UNI-T UDP5000 Series Programmable DC Power Supplies
Data Sheet
Product Overview
The UDP5000 series power supplies feature a compact design, high performance, versatile functionality, a wide output range, user-friendly operation, and flexible usage. The series includes four rated output voltages (40 V, 80 V, 160 V, 250 V) and four rated output power levels (400 W, 800 W, 1200 W, 2000 W), forming a total of 16 models.
Each model is equipped with:
- USB and LAN communication interfaces
- Output terminals on both the front and rear panels (maximum front panel output current: 10 A)
- Internal variable resistance function
- Discharge circuit control function
- CC/CV mode
- Various protection functions
Product Specifications
Model | Voltage Output | Current Output | Output Power |
---|---|---|---|
40 V | |||
UDP5040-40 | 0-40 V | 0-40 A | 400 W |
UDP5040-80 | 0-80 A | 800 W | |
UDP5040-120 | 0-120 A | 1200 W | |
UDP5040-200 | 0-200A | 2000 W | |
80 V | |||
UDP5080-20 | 0-80 V | 0-20 A | 400 W |
UDP5080-40 | 0-40 A | 800 W | |
UDP5080-60 | 0-60 A | 1200 W | |
UDP5080-100 | 0-100 A | 2000 W | |
160 V | |||
UDP5160-8 | 0-160 V | 0-8 A | 400 W |
UDP5160-16 | 0-16 A | 800 W | |
UDP5160-24 | 0-24 A | 1200 W | |
UDP5160-40 | 0-40 A | 2000 W | |
250 V | |||
UDP5250-6 | 0-250 V | 0-6 A | 400 W |
UDP5250-12 | 0-12 A | 800 W | |
UDP5250-18 | 0-18 A | 1200 W | |
UDP5250-30 | 0-30 A | 2000 W |
Features
- TFT-LCD Screen: Provides a richer display with high visibility.
- Intuitive and Concise Interface: A simple, user-friendly interface allows users to quickly familiarize themselves with the system.
- Wide-Range Output: A broad variable range of output voltage and current enables various voltage/current combinations.
- Discharge Load Function: Allows users to set the discharge load state within the power supply based on actual needs.
- List Mode: Enables presetting multiple sets of output voltages and currents for automatic output.
- Output Terminals on Front and Rear Panels: Supports output from both the front and rear panels (maximum front panel output current: 10 A).
- Internal Variable Resistance Function: Allows users to set the internal resistance to simulate a power supply with internal resistance.
- External Analog Voltage Control: Enables power supply output control using an external circuit.
- USB and LAN (SCPI) Interfaces: Support communication and remote control with external devices.
- Multi-Unit Series/Parallel Operation: Allows multiple units to be used together externally to increase output power.
Design Highlights
1.1 TFT-LCD, Multifunction Buttons
The 2.4-inch TFT-LCD screen delivers a vibrant color display, offering a comfortable visual experience. Intuitive multifunction buttons enhance usability, making operation more convenient.
1.2 Intuitive Setting Interface
The Menu key allows configuration of output, system, and network settings, or viewing product information.
Function Highlights
Slope Mode Setting
To prevent overshoot when output is enabled:
- Constant voltage (CV) mode: Prioritizes voltage control.
- Constant current (CC) mode: Prioritizes current control.
Slope Mode
Slope Mode | Description |
---|---|
CV High-Speed Priority (VHS) | Default to high-speed rise and fall rates |
CC High-Speed Priority (IHS) | |
CV Slope Priority (VSR) | Rise and fall rates can be set as needed |
CC Slope Priority (ISR) |
OVP/OCP Settings
Set upper limits for output voltage and current. The output set value cannot exceed 95% of the protection value. The protection value must be at least 105% of the output set value. Configure a delay time for protection activation, adjustable between 10 ms and 10,000 ms.
Auto Rage Function
UDP5000 series limits output voltage and current based on the rated power. As a result, voltage and current automatically adjust proportionally.
List Mode
Users can configure 128 groups of output voltage, current, and output time data combinations. Additionally, the number of cycles and stop status can be configured. Up to 32 data sets can be stored in the power supply's local memory or exported to external devices. Once the list is configured and output is activated, the power supply will automatically execute the programmed combinations.
Discharge Load Function
In the UDP5000 power supplies, a capacitor is connected to the output terminal, and the power supply is equipped with an internal discharge load that discharges the capacitor when the output is turned off. When the discharge load is disabled, a residual voltage that decreases slowly will remain on the output terminals after the output stops. Conversely, when the discharge load is enabled, the residual voltage will disappear rapidly.
Internal Variable Resistance
The internal variable resistance function can adjust the power supply's internal resistance, simulating real power sources such as secondary batteries, solar cells, and fuel cells. The internal variable resistance can only be set in constant voltage mode.
Power Internal Resistance Settings
Model | Vrtg[V] | Irtg[A] | Rint[Ω] | Resolution*1 | Power |
---|---|---|---|---|---|
UDP5040-40 | 40 | 40 | 0.000-1.000 | 0.001 | 400 W |
UDP5040-80 | 40 | 80 | 0.000-0.500 | 0.001 | 800 W |
UDP5040-120 | 40 | 120 | 0.000-0.333 | 0.001 | 1200 W |
UDP5040-200 | 40 | 200 | 0.000-0.200 | 0.001 | 2000 W |
UDP5080-20 | 80 | 20 | 0.000-4.000 | 0.001 | 400 W |
UDP5080-40 | 80 | 40 | 0.000-2.000 | 0.001 | 800 W |
UDP5080-60 | 80 | 60 | 0.000-1.333 | 0.001 | 1200 W |
UDP5080-100 | 80 | 100 | 0.000-0.800 | 0.001 | 2000 W |
UDP5160-8 | 160 | 8 | 0.000-15.000 | 0.001 | 400 W |
UDP5160-16 | 160 | 16 | 0.000-7.500 | 0.001 | 800 W |
UDP5160-24 | 160 | 24 | 0.000-5.000 | 0.001 | 1200 W |
UDP5160-40 | 160 | 40 | 0.000-3.000 | 0.001 | 2000 W |
UDP5250-6 | 250 | 6 | 0.000-31.250 | 0.001 | 400 W |
UDP5250-12 | 250 | 12 | 0.000-15.625 | 0.001 | 800 W |
UDP5250-18 | 250 | 18 | 0.000-10.416 | 0.001 | 1200 W |
UDP5250-30 | 250 | 30 | 0.000-6.250 | 0.001 | 2000 W |
Vrtg[V]: Rated voltage
Irtg[A]: Rated current
Rint[Ω]: Internal resistance
*1. Setting resolution for the power supply's internal resistance.
External Voltage Control
All UDP5000 series models support external analog control and multi-unit combined operation. When using the external analog interface, ensure that the appropriate cable is used for wiring.
For external voltage control, the power supply's operation mode must be set to External Voltage Control.
For detailed instructions, refer to the UDP5000 Series Programmable DC Power Supply-User Manual.
Circuit Connection Diagram
Using an external switch to control power output ON/OFF: By adjusting the ON/OFF state, the power supply output is regulated accordingly.
Adjust an external potentiometer to control the output voltage/current
Adjust Rvs and Ris to regulate the output voltage and current levels.
Rear Panel Analog Interface
Interface | Definition
GND | Ground terminal
VREF_V | The power supply internally provides a 10 V reference voltage for external voltage control output.
Vs/DO | Voltage analog control: Analog voltage setting (0-10 V) corresponds to 0-VoutMax.
Is/D1 | Current analog control: Current voltage setting (0-10 V) corresponds to 0-loutMax.
D2 | External digital control
ON/OFF | External Digital/Analog Control for Power Output ON/OFF: High level: Enables the output; Low level: Disables the output
Vm | Voltage output analog channel (0-10 V) corresponds to 0-VoutMax.
Im | Current output analog channel (0-10 V) corresponds to 0-loutMax.
NC | Not connected (not in use)
EN_BUS | Coordinates and synchronizes start/stop, protection, and operation modes in multi-unit series/parallel operation, ensuring stable and reliable collaboration between units.
Series/Parallel Operation
UDP5000 series power supplies support external multi-unit combined operation with identical models to increase output power. This includes both external multi-unit parallel and external multi-unit series connections.
Before use, select and configure a master unit and slave units, then connect the analog interfaces using the appropriate cables.
Single-control operation refers to setting one instrument as the master unit, while other identical models function as slave units. The entire system can then be controlled through operations on the master unit.
Single-Control Parallel Operation
Single-control parallel operation increases the output current. Maximum output current (loutMax) = Rated output current of power supply x Number of parallel-connected units. For example, if three UDP5040-40 models are paralleled, the output current can reach 40 A × 3 = 120 A. The difference in output current between the master unit and slave units must remain within approximately 5% of the rated value.
For single-control parallel operation, the maximum number of identical models that can be paralleled (including the master unit) is as shown in the following table:
Model | Maximum Parallel-Connected Units |
---|---|
400 W | 4 Units |
800 W | 4 Units |
1200 W | 2 Units |
2000 W | 2 Units |
Single-Control Series Operation
Single-control series operation increases the output voltage. Maximum output voltage (Voutmax)= Rated output voltage of power supply x Number of series-connected units. For example, if two UDP5250-6 models are series-connected, the maximum voltage can reach 250 V x 2=500 V.
Maximum Series-Connected Units: 2
The voltage setting accuracy is equivalent to the single-unit accuracy of the master unit. The difference in output voltage between the master unit and slave units must remain within approximately 5% of the rated value.
Model | Maximum Series-Connected Units |
---|---|
400 W | 2 Units |
800 W | 2 Units |
1200 W | 2 Units |
2000 W | 2 Units |
Technical Specifications
Specification Conditions
This section outlines the technical specifications for the 400 W, 800 W, 1200 W, and 2000 W models of the UDP5000 series. It includes detailed information on input/output specifications, voltage and current specifications, output characteristics, noise and ripple, and display characteristics.
Unless otherwise specified, technical specifications are based on the following settings and conditions (applicable to all models):
- Load Type: Purely resistive load
- Preheating Time: 30 minutes
- Post-Preheating Calibration: The product must be properly calibrated in an environment of 23°C ±5°C following the appropriate calibration procedures.
Technical Terminology
- Typical value (Typ.): These represent typical values when the product operates at an ambient temperature of 23°C (73.4°F). These values do not guarantee the product's performance.
- Rated value: The specified nominal value.
- Reading: The measured value displayed.
- Rated load and no-load:
- Constant voltage mode (when the output current is set to ≥ the maximum output current at the rated output voltage):
- Rated load: Using a purely resistive load, the output current reaches 95% to 100% of the maximum output current when the rated output voltage is applied.
- No-load: No output current flows through the load (i.e., an open-circuit state).
- Constant current mode (when the output voltage is set to ≥ the maximum output voltage at the rated output current)
- Rated load: Using a purely resistive load, the output voltage drops to 95% to 100% of the maximum output voltage when the rated output current is applied. Including the voltage drop in the load cable, the UDP5000 series power supply output voltage must not exceed the maximum output voltage at the rated output current.
- No-load: Using a purely resistive load, the output voltage drops to 10% of the maximum output voltage or 1 V (whichever is higher) when the rated output current is applied.
- Rise time: The time taken for the output voltage or current to increase from 10% to 90% of the rated value when the power supply output is turned on.
- Fall time: The time taken for the output voltage or current to decrease from 90% to 10% of the rated value when the power supply output is turned off.
- UDP5000 series power supply operates within the rated output power range for both voltage and current. However, the available current at the rated output voltage and the available voltage at the rated output current are limited by the rated output power:
- Maximum output current at the rated output voltage = Rated output power ÷ Rated output voltage
- Maximum output voltage at the rated output current = Rated output power ÷ Rated output current
3.1 400 W Model
AC Input Characteristics
Specification/Model | UDP5040-40 | UDP5080-20 | UDP5160-8 | UDP5250-6 |
---|---|---|---|---|
AC Input | 100 Vac - 240 Vac, 50Hz - 60Hz, single-phase | |||
Nominal input rating | 85 Vac to 265 Vac | |||
Input Voltage Range | 47 Hz to 63 Hz | |||
Input Voltage Frequency Range | 100 Vac | 200 Vac | ||
Current (Typ.)*1 | 5.6 A | 2.8 A | ||
Surge Current (MAX) *2 | Below 25 Apeak | |||
Power (MAX)*3 | 560 W | |||
Power Factor (Typ.)*1 | 0.99 (Input voltage: 100 V), 0.97 (Input voltage: 200 V) | |||
Efficiency (MIN)*1 | 75% (Typ.) | |||
Output Hold Time (MIN)*3 | Above 20 ms |
*1. Rated output current at output rated power.
*2. After the power switch is turned on (within approximately 1 ms), the charge current component that flows through the internal EMC filter circuit capacitor is not included.
*3. At output rated power.
Output Characteristics
Specification/Model | UDP5040-40 | UDP5080-20 | UDP5160-8 | UDP5250-6 | |
---|---|---|---|---|---|
Output Voltage*1 | 40 V | 80 V | 160 V | 250 V | |
Rated Value | Output Current*1 | 40 A | 20 A | 8 A | 6 A |
Output Power | 400 W | ||||
Maximum Adjustable Voltage *2 | 42 V | 84 V | 168 V | 262.5 V | |
Output Accuracy *3 | ±(0.05% of the set value + 0.05% of the rated value) | ||||
Setting Resolution | 1 mV | 1mV | 10 mV | 10 mV | |
Display Resolution | 1 mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Power Regulation*4 | ±6 mV | ±10 mV | ±12.2 mV | ±26 mV | |
Load Regulation*5 | ±6 mV | ±10 mV | ±12.2 mV | ±26 mV | |
Ripple and Noise *6 | p-p*7 | 50 mV | 100 mV | 100 mV | 100 mV |
Voltage | rms*7 | 5 mV | 12 mV | 20 mV | 20 mV |
Rise Time | Rated Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms |
No Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms | |
Fall Time | Rated Load | Below 50 ms | 150 ms | 250 ms | 250 ms |
No Load | Below 500 ms | 1200 ms | 2000 ms | 2000 ms | |
Maximum Remote Sense Compensation Voltage (MAX) | 1.5 V | 4 V | 5 V | 5 V | |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Current*2 | 42 A | 21 A | 8.4 A | 6.3 A | |
Output Accuracy*9 | ±(0.5% of the set value + 0.1% of the rated value) | ||||
Setting Resolution | 1 mA | 1 mA | 1 mA | 1 mA | |
Display Resolution | 1 mA | 1 mA | 1 mA | 1 mA | |
Power Regulation*4 | ±6 mA | ±4 mA | ±3 mA | ±2.5 mA | |
Current | Load Regulation*5 | ±13 mA | ±9 mA | ±7 mA | ±6 mA |
Ripple and Noise*6 | rms*7 | 80 mA | 40 mA | 30 mA | 12 mA |
Rise Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Fall Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Internal Resistance | 1.000 Ω | 4.000 Ω | 15 Ω | 31.25 Ω |
*1. The maximum output voltage and maximum output current are constrained by the maximum output power.
*2. OVP and OCP limits are set at approximately 95%.
*3. Applies to a 1% to 100% range of the rated voltage.
*4. For input voltages of 90 VAC to 121 VAC or 198 VAC to 242 VAC under constant load.
*5. When the power supply outputs the rated voltage and operates from no-load to rated load (i.e., outputs the rated current), voltage regulation is calculated as: Output Voltage Regulation= (Load Voltage at No-Load Current - Load Voltage at Rated Load Current) ÷ Load Voltage at Full Load Current ×100%
*6. In constant voltage mode, with the power supply delivering the rated current at rated power, measure the ripple and noise at the output using an oscilloscope
*7. Measurement bandwidth: 20 MHz.
*8. Ambient temperature range: 0°C to 50°C.
*9. Applies to a 1% to 100% range of the rated current. For 0% to 1%, values are typical (Typ.)
Display Characteristics
Specification/Model | UDP5040-40 | UDP5080-20 | UDP5160-8 | UDP5250-6 | |
---|---|---|---|---|---|
Maximum Display | 99.999 | 99.999 | 999.99 | 999.99 | |
Voltage Display | Minimum Display*1 | 0.100 | 0.100 | 0.50 | 0.50 |
Display Accuracy | 1 mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Current Display | Maximum Display | 99.999 | 99.999 | 9.999 | 9.999 |
Display Accuracy | 1 mA | 1 mA | 1 mA | 1 mA | |
Power Display | Maximum Display | 999.99 | 999.99 | 999.99 | 999.99 |
Display Accuracy | 1 mW (<100 W), 10 mW (≥100 W) |
*1. For the 5040 and 5080 models, the output voltage is 0.1 V, while for the 5160 and 5250 models, the output voltage is 0.5 V.
3.2 800 W Model
AC Input Characteristics
Specification/Model | UDP5040-80 | UDP5080-40 | UDP5160-16 | UDP5250-12 |
---|---|---|---|---|
AC Input | 100 Vac - 240 Vac, 50Hz - 60Hz, single-phase | |||
Nominal input rating | 85 Vac to 265 Vac | |||
Input Voltage Frequency Range | 47 Hz to 63 Hz | |||
Current (Typ.)*1 | 11.2 A | 5.6 A | ||
Surge Current (MAX) *2 | Below 50 Apeak | |||
Power (MAX)*3 | 1120 W | |||
Power Factor (Typ.)*1 | 0.99 (Input voltage: 100 V), 0.97 (Input voltage: 200 V) | |||
Efficiency (MIN)*1 | 75% (Typ.) | |||
Output Hold Time (MIN)*3 | Above 20 ms |
*1. Rated output current at output rated power.
*2. After the power switch is turned on (within approximately 1 ms), the charge current component that flows through the internal EMC filter circuit capacitor is not included.
*3. At output rated power.
Output Characteristics
Specification/Model | UDP5040-80 | UDP5080-40 | UDP5160-16 | UDP5250-12 | |
---|---|---|---|---|---|
Output Voltage*1 | 40 V | 80 V | 160 V | 250 V | |
Rated Value | Output Current*1 | 80A | 40 A | 16 A | 12 A |
Output Power | 800 W | ||||
Maximum Adjustable Voltage *2 | 42 V | 84 V | 168 V | 262.5 V | |
Output Accuracy*3 | ±(0.05% of the set value + 0.05% of the rated value) | ||||
Setting Resolution | 1mV | 1 mV | 10 mV | 10 mV | |
Display Resolution | 1mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Power Regulation*4 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Load Regulation*5 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Ripple and Noise *6 | p-p*7 | 50 mV | 100 mV | 100 mV | 100 mV |
Voltage | rms*7 | 5 mV | 12 mV | 20 mV | 20 mV |
Rise Time | Rated Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms |
No Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms | |
Fall Time | Rated Load | Below 50 ms | 150 ms | 250 ms | 250 ms |
No Load | Below 500 ms | 1200 ms | 2000 ms | 2000 ms | |
Maximum Remote Sense Compensation Voltage (MAX) | 1.5 V | 4 V | 5 V | 5 V | |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Current*2 | 84A | 42 A | 16.8 A | 12.6 A | |
Output Accuracy*9 | ±(0.5% of the set value + 0.1% of the rated value) | ||||
Setting Resolution | 1 mA | 1 mA | 1 mA | 1 mA | |
Display Resolution | 1 mA | 1 mA | 1 mA | 1 mA | |
Power Regulation*4 | ±10 mA | ±6 mA | ±5 mA | ±3 mA | |
Current | Load Regulation*5 | ±21 mA | ±13 mA | ±10 mA | ±7 mA |
Ripple and Noise*6 | rms*7 | 160 mA | 80 mA | 40 mA | 24 mA |
Rise Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Fall Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Internal Resistance | 0.500 Ω | 2.000 Ω | 7.5 Ω | 15.625 Ω |
*1. The maximum output voltage and maximum output current are constrained by the maximum output power.
*2. OVP and OCP limits are set at approximately 95%.
*3. Applies to a 1% to 100% range of the rated voltage.
*4. For input voltages of 90 VAC to 121 VAC or 198 VAC to 242 VAC under constant load.
*5. When the power supply outputs the rated voltage and operates from no-load to rated load (i.e., outputs the rated current), voltage regulation is calculated as: Output Voltage Regulation= (Load Voltage at No-Load Current - Load Voltage at Rated Load Current) ÷ Load Voltage at Full Load Current ×100%
*6. In constant voltage mode, with the power supply delivering the rated current at rated power, measure the ripple and noise at the output using an oscilloscope
*7. Measurement bandwidth: 20 MHz.
*8. Ambient temperature range: 0°C to 50°C.
*9. Applies to a 1% to 100% range of the rated current. For 0% to 1%, values are typical (Typ.)
Display Characteristics
Specification/Model | UDP5040-80 | UDP5080-40 | UDP5160-16 | UDP5250-12 | |
---|---|---|---|---|---|
Maximum Display | 99.999 | 99.999 | 999.99 | 999.99 | |
Voltage Display | Minimum Display*1 | 0.100 | 0.100 | 0.50 | 0.50 |
Display Accuracy | 1 mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Current Display | Maximum Display | 99.999 | 99.999 | 99.999 | 99.999 |
Display Accuracy | 1 mA | 1 mA | 1 mA | 1 mA | |
Power Display | Maximum Display | 999.99 | 999.99 | 999.99 | 999.99 |
Display Accuracy | 1 mW (<100 W), 10 mW (≥100 W) |
*1. For the 5040 and 5080 models, the output voltage is 0.1V, while for the 5160 and 5250 models, the output voltage is 0.5 V.
3.3 1200 W Model
AC Input Characteristics
Specification/Model | UDP5040-120 | UDP5080-60 | UDP5160-24 | UDP5250-18 |
---|---|---|---|---|
AC Input | 100 Vac - 240 Vac, 50Hz - 60Hz, single-phase | |||
Nominal input rating | 85 Vac to 265 Vac | |||
Input Voltage Frequency Range | 47 Hz to 63 Hz | |||
Current (Typ.)*1 | 16.8 A | 8.4 A | ||
Surge Current (MAX) *2 | Below 75 Apeak | |||
Power (MAX)*3 | 1680 W | |||
Power Factor (Typ.)*1 | 0.99 (Input voltage: 100 V), 0.97 (Input voltage: 200 V) | |||
Efficiency (MIN)*1 | 75% (Typ.) | |||
Output Hold Time (MIN)*3 | Above 20 ms |
*1. Rated output current at output rated power.
*2. After the power switch is turned on (within approximately 1 ms), the charge current component that flows through the internal EMC filter circuit capacitor is not included.
*3. At output rated power.
Output Characteristics
Specification/Model | UDP5040-120 | UDP5080-60 | UDP5160-24 | UDP5250-18 | |
---|---|---|---|---|---|
Output Voltage*1 | 40 V | 80 V | 160 V | 250 V | |
Rated Value | Output Current*1 | 120 A | 60 A | 24 A | 18 A |
Output Power | 1200 W | ||||
Maximum Adjustable Voltage *2 | 42 V | 84 V | 168 V | 262.5 V | |
Output Accuracy*3 | ±(0.05% of the set value + 0.05% of the rated value) | ||||
Setting Resolution | 1mV | 1mV | 10 mV | 10 mV | |
Display Resolution | 1mV | 1 mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Power Regulation*4 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Load Regulation*5 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Ripple and Noise *6 | p-p*7 | 50 mV | 100 mV | 100 mV | 100 mV |
Voltage | rms*7 | 5 mV | 12 mV | 20 mV | 20 mV |
Rise Time | Rated Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms |
No Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms | |
Fall Time | Rated Load | Below 50 ms | 150 ms | 250 ms | 250 ms |
No Load | Below 500 ms | 1200 ms | 2000 ms | 2000 ms | |
Maximum Remote Sense Compensation Voltage (MAX) | 1.5 V | 4 V | 5 V | 5 V | |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Current*2 | 126 A | 63 A | 25.2 A | 18.9 A | |
Output Accuracy*9 | ±(0.5% of the set value + 0.1% of the rated value) | ||||
Setting Resolution | 10 mA | 1 mA | 1 mA | 1 mA | |
Display Resolution | 1 mA | 1 mA | 1 mA | 1 mA | |
Power Regulation*4 | ±14 mA | ±8 mA | ±6 mA | ±3.5 mA | |
Current | Load Regulation*5 | ±29 mA | ±17 mA | ±12 mA | ±8.0 mA |
Ripple and Noise*6 | rms*7 | 240 mA | 120 mA | 60 mA | 36 mA |
Rise Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Fall Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Internal Resistance | 0.333 Ω | 1.333 Ω | 5.00 Ω | 10.416 Ω |
*1. The maximum output voltage and maximum output current are constrained by the maximum output power.
*2. OVP and OCP limits are set at approximately 95%.
*3. Applies to a 1% to 100% range of the rated voltage.
*4. For input voltages of 90 VAC to 121 VAC or 198 VAC to 242 VAC under constant load.
*5. When the power supply outputs the rated voltage and operates from no-load to rated load (i.e., outputs the rated current), voltage regulation is calculated as: Output Voltage Regulation= (Load Voltage at No-Load Current - Load Voltage at Rated Load Current) ÷ Load Voltage at Full Load Current ×100%
*6. In constant voltage mode, with the power supply delivering the rated current at rated power, measure the ripple and noise at the output using an oscilloscope
*7. Measurement bandwidth: 20 MHz.
*8. Ambient temperature range: 0°C to 50°C.
*9. Applies to a 1% to 100% range of the rated current. For 0% to 1%, values are typical (Typ.)
Display Characteristics
Specification/Model | UDP5040-120 | UDP5080-60 | UDP5160-24 | UDP5250-18 | |
---|---|---|---|---|---|
Maximum Display | 99.999 | 99.999 | 999.99 | 999.99 | |
Voltage Display | Minimum Display*1 | 0.100 | 0.100 | 0.50 | 0.50 |
Display Accuracy | 1mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Current Display | Maximum Display | 99.999 | 99.999 | 99.999 | 99.999 |
Display Accuracy | 1 mA (<100 A) 10 mA (≥100 A) | 1 mA | 1 mA | 1 mA | |
Power Display | Maximum Display | 999.99 | 999.99 | 999.99 | 999.99 |
Display Accuracy | 1 mW (<100 W), 10 mW (≥100 W) |
*1. For the 5040 and 5080 models, the output voltage is 0.1V, while for the 5160 and 5250 models, the output voltage is 0.5 V.
3.4 2000 W Model
AC Input Characteristics
Specification/Model | UDP5040-200 | UDP5080-100 | UDP5160-40 | UDP5250-30 |
---|---|---|---|---|
AC Input | 100 Vac - 240 Vac, 50Hz - 60Hz, single-phase | |||
Nominal input rating | 85 Vac to 265 Vac | |||
Input Voltage Frequency Range | 47 Hz to 63 Hz | |||
Current (Typ.)*1 | 28 A | 14 A | ||
Surge Current (MAX) *2 | Below 125 Apeak | |||
Power (MAX)*3 | 2800 W | |||
Power Factor (Typ.)*1 | 0.99 (Input voltage: 100 V), 0.97 (Input voltage: 200 V) | |||
Efficiency (MIN)*1 | 75% (Typ.) | |||
Output Hold Time (MIN)*3 | Above 20 ms |
*1. Rated output current at output rated power.
*2. After the power switch is turned on (within approximately 1 ms), the charge current component that flows through the internal EMC filter circuit capacitor is not included.
*3. At output rated power.
Output Characteristics
Specification/Model | UDP5040-200 | UDP5080-100 | UDP5160-40 | UDP5250-30 | |
---|---|---|---|---|---|
Output Voltage*1 | 40 V | 80 V | 160 V | 250 V | |
Rated Value | Output Current*1 | 200 A | 100 A | 40 A | 30 A |
Output Power | 2000 W | ||||
Maximum Adjustable Voltage *2 | 42 V | 84 V | 168 V | 262.5 V | |
Output Accuracy*3 | ±(0.05% of the set value + 0.05% of the rated value) | ||||
Setting Resolution | 1 mV | 1 mV | 10 mV | 10 mV | |
Display Resolution | 1mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Power Regulation*4 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Load Regulation*5 | ±6 mV | ±10 mV | ±20 mV | ±26 mV | |
Ripple and Noise *6 | p-p*7 | 50 mV | 100 mV | 100 mV | 100 mV |
Voltage | rms*7 | 5 mV | 12 mV | 20 mV | 20 mV |
Rise Time | Rated Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms |
No Load | Below 50 ms | Below 100 ms | Below 100 ms | Below 100 ms | |
Fall Time | Rated Load | Below 50 ms | 150 ms | 250 ms | 250 ms |
No Load | Below 500 ms | 1200 ms | 2000 ms | 2000 ms | |
Maximum Remote Sense Compensation Voltage (MAX) | 1.5 V | 4 V | 5 V | 5 V | |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Current*2 | 210 A | 105 A | 42 A | 31.5 A | |
Output Accuracy*9 | ±(0.5% of the set value + 0.1% of the rated value) | ||||
Setting Resolution | 10 mA | 10 mA | 1 mA | 1 mA | |
Display Resolution | 1 mA (<100 A) 10 mA (≥100 A) | 1 mA (<100 A) 10 mA (≥100 A) | 1 mA | 1 mA | |
Power Regulation*4 | ±22 mA | ±12 mA | ±8 mA | ±4.5 mA | |
Current | Load Regulation*5 | ±45 mA | ±25 mA | ±18 mA | ±10.0 mA |
Ripple and Noise*6 | rms*7 | 400 mA | 200 mA | 100 mA | 60 mA |
Rise Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Fall Time (Typ.) | Rated Load | 50 ms | 100 ms | 100 ms | 100 ms |
Temperature Coefficient*8 | 100 ppm/°C | ||||
Maximum Adjustable Internal Resistance | 0.20 Ω | 0.800 Ω | 3.00 Ω | 6.250 Ω |
*1. The maximum output voltage and maximum output current are constrained by the maximum output power.
*2. OVP and OCP limits are set at approximately 95%.
*3. Applies to a 1% to 100% range of the rated voltage.
*4. For input voltages of 90 VAC to 121 VAC or 198 VAC to 242 VAC under constant load.
*5. When the power supply outputs the rated voltage and operates from no-load to rated load (i.e., outputs the rated current), voltage regulation is calculated as: Output Voltage Regulation= (Load Voltage at No-Load Current – Load Voltage at Rated Load Current) ÷ Load Voltage at Full Load Current ×100%
*6. In constant voltage mode, with the power supply delivering the rated current at rated power, measure the ripple and noise at the output using an oscilloscope
*7. Measurement bandwidth: 20 MHz.
*8. Ambient temperature range: 0°C to 50°C.
*9. Applies to a 1% to 100% range of the rated current. For 0% to 1%, values are typical (Typ.)
Display Characteristics
Specification/Model | UDP5040-200 | UDP5080100 | UDP5160-40 | UDP5250-30 | |
---|---|---|---|---|---|
Maximum Display | 99.999 | 99.999 | 999.99 | 999.99 | |
Voltage Display | Minimum Display™™ | 0.100 | 0.100 | 0.50 | 0.50 |
Display Accuracy | 1 mV | 1mV | 1 mV (<100 V) 10 mV (≥100 V) | 1 mV (<100 V) 10 mV (≥100 V) | |
Current Display | Maximum Display | 999.99 | 999.99 | 99.999 | 99.999 |
Display Accuracy | 1 mA (<100 A) 10 mA (≥100 A) | 1 mA (<100 A) 10 mA (≥100 A) | 1 mA | 1 mA | |
Power Display | Maximum Display | 999.99 | 999.99 | 999.99 | 999.99 |
Display Accuracy | 1 mW (<100 W), 10 mW (≥100 W) |
*1. For the 5040 and 5080 models, the output voltage is 0.1V, while for the 5160 and 5250 models, the output voltage is 0.5 V.
3.5 Common Specifications
Protection Function
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model |
---|---|---|---|---|
Protection Function | ||||
Over-Voltage Protection (OVP) | Automatically turns off*¹ when triggered, and a window display: "OVP is triggered, the output will be turned off." | |||
Setting Range | 0% to 105% of the rated output voltage. | |||
Setting Accuracy | 1 mV (for single-unit rated values < 100 V) 10 mV (for single-unit rated values ≥ 100 V) | |||
Over-Current Protection (OCP)*2 | Automatically turns off*¹ when triggered, and a window display: "OCP is triggered, the output will be turned off." | |||
Adjustable Range | 0% to 105% of the rated output current. | |||
Adjustable Accuracy | 1 mA (for single-unit rated values < 100 A) 10 mA (for single-unit rated values ≥ 100 A | |||
Front Output Terminal Over-Current Protection (FOCP)*3 | Automatically turns off*¹ when triggered, and a window display: "FOCP is triggered, the output will be turned off." | |||
Value (fixed) | > 10A (Typ.) | |||
Over-Temperature Protection (OTP) | Automatically turns off*¹ when triggered, and a window display: "OTP is triggered, the output will be turned off." | |||
Shutdown (SD) | Turn off the power supply | |||
Power Limit | Automatically turns off*¹ when triggered, and a window display: "Power limit is triggered, the output will be turned off." | |||
Value (fixed) | Close to 105% of the rated output power. | |||
Communication Monitoring (Watchdog) | Restarts automatically, and a window display: "The watchdog failed. It will automatically restart." | |||
Single-Control Series/Parallel Operation Protection (PRL ALM) | Automatically turns off the output, and a window display: "Master/slave failed, the output will be turned off." |
*1. For 2000 W models, this may result in output disconnection or circuit breaker tripping.
*2. This does not prevent peak discharge currents in the internal capacitors of the UDP5000 power supply's output section during sudden load changes.
*3. The front-panel output current is limited to a maximum of 10 A. Values exceeding 10 A will trigger protection. If the OCP (Over-Current Protection) threshold is lower than the FOCP (Fast Over-Current Protection) threshold, OCP takes precedence.
Signal Input/Output
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model | |
---|---|---|---|---|---|
Signal Input/Output | |||||
Monitoring | Voltage Monitor (VMON) | Voltage Monitor Range: 0 V to 10 V | |||
Signal Output | Current Monitor (IMON)*1 | Voltage Monitor Range: 0 V to 10 V | |||
Output State | This function is enabled when the power supply output is active. | ||||
State Output | CV Mode | This function is enabled when the power supply operates in constant voltage (CV) mode. | |||
CC Mode | This function is enabled when the power supply operates in constant current (CC) mode. | ||||
Alarm State | This function is enabled when an alarm is triggered. | ||||
Power State | Power State: The power is on when the switch is turned on. | ||||
Input (Trigger Input) | Logic: Low level (0 V to 1.5 V), high level (3.5 V to 5 V) Input resistance: 10 ΚΩ (Typ.) | ||||
Trigger Signal | Logic: Low level (0 V to 0.6 V), high level (4.2 V to 5 V) | ||||
Output (Trigger Output) | Pulse Width: 100 µs (Typ.) |
*1. Control the voltage on the current sampling path to regulate the output current.
Control Function
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model |
---|---|---|---|---|
Control Functions | ||||
Output Voltage Control (VPGM) | 0% to 100% of the rated output voltage Voltage Control Range: 0 V to 10 V | |||
Accuracy | ±5% | |||
External Voltage Control Output Current Control (IPGM) | 0% to 100% of the rated output voltage Voltage Control Range: 0 V to 10 V | |||
Accuracy | ±5% |
Output ON/OFF Control: When the switch is closed, a high-level signal enables the output. When the switch is open, a low-level signal disables the output.
Other Functions
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model |
---|---|---|---|---|
Other Functions | ||||
Output ON/OFF Delayer | Setting Range: 0.1 s to 9999.9 s*¹ Resolution: 0.1s, 1 s, 10 s, 100 s, 1000 s, 10000 s | |||
OVP/OCP Enable Delayer *2 | Setting Range: 10 ms to 10000 ms Resolution: 10 ms, 100 ms | |||
Preset Value | Group Number: 32*3 groups | |||
Keypad Lock | Unlocks all keys except the Output key | |||
List Mode | Setting Group: 0 to 128 groups Cycle Count: 1 to 9,999, infinite*3 | |||
Synchronous Operation Mode | Step Time: 0.1 s to 99,999.9 s Synchronizes voltage and current settings, recovery steps, and sequence execution in a P-sequence program. | |||
Multi-Unit External Parallel Operation Mode *4 | Supports up to four parallel-connected units (same model), including the master unit. | Supports up to two parallel-connected units (same model), including the master unit. | ||
Multi-Unit External Series Mode | Supports up to two series-connected units (same model). | |||
Multi-Communication (VMCB) | Master Connects to PC: Ethernet, USB Master Connects to Slave: CAN |
*1. Factory default setting is 0.1 s.
*2. Factory default setting is 10 ms.
*3. Set it to 0 for infinite loop.
*4. The difference between the master and slave is 5% (Typ.)
Operation Display
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model |
---|---|---|---|---|
Operation Display | ||||
Output ON/OFF | The Output button lights up green when the output is active. | |||
CV Operation | The display shows "CV" in green when in constant voltage mode. | |||
CC Operation | The display shows "CC" in red when in constant current mode. | |||
Alarm | When a protective function is activated, a window displays detailed error information and automatically turns off the output | |||
Key Lock | The Lock button lights up green when locked. | |||
Remote Control | The operation mode is displayed as "External Control." |
Interface
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model | |
---|---|---|---|---|---|
Interface | |||||
General Specification | Software Protocol | IEEE Std 488.2-1992 | |||
Command | SCPI Specification 1999.0 | ||||
Hardware | USB | USB 2.0 Data Rate: 480 Mbps (high speed) B-type outlet Receiving: The program message terminator is LF (Line Feed) or EOM. Sending: The program message terminator is LF and +EOM. | |||
Device Category | USBTMC-USB48 | ||||
Hardware | LAN | IEEE 802.3 100Base-TX/10Base-T LXI 2011 Ver.1.4 LXI HISLIP Rev.1.01 IPv4, RJ-45 connector VXI-11, SCPI-RAW, HISLIP For VXI-11 and HiSLIP: When receiving, the message terminator can be LF or END. For SCPI-RAW: When receiving, the terminator is a low-level signal. When transmitting, the terminator is also a low-level signal. |
General Specifications
Specification/Model | 400 W Model | 800 W Model | 1200 W Model | 2000 W Model | |
---|---|---|---|---|---|
Dimensions | |||||
Weight (Main Unit Only) | Approx. 3 kg | Approx. 5.5 kg | Approx. 7.5 kg | Approx.13 kg | |
Dimensions | Refer to the outline dimension drawing. | ||||
Environmental Conditions | Operating Environment | Indoor use, Overvoltage Category II | |||
Operating Temperature | 0°C to +50°C | ||||
Operating Humidity | 20% R.H to 85% R.H (non-condensing) | ||||
Storage Temperature | -20°C to +60°C | ||||
Storage Humidity | 90% R.H or below (non-condensing) |
Cooling Method | Fan-forced air cooling | ||||
Altitude | Below 2000 meters | ||||
Grounding Polarity | Both positive ground and negative ground | ||||
Grounding Voltage | Applicable models: UDP5040, UDP5080, UDP5160, and UDP5250 Maximum: ±500 Vmax | ||||
Voltage Withstand | Through Main Circuit and Chassis | 1500 VAC applied for 1 minute with no abnormalities. | |||
Through Main Circuit and Secondary Circuit | UDP5040, UDP5080, UDP5160, and UDP5250 models: 1650 VAC applied for 1 minute with no abnormalities. | ||||
Insulation Resistance | Through Main Circuit and Secondary Circuit Across Secondary Circuit and Chassis | 100 ΜΩ or more (at ≤70% R.H), measured at 500 VDC. UDP5040, UDP5080, UDP5160, and UDP5250 models: 100 ΜΩ or more (at ≤70% R.H), measured at 500 VDC. UDP5040, UDP5080, UDP5160, and UDP5250 models: 40 ΜΩ or more (at ≤70% R.H), measured at 500 VDC. | |||
Electro Magnetic Compatibility (EMC)*1*2 | EMC 2014/30/EU EN61326-1(Class A*3) EN55011(Class A*3, Group I) EN61000-3-2 EN61000-3-3 Applicable Conditions: Cables and wires connected to this product must be ≤3 meters in length. | ||||
Safety *1 | Low Voltage Directive 2014/35/EU*2 EN61010-1(Class I*4, Pollution degree 2*5) |
*1. Not applicable to customized or modified products.
*2. Applies only to products with the CE marking.
*3. Class A device: Designed for industrial environments. Operation in residential areas may cause interference. To prevent radio/TV signal disruption, users should implement measures to reduce electromagnetic emissions if necessary.
*4. Class I device: Requires proper grounding of the protective earth terminal. Proper grounding is essential for safety compliance.
*5. Pollution Degree 2: Defined as non-conductive pollution (e.g., dust, oil mist) that may temporarily become conductive due to condensation, reducing insulation withstand capacity and surface resistivity.
Limited Warranty and Liability
UNI-T guarantees that the Instrument product is free from any defect in material and workmanship within three years from the purchase date. This warranty does not apply to damages caused by accident, negligence, misuse, modification, contamination, or improper handling. If you need a warranty service within the warranty period, please contact your seller directly. UNI-T will not be responsible for any special, indirect, incidental, or subsequent damage or loss caused by using this device. For the probes and accessories, the warranty period is one year. Visit instrument.uni-trend.com for full information.
Learn more at: www.uni-trend.com
UNI-T is the licensed trademark of UNI-TECHNOLOGY (CHINA) CO., Ltd.
Register your product to confirm your ownership. You will also get product notifications, update alerts, exclusive offers and all the latest information you need to know.
UNI-T products are protected under patent laws in China and internationally, covering both granted and pending patents. Licensed software products are the properties of UNI-Trend and its subsidiaries or suppliers, all rights reserved. This manual contains information that replaces all earlier published versions. The product information in this document subject to update without notice. For more information on UNI-T Test & Measure Instrument products, applications, or service, please contact UNI-T instrument for support, the support center is available on www.uni-trend.com ->instruments.uni-trend.com
https://instruments.uni-trend.com/ContactForm/
Headquarter
UNI-TREND (CHINA) CO., Ltd.
Address: No.6, Industrial North 1st Road, Songshan Lake Park, Dongguan City, Guangdong Province, China
Tel: (86-769) 8572 3888
Europe
UNI-TREND TECHNOLOGY EU GmbH
Address: Affinger Str. 12
86167 Augsburg Germany
Tel: +49 (0) 821 8879980
North America
UNI-TREND TECHNOLOGY US INC.
Address: 3171 Mercer Ave STE 104, Bellingham, WA 98225
Tel: +1-888-668-8648
Related Documents
![]() |
UNI-T UDP5000 Series Programmable DC Power Supplies User Manual User manual for the UNI-T UDP5000 Series Programmable DC Power Supplies. This manual provides detailed information on product overview, features, specifications, safety instructions, and installation procedures for these versatile DC power supplies. |
![]() |
UNI-T UDP5000 Series Programmable DC Power Supplies QuickStart Guide This guide provides a quick start to using the UNI-T UDP5000 Series Programmable DC Power Supplies, covering product overview, features, basic configuration, key functions, troubleshooting, and advanced operations like parallel and series connections, and remote control. |
![]() |
UNI-T UDP3305S-U Programmable DC Power Supply Datasheet Datasheet for the UNI-T UDP3305S-U Programmable DC Power Supply, detailing its features, technical specifications, and applications in research, industry, and education. Discover its multi-channel output, high precision, and connectivity options. |
![]() |
UNI-T UDP3303C-U Linear DC Power Supply User Manual This user manual provides comprehensive instructions for the UNI-T UDP3303C-U Linear DC Power Supply, covering its features, specifications, safety guidelines, operation, and maintenance. It details how to set voltage and current, use series and parallel modes, manage protection features, and utilize various interfaces like USB and RS232. |
![]() |
UNI-T UDP3303C-U Linear DC Power Supply User Manual User manual for the UNI-T UDP3303C-U Linear DC Power Supply, detailing its features, specifications, operation, and safety guidelines. |
![]() |
UDP3305S-U Programmable DC Power Supply User Manual This user manual provides comprehensive instructions for the UDP3305S-U Programmable DC Power Supply, covering its features, specifications, front and rear panel operations, inspection, installation, preparation for start-up, and advanced functions like list mode, delayer, preset, monitor, and trigger. |
![]() |
UNI-T UDP4303S Programmable Linear DC Power Supply User's Manual Comprehensive user's manual for the UNI-T UDP4303S Programmable Linear DC Power Supply, covering features, operation, safety, and technical specifications. |
![]() |
UDP5000 Series Programmable Digital Power Supply SCPI Programming Manual This manual provides comprehensive SCPI programming instructions for the UNI-T UDP5000 Series Programmable Digital Power Supply, covering command formats, parameters, and various functions for controlling and monitoring the power supply. |