User Guide for MEAN WELL models including: RPS-30-3.3, RPS-30-5, RPS-30-7.5, RPS-30-12, RPS-30-15, RPS-30-24, RPS-30-48, RPS-30 Series 30W Reliable Green Medical Power Supply, RPS-30 Series, RPS-30 Series Power Supply, 30W Reliable Green Medical Power Supply, 30W Power Supply, Reliable Green Medical Power Supply, Reliable Power Supply, Green Medical Power Supply, Power Supply

RPS-30-20240415

wangj

Блок живлення що вбудовується відкритий 30W 3,3V RPS-30-3.3 MeanWell за ціною від 512 грн - РКС Компоненти - РАДІОМАГ


File Info : application/pdf, 4 Pages, 152.27KB

PDF preview unavailable. Download the PDF instead.

rps-30-spec
30W Reliable Green Medical Power Supply

RPS-30 series
User's Manual

EN 60601-1 IEC 60601-1
8
ANSI/AAMI ES60601-1 BS EN/EN60601-1 IEC60601-1 TPTC004
 Features
3"×2" compact size Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1 Suitable for BF application with appropriate system consideration Cooling by free air convection EMI class B for class  configuration No load power consumption<0.1W Extremely low leakage current Protections: Short circuit / Overload / Over voltage Lifetime > 105K hours Operating altitude up to 4000 meters 3 years warranty

++
 Applications
Oral irrigator Hemodialysis machine Medical computer monitors Sleep apnea devices
 GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description
RPS-30 is a 30W highly reliable green PCB type medical power supply with a high power density on the 3" by 2" footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 92% and the extremely low no load power consumption is down below 0.1W. RPS-30 is able to be used for Class(no FG) system design. The extremely low leakage current is
less than 80A. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS
EN/EN55011, perfectly fitting all kinds of BF rated "patient contact" medical system equipment.a

Model Encoding RPS- 30 - 3.3
Output voltage Rated wattage Series name
Downloaded from Arrow.com.

File Name:RPS-30-SPEC 2024-04-15

30W Reliable Green Medical Power Supply

RPS-30 series

SPECIFICATION

ORDER NO.

RPS-30-3.3

RPS-30-5

RPS-30-7.5 RPS-30-12 RPS-30-15 RPS-30-24 RPS-30-48

DC VOLTAGE

3.3V

5V

7.5V

12V

15V

24V

48V

RATED CURRENT

6A

6A

4A

2.5A

2A

1.25A

0.625A

CURRENT RANGE

0 ~ 6.6A

0 ~ 6.6A

0 ~ 4.4A

0 ~ 2.75A

0 ~ 2.2A

0 ~ 1.375A

0 ~ 0.687A

RATED POWER

19.8W

30W

30W

30W

30W

30W

30W

OUTPUT PEAK LOAD(10sec.) Note.2 21.8W

33W

33W

33W

33W

33W

33W

RIPPLE & NOISE (max.) Note.3 80mVp-p

80mVp-p

80mVp-p

100mVp-p

100mVp-p

150mVp-p

150mVp-p

VOLTAGE ADJ.RANGE VOLTAGE TOLERANCE
LINE REGULATION LOAD REGULATION

3.1~3.6V ±2.0% ±0.5% ±1.0%

4.7~5.5V ±2.0% ±0.5% ±1.0%

7.12~8.3V ±2.0% ±0.5% ±1.0%

11.4~13.2V ±2.0% ±0.5% ±1.0%

13.5~16.5V ±2.0% ±0.5% ±1.0%

22.8~27.6V ±1.0% ±0.5% ±1.0%

45.6~52.8V ±1.0% ±0.5% ±1.0%

SETUP, RISE TIME

200ms, 30ms / 230VAC 200ms, 30ms / 115VAC at full load

HOLD UP TIME (Typ.)

30ms / 230VAC 16ms / 115VAC at full load

VOLTAGE RANGE Note.5 80 ~ 264VAC

FREQUENCY RANGE

47 ~ 63Hz

INPUT EFFICIENCY (Typ.)

80%

82%

84%

88%

89%

89.5%

92%

AC CURRENT (Typ.)

1A / 115VAC 0.5A / 230VAC

INRUSH CURRENT (Typ.) OLD STAR 30A/115VAC 60A/230VAC LEAKAGE CURRENT(max.) Note.6 Touch current< 80A/264VAC

OVERLOAD

115 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed

PROTECTION OVER VOLTAGE WORKING TEMP.

3.8~5V

5.7~6.8V

8.6~11.3V

13.8~16.2V

Protection type : Shut down o/p voltage, re-power on to recover

-30 ~ +70 (Refer to "Derating Curve")

17.2~20.3V 28.4~32.4V

55.2~64.8V

WORKING HUMIDITY

20% ~ 90% RH non-condensing

ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing

TEMP. COEFFICIENT

±0.03% /  (0 ~ 50)

VIBRATION

10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

OPERATING ALTITUDE Note.7 4000 meters

SAFETY STANDARDS

IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)

ISOLATION LEVEL

Primary-Secondary: 2xMOPP

WITHSTAND VOLTAGE ISOLATION RESISTANCE

I/P-O/P: 4KVAC I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH

Parameter

Standard

Test Level / Note

EMC EMISSION SAFETY &
EMC (Note. 8)

Conducted emission Radiated emission Harmonic current Voltage flicker BS EN/EN55035, BS EN/EN60601-1-2

BS EN/EN55011 (CISPR11) BS EN/EN55011 (CISPR11) BS EN/EN61000-3-2 BS EN/EN61000-3-3

Class B Class B Class A -----

Parameter

Standard

Test Level / Note

EMC IMMUNITY

ESD
RF field susceptibility
EFT bursts Surge susceptibility Conducted susceptibility Magnetic field immunity

BS EN/EN61000-4-2
BS EN/EN61000-4-3
BS EN/EN61000-4-4 BS EN/EN61000-4-5 BS EN/EN61000-4-6 BS EN/EN61000-4-8

Level 4, 15KV air ; Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz ) Level 3, 2KV Level 4, 2KV/Line-Line Level 3, 10V Level 4, 30A/m

MTBF

Voltage dip, interruption

BS EN/EN61000-4-11

3550.0K hrs min. Telcordia SR-332 (Bellcore) ; 628.7K hrs min.

100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods MIL-HDBK-217F (25)

OTHERS DIMENSION (L*W*H) PACKING

76.2*50.8*24mm or 3" * 2" *0.945" inch 0.09Kg; 120pcs/11.8Kg/0.94CUFT

NOTE

1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 3. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1F & 47F parallel capacitor. 4. Tolerance : includes set up tolerance, line regulation and load regulation. 5. Derating may be needed under low input voltages. Please check the derating curve for more details. 6. Touch current was measured from primary input to DC output. 7. The ambient temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan models for operating altitude higher than 2000m(6500ft). 8. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still
meets EMC directives. For guidance on how to perform these EMC tests, please refer to "EMI testing of component power supplies." (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
 Product Liability DisclaimerFor detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Downloaded from Arrow.com.

File Name:RPS-30-SPEC 2024-04-15

30W Reliable Green Medical Power Supply

RPS-30 series

Block Diagram

EMI FILTER

RECTIFIERS

POWER

I/P

&

&

SWITCHING

RECTIFIER

FILTER

O.L.P.

CONTROL

Derating Curve

100
80
60 50 40
20

-30

0

10 20

30

40

50

60

70 (HORIZONTAL)

AMBIENT TEMPERATURE ()

Static Characteristics
100 90 80 70 60 50 40
80 100 115 120 140 160 180 200 220 240 264
INPUT VOLTAGE (V) 60Hz

LOAD (%)

fosc : 65KHz

RECTIFIERS

+V

& FILTER

-V

DETECTION CIRCUIT
O.V.P.

LOAD (%)

Downloaded from Arrow.com.

File Name:RPS-30-SPEC 2024-04-15

30W Reliable Green Medical Power Supply

RPS-30 series

50.8 44.45

3.175

Mechanical Specification
3.175
FS1
AC FUSE T2.5A/250V
3 2 1
CN1

Top View 76.2 69.85

Case No. Unit:mm
HS2
CN2 1 2
LED SVR1
4-3.2

3 max. 24

Side View

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing

Terminal

1

AC/N

2

No Pin

JST VHR or equivalent

JST SVH-21T-P1.1 or equivalent

3

AC/L

DC Output Connector (CN2) : JST B2P-VH or equivalent

Pin No. Assignment Mating Housing

Terminal

1

+V

2

-V

JST VHR JST SVH-21T-P1.1

or equivalent

or equivalent

Installation Manual
Please refer to : http://www.meanwell.com/manual.html

Downloaded from Arrow.com.

File Name:RPS-30-SPEC 2024-04-15



References

Acrobat Distiller 9.0.0 (Windows); modified using iText 5.5.6 ©2000-2015 iText Group NV (AGPL-version)