Two (2) year from date of purchase. Your product will be duly repaired upon prompt notification in compliance with the following conditions:

Serial No. Purchase Date

For your reference, make a note of the serial number, date of purchase and supplier here.

13. Product Disposal

Information on the disposal of electrical and electronic devices in the European Community.

The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business to business sphere, to which this product is assigned, may no longer be disposed off in municipal or domestic waste. They are marked with the following symbol to indicate this.

[WEEE symbol: Crossed-out wheeled bin]

As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.

Models: iFuge L1000, Centrifuge Device, iFuge L1000 Centrifuge Device