Safe-Accu 2 Blood Glucose Meter User Manual

Manufacturer: Sinocare

Version: 181107-1

Introduction to the Your Blood Glucose Monitoring System

Intended Use

The Safe-Accu 2 Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.

The Safe-Accu 2 Blood Glucose Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes.

The Safe-Accu 2 Blood Glucose Monitoring System includes: Safe-Accu 2 blood glucose meter, Safe-Accu 2 blood glucose test strip, blood glucose control solution.

Note: The blood glucose control solution is optional.

The Safe-Accu 2 meter displays blood glucose results in mmol/L or mg/dL. Each meter only displays one unit of measurement. The unit is preset and cannot be changed. Please confirm the unit before testing.

Test Principle

A blood glucose test is based on the measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.

Important Safety Information

  • For self-testing customers, your Safe-Accu 2 meter and lancing device are for single person use only. Do not share them with anyone else, even family members.
  • For professional use, if the meter is being used on a second person, the meter and lancing device should be disinfected first.
  • Safe-Accu 2 test strip and lancet are for single use only. DO NOT REUSE.
  • Do not use other test strips and control solutions with the Safe-Accu 2 meter.
  • Do not use the meter if it is not working properly, or if there is any damage.
  • Keep the Safe-Accu 2 test strip vial away from children. The strip vial, test strips, and control solution bottles can be choking hazards. DO NOT drink the control solution.
  • Remove the batteries if the meter is not likely to be used for some time.

DO NOT CHANGE YOUR TREATMENT BASED ON A SINGLE RESULT THAT DOES NOT MATCH HOW YOU FEEL OR IF YOU BELIEVE THAT YOUR TEST RESULT COULD BE FALSE.

Patient Cautions

  • Not for neonatal (newborn or infant) use.
  • Not for screening or diagnosis of diabetes mellitus.
  • Not for use on critical patients.
  • For in vitro diagnostic use only.

Important Health-Related Information

The American Diabetes Association (ADA) suggests the following targets for normal blood glucose range. More or less stringent goals may be appropriate for each person. If your test result is out of the range or your test result is not in accordance with your feeling, retest your blood glucose. If still out of the range, please contact a healthcare professional.

Expected Results for people without diabetes:

Plasma Blood Glucose ResultFasting Plasma Glucose
<100 mg/dl (<5.55 mmol/L)<100 mg/dl (<5.55 mmol/L)

FOR Safe-Accu 2 meter:

  1. If test results are greater than 600 mg/dL (33.3 mmol/L), the meter will display "HI". Retest your blood glucose immediately with a new test strip. If your reading is still "HI", contact a healthcare professional immediately.
  2. If test results are lower than 20 mg/dL (1.1 mmol/L), the meter will display "LO". Retest your blood glucose immediately with a new test strip. If your reading is still "LO", contact a healthcare professional immediately.
  3. If test result is >70 mg/dL (3.9 mmol/L) and < 126 mg/dL (7.0 mmol/L), the meter will display the test result with a green light on.
  4. If test result is >126 mg/dL (>7.0 mmol/L) and < 250 mg/dL (13.9 mmol/L), the meter will display the test result.
  5. If test result is >70 mg/dL (3.9 mmol/L) or >250 mg/dL (>13.9 mmol/L), the meter will display the test result with a red light on.

Note: Do not change your treatment solely based on test results and light indications. Please consult a healthcare professional for your treatment.

About Your Safe-Accu 2 Meter

Full Screen Display

The meter display shows various indicators:

  • Blood drop symbol: ? Indicates readiness to apply blood sample.
  • Battery symbol: ? Indicates battery status.
  • Unit of measure: Displays 'mmol/L' or 'mg/dL'.
  • Temperature symbol: ?️ Indicates ambient temperature.

Button Functions

ButtonFunctionAction
M and ^/vTo turn the meter on and enter the blood glucose test result review mode. To turn the meter off. To delete the blood glucose test results. To confirm date/time settings. To adjust settings for date-time/ scroll through test results.Press and hold M for 3 seconds. Press and hold M for 3 seconds to turn off. In test results review mode, press and M together to delete all blood glucose test results. Press and hold M. Press and release ^ or v.
MTo enter date-time setting mode.Press and hold M for 3 seconds.

Setting Up Your Safe-Accu 2 Blood Glucose Meter

Setup Steps for Date/Time

The Safe-Accu 2 blood glucose meter is preset with the date and time. You may need to adjust it to your local time zone. Check the date and time each time the batteries are replaced. Reset the date and time if they are not correct.

STEP 1: Insert the Batteries

Open the battery cover on the back side of the Safe-Accu 2 blood glucose meter. Insert two AAA alkaline batteries as indicated by the '+' and '-' symbols.

STEP 2: How to Set the Date and Time

  1. Starting with the meter turned off, press and hold the M button for 3 seconds.
  2. The meter enters the Date and time setting mode.
  3. It displays "d-1" and the year flashes.

STEP 3: Setting the Year

The year appears flashing at the top right side of the meter display.

  1. Press and release ^ or v to increase or decrease the year.
  2. Press and release M to confirm the year.

STEP 4: Setting the Month

The month appears flashing at the lower left side of the meter display.

  1. Press and release ^ or v to increase or decrease the month.
  2. Press and release M to confirm the month.

STEP 5: Setting the Day

The day appears flashing at the top right side of the meter display.

  1. Press and release ^ or v to increase or decrease the day.
  2. Press and release M to confirm the day.

STEP 6: Setting the Hour

The hour appears flashing at the lower left side of the meter display.

  1. Press and release ^ or v to increase or decrease the hour.
  2. Press and release M to confirm the hour.

STEP 7: Setting the Minutes

The minutes appear flashing at the top right side of the meter display.

  1. Press and release ^ or v to increase or decrease the minutes.
  2. Press and release M to confirm the minutes.

STEP 8

Press and release M. The date and time are set, and the meter turns off.

Note: Time is displayed in 24-hour format only. If the battery is replaced, it may cause the date and time to reset. Resetting the date and time does not change the timestamps of previous test results.

Note: Unit of Measure is factory preset to mg/dL or mmol/L. It cannot be changed by the user.

Control Solution Testing

Why Perform a Control Solution Test

  • Ensures that your meter and test strips are working properly.
  • Allows you to practice testing without using blood.

When Should You Perform a Control Solution Test

  • When you begin using a new vial of Test Strips.
  • When the vial cap is opened for a long time.
  • When Test Strips have been exposed to extreme environmental conditions.
  • When you want to check whether the meter or test strips are working properly.
  • When you drop or break the meter.
  • When you want to check whether the test steps are correct.

Important Information

  • Use only Sinocare blood glucose control solution with your Meter.
  • Check the expiration date on the control solution bottle. Do not use if expired.
  • Use the test strips within a period of 6 months from the date that you first open the vial.
  • Record the opened date on the control solution bottle, which is 3 months from when you first open it. Discard the bottle after 3 months.
  • Discard any bottle that appears to be cracked or leaking.
  • For in vitro diagnostic use only.

Performing a Control Solution Test

Start with the blood glucose meter off. Prepare the following items before control solution test: a. blood glucose meter; b. blood glucose test strip; c. blood glucose control solution.

STEP 1: Wash your Hands

Wash your hands with mild soap and water. Be sure to dry your hands before performing a test.

STEP 2: Insert a Safe-Accu 2 Test Strip

Insert a Safe-Accu 2 Test Strip with the printing face up and the contact end into the test strip port until it cannot go any further. The blood glucose meter will turn on.

Attention: Only when the meter displays a flashing blood drop symbol "?", can you apply control solution. If there is no blood drop symbol, restart the meter.

STEP 3: Apply the Control Solution

  1. Check the expiration date and discard dates on your control solution and Safe-Accu 2 test strip vials. Do not use control solution or test strip which expired.
  2. Shake the control solution bottle well, then remove the cap. Squeeze the bottle and discard the first drop. Squeeze the bottle again to get a second drop and bring the tip of the Safe-Accu 2 test strip to touch the drop of solution until the blood glucose meter beeps. After a 10-second countdown, the meter will display the control solution test result.
  3. Control test results must be within the range of the control solution level printed on the test strip vial label. If the result is out of range, test again. If still out of range, do not use test strips or meter for blood glucose testing. Call customer service at +86-731-89935581/+86-731-89935582 or contact local dealers for help.
  4. Eject the test strip.

Attention: Discard used test strip and control solution according to local regulations.

Keep control solution away from children and pets.

Questionable Results:

Probable CauseAction
Operation mistakeRetest according to correct steps.
Control solution not being shaken wellShake control solution well and retest with a new test strip.
Use the first drop of control solutionRetest with a new test strip.
Expired or contaminated control solutionChange to a new vial of control solution and retest.
Expired or contaminated test stripChange to a new vial of test strip and retest.
Temperature is too high or too lowEquilibrate meter, test strip, and control solution at room temperature for 30 minutes and retest.
Meter malfunctionContact customer service or local dealers.

Blood Glucose Testing

Prepare meter, strip, lancing device, and lancet before testing.

STEP 1: Prepare Finger

Wash your hands or use an alcohol swab to clean the fingertip. MAKE SURE YOUR HAND IS DRY BEFORE TESTING.

STEP 2: Prepare Lancing Device

Unscrew the lancing device cover.

STEP 3: Insert Lancet

Insert a lancet into the lancing device until it comes to a full stop.

STEP 4: Remove Lancet Protector

Twist and pull to remove the lancet protector. Retain the lancet protector for safe lancet disposal.

STEP 5: Reassemble Lancing Device

Screw the lancing device cover back on.

STEP 6: Adjust Puncture Depth

Twist the adjustable comfort tip to adjust the puncture depth. Tips: 1-2 (least depth) for soft or thin skin, 3-4 (average depth) for average skin, 5 (deepest depth) for thick or calloused skin.

STEP 7: Arm Lancing Device

Pull the spring cover back to arm the lancing device. The Sinocare lancing device is prepared and ready to lance your finger for a blood sample. The lancet is for single use only. To prevent the risk of bloodborne pathogen transmission, it shall not be shared with other people.

STEP 8: Prepare Test Strip

Take a strip out of the strip vial. Immediately close the vial lid.

Attention: Finish test within 3 minutes after taking out the strip. Otherwise, the test result may be inaccurate.

STEP 9: Insert Test Strip

Insert strip into the test strip port. For Safe-Accu 2, the meter will turn on.

Tips: Insert the strip with the printed side facing up and contact end into the strip port.

Attention: If blood is not applied within 3 minutes, the meter will automatically turn off. Reinsert the strip and begin the test.

If the meter displays "E-2", it means the environment temperature is beyond 40°C or below 5°C. Please place the meter, strip, and control solution in an appropriate place (10°C~35°C), wait at least 30 minutes, and test again.

STEP 10: Lance Finger

Make sure the lancing device is pressed firmly to your finger. Press the button to lance.

Attention: If you did not get enough blood sample, please gently massage your finger. DO NOT SQUEEZE YOUR FINGER. Refer to Step 6 to adjust puncture depth and lance again.

STEP 11: Apply Blood Sample

When the meter screen displays a flashing blood symbol (?), apply the blood sample to the reaction chamber.

Attention: Hold the tip of the test strip to the blood drop until the meter beeps. Visually confirm that the blood sample fills the entire reaction chamber at the end of the test strip. Do not move the meter or strip and press the main button when testing. If you did not apply enough blood sample to the test strip, use a new test strip to test again. Do not use the test strip by pressing the finger firmly, otherwise it may cause inaccurate test results.

Caution:

  • DO NOT smear or scrape the blood onto the Test Strip.
  • DO NOT apply blood to the Test Strip when the Test Strip is not in the test strip port.
  • DO NOT put blood or foreign objects into the test strip port.

STEP 12: View Test Result

After a 10-second countdown, the meter displays the test result in mg/dL or mmol/L.

Example: 90 mg/dL

If the meter displays "E-3", please read the blood glucose testing section carefully and use a new test strip to test again.

STEP 13: Eject Strip

Eject the strip.

STEP 14: Dispose Lancet

Remove the lancing device cover. With the lancet protector on the table, stick the needle of the lancet into the protector. Eject the lancet and screw the lancet cover back on.

Used test strip and lancet may be biohazardous. Please discard them carefully according to local regulation requirements. If the meter is used by multiple people, use 75% alcohol to clean the meter surface after testing to avoid pathogen infection.

Viewing Test Result

After the blood glucose meter counts down from 10, your blood glucose result appears along with the unit of measure. This blood glucose result is stored in the meter memory. Turn the meter off by removing the Safe-Accu 2 test strip. Discard the used test strip carefully to avoid contamination.

  • If the test result is above 600 mg/dL (33.3 mmol/L), the screen displays "HI". Use a new test strip to test again. If it still displays "HI", please contact a healthcare professional immediately.
  • If the test result is below 20 mg/dL (1.1 mmol/L), the screen displays "LO". Use a new test strip to test again. If it still displays "LO", please contact a healthcare professional immediately.

Attention: The test results of "HI" or "LO" will not be stored in memory.

Example Displays:

mg/dL unit display example

mmol/L unit display example

Memory

The Safe-Accu 2 meter stores 200 blood glucose test results. You can review individual results by entering the memory mode. Once 200 results are in memory, adding a new result causes the oldest one to be deleted. It also provides 7-, 14-, and 28-day averages of your blood glucose test results. Recall test results in memory with these easy steps.

Step 1: Enter Memory Mode and Review Averages

With the meter off, press and hold the M button for 3 seconds to turn the meter on. The meter will beep and display the 7-day average of your blood glucose test result.

Press M once, the screen will display the 14-day average of your blood glucose test result. Press M twice, the screen will display the 28-day average of your blood glucose test result. Press M three times, you will see the last test result displayed. Press ^ or v to view all the blood glucose results in memory.

If no test results are in memory, the meter will display 3 dashes with a beep.

Step 2: Delete Memory

How to delete blood glucose test results:

  1. Press and hold M for 3 seconds to turn the meter on. The meter will enter the test results review mode.
  2. Press and release ^ and v at the same time. The screen displays 3 dashes and the meter makes a short beep sound at the same time. All stored blood glucose test results are deleted.

Press M button for 3 seconds to turn the meter off.

Note: You cannot review your test results when the test strip is in the test strip port.

Information for Healthcare Professional Use

  • Only trained healthcare professionals should be allowed to use the Safe-Accu 2 blood glucose monitoring system on multiple patients.
  • Only trained healthcare professionals should be allowed to collect venous blood samples.
  • Any patient who has infectious diseases must use their own meter.
  • If the same meter is used on multiple patients, there is a risk of infection between both patients and healthcare professionals.
  • Water or other disinfection solutions remaining on the skin may dilute blood and cause inaccurate test results.
  • Please dispose of used lancets, test strips, and expired control solution as medical waste.

Perform Test:

  1. Wear disposable gloves.
  2. Healthcare professionals should use an alcohol swab to clean the lance site. Then ensure the site is dry before testing.
  3. Use a qualified lancing device and choose a suitable penetration depth for patients.
  4. Using the lancing device to lance.
  5. Hold the tip of the test strip to the blood sample until the test strip reaction zone is full and the meter begins countdown.
  6. Wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips.
  7. Clean and disinfect the meter and lancing device before use on the next patient.

Comparing Meter with Laboratory Results

When comparing results between the Safe-Accu 2 blood glucose monitoring system and a laboratory system, Safe-Accu 2 blood glucose monitoring system blood tests should be performed within 30 minutes of a laboratory test.

Cleaning and Disinfecting

  • Clean and disinfect the meter immediately after getting any blood on the meter or if the meter is dirty.
  • If the meter is being used on a second person, the meter and lancing device should be cleaned and disinfected first. Do not clean the meter during the test.

To clean the meter

  1. Wash hands thoroughly with soap and water.
  2. Make sure the meter is off and a test strip is not inserted. Use 75% alcohol to rub the entire outside of the meter. Make sure no liquids enter the test strip port or other openings in the meter.
  3. Let the meter air dry thoroughly before using to test.
  4. Wash hands thoroughly again after handling the meter.
  5. Verify the meter is working properly by performing a control solution test. If your meter cannot work properly, please contact customer service at +86-731-89935581/+86-731-89935582 or contact local dealers.

To disinfect the meter

Please follow clinic institution disinfection methods to disinfect the meter.

Power Saving

  • If there is no blood applied to the test strip within 3 minutes, the meter will automatically turn off.
  • After testing, the test results will be displayed on the screen with the test strip in the strip port. If there is no operation on the meter within 3 minutes, the meter will automatically turn off.
  • If there is no other operation on the meter within 1 minute, such as entering the test results review mode, the meter will automatically turn off.

Performance Characteristics

Accuracy:

95% of test results from the Safe-Accu 2 meet the requirement below:

Concentration RangeBias%
<5.5 mmol/L (100 mg/dL)Within 0.83 mmol/L(15 mg/dL)
5.5 mmol/L (100 mg/dL)Within 15%

Precision:

Test results from the Safe-Accu 2 meet the requirement below:

Concentration RangeRequirement
<5.5 mmol/L (100 mg/dL)SD< 0.34 mmol/L (6.0 mg/dL)
5.5 mmol/L (100 mg/dL)CV<6.0%

Details please read the Safe-Accu 2 blood glucose test strip package insert.

Specifications

ParameterValue
Blood volume0.6 µL
Sample typeCapillary whole blood, Venous whole blood
CalibrationPlasma equivalent
Measuring time10s
Meter storage/transportation conditions-20~55 °C, 80% RH
Dimension105*60*23mm
Weight77.2g, with battery
Power source3V DC, 10mA, 2 AAA alkaline batteries
Memory200 blood glucose test results
Operating condition10~35 °C, 80% RH
ConstructionHand-held
Measurement unitsmg/dL or mmol/L
Measurement range20~600 mg/dL or 1.1~33.3 mmol/L
Software Version (mmol/L)20160112(A.2.1)
Software Version (mg/dL)20141014(A.1.1)

Additional features

  • Auto electrode inserting detection
  • Auto sample loading detection
  • Auto reaction time count-down
  • Temperature Warning
  • Battery power Warning
  • Used test strips Warning

Maintenance

  • Use of this instrument in a dry environment, especially if synthetic materials are present (synthetic clothing, carpets etc.) may cause damaging electrostatic discharges that may cause erroneous results.
  • Electromagnetic Compatibility (EMC): The meter complies with the electromagnetic requirements IEC 61326-2-6: 2012 specified in ISO 15197-2013. Electromagnetic emissions are low and unlikely to interfere with other nearby electronic equipment, nor are emissions from nearby electronic equipment likely to interfere with the meter.
  • Immunity to electrostatic discharge meets the requirements of IEC 61326-2-6: 2012. The meter has been tested for radio frequency interference at the frequency range and test levels specified by IEC 61326-2-6: 2012.
  • Do not use this instrument in close proximity to sources of strong electromagnetic radiation, as these may interfere with the proper operation.
  • Avoid dirt, dust, blood, control solution, or liquids entering the meter test port.
  • Do not keep the meter in an area where it may be crushed.
  • Store the blood glucose monitoring system (meter, test strip, control solution) in a dry place.
  • Do not freeze.
  • Do not store in the kitchen or bathroom.

Changing the Batteries

Make sure your meter is off when you change the batteries.

Caution: Keep batteries away from children. If a battery is swallowed, call a healthcare professional immediately.

Step 1: Slide the battery cover out of the meter.

Step 2: Remove the old batteries. Place the new ones in the drawer according to the inner compartment directions of "+" and "-" sides.

Step 3: Slide the battery cover back into position until it locks into place.

Troubleshooting

Display

DisplayWhat it meansWhat to do
"LO"Your result is below the measurement limit, which is less than 20 mg/dL or less than 1.1 mmol/L.Use a new test strip to test again. If the meter still displays "LO", call a healthcare professional immediately.
"HI"Your result is above the measurement limit, which is higher than 600 mg/dL or higher than 33.3 mmol/L.Use a new test strip to test again. If the meter still displays "HI", call a healthcare professional immediately.
Low battery iconLow battery power.Replace the batteries immediately.
Temperature warning iconMeter is out of the required testing temperature range.Move meter, strip, and control solution to an appropriate place. Wait at least 30 minutes and test again.

Display

DisplayWhat it meansWhat to do
"E-1"Test method error: insert strip into meter after applying sample to strip.Insert test strip into meter before applying sample to strip.
"E-4"Used test strip / Expired test strip or damped test stripUse a new test strip to test again.
"E-5"Test strip exposed to air for more than 3 minutes.Use a new test strip to test again.
"E-6"Parameter error.Call Customer Service at +86-731-89935581 / +86-731-89935582 or contact your local dealers.

Display

DisplayWhat it meansWhat to do
Meter does not turn on after inserting a test stripBattery is installed incorrectly; Low battery power; Test strip is inserted incorrectly or not completely in;Check that the battery is correctly installed with the right position of sign "+" and "-"; Replace the battery; Insert the test strip correctly with the printing facing up and push it into the strip port until it stops;
No display.Remove the battery, wait for 3 minutes, reinstall the battery. If still no display, call Sinocare customer service at +86-731-89935581/ +86-731-89935582 or contact your local dealers.
Test strip port malfunctionCall Sinocare customer service at +86-731-89935581/ +86-731-89935582 or contact your local dealers.

Display

DisplayWhat it meansWhat to do
Meter does not begin test after applying a blood sampleNot enough blood sample / Sample is applied to wrong area of test strip / Used test strip / Test strip exposed in air too long after opening the vial or package / The end of test strip may press too firmly to your finger, blood sample cannot be drawn into test strip reaction chamber;Repeat the test with a new test strip with enough sample / Repeat the test with a new test strip with blood sample drawn into right area. Please refer to "Blood glucose testing" section. / Repeat the test with a new test strip / Finish the test immediately after taking out a test strip from a vial or package; / Use the side of test strip reaction chamber gently touch against the fingertip to draw a drop of blood;

Display

DisplayWhat it meansWhat to do
Meter displays incorrect test resultThe test strip package is damaged, and desiccant tablet becomes soft; the test strip vial open for too long; / Expired test strip; / Meter or test strip malfunctionUse test strip from a new vial. / Use test strip from a new vial within USE-BY date. / Call Sinocare customer service at +86-731-89935581/+86-731-89935582 or contact your local dealers.
Meter begins to countdown after inserting a test strip without sampleTest strip may be affected with damp / Test strip inserted incorrectly or incompletelyUse a new test strip to test. / Insert the test strips with the arrow side facing up and in the direction of the arrows until it will go no further.

Display

DisplayWhat it meansWhat to do
Test result is "HI" or "LO", but no red light on.LED light malfunctionCall Sinocare customer service at +86-731-89935581/+86-731-89935582 or contact your local dealers.
Test result is >70mg/dL (>3.9 mmol/L) and < 126 mg/dL (7.0 mmol/L), but no green light onLED light malfunctionCall Sinocare customer service at +86-731-89935581/+86-731-89935582 or contact your local dealers.
Test result is 70mg/dL (3.9 mmol/L) or >250 mg/dL (>13.9 mmol/L), but no red light on.LED light malfunctionCall Sinocare customer service at +86-731-89935581/+86-731-89935582 or contact your local dealers.

If you follow the actions recommended but the problem is not resolved, please contact Customer Service at +86-731-89935581/+86-731-89935582 or contact your local dealers for assistance.

Explanation of Symbols

  • Consult instructions for use: ℹ️
  • Caution: ⚠️
  • Storage and transportation Temperature limit: ?️ -20 to 55 °C
  • Manufacturer: [Manufacturer]
  • In vitro diagnostic medical device: [IVD]
  • Fragile, handle with care: fragile
  • Keep dry: ?
  • Keep away from sunlight: ☀️
  • Serial number: [SN]
  • Authorized representative in the European Community: [EC Rep]
  • CE marking and notify body No.: CE
  • Biological risks: ☣️
  • WEEE symbol (Waste Electrical and Electronic Equipment): ♻️ Symbol for the marking of electrical and electronics devices according to Directive 2012/19/EC. The device, accessories, and packaging must be disposed of correctly at the end of their usage; please follow Local Ordinance or Regulatory for disposal.

Warranty

Changsha Sinocare Inc. provides the following Warranty to the original purchaser of the Safe-Accu 2 blood glucose monitoring system:

  1. Changsha Sinocare Inc. warrants this meter to be free of defects in materials and workmanship from the date of purchase. If the meter does not work properly, Sinocare will replace it with a new meter or equivalent product free of charge. Failure of the meter due to abuse or use not in accordance with the instructions for use is not covered by this Warranty.
  2. This warranty does not include the battery supplied with the meter.
  3. Do not take the meter apart. This action will void the warranty and may cause the meter to display false results.
  4. If your meter does not work properly, please fill in the warranty card carefully. Use 75% alcohol to clean your meter and send your meter with the warranty card to your local dealer.

References

  1. American Diabetes Association Standards of medical care in diabetes-2016.
  2. Larsson-Cohn U: Difference between capillary and venous blood glucose during oral glucose tolerance tests. Scand J Clin Lab Invest 36:805-808, 1976

Index

  • Normal blood glucose range...4
  • Control solution, test...16
  • Prepare, lancing device...19
  • Test result, reviewing...8
  • Test result, memory ...26
  • Healthcare professional, test......35
  • Clean and disinfect, meter......30
  • Storage condition......41
  • Operating, condition......32
  • Specifications......32
  • Maintenance...33
  • Change, battery......34
  • Troubleshooting......35
  • Symbol......41
  • Warranty...43

Thank You

Thank you for choosing the Safe-Accu 2 blood glucose meter. A free replacement under normal operation is provided. Please complete the warranty card and send it back to our agent or our factory. We will inform you with information of the meter. Thank you for supporting our product and us!

Purchase date: _______________ Meter No.: _______________ (Please reserve this part, and show it when in need)

Models: Accu2 Kit-100, Blood Glucose Monitoring System

File Info : application/pdf, 26 Pages, 531.66KB

PDF preview unavailable. Download the PDF instead.

181107 Safe-Accu2 0f210d40-a81b-4dfc-81a0-03da6a227d62 ?v=1626227831

References

iLovePDF

Related Documents

Preview Sinocare Safe AQ Smart, Safe AQ Voice, Safe AQ Air Blood Glucose Meter User Manual
Comprehensive user manual for the Sinocare Safe AQ Smart, Safe AQ Voice, and Safe AQ Air blood glucose monitoring systems. Learn about intended use, safety information, test procedures, troubleshooting, and specifications.
Preview Safe-Accu Blood Glucose Monitoring System User Manual
User manual for the Sinocare Safe-Accu Blood Glucose Monitoring System, providing instructions on setup, testing, maintenance, and troubleshooting.
Preview Safe AQ Smart, Safe AQ Voice, Safe AQ Air Blood Glucose Meter User Manual
Comprehensive user manual for the Sinocare Safe AQ Smart, Safe AQ Voice, and Safe AQ Air blood glucose monitoring systems. Learn about intended use, safety information, test principles, operation, troubleshooting, and maintenance.
Preview Sinocare Safe AQ Pro Series Blood Glucose Monitoring System User Manual
This user manual provides comprehensive instructions for the Sinocare Safe AQ Pro Series Blood Glucose Monitoring System, covering setup, operation, maintenance, troubleshooting, and important safety information for accurate blood glucose level monitoring.
Preview Sinocare Safe AQ Voice Quick Guide: Easy Blood Glucose Monitoring
A concise guide to using the Sinocare Safe AQ Voice blood glucose meter for quick and accurate blood sugar testing. Learn how to prepare materials, perform the test, and interpret results.
Preview Sinocare Safe-Accu 2 Blood Glucose Test Strips: User Guide and Specifications
Comprehensive package insert for Sinocare Safe-Accu 2 Blood Glucose Test Strips, covering intended use, test principles, performance, and important warnings for self-testing.
Preview Sinocare Safe AQ UG Blood Glucose and Uric Acid Meter User Manual
Comprehensive user manual for the Sinocare Safe AQ UG Blood Glucose and Uric Acid Monitoring System, detailing setup, operation, testing procedures, and troubleshooting for accurate health monitoring.
Preview Sinocare Safe AQ Smart Blood Glucose Meter - Operation Manual
Concise operation manual for the Sinocare Safe AQ Smart blood glucose meter, detailing preparation, usage steps, and important notes for accurate blood sugar testing.