Fluke 62 MAX / 62 MAX + Infrared Thermometer
Users Manual
Introduction
The Fluke 62 MAX and 62 MAX + Infrared Thermometers (the Product) can determine the surface temperature by measuring the amount of infrared energy radiated by the target's surface. Note that the Japanese models indicate Celsius only.
[Warning] Read all safety information before you use the Product.
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
- Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
- Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
- Canada: 1-800-36-FLUKE (1-800-363-5853)
- Europe: +31 402-675-200
- Japan: +81-03-6714-3114
- Singapore: +65-6799-5655
- Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
Safety Information
A [Warning] identifies conditions and procedures that are dangerous to the user. A [Caution] identifies conditions and procedures that can cause damage to the Product or the equipment under test.
Table 1 describes symbols used on the Product and in this manual.
[Warning] To prevent eye damage and personal injury:
- Read all safety information before you use the Product.
- Do not use the Product if it operates incorrectly.
- Use the Product only as specified, or the protection supplied by the Product can be compromised.
- Before you use the Product, inspect the case. Do not use the Product if it appears damaged. Look for cracks or missing plastic.
- See emissivity information for actual temperatures. Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard.
- Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
- Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
- Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
- Do not use the Product around explosive gas, vapor, or in damp or wet environments.
- Use the Product only as specified or hazardous laser radiation exposure can occur.
Table 1. Symbols
Symbol | Meaning | Symbol | Meaning |
---|---|---|---|
[Warning symbol] | Risk of danger. Important information. See Manual. | [WEEE symbol] | This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke's website for recycling information. |
[Laser Warning symbol] | Warning. Laser. | CE | Conforms to European Union directives. |
[Battery symbol] | Battery | [Australian standards symbol] | Conforms to relevant Australian standards. |
[China Metrology Certification symbol] | Conforms to China Metrology Certification |
Maintenance
[Caution] To avoid damage to the Product, do not leave the thermometer on or near objects of high temperature.
How to Change the Battery
To install or change the AA IEC LR06 battery, open the battery compartment and replace the battery as shown in Figure 16.
How to Clean the Product
Use soap and water on a damp sponge or soft cloth to clean the Product case. Carefully wipe the surface with a moist cotton swab. The swab may be moistened with water. See Figure 17.
Specifications
62 MAX | 62 MAX + | |
---|---|---|
Temperature Range | -30 °C to 500 °C (-22 °F to 932 °F) | -30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy (Calibration geometry with ambient temperature 23 °C ±2 °C) | ≥0 °C: ±1.5 °C or ±1.5% of reading, whichever is greater (≥32 °F: ±3 °F or ±1.5% of reading, whichever is greater) > -10 °C to <0 °C: ±2 °C (≥14 °F to <32 °F: ±4 °F) < -10 °C: ±3 °C (<14 °F: ±6 °F) |
≥0 °C: ±1 °C or ±1% of reading, whichever is greater (≥32 °F: ±2 °F or ±1% of reading, whichever is greater) > -10 °C to <0 °C: ±2 °C (≥14 °F to <32 °F: ±4 °F) < -10 °C: ±3 °C (<14 °F: ±6 °F) |
Response Time (95 %) | <500 ms (95 % of reading) | <300 ms (95 % of reading) |
Spectral Response | 8 to 14 microns | 8 to 14 microns |
Emissivity | 0.10 to 1.00 | 0.10 to 1.00 |
Temperature Coefficient | ±0.1 °C/°C or ±0.1%/°C of reading (whichever is greater) | ±0.1 °C/°C or ±0.1%/°C of reading (whichever is greater) |
Optical Resolution | 10:1 (calculated at 90 % energy) | 12:1 (calculated at 90 % energy) |
Display Resolution | 0.1 °C (0.2 °F) | 0.1 °C (0.2 °F) |
Repeatability (% of reading) | ±0.8 % of reading or ±1.0 °C (2 °F), whichever is greater | ±0.5 % of reading or ±0.5 °C (1 °F), whichever is greater |
Power | 1 AA IEC LR06 Battery | 1 AA IEC LR06 Battery |
Battery Life | 10 hours with laser and backlight on | 8 hours with laser and backlight on |
Weight | 255 g (8.99 oz) | 255 g (8.99 oz) |
Size | (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches | (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches |
Operating Temperature | 0 °C to 50 °C (32 °F to 122 °F) | 0 °C to 50 °C (32 °F to 122 °F) |
Storage Temperature | -20 °C to 60 °C (-4 °F to 140 °F), (without battery) | -20 °C to 60 °C (-4 °F to 140 °F), (without battery) |
Operating Humidity | 10 % to 90 % RH non-condensing @ 30 °C (86 °F) | 10 % to 90 % RH non-condensing @ 30 °C (86 °F) |
Operating Altitude | 2000 meters above mean sea level | 2000 meters above mean sea level |
Storage Altitude | 12,000 meters above mean sea level | 12,000 meters above mean sea level |
Standards and Agency Approval
Ingress Protection Rating | IP 54 per IEC 60529 |
Drop Test | 3 meters |
Vibration and Shock | IEC 60068-2-6 2.5 g, 10 to 200 Hz, IEC 60068-2-27, 50 g, 11 ms |
EMC | EN 61326-1:2006 EN 61326-2:2006 |
Compliance Details
- Compliance: EN/IEC 61010-1
- Laser Safety: EN 60825-1 Class 2
- Implement Standard: Q/SXAV37
The Product
Diagram 1: Product Overview
Shows the Fluke 62 MAX/MAX+ infrared thermometer with a laser warning label. The label indicates 'LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT λ=650nm,<1mW,EN60825-1:2007'. Two views of the device are presented.
Diagram 2: Device Display
Illustrates the device's LCD display in 'SCAN' mode. It shows temperature units (°C/°F), a large temperature reading (e.g., 1888.8), and indicators for Max, Alarm, HILO, EMS, Min, Avg, and Diff. An example reading of '1880' is also shown.
Diagram 3: Laser Operation
Demonstrates how to turn the laser pointer on and off using the 'SET' button. The display shows 'LAS ON' and 'LAS OFF'.
Diagram 4: Temperature Units
Shows how to switch between Fahrenheit (°F) and Celsius (°C) temperature units using the 'SET' button.
Diagram 5: Emissivity Adjustment
Illustrates adjusting the Emissivity (EMS) setting. The display shows example EMS values like '0.98' and '0.58', with a current reading of '0.88'.
Diagram 6: Measurement Modes
Explains how to access and view Maximum (Max), Minimum (Min), Average (Avg), and Differential (Diff) temperature readings using the 'SEL' button. Example readings are shown: Max 68.0, Min 68, Avg 76, Diff 22.
Diagram 7: Auto-Off Timer
Depicts setting the auto-off timer. Options shown include 'tr19 On' and '10 min', and the ability to turn it 'OFF'.
Diagram 8: Alarm Settings
Shows how to set High (HI) and Low (LO) alarms. Example setpoints displayed are 'Alarm HI 357', 'Alarm HI 358', 'Alarm LO 20', 'Alarm LO 19', and 'OFF'.
Diagram 9: Alarm Indication
Illustrates how the device indicates when a set alarm threshold is met. It displays the alarm type (HI/LO) and the measured temperature, e.g., 'Alarm HI 357', 'Alarm HI 356', 'Alarm LO 20', 'Alarm LO 19'.
Diagram 10: Pipe Temperature Measurement
Shows measuring the temperature of a pipe. It indicates a 30-minute timer and displays ambient temperatures (0°C/32°F) and target temperatures (22°C/72°F).
Diagram 11: Safety Warnings
Displays warnings related to laser safety and operating near high temperatures (125°F). It shows a laser beam pointing at a person's eye with a warning symbol.
Diagram 12: 62 MAX Distance-to-Spot Ratio
Illustrates the Distance-to-Spot (D:S) ratio for the 62 MAX model (10:1). It shows how the measurement spot size increases with distance: 1.5" spot at 12" distance, 2.4" spot at 24" distance, and 4" spot at 36" distance. The laser beam spread is also depicted.
Diagram 13: 62 MAX+ Distance-to-Spot Ratio
Illustrates the Distance-to-Spot (D:S) ratio for the 62 MAX+ model (12:1). It shows how the measurement spot size increases with distance: 1.4" spot at 12" distance, 2" spot at 24" distance, and 3.4" spot at 36" distance. The laser beam spread is also depicted.
Diagram 14: Scanning and Readings
Demonstrates scanning a surface and displaying various readings like Max, Min, Avg, and Diff. It shows how to use the 'SEL' button to cycle through these modes. Example readings shown include Max 90.0, Min 70, Avg 76, Diff 22.
Diagram 15: Measuring Through Obstacles
Shows measuring the temperature of a pipe through a wall. It indicates measurements at different angles (75°, 225°, 75°) and displays readings like 125.2°F (Max) and 225.0°F (Max), with an 'OK' confirmation and a reading of 236.
Diagram 16: Battery Replacement
Shows the process of replacing the AA IEC LR06 battery in the device, indicating the positive (+) and negative (-) terminals.
Diagram 17: Cleaning Instructions
Illustrates cleaning the device with a damp cotton swab and cloth. A symbol indicates not to use a bucket of water.
Limited Warranty and Limitation of Liability
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
(Document Revision: 11/99)