User Guide for hygiena models including: KIT230219, MON88017, MON88017 GMO Maize Multiplex Detection Kit, GMO Maize Multiplex Detection Kit, Maize Multiplex Detection Kit, Multiplex Detection Kit, Detection Kit, Kit

Guide de référence

Identification des OGM | Hygiena

Guía de referencia

GMO Maize Multiplex Detection Kit (MON88017, NK603, MIR162) | Hygiena

Guia de Referência


File Info : application/pdf, 2 Pages, 344.50KB

PDF preview unavailable. Download the PDF instead.

foodproof-sl-gmo-maize---multiplex-detection-mon88017-nk603-mir162-ready-reference-guide
foodproof® SL
GMO Maize Multiplex Detection Kit (MON88017, NK603, MIR162) Ready Reference Guide
Revision A, December 2023 Product No. KIT230219
PCR kit for the qualitative detection of MON88017, NK603 and MIR162 DNA using real-time PCR instruments. Before starting, it is strongly recommended to read the entire product manual available on our website.
PROGRAM SETUP
Program your real-time PCR instrument before setting up the PCR reactions. Select the following channels: u FAM (MON88017), VIC/HEX (NK603), ROX (MIR162) and Cy5 (Internal Control).

95 °C

60 °C

50 °C

2 min |

10 min

1 cycle

15 sec | 60 sec*
40 cycles

Pre-incubation: 1 cycle Step 1: 50 °C for 2 min Step 2: 95 °C for 10 min Amplification: 40 cycles Step 1 : 95 °C for 15 sec
Step 2*: 60 °C for 60 sec
* Fluorescence detection

For some real-time PCR instruments the probe quencher as well as the use of a passive reference dye must be specified. This kit contains probes with a non-fluorescent "dark" quencher and no passive reference dye.

DATA INTERPRETATION
Verify results of positive (Control Template) and negative (H2O) controls, before interpreting the sample results. Always compare samples to positive and negative controls. Review data from each channel and interpret results as described in the table.

FAM + + + + -

VIC/HEX + + + + -

ROX + + + + -

Cy5 + or + or + or + or + or + or + or -
+ -

Result Interpretation Positive for MON88017, NK603 and MIR162 Positive for NK603 and MIR162 Positive for MON88017 and MIR162 Positive for MON88017 and NK603 Positive for NK603 Positive for MON88017 Positive for MIR162 Negative for MON88017, NK603 and MIR162 Invalid

RR-KIT230219-RevA

foodproof® is a registered trademark of Hygiena. The above mentioned real-time PCR instruments are registered trademarks of their respective holders.

PREPARATION OF THE PCR MIX
Take appropriate precautions to prevent contamination, e.g., by using filter tips and wearing gloves. Thaw reagents, mix (do not vortex!) and briefly spin vials before opening.

+12.5 µL + 4 µL + 3.5 µL

1. PREPARE PCR MIX

Add 12.5 µL Master Mix (2XM), 4.0 µL Primer/Probe Mix (P) and 3.5 µL PCR-grade H2O (not included)

for each reaction to a suitable tube.

(n samples + 2 controls + at least one additional reaction to cover pipetting loss).

Mix carefully but thoroughly by pipetting up and down.

+20 µL

2. ADD PCR MIX
Pipette 20 µL of prepared PCR mix into each strip or plate well.

+5 µL

3. ADD SAMPLES AND CONTROLS
Pipette 5 µL of samples, negative control (PCR-grade H2O) or Control Template (C) into respective wells.

4. SEAL
Carefully seal strips/plate.

5. CENTRIFUGE
Briefly spin strips/plate in a suitable centrifuge.

6. START REAL-TIME PCR RUN
Cycle samples as described above.

foodproof® SL GMO

KIT230219

Maize Multiplex Detection Kit Kit for 50 reactions

(MON88017, NK603, MIR162) Store kit at -15 to -25 °C

For food testing purposes FOR IN VITRO USE ONLY

Hygiena® | Camarillo, CA 93012 USA | diagnostics.support@hygiena.com | www.hygiena.com



References

Adobe PDF Library 17.0 Adobe InDesign 19.0 (Windows)