SERVOspin PLUS OPERATIONS MANUAL


File Info : application/pdf, 8 Pages, 1.22MB

PDF preview unavailable. Download the PDF instead.

136119-servospin-plus-benchtop-centrifuge-servoprax-operations-manual
OPERATIONS MANUAL
SERVOspin PLUS
LOW SPEED BENCHTOP CENTRIFUGE g H7 800
9

CONTENTS

1. Introduction

10

1.1 Delivery package

10

1.2 Structure Description

11

1.3 Installing the device

12

2. Technical data

12

3. Safety precautions

12

4. Operation guide

13

4.1 Control elements

13

4.2 Rotor insertion and removal

13

4.3 Loading the rotor

13

4.4 Centrifugation with a preset time

13

4.5 Short-spin centrifugation

14

4.6 Switching to the rcf display

14

4.7 Open the centrifuge in the event of a power failure

14

5. Maintenance and cleaning

14

6. Troubleshooting

14

FOREWORD
Thank you for purchasing our SERVOspin PLUS centrifuge, This Manual for users contains function and operation of the Instrument. In order to use the instrument properly, please read this manual carefully before using the instrument.

OPENING CHECK
Please check the Instrument and appendix with the packing list when you first open the instrument packing case. If you find there is something wrong with the instrument and the appendix, do contact the vendor or the producer.

1 INTRODUCTION
This centrifuge is suitable for 15 ml, 7,5 ml and 5 ml centrifuge, Blood Collection Tubes and 100 mm tube for centrifuging.
Before starting up this centrifuge for the first time, please read the rest of this operations manual.

1.1 DELIVERY PACKAGE
SERVOspin PLUS centrifuge Rotor Centrifuge tube adapter Operations manual

1 unit 1 unit 1 unit 1 unit

10

1.2 STRUCTURE DESCRIPTION
1
2

1 Cover lock 2 Mater rotor 3 Safety switch 4 Emergency door handle 5 Time setting (   ) 6 Speed setting (   ) 7 Lid open button 8 Short-Spin 9 Start/Stop

3

5 98

6 7
4

 Power switch



 Power connector



11

1.3 INSTALLING THE DEVICE
Place the centrifuge onto a level , horizontal surface. Make sure that the ventilation slits are not blocked.
Insert the column connector of the adapter to power connector of the device ,and insert another connector of the adapter to mains power supply.
A safety distance of 30 cm should be observed around the centrifuge during operation. No objects which could cause additional damage in the event of a centrifuge crash should be positioned in this space.
Power on the main switch. The centrifuge is ready to operate when the display becomes visible. Place the rotor onto the rotor axle and tighten using the rotor nut.
Before starting up the centrifuge for the first time, make sure the rotor nut is securely fastened.

2 TECHNICAL DATA
Model Power supply Power Max. speed Max. centrifugal force rpm/rcf Time Max. load Max. permitted density Acceleration time to max. speed Braking time to max. speed Ambient temperature Abmessungen (B xT x H) Gewicht mit Rotor

SERVOspin PLUS 100 - 240 V ~ 50 - 60 Hz 40 Watt 5000 U/min 3074 rcf Ja 10 sec. ~ 99 min. 8 x 15 ml 1,2 g/ml 20 - 80 seconds 20 - 80 seconds 5 °C - 35 °C 35 x 40 x 32 cm 5 kg

3 SAFETY PRECAUTIONS
d The rotor and the rotor lid must always be securely fastened. If the centrifuge
makes unusual noises when started , the rotor or rotor lid is not fastened correctly. Switch the device off immediately by pressing "START/STOP" key.
d Do not use damaged rotors!
d Do not move the centrifuge during the run!
12

PLACE THE TUBES A liquid density of 1.2 g/ml must not be exceeded at the max. speed.
Damaged tubes can not be centrifuged. This is because broken tubes can, in addition to sample loss, result in further damage to the centrifuge.
Close the test tube lids before centrifuging. Open lids can be ripped off during centrifuging and damage the centrifuge. Must load the test tubes symmetrically.

MAINTAIN OF ROTOR Even slight scratches and tears of the rotor can lead to serious internal material damage.
Avoid damage caused by aggressive chemicals, including among others: strong and weak alkali, strong acids.
In the event of contamination caused by aggressive agents, The rotor must be cleaned immediately using a neutral cleaning liquid.

4. OPERATION GUIDE

4.1 CONTROL ELEMENTS

START/STOP

for start or stop centrifugation

SHORT-SPIN

for reduced-time centrifugation

OPEN

to open the lid

 

for setting the time and speed

4.2 ROTOR INSERTION AND REMOVAL Fit the rotor onto the motor shaft, followed by the rotor nut. Tighten the rotor nut by turning clockwise. To release the rotor, turn the rotor nut counterclockwise. Before each start, check that the rotor is firmly tightened!

4.3 LOADING THE ROTOR
Rotor must always be loaded symmetrically. Minimize differences in weight between the filled sample tubes-taring with a scale is recommended. This will reduce wear on the drive and cut running noise.

4.4 CENTRIFUGATION WITH A PRESET TIME
Turn on the mains switch, pressing "OPEN" to open the lid, load the rotor symmetrically. Fasten the rotor lid and close the centrifuge lid.
Pressing Time's   change the run time, the time can be pre-selected between 20 seconds and 99 minutes, and pressing speed's   to change the run speed the max. speed is 5000 rpm.
Pressing the first "START/STOP" to start the run. Pressing the second "START/STOP" to end the run prematurely. The remaining run time appears in the display.
After the run, the lid lock of centrifuge opens automatically.
The time setting and the rotational speed may be changed during the run. The remaining run time appears in the display.
13

SHORT-SPIN CENTRIFUGATION
Turn on the mains switch, pressing "OPEN" to open the lid, load the rotor symmetrically. Fasten the rotor lid and close the centrifuge lid.
Pressing "SHORT-SPIN". Short-spin centrifugation is possible for as long as this key is held down. The maximum rpm is 5000

4.6 SWITCHING TO THE RCF DISPLAY Pressing speed's   simultaneously the display switches from rpm to rcf. Also pressing speed's   simultaneously again, the display switches from rcf to rpm. You can calculate with the following formula: rcf = 1118 * 10-5 * n2 * rmax n: rotational speed in 1/min rmax = 6 cm, max. centrifuging radius in cm

4.7 OPEN THE CENTRIFUGE IN THE EVENT OF A POWER FAILURE
Disconnect the centrifuge from the mains supply. Wait until the rotor has come to a standstill, Then left up the device, remove the lock pole to right , the lid lock will be opened.

5. MAINTENANCE AND CLEANING
d The rotor and the outside of centrifuge should be cleaned regularly with a moist
cloth. Disconnect the centrifuge from the mains supply, remove the rotor and clean it separately. Only neutral agents may be used for cleaning.
d Please check the rotor and especially the rotor bores regularly for deposits or
damage, then reinsert the rotor and tighten the rotor nut.

6 TROUBLE SHOOTING

Error

Cause

Solution

No display

No main power

Plug in mains cable both

connection, Power failure sides, check mains fuse

Lid can´t be opened

Power failure Rotor is still spinng. Lid lock failure

Emergency lid release (4.7) Stop centrifuge, Contact service

Centrifuge shaked during Rotor not loaded

acceleration

symmetrically

Stop centrifuge and load centrifuge symmetrically

Er=01

Open lock failure

Contact service

Er=02

Close lock failure

Contact service

Er=08

Motor failure

Contact service

14

15

i servoprax GmbH

Am Marienbusch 9 · D-46485 Wesel/Germany

Tel. +49 281 95283 0· Fax +49 281 56071

info@servoprax.de · www.servoprax.de

16

Edited at 07.12.2015

1-H7 800-239-2-0001-1512

