UNI-T UT181A OPERATING MANUAL
Model: UT181A
Preface
Dear users,
Thank you for selecting new UNI-T meters. In order to use this meter properly, please read through this manual carefully before you use it, especially the section on "Safety Precautions". If you have finished reading this manual, we recommend that you properly keep this manual for reference in the future.
Limited Warranty and Limited Liability
UNI-T company guarantees that the product will be free from any defects in materials and workmanship from the date of purchase within one year. This warranty does not apply to fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal operation or handling. Resellers are not authorized to give any other guarantee in the name of UNI-T. If any warranty service is required during the warranty period, please contact your nearest authorized UNI-T service center to obtain return authorization information; then send the product to that service center with a description of the problem. This warranty is the only compensation that you can get. In addition, UNI-T does not provide any warranty, express or implied, for example, implied warranty applicable for a particular purpose. Meanwhile, UNI-T is not responsible for any special, indirect, incidental, or consequential damage or loss as a result of any reason or speculation. Because some regions or countries do not allow limitations on implied warranties and incidental or consequential damages, so the above limitations of liability and provisions may not apply to you.
I. Overview
The Model UT181A is a 60000 counts, 4 5/6 digits, handheld auto-range true RMS intelligent Multimeter (hereinafter referred to as "the meter"). The overall circuit is designed with LSI ADC as the core. The full-range overload protection circuit and unique appearance design make it a special electrical meter with superior performance. It can be used to measure parameters such as AC and DC voltage, AC and DC current, resistance, conductance, diode, continuity, capacitance, temperature, frequency, and pulse width. It is provided with functions for data hold, maximum/minimum/average measurement, comparison measurement, relative measurement, peak detection, trends capture, and data record/readback for as many as 20,000 pieces.
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly.
⚠️ Warning: Please read the "Rules for Safe Operation" before using the meter.
II. Unpacking Inspection
Open the package case and take out the meter. Please check the following items to see any missing or damaged part:
- One operating manual
- A pair of test leads
- Two K type temperature probes
- One temperature connector
- One charging adapter
- One charging connector
- One USB cable
- One CD
- One cloth bag
- One warranty card
If you find any missing or damage, please contact your supplier immediately.
III. Rules for Safe Operation
Please note the "Warning Signs and Words." Warnings indicate the conditions and actions which pose hazards to users or may damage the meter or equipment under test.
This Meter is designed to comply with the safety standards: Conforms to UL STD. 61010-1, 61010-2-030, 61010-2-032, 61010-2-033; Certified to CSA STD. C22.2 NO. 61010-1, 61010-2-030; IEC STD 61010-2-032, 61010-2-033 in pollution degree 2, measurement category (CAT III 1000V, CAT IV 600V) and double insulation, as well as with the IP65 standards for waterproof and dustproof. If you use the meter without following the operating instructions, the protection provided by the meter may be impaired or lost.
Measurement Category III (CAT III) is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation.
Measurement Category IV (CAT IV) is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
- Never use a damaged meter. Before using the meter, check the meter case to see cracks or any missing plastic part. Pay special attention to the insulation around the connectors.
- Before using the meter, ensure the battery cover is closed and latched. Before opening the battery cover, remove the test leads from the meter.
- Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. If any test lead is damaged, replace it before using the meter.
- Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and grounding.
- Do not use the meter when removing the cover or opening the case.
- When the meter works at a voltage 30V (rms) in AC, 42V (peak) in AC, or over 60V in DC, special care should be taken for there is danger of electric shock.
- Replacement fuse must comply with the specifications in this Operating Manual.
- Use proper terminals, function, and range for your measurement. Do not operate the meter separately.
- When measuring the current, shut off the power to the circuit then connect the meter to the circuit. Remember: The meter and the circuit must be connected in series.
- When making electrical connections, connect the common test lead before connecting the main test lead; when disconnecting, disconnect the main test lead before disconnecting the common test lead.
- If the meter works improperly, do not use it. The protection measures of the meter may have failed. If in doubt, send the meter for repair.
- Do not store or use the meter in an environment of high temperature, humidity, inflammable, explosive, and strong magnetic field.
- When using the probes, keep your fingers behind the finger guards.
- Do not use the low-pass filter to verify hazardous voltage, for there may be a voltage over the indicated value. First, measure the voltage to check whether it is hazardous in the case of not connecting a filter, then select low-pass filtering.
- Before testing resistance, continuity, conductance, diode, or capacitor, shut off the power to the circuit and discharge all high-voltage capacitors.
- Before measuring the current, check the fuses of the meter.
- Do not measure the voltage or current higher than the allowed input values. When the range of measured values cannot be determined, set the functional range switch at the maximum-range position.
- When the symbol [battery symbol] shows on the LCD display, the battery should be promptly charged to ensure measurement accuracy.
- Do not arbitrarily change the wiring within the meter to avoid damaging the meter and endangering the safety.
- Soft cloth and mild detergent should be used to clean the surface of the meter when servicing. No abrasive and solvent should be used to prevent the surface of the meter from corrosion, damage, and accident.
- To test the known voltage before use to confirm the product function is correct.
- If you want to replace the test lead, you need to replace it with a test lead of the same or higher grade of CAT III 1000V/CAT IV 600V.
⚡ Dangerous Voltage: When the meter detects the voltage ≥30V or overload (OL), the symbol ⚡ will display as a warning for potential hazardous voltage.
IV. Electrical Symbols
? | Double Insulated |
⏚ | Grounding |
⚠️ | Warning |
~ | AC (Alternating Current) |
- | DC (Direct Current) |
? | Continuity Buzzer |
➡️| | Diode |
☍ | Capacitance |
~/- | AC or DC (Alternating Current or Direct Current) |
⚡ | Danger High Voltage |
CE | Conforms to Standards of European Union. |
[Intertek logo] | This symbol signifies the product complies with both USA and Canada requirement. |
V. Meter Structure (see Figure 1)
A diagram showing the front view of the UNI-T UT181A multimeter. It labels the Case (1), LCD Display (2), Function Keys (3), Rotary Switch (4), and Input Terminals (5).
- 1. Case: The outer housing of the meter.
- 2. LCD: The Liquid Crystal Display screen for showing measurements and status.
- 3. Function Keys: Buttons (F1-F4, HOLD, RANGE, MAX MIN, SELECT) used to control meter functions.
- 4. Rotary Switch: A dial used to select measurement modes and ranges.
- 5. Input Terminals: Sockets for connecting test leads and probes.
VI. LCD Display (see Figure 2)
A diagram illustrating the LCD display of the UNI-T UT181A multimeter, detailing various indicators:
- 1. Label of Function Keys: Indicates auxiliary functions under the current measuring interface.
- 2. Simulation Bar: Analog display of input signals.
- 3. Minus Sign: Indicates a negative reading.
- 4. Lightning Symbol: Indicates Danger - High Voltage (⚡).
- 5. Time and Date: Displays the time and date set in the internal clock.
- 6. Small Measurements: Real-time input values displayed when primary and secondary displays are covered by menus or pop-up messages.
- 7. Telecommunication: Indicates the USB/Bluetooth function is enabled (? or ?).
- 8. Buzzer: Indicates the button sound is enabled (?).
- 9. Battery Capacity: Indicates the capacity level of the rechargeable battery (?).
- 10. Range Indicator: Shows the current range and range mode (auto or manual) for the meter.
- 11. Unit: Indicates the measurement unit (e.g., V, A, Ω).
- 12. Aided Function Display: Indicates an aided measuring function, such as LPF (Low Pass Filter).
- 13. Aided Display: Indicates aided measuring information on input signals.
VII. Keys, Rotary Switch and Input Terminals
(1) Keys
The 14 keys on the meter are used to activate selected features of the extensible rotary switch, browse menus, or control the meter power.
Key | Function |
---|---|
? (Power) | Turn on or turn off the meter power. |
F1, F2, F3, F4 | Select the sub-functions and modes related to the functions of the rotary switch. |
◀▶ (Cursor keys) | Used to select menu items, scroll reading information, and input data. |
HOLD | Keep the current reading on the display. |
RANGE | Switch the range mode of the meter to manual mode, then cycle through available ranges. Long press to return to automatic range measurement. |
MAX MIN | Starts to record MAX MIN values. |
SELECT ⚙️ | Press to select complex functions. Long press to enter Help Menu. |
? (Backlight) | Press to switch the backlight brightness. Long press to turn off the backlight. |
(2) Rotary Switch
Knob | Function |
---|---|
V | Measurement of AC Voltage |
mV | Measurement of mV in AC and mV in AC+DC |
V | Measurement of voltage in DC and AC+DC |
mV °C°F | Measurement of mV in AC and temperature |
Ωns)) | Measurement of resistance, continuity, and specific conductance |
➡️| | Diode test and capacitance measurement |
Hz% ms-Pulse | Measurement of frequency, duty cycle, and pulse width |
μA | Measurement of µA in AC, DC, and AC+DC |
mA | Measurement of mA in AC, DC, and AC+DC |
A | Measurement of ampere in AC, DC, and AC+DC |
? (CHG) | Battery charge |
(3) Input Terminals
Terminal | Description |
---|---|
A | Input terminals for measuring current (maximum duration is 30 seconds for 20A overload and suspend for 10 minutes) and frequency from 0A to 10.00A. |
mA μA | Input terminals for measuring current and frequency from 0A to 600mA. |
COM | Common terminals for all measurements. |
VΩCHz | Input terminals for measuring voltage, continuity, resistance, diode, conductance, capacitance, frequency, period, and duty cycle. |
In addition to the above, four terminals are used for charging and temperature measurement through corresponding connectors. "Lead Error!" will show on the display for warning if the probes are inserted improperly.
VIII. Technical Index
1. General Specifications
Maximum voltage between any terminal and ground | 1000 V |
Fuse protection of mA or μA input terminals | 0.8A H 1000V Fuse Type 6X32mm |
Fuse protection of A input terminals | 10A H 1000V Fuse Type 10X38mm |
Max. display | 60000 |
Range | Auto/Manual |
Polarity | Auto |
Operating temperature | -20°C ~ 50°C |
Storage temperature | -30°C ~ 60°C |
Relative humidity | ≤80% (0°C~30°C), ≤75% (30°C~40°C), ≤45% (40°C~50°C) |
Electromagnetic compatibility | In the RF-field of 1V/m: Overall accuracy = specified accuracy + 5% of range. No specified index for the RF-field over 1V/m. |
Operating altitude | 0~2000m |
Temperature coefficient | 0.1X (specified accuracy)/ °C (<18°C or >28°C) |
Internal battery | Lithium battery of 7.4V 2200mAh |
Power adapter | Input: 100V~240V, 50/60Hz, 0.2A max; Output: DC10V 500mA (short-circuit protection for output). External diameter of 5.5 mm and internal diameter of 2.5mm. |
Low battery indicator | The symbol [battery symbol] shows on the LCD. |
Dimension | About (225 X 100 X 60) mm |
Weight | About 608g (including battery) |
Safety standards | IEC/EN61010-1, EN61010-2-030, EN 61010-2-033 in pollution degree 2, CAT III 1000V, CAT IV 600V |
Standards for waterproof and dustproof | IP65 |
2. Electrical Specifications
Accuracy: ±(% Reading + Digits), one-year calibration period; If the temperature variation of environment reaches ±5°C, the accuracy can be adopted after two hours. The accuracy can be adopted after two hours when the battery charging is completed. Ambient temperature: 23°C ± 5°C; Ambient humidity: ≤75%RH; Temperature coefficient: 0.1x (Accuracy)/ °C (<18 °C or >28°C)
(1) AC Voltage
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) | |||
---|---|---|---|---|---|
45~1kHz | 1k~10kHz | 10k~20kHz | 20k~100kHz | ||
60mV | 0.001mV | ±(0.6%+60) | ±(1.2%+60) | ±(3%+60) | ±(4%+60) |
600mV | 0.01mV | ±(0.3%+30) | ±(1.2%+40) | ±(3%+40) | ±(4%+40) |
6V | 0.0001V | ±(0.3%+30) | ±(1.2%+40) | ±(3%+40) | ±(4%+40) |
60V | 0.001V | ±(0.3%+30) | ±(1.2%+40) | ±(3%+40) | ±(4%+40) |
600V | 0.01V | ±(0.3%+30) | ±(1.2%+40) | ±(3%+40) | Only for reference |
1000V | 0.1V | ±(0.6%+30) | ±(3%+40) | ±(6%+40) | Only for reference |
- Input impedance: About 10MΩ
- Overload protection: 1000V
- Display: True virtual value for 10% to 100% of the range.
(2) DC Voltage
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
60mV | 0.001mV | ±(0.025%+20) |
600mV | 0.01mV | ±(0.025%+5) |
6V | 0.0001V | ±(0.025%+5) |
60V | 0.001V | ±(0.03%+5) |
600V | 0.01V | ±(0.03%+5) |
1000V | 0.1V | ±(0.03%+5) |
- Input impedance: About 10MΩ
- Overload protection: 1000V
- Relative mode (REL) is required to compensate bias voltage for 60mV.
(3) AC Voltage + DC Voltage
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) | ||
---|---|---|---|---|
50~1kHz | 1k~10kHz | 10k~35kHz | ||
60mV | 0.001mV | ±(1%+80) | ±(3%+40) | ±(6%+40) |
600mV | 0.01mV | ±(1%+80) | ±(3%+40) | ±(6%+40) |
6V | 0.0001V | ±(1%+80) | ±(3%+40) | ±(6%+40) |
60V | 0.001V | ±(1%+80) | ±(3%+40) | ±(6%+40) |
600V | 0.01V | ±(1%+80) | Only for reference | Only for reference |
1000V | 0.1V | ±(1.2%+80) | Only for reference | Only for reference |
- Input impedance: About 10MΩ
- Overload protection: 1000V
- Display: True virtual value for 10% to 100% of the range.
(4) AC Current
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) | |
---|---|---|---|
45~1kHz | 1k~10kHz | ||
600 μA | 0.01 μA | ±(0.6%+40) | ±(1.2%+40) |
6000 μA | 0.1 μA | ±(0.6%+20) | ±(1.2%+40) |
60mA | 0.001mA | ±(0.6%+40) | ±(1.2%+40) |
600mA | 0.01mA | ±(0.6%+20) | ±(1.2%+40) |
10A | 0.001A | ±(1%+20) | ±(3%+40) |
- Display: True virtual value for 10% to 100% of the range.
- Overload protection: μA/mA range: 0.8A H 1000V Fuse Type Φ 6x32 mm; 10 A range: 10A H 1000V Fuse Type 10x38mm.
- Switch on for 30 seconds and suspend for 10 minutes for 20A. Not specified for over 10A.
(5) DC Current
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
600 μA | 0.01 μA | ±(0.08%+20) |
6000 μA | 0.1 μA | ±(0.08%+10) |
60mA | 0.001mA | ±(0.08%+20) |
600mA | 0.01mA | ±(0.15%+10) |
10A | 0.001A | ±(0.5%+10) |
- Overload protection: μA/mA range: 0.8A H 1000V Fuse Type Φ 6x32 mm; 10 A range: 10A H 1000V Fuse Type 10x38mm.
- Switch on for 30 seconds and suspend for 10 minutes for 20A. Not specified for over 10A.
(6) AC Current + DC Current
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) | |
---|---|---|---|
50~1kHz | 1k~10kHz | ||
600 μ A | 0.01 μ A | ±(0.8%+40) | ±(2.0%+40) |
6000 μ A | 0.1 μ A | ±(0.8%+20) | ±(2.0%+40) |
60mA | 0.001mA | ±(0.8%+40) | ±(2.0%+40) |
600mA | 0.01mA | ±(0.8%+20) | ±(2.0%+40) |
10A | 0.001A | ±(1.2%+20) | ±(3%+40) |
- Display: True virtual value for 10% to 100% of the range.
- Overload protection: μA/mA range: 0.8A H 1000V Fuse Type Φ 6x32 mm; 10 A range: 10A H 1000V Fuse Type 10x38mm.
- Switch on for 30 seconds and suspend for 10 minutes for 20A. Not specified for over 10A.
(7) Resistance
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
600Ω | 0.01Ω | In REL state: ±(0.05%+10) |
6KΩ | 0.0001KΩ | ±(0.05%+2) |
60KΩ | 0.001KΩ | ±(0.3%+10) |
600KΩ | 0.01KΩ | ±(2%+10) |
6MΩ | 0.0001MΩ | ±(2%+10) |
60MΩ | 0.001MΩ | ±(2%+10) |
- Overload protection: 1,000V
- Humidity for 60 MΩ: <50%
(8) Conductance
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
60nS | 0.01nS | ±(2%+10) |
- Overload protection: 1000V
- Humidity: <50%
(9) Capacitance
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
6nF | 0.001 nF | ±(3%+10) |
60nF | 0.01nF | ±(2.5%+5) |
600nF | 0.1nF | ±(2%+5) |
6μF | 0.001μF | ±(2%+5) |
60μF | 0.01μF | ±(5%+5) |
600μF | 0.1μF | Not specified |
6mF | 1μF | Not specified |
60mF | 10μF | Not specified |
- Overload protection: 1000V
- Display digits: 6000
(10) Temperature
Range | Resolution | Accuracy |
---|---|---|
-40℃~40℃ | 0.1°C | ±(2.0%+30) |
40℃~400℃ | ±(1.0%+30) | |
400℃~1000℃ | ±2.5% | |
-40°F~104°F | 0.2°F | ±(2.5%+50) |
104°F~752°F | ±(1.5%+50) | |
752°F~1832°F | ±2.5% |
- Overload protection: 1000V
- Two-channel temperature measurement can be performed via temperature connectors.
- Temperature sensor: Applicable to K type (chromel-silicel) thermocouple. Spare parts are point contact K type (chromel-silicel) thermocouple (only applicable to the measurement when temperature is below 230 ° C).
(11) Frequency
Range | Resolution | Accuracy |
---|---|---|
60Hz | 0.001 Hz | ±(0.02%+8) |
600Hz | 0.01 Hz | |
6kHz | 0.0001kHz | |
60kHz | 0.001kHz | |
600kHz | 0.01kHz | ±(0.01%+5) |
6MHz | 0.0001MHz | |
60MHz | 0.001MHz |
- Overload protection: 1000V
- Input amplitude: 10Hz~30MHz: 600mV ≤ a ≤ 30Vrms. Greater than 30MHz: Not specified.
(12) Duty Cycle
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
10%~90%(10Hz~2kHz) | 0.01% | ±(1.2%+30) |
- Overload protection: 1000V
- When the rise time is less than 1 μs, the signals center on trigger level.
(13) Pulse Width
Range | Resolution | Accuracy Tolerance: ±(% Reading + Digits) |
---|---|---|
250mS | 0.001mS~0.01mS | ±(1.2%+30) |
- Overload protection: 1000V
- When the rise time is less than 1 μs, the signals center on trigger level.
- The pulse width is greater than 2 μs for 10Hz to 200kHz. The pulse width depends on signal frequency.
(14) Continuity Test
Range | Resolution | Remark |
---|---|---|
? (Continuity) | 0.01Ω | Open circuit voltage is around 3V; when the buzzer selects short circuit for sound warning, the resolution is less than 10Ω. The buzzer continuously sounds. If the resolution is greater than 50Ω, the buzzer does not sound. When the buzzer selects open circuit for sound warning, the resolution is greater than 50Ω. The buzzer continuously sounds. If the resolution is less than 10Ω, the buzzer does not sound. |
- Overload protection: 1000V
(15) Diode Test
Range | Resolution | Remark |
---|---|---|
➡️| (Diode) | 0.0001V | Open-circuit voltage is around 3V. The forward voltage drop value of the measured PN junction is approximately ≤ 3V. When the buzzer activates, it will beep briefly for the normal semiconductor junction; if the semiconductor junction shorts out, it will beep continuously. Silicon PN junction drops between 0.5~0.8V as the normal value. |
- Overload protection: 1000V
IX. Measurement Operation
1. Meter Power Control
1) Manually start up and shut down the meter power
When the meter is off, long press the power button (?) to start the meter. When the meter is on, long press to shut it off. The meter cannot be powered off when charging.
2) Indicators for Battery Capacity
The meter is powered by lithium battery. The indicators for battery capacity are in the upper right corner of the display to indicate the relative conditions of the battery. Various indicators for battery capacity are described in the following table:
Meaning | Battery Capacity |
---|---|
? [Full Battery Symbol] | Full capacity |
? [Half Battery Symbol] | Half capacity |
? [Empty Battery Symbol] | Empty |
When the battery capacity is lower than 3% of full capacity, the meter will automatically shut down.
3) Backlight Control
If the display is not visible in low-light situations, press the backlight key (?) to switch the backlight brightness. Long press the key to turn off the backlight and enter power saving mode. When the backlight is off, the green light flashes to indicate that the meter is still collecting data. Press any key or turn the rotary switch to turn on the backlight again.
4) Auto Power-off
If there is no turning of the rotary switch or key action within the time set in the menu item "POWE OFF" under the meter setting "AUTO POWER SAVE", the meter will automatically shut down. Long press the power button (?) to restart the meter. When the item "POWE OFF" is set as OFF, Auto power-off will be completely prohibited.
5) Power Saving Mode
Set the control time of automatic reduction for backlight brightness and off time for display via the meter menu bar "AUTO POWER SAVE" to enter power-saving mode. Please refer to the detailed description about the meter settings.
2. Meter settings
Press the function key SETUP (Menu) to set and view the relevant information on the meter. Press the cursor keys (◀▶) to select relevant menu items of the meter as follows.
1) Keypad Tone
Set as ON to enable key sound and OFF to disable key sound. Meanwhile, the buzzer symbol ? in the upper right corner will appear or disappear for its ON and OFF.
2) Lead Alarm Buzzer
Set as ON to enable the alarm sound for wrong insertion by probe and OFF to disable the alarm sound for mis-inserted probe.
3) Communication
Set as ON to enable USB or Bluetooth communication while a symbol ? or ? will appear in the upper left corner. Set as OFF to disable USB or Bluetooth communication while the symbol disappears.
4) Date & Time
Press the function key SET (Menu) to set the date and time within the meter. Press ◀ or ▶ to select the required edit position, and press ◀ or ▶ to enter different numbers, then press the function key OK (Menu) to confirm. To cancel the settings, press the function key CANCEL (Menu).
5) AUTO POWER SAVE
Press the function key SET (Menu) to set the control time of automatic reduction for backlight brightness, off time of display, and auto power-off time. Press ◀ or ▶ to move the cursors to select different items. Press ◀ or ▶ to enter the time for power saving mode of this item with minute unit.
Menu Item | Description | Set Value |
---|---|---|
Brightness Down | Control time of automatic reduction for backlight brightness | ON: 1-60 Min OFF: This function is disabled |
Display Off | Off time of display | ON: 1-60 Min OFF: This function is disabled |
Power Off | Auto power-off time | ON: 1-60 Min OFF: This function is disabled |
Press the function key OK (Menu) to confirm the above settings. To cancel the settings, press the function key CANCEL (Menu).
6) More Settings
Press the function key ENTER (Menu) to set the languages of help information, format memory, reset the meter settings, check product model, serial number, and available memory space. Press the cursor keys (◀▶) to select the relevant menu items of the meter as follows.
- Help Menu Language
Press the function key SET (Menu) to set the language for help information. Press ◀ or ▶ to select a different language. Then press the function key OK (Menu) to confirm. To cancel the settings, press the function key CANCEL (Menu). - Memory Format
Press the function key FORMAT (Menu) to enter memory format, then press the function key YES (Menu) to confirm. To cancel the format, press the function key NO (Menu). - Reset All Setting
Press the function key RESET (Menu) to reset the menu with default settings, then press the function key YES (Menu) to confirm. To cancel the reset, press the function key NO (Menu). - About
Press the function key ABOUT (Menu) to check product model, serial number, and available memory space.
3. AC Voltage
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to V or mV as shown in Figure 5. Connect the test leads to the power or load under test in parallel.
- Directly read the measured voltage values on the display. True virtual values are displayed for AC measurement.
- Press the function key MENU (Menu) to enter one menu item in which basic AC voltage measurement can be modified. Press the cursor keys (◀▶) to select menu items. The red cursor key indicates the selected item, then press F1 to enter the corresponding measuring mode, press F2 to enter relative value measurement, press F3 to set dBm resistance, and press F4 to close the window of additional function.
▲ Attention:
- Do not input a voltage higher than 1000V. Higher voltage may be measured but it poses risk to damage the meter.
- When measuring high voltage, special care should be taken to avoid electric shock.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
- The response mode of AC coupled true RMS is adopted for the conversion of AC with sinusoidal input calibration. The accuracy of non-sinusoidal wave must be adjusted based on the following:
- For crest of 1.4~2.0, the accuracy shall be added 1.0%.
- For crest of 2.0~2.5, the accuracy shall be added 2.5%.
- For crest of 2.5~3.0, the accuracy shall be added 4.0%.
Figure 5
4. DC Voltage
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to DC V or DC mV as shown in Figure 6. Connect the test leads to the power or load under test in parallel.
- Directly read the measured voltage values on the display.
- Press the function key MENU (Menu) to enter one menu item in which basic DC voltage measurement can be modified. Press the cursor keys (◀▶) to select menu items. The red cursor key indicates the selected item, then press F1 to enter the corresponding measuring mode, press F2 to enter relative value measurement, and press F4 to close the window of additional function.
▲ Attention:
- Do not input a voltage higher than 1000V. Higher voltage may be measured but it poses risk to damage the meter.
- When measuring high voltage, special care should be taken to avoid electric shock.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
Figure 6
5. AC and DC Current
- Insert the red test lead into the μA/mA or A terminal and the black test lead into the COM terminal.
- Set the rotary switch to the position as shown in Figure 7. Press the key SELECT to select the required AC or DC to be measured. Connect the test leads to the test circuit in series.
- Directly read the measured current values on the display. True virtual values are displayed for AC measurement.
- Press the function key MENU (Menu) to enter one menu item in which basic AC or DC current measurement can be modified. Press the cursor keys (◀▶) to select menu items. The red cursor key indicates the selected item, then press F1 to enter the corresponding measuring mode, press F2 to enter relative value measurement, and press F4 to close the window of additional function.
⚠️ Warning:
- Before connecting to the test circuit in series, turn off the power to the circuit first and discharge all the high-voltage capacitors.
- Use proper input terminals and functions for measurement. If the current size cannot be estimated, the range of large current should be measured first.
- When the test lead is inserted in the input terminal of current, do not connect its test prod to any circuit in parallel; it will blow the fuses within the meter and damage the meter.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
Figure 7
6. Resistance
- Insert the red test lead into the Ω terminal and the black test lead into the COM terminal.
- Set the rotary switch to the measurement Ωns)), and current Ω is the default resistance measurement as shown in Figure 8. Connect the test leads to both ends of the measured resistance.
- Directly read the measured resistance values on the display.
▲ Attention:
- If the open circuit of measured resistance or resistance value exceeds the maximum range of the meter, "OL" will show on the display.
- When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement.
- When measuring low resistance, the test lead will bring about measurement errors of resistance between 0.10 to 0.20. To obtain accurate readings, relative measurement can be adopted. First short the input test lead, then press the key MENU and press F2 to enter the relative measurement. Perform the measurement of low resistance after the meter automatically subtracts the display value of shorted test lead.
- If the resistance value is greater than 0.50 for a shorted test lead, the test lead should be checked to see any loose or other factors.
- When measuring the resistance above 1MΩ, the readings require a few seconds to be stable. It is normal for the measurement of high resistance. In order to obtain stable readings, short test lines can be used for the measurement.
- Do not input a voltage higher than 30V in AC(rms), AC(peak 42V) or 60V in DC to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
Figure 8
7. Conductance
- Insert the red test lead into the Ω terminal and the black test lead into the COM terminal.
- Set the rotary switch to the measurement Ωns)), press the key SELECT to select conductance measurement 60nS. Connect the test leads to both ends of the measured resistance as shown in Figure 8.
- Directly read the measured conductance value on the display.
▲ Attention:
- When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement.
- Do not input a voltage higher than 30V in AC(rms), AC(peak 42V) or 60V in DC to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
8. Capacitance
- Insert the red test lead into the ☍ terminal and the black test lead into the COM terminal.
- Set the rotary switch to the measurement ☍, press the key SELECT to select capacitance measurement. Connect the test leads to both ends of the measured capacitance as shown in Figure 9.
- Directly read the measured capacitance value on the display.
▲ Attention:
- If the measured capacitance shorts or capacitance value exceeds the maximum range of the meter, "OL" will show on the display.
- For the measurement of capacitance within small range, relative measurement REL should be adopted to avoid the influence of distributed capacitance for correct readings.
- For the measurement of capacitance greater than 600 μF, it needs longer time for correct readings.
- To ensure the measuring accuracy, the capacitor should be discharged completely then put into the meter to measure especially for a capacitor with high voltage thus to avoid damage to the meter and personal injury.
- Do not input a voltage higher than 30V in AC(rms), AC(peak 42V) or 60V in DC to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the capacitor under test.
Figure 9
9. Continuity test
- Insert the red test lead into the Ω terminal and the black test lead into the COM terminal.
- Set the rotary switch to the measurement Ωns)), press the key SELECT to select the continuity test ?. Connect the test leads to both ends of the measured resistance as shown in Figure 8. Press the function key MENU (Menu) to enter the menu items. If pressing the key SHORT (Menu), then the buzzer selects short circuit for alarm sound, the measured resistance between both ends <10Ω, the buzzer continuously sounds. If the resolution is greater than 50Ω, the buzzer does not sound. If pressing the key OPEN (Menu), then the buzzer selects open circuit for alarm sound, the measured resistance between both ends > 50Ω, the buzzer continuously sounds. If the resolution is less than 10Ω, the buzzer does not sound.
- Directly read the measured resistance value on the display.
⚠️ Warning:
- When measuring the in-circuit resistance, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged to ensure a correct measurement.
- Do not input a voltage higher than 30V in AC(rms), AC(peak 42V) or 60V in DC to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
10. Diode
- Insert the red test lead into the + terminal and the black test lead into the COM terminal. The polarity of red test lead is "+" and "-" for black test lead.
- Set the rotary switch to ➡️| and the diode measuring mode ➡️| is default. Connect the test leads to both ends of the measured diode as shown in Figure 10. Directly read the approximate forward PN junction voltage of the measured diode on the display.
- Press the function key MENU (Menu) to enter menu items. If pressing the key ALARM, the buzzer starts up. It will beep briefly for the normal semiconductor junction; if the semiconductor junction shorts out, it will beep continuously. Silicon PN junction drops between 0.5~0.8V as the normal value. If pressing the key NORMAL, the buzzer will not start up.
▲ Attention:
- If the measured diode is in open circuit or the polarity is reversed, "OL" will display.
- When measuring the in-circuit diode, all the power within the measured circuit must be shut off before measurement and all the capacitors must be discharged.
- Open-circuit voltage of diode test is around 3V.
- Do not input a voltage higher than 30V in AC(rms), AC(peak 42V) or 60V in DC to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
Figure 10
11. Frequency/Duty Cycle Measurement /Pulse Width
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to the measurement Hz% ms-Pulse, press the key SELECT to select Hz or Duty% or ms-Pulse. Connect the test leads to the signal source under test in parallel as shown in Figure 11.
- Directly read the measured values of frequency or duty cycle or pulse width on the display.
▲ Attention:
- Simulation bar displays the frequency of the measured signal for duty cycle and pulse width.
- Do not input a voltage with higher frequency than 30Vrms to avoid personal injury.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
Figure 11
12. Temperature
- Set the rotary switch to the measurement mV °C°F, then press the key SELECT to select Celsius °C or Fahrenheit °F. Insert the temperature connector into four terminals and connect two temperature probes to the temperature connector. The probes detect the surface of the object under test as shown in Figure 12.
- Directly read Celsius or Fahrenheit temperature values of the two surfaces under test on the display.
- Press the function key MENU (Menu) to open one menu item F2 in which basic temperature measurement can be modified to enter the relative measuring mode. Press F4 to close the window of additional function. Press the cursor keys (◀▶) to select menu items. The red cursor key indicates the selected item, then press F1 to enter corresponding value measurement. Press F2 to enter relative value measurement, and press F4 to close the window of additional function.
▲ Attention:
- The ambient temperature for the meter shall not exceed the range from 18° C to 28 °C, otherwise it will cause measurement errors. The measurement effects are more obvious at low temperatures.
- Remove the temperature probes after completing all the measuring operations.
- Point contact K type (chromel-silicel) thermocouple (only applicable to the measurement when temperature is below 230° C).
Figure 12
13. LPF Measurement
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to V. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
- Press the function key MENU (Menu) to enter one menu item. Press the cursor keys (◀▶) to select Low Pass, then press the function key (Menu) Low Pass to enter LPF measuring mode.
- The meter performs the measurement in AC mode. The AC signals go through a filter which holds up the voltage higher than 1KHz. As shown in the following figure, the low-pass filter can measure the composite signals of sinusoidal wave generated by inverter and variable-frequency motor.
▲ Attention:
- To avoid electric shock or personal injury, do not use low-pass filter to verify hazardous voltage, for the voltage over the indicated value may exist. First, measure the voltage in the case of disconnecting the filter to see any hazardous voltage. Then select the filter function.
- In the LPF measuring mode, the meter will turn to manual mode. Press the key RANGE to select a range. When the low-pass filter is enabled, the automatic range is unavailable.
- Do not input a voltage above 1000V. Higher voltage may be measured but it poses the risk of damaging the meter.
- After completing all the measuring operations, disconnect the connection between the test leads and the circuit under test.
14. dBV
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to V. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
- Press the function key MENU (Menu) to enter one menu item. Press the cursor keys (◀▶) to select the menu item dBV, then press the function key (Menu) dBV to enter dBV measurement.
- dBV is primarily shown and AC voltage value is secondly shown on the display. AC voltage of the measured signal is displayed on the simulation bar.
15. dBm
- Insert the red test lead into the V terminal and the black test lead into the COM terminal.
- Set the rotary switch to V. Connect the test leads to the power or load to be tested in parallel as shown in Figure 5.
- Press the function key MENU (Menu) to enter one of the menu items, press the cursor keys (◀▶) to select the menu item dBm. A reference impedance (resistance) must be used for the measurement of dBm to calculate dB value on the basis of 1mW. Press the function key RES (Menu) to select a reference impedance value. Press ◀ or ▶ to scroll between ten defined reference values: 4,8,16,25,32,50,75,600, 1000 and 1200. Press the function key EDIT (Menu) and the cursor keys (◀▶) to select any one of the reference impedance values between 4Ω~1200Ω. Press the function key OK (Menu) to set reference values. Press the function key dBm (Menu) to enter dBm measurement.
- dBm is primarily shown and AC voltage value is secondly shown on the display. AC voltage of the measured signal is displayed on the simulation bar.
16. Maximum and Minimum Value
Press MAX MIN to activate the measurement of maximum and minimum values. Real-time measurements are primarily displayed. Maximum, average, and minimum values, elapsed time, start date and time corresponding to the three values are secondly displayed. Press the function key RESTART (Menu) to re-activate the measurement of maximum and minimum values. Press the function key EXIT (Menu) to exit the measurement of maximum and minimum values.
17. Relative value
Press the function key MENU (Menu) to enter a next interface, press the function key REL (Menu) to enter the measuring mode of relative values, then press the function key REL (Menu) to activate the measurement of relative values. By this time, the measurements _ basic values are primarily displayed and the relative values and real-time measurements are secondly displayed. Press the function key MENU (Menu) to select basic measurement to exit the measurement of relative values.
18. Peak detection
Press the function key MENU (Menu) to select Peak item, then press the function key PEAK (Menu) to activate peak detection. Response time is 1ms. Transient values can be more accurately measured by using peak recording function.
19 Compare Mode COMP
Press the function key COMP (Menu) to enter the measurement of compare mode. Press ◀ or ▶ to select the settings of compare mode as follows.
1) Pass Mode
Press the function key EDIT (Menu) to enter the setting mode of compare types. Press ◀ or ▶ to set one of the four types as follows:
- INNER (Low Value ≤ Input Value ≤ High Value)
- OUTER (Input Value < Low Value or Input Value > High Value)
- <Value
- >Value
Press the function key OK (Menu) to confirm the above settings. To cancel the settings, press the function key CANCEL (Menu).
2) Beep Mode
Press the function key EDIT (Menu) to enter buzzer start mode. Press ◀ or ▶ to select one of the three items as follows:
- PASS ON: It indicates that the buzzer starts to sound when the compare result shows as PASS.
- FAIL ON: It indicates that the buzzer starts to sound when the compare result shows as FAIL.
- OFF: Buzzer close.
Press the function key OK (Menu) to confirm the above settings. To cancel the settings, press the function key CANCEL (Menu).
3) Low Value or High Value or Value
Press the function key EDIT (Menu) to set compare reference values. Press ◀ or ▶ to move the cursors to select edit locations. Press ◀ or ▶ to enter different numbers. Press the function key OK (Menu) to confirm. To cancel the settings, press the function key CANCEL (Menu).
20. Recording Measurement Data
After completing the settings, press the function key START (Menu) to start the measurement of compare mode. Press the function key EXIT (Menu) to exit the measurement of compare mode.
Press the function key SAVE (Menu) to enter the modes of single recording, continuous recording, and inquire recording. Press the cursor keys (◀▶) to select the following modes.
1) Save
Press the function key SAVE (Menu) to record the current measuring data for once up to 20000 pieces.
2) View Save
Press the function key VIEW (Menu) to view one-time recorded data. Press ◀ or long press the function key PREV (Menu) to view the previous recorded data. Press ▶ or long press the function key NEXT (Menu) to view the next recorded data. Press the function key DELETE (Menu) to delete the current recorded data. Press the function key RETURN (Menu) to return to the last menu. In addition to displaying the recorded data, the location and total number of current recorded data are displayed in the lower left corner. The date and time of current data recording are displayed in the lower right corner.
3) Delete All Save
Press the function key DELETE (Menu) to delete all one-time recorded data. Press the function key YES (Menu) to confirm. To cancel the operation, press the function key NO (Menu).
4) Record
Press the function key ENTER (Menu) to enter the continuous recording mode with maximum recording pieces up to 10000. Press ◀ or ▶ and move cursors to select the three settings as follows.
- Edit Name
Press the function key EDIT (Menu) to name the continuous recording items. Press ◀ or ▶ to move the cursor and select edit locations. Press the F1 key to switch the input mode to "Uppercase" mode, then press ◀ or ▶ to input uppercase letters. Press the F2 key to switch the input mode to "Lowercase" mode, then press ◀ or ▶ to input lowercase letters. Press the F3 key to switch the input mode to "Digital" mode, then press ◀ or ▶ to enter numbers or symbols. Press the F4 key to save exit and exit edit mode. - Set Interval
Press the function key EDIT (Menu) to set the interval time of continuous recording. Press ◀ or long press ◀ to move the cursors to select edit locations. Press ▶ or long press ▶ to enter different numbers. The interval time can be set as 1Sec ~ 60Min. - Set Duration
Press the function key EDIT (Menu) to set the duration time of continuous recording. Press ◀ or long press ◀ to move the cursors to select edit locations. Press ▶ or long press ▶ to enter different numbers. The duration time can be set as days, hours and minutes. - Max Duration
Indicates the maximum time for continuous recording. After completing the settings, press the function key START (Menu) to start a continuous recording as shown in Figure 13. The character REC shows on the display and flashes with a red dot. The relevant display information is shown in the following table:
Information | Description |
---|---|
Elapsed Time | Run time displayed in hours:minutes:seconds |
Remaining Time | Time left displayed in hours:minutes:seconds |
Samples | Total recording points for present recorded events |
Start | Time and date started from recording period |
Figure 13
To manually stop recording, press the function key STOP (Menu). When stop recording, return to the menu View Record to view the record events. See the relevant operations in the following menu View Record.
5) View Record
Press the function key (Menu) to view the information on record events as shown in Figure 14. The basic display information is in the following table:
Information | Description |
---|---|
Name | Name of record event |
Interval | Interval time |
Duration | Actual duration time after continuous recording stops |
Samples | Total recording points for present record events |
Maximum | Maximum value for present record data |
Average | Average value for all data of present record |
Minimum | Minimum value for present record data |
REC | Position of present record incident and total number of record events |
Start | Time and date started from recording period |
Figure 14
Press the function key PREV (Menu) to display basic information on the previous record. Press the function key NEXT (Menu) to display basic information on the next record. Press the function key RETURN (Menu) to return to the previous menu. Press the function key TREND (Menu) to view the trend chart of present record as shown in Figure 15. The display information of trend data is in the following table:
No. | Description |
---|---|
① | Measured value corresponding to cursor |
② | Date and time for measurement corresponding to cursor |
③ | Cursor |
④ | Trend line |
⑤ | Time label of X-axis to display during elapsed time |
⑥ | Name of record event |
⑦ | Time and date started from recording period |
Figure 15
View the trend chart. Press ◀ or long press the F2 key to move the cursor left. The cursor moves one data to the left for each pressing. Long pressing, the cursor speeds up to move left. Press ▶ or long press the F3 key to move the cursor right. The cursor moves one data to the right for each pressing. Long pressing, the cursor speeds up to move right. Press ⬆ or ⬇ to scale the graph vertically. Press ◀ or ▶ to scale the graph horizontally.
Press the F1 key to delete this record. Press the function key YES (Menu) to confirm the deletion. To cancel the deletion, press the function key N0 (Menu).
6) Delete All Record
Press the function key DELETE (Menu) to delete all record events. Press the function key YES (Menu) to confirm. To cancel the operation, press the function key N0 (Menu).
X. Maintenance and Repair
21. Communication
Figure 16
USB Communication Bluetooth Communication (optional)
Turn on communication via settings (see detailed operations in the meter settings). The symbol ? or ? will appear in the upper left corner on the display as shown in Figure 16. The meter carries on USB communication with USB cable (standard accessory) connecting to PC. The meter can use Bluetooth module (optional) and mobile phone for Bluetooth communication.
1. General Maintenance and Repair
Regularly clean the meter case with damp cloth and mild detergent. Do not use abrasives, isopropyl alcohol or solvents. Dirt or moisture on the terminals can affect readings but also enable the warning function mistakenly due to mis-insertion. Clean the terminals according to the following steps:
- Turn off the meter and remove all test leads.
- Clean up the dirt on the terminals.
- Soak a clean cotton swab in mild detergent and water. Clean each terminal with the cotton swab. Dry each terminal with canned compressed air to force water and detergent to flow out of the terminals.
- In case of anything abnormal for the meter, stop using it and send it for repair.
- When the meter needs to be verified or repaired, qualified service personnel or designated maintenance department are required to repair it.
2. Testing Fuse
As shown in Figure 17, the measurement of the meter is for resistance function. Insert a test lead into the terminal as shown in Figure 17 and contact the probe tip at the other end of test lead with the metal in the terminal of current input. If the message "Lead Error!" (connection error for test lead) appears, it shows that the probe tip is inserted too deeply in the terminal of current input. Draw out the test lead a little until the error message disappears and OL (overload) or resistance reading appears on the display. If the resistance reading for A terminal is less than 0.50Ω, it shows that the fuse F2 is normal. If the reading is OL, you need to replace F2; If the resistance reading of μA m is less than 1.2MΩ, it shows that the fuse F1 is normal. If the reading is OL, F1 needs to be replaced;
Figure 17
3. Repalcing Fuse
Inspect or replace the meter fuse as shown in Figure 18 according to the following steps:
- Turn off the meter and remove the test leads from the terminals.
- Use a flat-head screwdriver to turn a counterclockwise semicycle for the screw on the battery cover then remove the battery cover.
- Gently pry one end of the fuse and then remove the fuse from the clip.
- The required fuse installed for mA or μA input terminal: 0.8A H 1000V Fuse Type 6X32mm (F1). The required fuse installed for A input terminal: 10A H 1000V Fuse Type 10X38 mm (F2).
- Reinstall the battery cover and then turn the screw clockwise semicycle to tighten the battery cover.
Figure 18
4. Battery charge
When the indicator of battery level in the upper right corner is less than 5% of full capacity, the meter should be immediately charged, otherwise it will affect the measurement accuracy. As shown in Figure 19, set the rotary switch to ? (CHG), the string "Please plug in AC adapter!" on the display prompts you to insert a power adapter for charging. Connect the power connector to the four terminals in the meter, then insert the power adapter into the power connector. The character "Charging" on the display indicates it is in charging with a red light 5% stepped in charging process. When the battery is fully charged, the red light is off and the charging process action stops.
▲ Attention:
- You must use the power adapter specified by the manufacturer.
Figure 19
Company Information
UNI-TREND TECHNOLOGY (CHINA) CO., LTD.
No6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial Development Zone, Dongguan City,
Guangdong Province, China
Tel: (86-769) 8572 3888
http://www.uni-trend.com