Nicwell Care Quick Start Guide

Please email us at nicwellcare@outlook.com with any additional questions and concerns. Your satisfaction is paramount for us!

Getting Started

  1. Connect Electrode Pads

    Connect electrode pads to respective electrode wire, ensuring no bare metal of the pins is exposed.

    Diagram Description: Two electrode wires are shown, each with a pin connector.

  2. Insert Electrode Wires

    Insert one electrode wire into each desired channel.

    Diagram Description: A device with two channels is shown. An electrode wire is being inserted into each channel.

  3. Prepare for Use

    Remove electrode pads from plastic film and place on skin. Ensure the skin is clean and dry. Pads should be replaced when they lose their adhesive quality or after using 10-15 times, or if you sense a change in stimulation sensation. If you're in doubt about the integrity of the pads, replace with new electrode pads.

  4. Power On

    Power on the device by pressing the [Power] button.

    Diagram Description: The device screen shows the power button being pressed. The screen displays initial settings with '88 88', 'P1 30', and 'TENS'.

  5. Select Treatment Mode

    Press the [M] button to select the treatment mode: TENS, EMS, or MASS.

    Diagram Description: A sequence of screen displays showing the selection of TENS, EMS, and MASS modes, each with program 'P1' and intensity '30'.

  6. Select Program

    Press the [Program] button to select the current treatment program.

    TENS: 9 programs (P1-P9)
    EMS: 8 programs (P1-P8)
    MASSAGE: 5 programs (P1-P5)

    Diagram Description: A sequence of screen displays showing program selection from P1 to P9 for TENS, P1 to P8 for EMS, and P1 to P5 for MASSAGE.

  7. Select Treatment Time

    Press the [Time] button to select treatment minutes (10-20-30-40-50-60-70-80-90 cycle).

    Diagram Description: A sequence of screen displays showing the time being adjusted from 10 minutes to 90 minutes.

Advanced Features

  1. Adjust Intensity

    To adjust the intensity level, press the [Increase Intensity] button [+] to increase, or the [Decrease Intensity] button [-] to decrease until the desired intensity level is reached (1-40 levels). When output intensity decreases to 00, the treatment will end.

    Diagram Description: Screen displays showing intensity levels being adjusted from 01 to 40.

  2. Safety Lock

    The safety lock activates automatically when no button is pressed for 20 seconds while in treating mode. The [Lock] indicator will display. This is a safety feature to prevent accidental touches. To unlock, press the [Unlock] button.

    Diagram Description: Screen displays showing the lock indicator and the unlock button.

  3. Pause and Stop Treatment

    At the treating mode, press the [Pause] button to pause treatment. The [Pause] symbol will display on the screen. Press the [Stop] button to stop treatment. If you don't like the current program, press the [Stop] button and then select another program.

  4. Electrode Detection Function

    When the electrode wire is not properly connected to the machine, the electrode wire is not properly connected to the electrode pad, or the electrode pad is not completely attached to the skin, the instrument determines that the circuit is short-circuited. The electric current level will be reset to 0 mA and shows 'ER' on the display. Please recheck the wire and pads are properly connected, or replace some new pads and stick them on the skin.

  5. Charging

    The [Low Battery] signal indicates when the battery is low. Turn off the device and charge the battery. When charging, the indicator light will be red. When charging is completed, the indicator light will be green. A 5V/1A Wall Charger is recommended (wall charger not included in this product).

  6. Automatic Shut Off

    The device turns off automatically when the treatment minutes have elapsed or when no button is pressed for 60 seconds.

  7. Memory

    The most recently set program and therapy time are stored. The last treatment program you used and the therapy time you set will appear on the display when you turn on the device.

Warranty Activation

ACTIVATE YOUR 1 YEAR WARRANTY WITHIN 14 DAYS by emailing nicwellcare@outlook.com. Please title your email "TENS UNIT Warranty". Your email will serve as your warranty application.

Models: DE-TENS-A Tens Unit Muscle Stimulator Machine, DE-TENS-A, Tens Unit Muscle Stimulator Machine, Unit Muscle Stimulator Machine, Muscle Stimulator Machine, Stimulator Machine, Machine

PDF preview unavailable. Download the PDF instead.