Owner's Manual for ARROW models including: PR-45541-HP, PB-40002-100B, PR-45541-HP ADVANCE Catheter, ADVANCE Catheter, Catheter

linux, amd64

4.5Fr. 55cm

4 days ago — Rx only. Warning: Read all package insert warnings, precautions, and instructions prior to use. Failure to do so may result in severe patient injury or ...


File Info : application/pdf, 1 Pages, 673.85KB

PDF preview unavailable. Download the PDF instead.

LBL070028
Dimensions: Width: 8-1/2" Length: 11

PB-40002-100B

ADVANCE
Antimicrobial Antithrombogenic

PR-45541-HP (IPNXXXXXX)

4.5 1 Fr.

Lumen

55 cm catheter length

.018 inch dia. spring-wire guide

Pressure Injectable Arrowg+ard Blue Advance® One-Lumen PICC

Contents:
1: One Lumen TaperFree® Catheter with Arrowg+ard Blue Advance® Antimicrobial/Antithrombogenic Protection1: 4.5 Fr. (1.58 mm OD) x 55 cm Pressure Injectable and Contamination Guard 
1: GlideThruTM Peel-Away Sheath: 4.5 Fr. x 4" (10 cm) Radiopaque over 4.5 Fr. Dilator  1: Spring-Wire Guide: .018" (0.46 mm) dia. x 17-3/4" (45 cm) (Straight Soft Tip on Both Ends)  1: Introducer Needle: Echogenic 21 Ga. x 2-3/4" (7 cm) TW  1: Syringe: 5 mL Luer-Lock  1: Dust Cap: Non-Vented  1: Patient ID Card  1: Patient Information Booklet  1: Paper Tape Measure  1: Dressing: STATLOCK®2 PICC Plus Catheter Stabilization Device 
1Licensed under US Patent No. 7,329,412 and 11,219,706. 2A registered trademark of C. R. Bard, Inc. All components are CE 2797 unless otherwise noted.

Rx only Warning: Read all package insert warnings, precautions, and instructions prior to use. Failure to do so may result in severe patient injury or death. www.teleflex.com/IFU California Prop. 65  WARNING: Cancer and Reproductive Harm. www.P65Warnings.ca.gov
MR Safe Not made with natural rubber latex. Store below 25°C (77°F). Avoid excessive heat above 30°C (86°F).
Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable Arrowg+ard Blue Advance antimicrobial/antithrombogenic catheter is contraindicated: · for patients with known hypersensitivity to chlorhexidine · in the presence of device related infection in the intended insertion vessel or catheter pathway · in the presence of thrombosis in the intended insertion vessel or catheter pathway

Lumen Pink (17 Ga.)

Priming Volume*
(mL) 0.66

Gravity Flow Rate
(mL/hr) 1090

Pump Flow Rate
(mL/hr) 5380

MAX Pressure Injection
Flow Rate** (mL/sec) 5

* Priming volumes are approximate and are done without accessories.  Flow rate values are approximate and are determined using deionized water at 100 cm head height.  Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities. ** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.

LotNumber YYYY-MM-DD

YYYY-MM-DD

(01)XXXXXXXXXXXXXX (17)YYMMDD (11)YYMMDD (10)LotNumber

CAS # 7440-48-4


EU Authorized Representative and Importer:
Teleflex Medical IDA Business and Technology Park Dublin Road, Athlone Co. Westmeath, Ireland
All components are CE 2797 unless otherwise noted.
Arrow International LLC Subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd. Morrisville, NC 27560 USA
Product of XXX Packaged in XXX

LBL070028 R00 (2024-05)

