Product Instructions
AlerTox® ELISA Sesame Kit
For the quantitative detection of sesame proteins in food and drinks
REF KIT3051 (96 reactions)
Table of Contents
1. Introduction
The prevalence of sesame allergies appears to be rising worldwide; it is currently the ninth most common food allergy in the US, where approximately 17% of children with food allergies are allergic to sesame seeds.
Sesame is considered a major food allergen (e.g., one of the "Big 9") that requires food labeling in many regions, including the EU, US, UK, Australia, New Zealand, Canada and Israel. Sesame seeds are approximately 20% protein, some of which are allergenic and heat-resistant to various production processes. Therefore, accurate and reliable detection of sesame is important for consumer safety and compliance with food labeling regulations.
Note: Read this manual carefully before starting the test. The test must be performed by thoroughly trained staff.
1.1 Test Sensitivity and Specificity
The AlerTox® ELISA Sesame Kit detects and quantifies sesame proteins in drinks and food, such as bread, cakes, pastries, snacks and other foodstuffs, which may be raw, heated or baked. The limit of detection (LOD) is 0.2 ppm (mg of sesame seeds per kg of sample), the limit of quantification (LOQ) is 2.0 ppm sesame seeds (mg/kg) and the detection is quantitative between 2.0 and 20 ppm sesame seeds (mg/kg; see Section 6.2.1, Summary of Specifications for more details). See Section 4, Results Calculations, for more details about the expression of the results.
As noted below, the cross-reactivity of black sesame is 16%.
Cross-Reactive Matrix | Percent Cross-Reactivity (%) |
---|---|
Black sesame | 16 |
Psyllium | 0.001 |
Rapeseed | 0.0003 |
Note: Poppy seed showed results between 0.5 LOD and 1 LOD and may provide values above the LOQ. See Section 6.2.2, Recovery and Section 6.2.3, Non-Cross Reactivity, for additional data.
Important: Do not modify the protocol with respect to the timing, temperatures, plate washing, pipetting volumes, types of buffers or pH values of the buffers. Any of these protocol modifications will invalidate the test system.
1.2 Sample Preparation
The AlerTox ELISA Sesame Kit is one in a series of related allergen test kits from Hygiena®. Sixteen different allergens, including sesame, can be detected and measured using a single sample extract with these different allergen-specific ELISA tests, while a few need individual extractions. See Section 6.1, Sample Extraction Compatibility, for more details.
1.3 Test Principle
The AlerTox ELISA Sesame Kit works on the principle of a quantitative sandwich ELISA. The antigen concentration is directly proportional to the color intensity of the test sample. Here is a brief overview of the sandwich ELISA test:
- Primary antibodies directed against sesame proteins are bound on the surface of a microtiter plate. Sesame-containing standards or test samples are placed into the wells of the microtiter plate. After a 20-minute incubation at room temperature (15 to 25 °C, 59 to 77 °F), the wells are washed with washing solution to remove unbound material.
- Peroxidase-conjugated secondary antibodies directed against the sesame protein are put into the wells, and after a second 20-minute incubation, the plate is washed again.
- The Substrate Solution is added, and the plate is incubated for another 20 minutes, resulting in the development of a blue color in positive wells. The addition of the Stop Solution inhibits further color development, and the color turns yellow. The yellow color is measured photometrically at 450 nm (OD450 nm).
2. Materials and Storage
2.1 Materials Supplied in the Kit
Item | Description | 96 wells |
---|---|---|
1 | Breakable strips of 8 wells, each coated with anti-sesame primary antibodies. In a re-sealable foil bag containing a frame and drying agent. Ready to use. | 12 strips |
2 | 5 AlerTox Sesame Standards, concentrations: 0 - 2 - 4 - 10 - 20 ppm. Ready to use. | 5 x 3 mL |
3 | Conjugate: Peroxidase-conjugated, anti-sesame secondary antibodies. Ready to use. | 1 x 15 mL |
4 | Substrate Solution, containing trimethylbenzene (TMB). Ready to use. | 1 x 15 mL |
5 | Stop Solution, containing sulfuric acid (H2SO4). Ready to use. | 1 x 15 mL |
6 | 10X Extraction & Sample Dilution Buffer. | 4 x 30 mL |
7 | 10X Washing Solution. | 2 x 60 mL |
2.2 Storage Conditions and Stability
- All kit components should be kept at 2 to 8 °C (36 to 46 °F) in the dark. DO NOT FREEZE.
- Return all reagents to 2 to 8 °C (36 to 46 °F) immediately after use.
- The diluted Washing Solution (1X) can be used for 4 weeks when stored at 2 to 8 °C (36 to 46 °F). Important: If needed, redissolve precipitants by warming the 10X Washing Solution at 37 °C (99 °F) for 15 minutes before dilution. Do not use the buffer if the precipitant does not redissolve.
- The diluted Extraction & Sample Dilution Buffer (1X) can be used for 1 week when stored at 2 to 8 °C (36 to 46 °F). Important: If crystals precipitate during storage of the 10X Extraction & Sample Dilution Buffer, warm it to 37 °C (99 °F) for 15 minutes before use. Do not use the buffer if the precipitant does not redissolve.
- The Sample Extracts are stable for at least 24 hours at 2 to 8 °C (36 to 46 °F) or longer when frozen.
2.3 Material Required but Not Provided
- AlerTox Polyphenol Additive (Product No. ASY3213), only for samples with polyphenols and antioxidants*
- Multi-channel pipettor: 50 – 200 μL
- Pipettors: 10 – 100 μL, 100 – 1,000 μL
- Water bath, adjustable to 60 °C (140 °F)
- 15-30 mL containers for the extractions
- ELISA Plate Reader with filter (450 nm) (Absorbance 96 ELISA Reader, Product No. MCH3005, or similar)
- Centrifuge
- Distilled water
- Stomacher, Mill, Mortar, Blender, etc.
- Vortex mixer
* Examples of foods rich in polyphenols, including tannins, and antioxidants are chocolate, tea, coffee, wine, purple corn and corn fiber, soy, berries and legumes such as chickpeas or lentils.
2.4 Optional Materials/Equipment
- Homogenizer for sample extraction
- Repeating pipettor to minimize assay drift
- Recommended: An ELISA plate washer system to reduce the washing time and improve consistency
AlerTox ELISA Kits have been validated on fully automated ELISA systems (such as the BEAR Automated ELISA Robot). For validation details, contact Hygiena at www.hygiena.com/support.
3. Test Procedure
3.1 Reagent Preparation
We advise preparing reagents immediately before use and only preparing the amount necessary for the number of samples plus the 5 standards. Duplicate measurements of each sample and standard are recommended based on good laboratory practices (GLP) and quality control requirements.
Important: All reagents must be at room temperature (15 to 25 °C, 59 to 77 °F) at the time of use.
3.1.1 Extraction & Sample Dilution Buffer
Dilute the 10X Extraction & Sample Dilution Buffer 1:10 with distilled water to create the 1X solution.
Important: If needed, redissolve precipitants by warming the 10X Extraction & Sample Dilution Buffer at 37 °C (99 °F) for 15 minutes before dilution. Do not use the buffer if the precipitant does not redissolve.
Note: You will need the following amounts for each sample in your test:
Sample Type | Amount of Sample | Amount of 1X Extraction & Sample Dilution Buffer |
---|---|---|
Solid | 0.5 g | 10 mL |
Liquid | 0.5 mL | 9.5 mL |
3.1.2 Washing Solution
Dilute the 10X Washing Solution 1:10 with distilled water to create the 1X solution.
Important: If needed, redissolve precipitants by warming the 10X Washing Solution at 37 °C (99 °F) for 15 minutes before dilution. Do not use the buffer if the precipitant does not redissolve.
Note: You will need approximately 2.5 mL of 1X Washing Solution per well.
3.1.3 ELISA Plate
To prepare the ELISA plate, open the foil bag, remove the number of strips required to run the tests (samples plus the 5 standards, all in duplicate) and put the strips into a frame.
Notes:
- When opening the foil bag for the first time, be careful not to cut the ziplock off the bag.
- Unused wells must be stored in the foil bag with the drying agent at 2 to 8 °C (36 to 46 °F). Ensure the ziplock on the foil bag is sealed tightly.
3.2 Sample Preparation
Important: See Appendix A for the sample preparation protocol for samples containing polyphenols, tannins or antioxidants. For other samples, follow the procedure below:
- Resuspend the sample in 1X Extraction & Sample Dilution Buffer based on sample type:
- For solid samples:
- Maximize the homogeneity of the sample by finely pulverizing a minimum of 5 g of sample in a mortar, impact mill or a similar device.
- Resuspend 0.5 g of the homogenized mixture in 10 mL of 1X Extraction & Sample Dilution Buffer.
- For liquid samples:
- Add 0.5 mL of the liquid sample to 9.5 mL of 1X Extraction & Sample Dilution Buffer.
- Mix well.
- Incubate the mixture for 15 minutes in a preheated water bath at 60 °C (140 °F), shaking samples every 2 minutes to ensure homogeneity.
- Centrifuge the mixture for 10 minutes at 2,000 x g at room temperature (15 to 25 °C, 59 to 77 °F). If the supernatant is still not completely separated from the precipitate, filter the supernatant.
- Cool the sample extract (supernatant or filtrate) to room temperature (15 to 25 °C, 59 to 77 °F).
3.2.1 Workflow Overview
Step 1: Add 0.5 mL of liquid sample to a tube OR Step 1a: Homogenize the solid sample. Step 1b: Weigh 0.5 g of sample in a tube.
Step 2: Add Extraction & Sample Dilution Buffer (9.5 mL for liquids, 10 mL for solids) and mix well. (*Add Extraction Additive to polyphenol-containing samples (except wine with the Casein, Lysozyme and Ovalbumin Kits) before adding buffer. ** Wine samples are ready for ELISA testing (skip steps 3-5).)
Step 3: Incubate 15 min at 60 °C with mixing.
Step 4: Centrifuge at 2,000 x g for 10 min at RT and, if needed, filter.
Step 5: Cool sample to RT.
Important: See special instructions for sample extraction for the AlerTox ELISA Casein, Crustacean, Fish, Histamine, Lysozyme and Milk Kits.
3.3 ELISA Procedure
Important: The most critical points of the ELISA procedure are the temperature, timing and plate washing. Insufficient washing will result in poor precision and false results.
Note: For higher reproducibility, we recommend using a well-maintained, automated plate washer in steps 3 and 6 below.
- Add 100 µL of the standards or sample extracts in duplicate into the appropriate wells of the microtiter plate.
- Incubate for 20 minutes at room temperature (15 to 25 °C, 59 to 77 °F).
- Wash plates three (3) times with 300 µL of 1X Washing Solution per well.
- Add 100 µL of Conjugate Solution into each well.
- Incubate for 20 minutes at room temperature (15 to 25 °C, 59 to 77 °F).
- Wash plates five (5) times with 300 µL of 1X Washing Solution per well.
- Pipette 100 µL of Substrate Solution into each well.
- Allow the reaction to develop in the dark (the substrate is light-sensitive) for 20 minutes at room temperature (15 to 25 °C, 59 to 77 °F).
- Stop the enzyme reaction by adding 100 µL of Stop Solution (0.5 M H2SO4) into each well.
- Gently shake the plate by hand and wait for 1 minute.
- To measure results, use an ELISA plate reader with a 450 nm filter (OD450 nm), following the instrument manufacturer's instructions.
Note: See Section 7, Example Assay Layout. If you have a large number of samples, pipette one set of standards before the samples and the duplicate set of standards after the samples and use the arithmetic mean values for calculations.
Important: Do not shake the plate during this incubation.
Note: At the end of the automated washing or between each manual wash, invert the plates and strike them against clean, dry paper towels to empty the wells and remove residual liquid.
Important: Do not shake the plate during this incubation.
Note: At the end of the automated washing or between each manual wash, invert the plates and strike them against clean, dry paper towels to empty the wells and remove residual liquid.
Important: Do not shake the plate during this incubation.
Note: Wells containing blue color turn yellow in the presence of sesame protein.
Note: Measure the color change within 30 minutes.
Important: If any sample results fall outside the range of the sesame protein standard curve, do not extrapolate the data. Instead, dilute the sample extract further with 1X Extraction & Sample Dilution Buffer and repeat the ELISA test using this diluted sample extract and standards, in duplicate.
3.3.1 Workflow Overview
Step 1: Add 100 µL of sample/standard to each well. Step 2: Incubate 20 min at RT. Step 3: Wash 3X with 300 µL of 1X Wash Solution. Step 4: Add 100 µL of Conjugate Solution. Step 5: Incubate 20 min at RT. Step 6: Wash 5X with 300 µL of 1X Wash Solution. Step 7: Add 100 µL of Substrate Solution. Step 8: Incubate 20 min in the dark at RT. Step 9: Add 100 µL of Stop Solution. Step 10: Mix & wait 1 min. Step 11: Read results using a plate reader (450 nm).
4. Results Calculations
The results are measured as the concentration of whole sesame seeds and not sesame protein. See Step 5 below for a conversion factor to calculate sesame protein concentrations.
The standards are prepared for a direct determination of sesame seed concentrations in samples. The dilution of samples in the extraction process, as described in the sample preparation procedures, is already taken into consideration when calculating levels. However, results must account for any additional dilution (e.g., due to high sample concentration or some alternative sample extraction procedures) (Step 4, notes below). Use the AlerTox ELISA Calculation Worksheet (available at www.hygiena.com/documents) or the following instructions to calculate results.
Important: Do not use the AlerTox ELISA Calculation Worksheet if the Zero Standard on the plate reader software is defined as the Blank for the calculation of B - Bo.
When interpreting the results, the arithmetic mean is used for calculations.
- Calculate the mean OD value (OD450 nm) for each set of duplicate reference standards and duplicate samples.
- Subtract the mean value of the Zero Standard from each mean OD value of standards or samples (OD - ODStandard 0 = B - Bo). See below, Example Assay Data.
- To create the standard curve, plot the adjusted OD values of standards 1 to 4 on the y-axis versus the concentration of sesame seeds in ppm on the x-axis. See below, Example of a Typical Standard Curve.
Important: If the Zero Standard on the plate reader software is defined as the Blank for the calculation of B – Bo, skip this step.
Example Assay Data
Standard | Target Antigen [ppm] | Mean OD 450nm | B - Bo |
---|---|---|---|
Zero | 0.0 | 0.108 | - |
1 | 2.0 | 0.265 | 0.157 |
2 | 10.0 | 0.606 | 0.498 |
3 | 25.0 | 1.193 | 1.085 |
4 | 50.0 | 1.928 | 1.820 |
Example of a Typical Standard Curve
Note: When using the standard sample preparation procedure (Section 3.2), it is not necessary to multiply the resulting concentration of the foodstuff sample by the dilution factor of 20.
5. To convert ppm of sesame seeds to ppm of protein, divide the results by 5.0 [1,2].
5. General Precautions
- If your skin comes in contact with toxic or irritating substances, rinse the affected area with plenty of water and seek medical attention if needed. Please refer to the SDS, available at www.hygiena.com/SDS.
- The Substrate Solution contains TMB, which is highly toxic if inhaled, ingested or contacts skin. Please refer to the SDS.
- The Stop Solution contains H2SO4, which is corrosive. Please refer to the SDS.
- Handle the test kit in accordance with GLP.
- Do not use reagents beyond the expiration date of the kit.
- Handle all solutions with gloves.
- During the sample extraction, avoid cross-contamination.
- Devices, such as a blender, must be cleaned after each sample preparation.
- Use sterile pipette tips.
- Do not exchange reagent vial caps.
- Do not interchange reagents between kits of different lot numbers.
- Do not alter reagents. Doing so can cause inaccurate results.
- All reagents must be equilibrated to room temperature (15 to 25 °C, 59 to 77 °F) before use.
- Do not use solutions if they become cloudy or precipitate. The only exceptions are 10X Washing Solution and 10X Extraction & Sample Dilution Buffer, which may have crystalline precipitants that must be completely dissolved before use (see Section 2.2).
- Substrate Solution is light sensitive. Avoid exposure to direct light and store in the dark.
- Use only distilled water for the dilution of concentrated buffers.
- Do not allow wells to dry completely.
- Avoid incubating microtiter plates on cold work benches.
6. Additional Information
6.1 Sample Extraction Compatibility
The following AlerTox ELISA kits share the same sample preparation protocol, meaning the sample extract can be tested using 16 different ELISA Assays:
Compatible Sample Extractions | |||
---|---|---|---|
AlerTox ELISA Almond | AlerTox ELISA BLG* | AlerTox ELISA Cashew | AlerTox ELISA Coconut |
AlerTox ELISA Egg | AlerTox ELISA Hazelnut | AlerTox ELISA Lupine | AlerTox ELISA Peanut |
AlerTox ELISA Macadamia | AlerTox ELISA Mustard | AlerTox ELISA Ovalbumin | AlerTox ELISA Walnut |
AlerTox ELISA Pistachio | AlerTox ELISA Sesame | AlerTox ELISA Soy (STI) | AlerTox ELISA Wnt |
* BLG = β-lactoglobulin
+ Only the wine extract is compatible. (Cheese and other food extracts are not compatible.)
+ STI = Soy trypsin inhibitor
Individual samples must be extracted separately when using the following kits:
Individual Sample Extractions Required | ||
---|---|---|
AlerTox ELISA Casein | AlerTox ELISA Crustacean | AlerTox ELISA Fish |
AlerTox ELISA Histamine* | AlerTox ELISA Lysozyme | AlerTox ELISA Milk |
* The AlerTox ELISA Histamine Kit is based on a competitive ELISA test, while all other AlerTox ELISA Kits are based on sandwich ELISA tests.
+ Cheese and other food samples, except for wine, must be extracted separately.
6.2 AlerTox ELISA Sesame Kit
6.2.1 Summary of Specifications
Specification | AlerTox ELISA Sesame* |
---|---|
Results | Concentration of whole sesame seeds |
Limit of Detection (LOD) | 0.2 ppm |
Limit of Quantification (LOQ) | 2.0 ppm |
Standard Range | 0.0–20 ppm |
Quantification Range | 2.0–20 ppm |
Calculation Factor+ | Sesame protein [1,2] Divide by 5.0 |
* ppm = mg of sesame seed per kg of sample
+ Use the calculation factor to convert the results to the concentration of sesame protein.
For lot-specific assay data and acceptance/rejection criteria for measured values, see the Certificate of Analysis (www.hygiena.com/COA).
6.2.2 Recovery
Matrix* | Recovery (%) |
---|---|
Cracker | 109 |
Salad Dressing | 93 |
Sausage | 92 |
Soup | 110 |
* Tested in typical matrices.
6.2.3 Non-Cross Reactivity
Of the matrices that were tested, the following were found to be non-cross-reactive with the AlerTox ELISA Sesame Kit:
Non-Cross-Reactive Matrices
Adzuki bean | Almond | Apricot | Barley |
Bean, white | Beef | Beef (cooked) | Brazil nut |
Buckwheat | Cabbage, white | Caraway | Cardamom |
Carob gum | Carrot | Cashew | Cayenne |
Celery | Cherry | Chervil | Chestnut |
Chia seeds | Chicken | Chickpea | Chili |
Cinnamon | Clove | Cocoa | Coconut |
Cod | Corn | Cumin | Cumin, black |
Dill | Duck | Egg dried | Fennel |
Fenugreek | Flaxseed | Garden cress | Garlic (fresh) |
Garlic (granulated) | Gelatin, cow | Gelatin, fish | Ginger (fresh) |
Ginger (ground) | Gliadin | Guar gum | Gum arabic |
Hazelnut | Horseradish | Isinglass | Kidney bean |
Kiwi | Lamb | Leek | Lentil |
Lupine | Macadamia | Milk, cow | Milk, goat |
Mustard, yellow | Nutmeg | Oats | Onion |
Paprika | Pea | Peach | Peanut |
Pecan | Pepper, black | Pine seed | Pistachio |
Plum | Poppy seed | Pork | Potato |
Prawn (cooked) | Prawn (raw) | Pumpkin seed | Radish |
Rice | Rye | Saccharose | Shrimp (cooked) |
Shrimp (raw) | Soy flour | Soy lecithin | Split pea |
Sunflower seed | Thyme | Tofu | Tomato |
Turkey | Turmeric | Walnut | Wheat |
7. Example Assay Layout
S0: Zero Standard (without antigen): the mean value = Bo.
S1-S4: Standards: the mean value = B.
SP: Samples: the mean value = B.
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | |
---|---|---|---|---|---|---|---|---|---|---|---|---|
A | S0 | S0 | SP4 | SP4 | SP12 | SP12 | ||||||
B | S1 | S1 | SP5 | SP5 | Etc. | Etc. | ||||||
C | S2 | S2 | SP6 | SP6 | Etc. | Etc. | ||||||
D | S3 | S3 | SP7 | SP7 | Etc. | Etc. | ||||||
E | S4 | S4 | SP8 | SP8 | Etc. | Etc. | ||||||
F | SP1 | SP1 | SP9 | SP9 | Etc. | Etc. | ||||||
G | SP2 | SP2 | SP10 | SP10 | Etc. | Etc. | ||||||
H | SP3 | SP3 | SP11 | SP11 | Etc. | Etc. |
8. References
- Pathak N, Rai AK, Kumari R and Bhat KV (2014) Value addition in sesame: A perspective on bioactive components for enhancing utility and profitability. Pharmacogn Rev 8(16):147 – 155. doi: 10.4103/0973-7847.134249
- Wei P, Zhao F, Wang Z, et al. (2022) Sesame (Sesamum indicum L.): A comprehensive review of nutritional value, phytochemical composition, health benefits, development of food, and industrial applications. Nutrients 14(19):4079. doi: 10.3390/nu14194079
9. Disclaimer
Field of use: Use the Hygiena product for research and development, quality assurance and quality control under supervision of technically qualified persons. The information generated from the Hygiena product is only to be used in conjunction with the user's regular quality assurance program. The Hygiena product should not be used as the sole basis for assessing the safety of products for release to consumers. Data obtained from the Hygiena product must not be used for human diagnostic or human treatment purposes. Before using product, read the Limitation of Warranty and Liability (available in the Hygiena General Terms and Conditions at www.hygiena.com/terms-and-conditions).
These products are made from high-quality raw materials. No warranty of any kind is made, either expressed or implied, as to their suitability other than to measure the target antigen content when used exactly in accordance with these instructions, except regarding the quality of these materials.
Use of the kit for any other purpose is outside its intended use. For matrices that have not been previously validated, Hygiena cannot guarantee that the kit is fit for purpose and that the results obtained for these matrices are accurate. Customers may choose to use the product on unvalidated food or surface matrices; however, Hygiena strongly recommends that users perform their own fit-for-use testing to confirm suitability and performance in their specific application. Any damages, including consequential or special damage or expense arising directly or indirectly from using this product, are limited to the replacement value of the kit.
For additional information or assistance with matrix validation, contact Hygiena at www.hygiena.com/support.
All Hygiena Terms and Conditions apply and can be found at: www.hygiena.com/terms-and-conditions.
10. Contact Information
For more information, visit www.hygiena.com/contact. For technical support, visit www.hygiena.com/support.
11. Change Index
INS3022 REVD, July 2020
Clarified parts of the conversion factors table.
INS-KIT3051-001-REVA, June 2025
Updated recovery data, selectivity data and document ID number. Included use of the AlerTox Polyphenol Additive for some sample preparations.
Appendix A. Specialized Sample Extraction Procedures
A.1 For Foods and Drinks Containing Polyphenols, Tannins or Antioxidants
Follow this sample preparation protocol when testing foods and drinks that are rich in polyphenols, including tannins, and antioxidants. Examples are listed in the following table:
Representative Matrices | ||
---|---|---|
Berries | Chocolate | Corn, purple |
Corn fiber | Coffee | Legumes (e.g., chickpeas, lentils) |
Soy | Tea | Wine |
Important: This procedure is not for use with the following kits:
- AlerTox ELISA Crustacean Kit
- AlerTox ELISA Histamine Kit
- AlerTox ELISA Lysozyme Kit
Note: Wine extracts for the following kits:
- AlerTox ELISA Casein Kit
- AlerTox ELISA Ovalbumin Kit
a. For solid samples, maximize the sample homogeneity by finely pulverizing a minimum of 5 g of sample in a mortar, impact mill or similar device.
Note: For liquid samples, proceed to Step b.
b. Mix the sample with AlerTox Polyphenol Additive (Product No. ASY3213) and 1X Extraction & Sample Dilution Buffer, based on the kit used:
- For AlerTox ELISA Kits except Hazelnut and Pistachio: mix the sample and AlerTox Polyphenol Additive first, then add 1X Extraction & Sample Dilution Buffer (see table below) and mix well.
- For the AlerTox ELISA Hazelnut and Pistachio Kits: Dissolve 1 g of AlerTox Polyphenol Additive in 100 mL of 1X Extraction & Sample Dilution Buffer before mixing with the specified amount of sample (see table below).
Kit | Sample | AlerTox Polyphenol Additive | 1X Extraction & Sample Dilution Buffer |
---|---|---|---|
AlerTox ELISA Kits* | 1 g (Step a, solid) | 2 g | 20 mL |
1 mL | 2 g | 19 mL | |
AlerTox ELISA Milk Kit | 0.5 g (Step a, solid) | 1 g | 10 mL |
0.5 mL | 1 g | 9.5 mL | |
AlerTox ELISA Hazelnut and Pistachio Kits | 0.5 g (Step a, solid) | 10 mL | |
0.5 mL | 9.5 mL |
* i.e., all AlerTox ELISA Kits except those specific for hazelnut, pistachio, milk or those excluded in the Important note above.
c. Incubate for 15 minutes in a preheated water bath at 60 °C (140 °F), shaking the samples every 2 minutes to ensure homogeneity.
d. Centrifuge for 10 minutes at ≤ 2,500 x g.
e. If the supernatant is still not completely separated from the particulates, filter the supernatant.
f. Proceed with the ELISA Procedure (Section 3.3).
Important: The results calculations will not require additional dilution-factor adjustments for this procedure.
Hygiena
Camarillo, CA 93012
USA
Manufactured by
Hygiena Diagnóstica España S.L.
P. I. Parque Plata
Calle Cañada Real 31 – 35
41900, Camas (Sevilla), Spain
File Info : application/pdf, 15 Pages, 1.39MB
DocumentDocumentReferences
Innovative Diagnostic Solutions for Hygiene & Food Safety | Hygiena
Innovative Diagnostic Solutions for Hygiene & Food Safety | Hygiena
Document Library | Hygiena
Contact Us | Hygiena
Document Library | Hygiena
Document Library | Hygiena
Technical Support | Hygiena
Terms and Conditions of Use | Hygiena
Value Addition in Sesame: A Perspective on Bioactive Components for Enhancing Utility and Profitability | Pharmacognosy Reviews
Related Documents
![]() |
AlerTox® ELISA Egg Kit - Quantitative Detection of Egg Proteins in Food This document provides detailed instructions for the Hygiena AlerTox® ELISA Egg Kit, designed for the quantitative detection of egg white proteins (specifically ovomucoid) in various food products. It covers the test principle, kit components, storage, sample preparation, ELISA procedure, result calculation, general precautions, and additional information for accurate allergen testing. |
![]() |
AlerTox® ELISA Soy Kit: Quantitative Detection of Soy Proteins in Food User manual for the AlerTox® ELISA Soy Kit (KIT3047) by Hygiena, detailing the quantitative detection of soy proteins (soy trypsin inhibitor) in food products. Covers test principle, procedure, materials, storage, results calculation, and safety guidelines. |
![]() |
AlerTox® ELISA Almond Kit Instructions Instructions for the AlerTox® ELISA Almond kit, designed for the quantitative detection of almond proteins in food products. This document outlines the test principle, materials, procedure, calculations, and precautions. |
![]() |
AlerTox® ELISA Pistachio Kit Instructions This document provides detailed instructions for using the AlerTox® ELISA Pistachio kit for the quantitative detection of pistachio proteins in food products. It covers test principles, materials, procedures, calculations, and precautions. |
![]() |
AlerTox® ELISA Cashew Kit: Product Instructions and Guide Detailed instructions for the Hygiena AlerTox® ELISA Cashew Kit, designed for the quantitative detection of cashew proteins in food products. Includes test principle, materials, procedures, calculations, and safety precautions. |
![]() |
AlerTox® ELISA Mustard Kit: Quantitative Detection of Mustard Proteins This document provides detailed instructions for the AlerTox® ELISA Mustard Kit, a quantitative assay for detecting mustard proteins in food products. Learn about test sensitivity, specificity, sample preparation, and ELISA procedures for accurate allergen detection. |
![]() |
AlerTox ELISA Milk Kit Product Instructions Instructions for the AlerTox ELISA Milk Kit, a quantitative test for milk proteins (casein and beta-lactoglobulin) in food products and rinse water. Includes test principle, materials, procedure, results calculation, and precautions. |
![]() |
Instructions du Kit AlerTox® ELISA Amande Manuel d'instructions détaillé pour le Kit AlerTox® ELISA Amande d'Hygiena, utilisé pour la détection quantitative des protéines d'amande dans les produits alimentaires. Ce guide couvre la préparation des réactifs et des échantillons, la procédure ELISA, le calcul des résultats, les précautions de sécurité et les informations complémentaires. |