GIMA ARM-30E+ Digital Blood Pressure Monitor

User Manual

Product Information

Model: ARM-30E+

Manual Version: A4A3

Publication Date: 2024-03

Table of Contents

  1. Checks upon opening the package
  2. Package contents
  3. Safety precautions
  4. Product composition
  5. Intended use / Instructions for use
  6. Contraindications
  7. Product components
  8. Preparation
  9. Function settings
  10. Procedure for correct measurement execution
  11. Warnings and precautions
  12. Frequently asked questions about blood pressure
  13. Anomaly management
  14. Cleaning and disinfection
  15. Care and maintenance
  16. Specifications
  17. Appendix 1 EMC Information

1. Checks upon opening the package

Before use, carefully open the package and verify that all parts listed in the following table of package contents are present and have not been damaged during transport. Proceed with installation and use strictly following the manual.

2. Package contents

No. Name Quantity
1 Sphygmomanometer 1
2 Cuff 22~42cm (8.6~16.5 inches) 1
3 User Manual 1
4 Quick Guide 1

3. Safety precautions

The warnings and illustrations in this manual are intended to allow the user to operate the device safely and correctly, avoiding harm to themselves and others. The specific meaning of each symbol is as follows:

4. Product composition

This product consists of a main unit and a cuff.

5. Intended use / Instructions for use

The Sphygmomanometer is intended for use in a domestic environment or in healthcare facilities for measuring systolic and diastolic blood pressure and pulse rate in adults using a non-invasive oscillometric technique.

Intended users

  1. Healthcare professional or lay user.
  2. Who has read and understood the user manual.

Clinical benefits

The device allows patients to monitor their systolic blood pressure, diastolic blood pressure, and pulse rate at any time in their own home. This allows for a significant reduction in hospital visits, and the consequent risks associated with travel, improving patients' quality of life.

6. Contraindications

To avoid inaccurate measurements or injuries, do not use this device in the presence of any of the following contraindications:

  1. This device is not suitable for patients with implanted electronic devices, such as pacemakers and defibrillators.
  2. Avoid measuring on the same side of the arm as a mastectomy or lymph node dissection.
  3. The device measures blood pressure using a pressurized cuff. Do not use the device if the limb chosen for measurement has lesions (e.g., open wounds), is affected by ongoing pathologies, or is used for administering therapy (e.g., phlebolysis); these conditions make it unsuitable for contact with or exposure to the pressure exerted by the cuff and may worsen the lesions or pathologies.
  4. Avoid taking measurements on patients with disorders, diseases, and sensitivity to certain environmental conditions that may render them unable to control their movements (e.g., tremors or shivers) and on patients unable to communicate clearly (e.g., infants and unconscious patients).
  5. To determine blood pressure, the device uses the oscillometric method. The arm subjected to measurement must have normal perfusion.

The device is not intended for use on limbs where blood circulation is limited or altered. If you suffer from perfusion or blood disorders, consult your doctor before using the device.

7. Product components

(1) Main unit

Display

User selection and date/time setting button

Power on/off button

Memory access button

Speaker (audio version only)

USB-C input for DC power

(2) Display screen

Date and time

User icon

Inflation icon

Deflation icon

Memory icon

Correct cuff placement indicator

Registration number

Movement detection indicator

Heartbeat icon

Blood pressure icon

Average blood pressure icon

Pulse rate value

Battery charge status indicator

Irregular heartbeat indicator

Pressure unit

8. Preparation

(1) Battery installation

  1. Open the battery compartment cover as illustrated in the figure.
  2. Insert 4 AAA batteries into the battery compartment, paying attention to the battery polarity indication. Refer to the image below for correct battery installation.

Warning: If you do not plan to use the device for a prolonged period (over 3 months), remove the batteries.

(3) USB-C power supply (cable not included in the package)

The product can be powered by connecting an external DC power adapter (5V, 1A) via the USB-C connector.

NOTE:

9. Function settings

(1) User selection

When the device is in standby, press the "?" button to access the user group selection interface. Then press the "?" button again to switch to other user groups.

[Image showing user selection]

(2) Date and time setting

When the device is in standby, press the "?" button for approximately 3 seconds to access the date setting; the "year" will flash. Press the "▶️" button to set the desired year, then use the "?" button to confirm the selection. Once the "year" is set, the month setting will be accessed automatically. At this point, the "month" icon will flash. Use the "▶️" button to scroll to the desired setting. Proceed analogously to set "day", "hour", and "minute".

[Image showing date and time setting progression]

(3) Display unit setting

When the device is in standby, press and hold the "⚙️" button for about 5 seconds to access the unit setting. Press the "⚖️" button to switch from mmHg to kPa. Press the "" button to confirm the selected unit. The default unit is mmHg.

10. Procedure for correct measurement execution

(1) Preparation for measurement

(2) Wearing the cuff correctly

  1. Connect the cuff to the device by firmly inserting the air tube plug into the designated port.

[Image showing cuff connection]

Cuff

Air tube plug

Air tube

Port for air tube

  1. Insert your hand inside the cuff. Pull the free end of the cuff until it reaches the upper left part of your arm.

[Image showing cuff application]

Note: The lower edge of the cuff should be about 2 - 3 cm (0.8-1.2 inches) above the inner elbow. The air tube should be positioned on the inside of the arm and aligned with the middle finger.

[Diagram showing cuff placement relative to the elbow]

Lower edge: 2 - 3 cm (0.8-1.2 inches)

Ensure that the air tube is positioned on the inner side of the arm and that the cuff is wrapped snugly so that it cannot move around the arm.

Note: Repeated measurements can cause blood congestion in the arm, which will affect the measurement results. Be careful not to rest your arm on the air tube.

To avoid potential blood congestion and ensure measurement accuracy, it is recommended to raise your left hand, open and close your fist several times, or remove the cuff and rest for at least 2-3 minutes before taking the measurement.

(3) Sitting correctly

Take the measurement when relaxed and comfortably seated in a comfortable environment with a comfortable temperature. Rest your arm on a flat surface, such as a table.

Warning: Be careful not to twist the connection tube, as the pressure exerted by the cuff could disturb blood flow and cause patient injury.

(4) Taking a measurement

  1. Fasten the cuff following the instructions in point 2 and start the measurement only after correctly wearing the cuff.
  2. Press the "▶️" button. All icons will be displayed, then the device will start inflating the cuff and "00" will be displayed on the screen.

[Image sequence showing measurement process]

Check the measured values at the end of the measurement.

Note: If you feel unwell during the measurement, press the "▶️" button immediately to stop the measurement.

When the correct inflation pressure is reached, the value displayed on the screen will gradually decrease, and the heartbeat symbol will flash. At the end of the measurement, the results of systolic blood pressure, diastolic blood pressure, and pulse will be displayed on the screen.

Note: In case of unexpected results, consult your doctor.

(5) Memory function

  1. Each measured value is automatically stored in the appropriate user group. This device can store up to 99 sets of measurements for each user. When the memory is full, older values will be replaced by new ones.

When the device is in standby mode, press the "?" button once to view the average value of the last 2 or 3 blood pressure measurements. Pressing the "?" button again will display the last measured value.

Pressing the "?" button again will allow you to scroll through the other results in memory.

(6) Memory deletion

When the device is in standby mode, press the "?" button to select the user group for which you want to delete the measured values.

Press the "▶️" button to turn off the device, then press the "▶️" button once to activate the screen. Repeatedly press the "" button for 3 seconds to delete the results in the selected user's memory; finally, the "" icon will be displayed on the screen.

(7) Correct cuff placement indication

When the cuff is worn correctly, the " (OK)" icon is always displayed on the screen. If the cuff is not tight enough, the "⚠️ o)" icon will flash to indicate this. If the "⚠️ o)" icon flashes, press the "▶️" button to stop the measurement.

(8) Movement detection indication

If the user moves their body or arms during measurement, incorrect results may be obtained. In this case, the "?" icon will flash, and it will be necessary to correct the posture and repeat the measurement.

(9) Device shutdown

To turn off the sphygmomanometer, simply press the "▶️" button. The device will automatically turn off after 1 minute.

11. Warnings and precautions

Warnings

Precautions

12. Frequently asked questions about blood pressure

Q1: Why can blood pressure readings taken at home be lower than those taken at the hospital?

Q2: Why can blood pressure readings taken at home be higher than those taken at the hospital?

Q3: What is the best time to take measurements?

Q4. Why is the blood pressure value different each time?

  1. Blood pressure varies slightly with each contraction. For example, in a subject with a pulse of 70 beats per minute, blood pressure will vary 100,800 times a day. Since blood pressure changes continuously, it is difficult to obtain the correct value by measuring it only once. It is advisable to repeat the measurement 2-3 times. Normally, the first measurement result is higher due to agitation or inadequate preparation, as in the second measurement you are more relaxed, the measurement result is usually 5-10 mmHg (0.7-1.3 kPa) lower than the first. This difference is more evident in subjects with higher blood pressure.

--When taking consecutive measurements, keep in mind that: Due to cuff compression, blood extravasation may occur, resulting in poor blood flow to the fingertips. Subsequent measurements after blood extravasation will not produce correct results. Loosen the cuff, raise your hand above your head, grasp and extend your palms 5 times. This will help drain the extravasated blood, and you can proceed with blood pressure measurement.

13. Anomaly management

* If the measurement does not occur correctly, one of the following symbols will be displayed. Use the correct measurement method.

Errors Cause
Er U Inflation does not reach 30 mmHg in 12 seconds.
Er H Inflation reaches 295 mmHg.
Er 1 Pulse rate is not detected correctly.
Er 2 Excessive disturbances (patient moves, talks, or there are electromagnetic disturbances during measurement).
Er 3 Measurement result is abnormal.
Er 23 SYS value is less than 57 mmHg.
Er 24 SYS value is greater than 255 mmHg.
Er 25 DIA value is less than 25 mmHg.
Er 26 DIA value is greater than 195 mmHg.

* Troubleshooting

Anomaly Possible fault Solution
Impossible to turn on the device Battery charge is insufficient. Replace batteries or connect the USB-C charging cable for power.
No pressurization Batteries inserted with reversed polarity. Insert batteries correctly.
The air tube plug may not be inserted securely. Insert the air tube plug securely into the socket.
Impossible to take measurement due to a display error The air tube is broken or has a leak. Contact the retailer to replace the cuff with a new one.
The arm was moved during pressurization. Keep the arm and body still.
Talking during measurement. Remain silent during blood pressure measurement.
Air leak from the cuff The cuff is not wrapped tightly enough. Tighten the cuff.
The cuff's air bladder is torn. Contact the retailer to replace the cuff with a new one.

If, after applying the solutions described above, it is still not possible to measure blood pressure, contact the retailer. DO NOT attempt to disassemble the device yourself.

14. Cleaning and disinfection

(1) Cleaning

The device can be cleaned with a soft, clean cloth dampened with a small amount of neutral detergent or water. It is recommended to clean the device before and after each use. Each cleaning should not last more than 3 minutes. Do not clean more than 3 times consecutively.

⚠️ Do not use corrosive detergents. During cleaning, be careful not to immerse any part of the device to prevent liquid from entering the instrument.

(2) Disinfection

Recommended disinfectant: 75% alcohol for medical use

Steps:

  1. Thoroughly clean the device with a soft, clean cloth dampened with a small amount of the disinfectant indicated above and immediately dry with a soft, clean, and dry cloth.
  2. The device body can also be cleaned and disinfected with a soft, clean cloth dampened with a small amount of 75% medical alcohol.

⚠️ Do not use steam disinfection, high temperature, or UV rays. These methods could damage the device and shorten its lifespan.

It is recommended to disinfect the device before and after use. Disinfection should not last more than 1 minute. Do not disinfect more than 2 times consecutively.

(3) Disposal

The disposal of the device, other components, and optional accessories must be carried out according to local regulations. Improper disposal can pollute the environment.

Notes:

15. Care and maintenance

[Image of cleaning supplies and device]

Water or neutral detergent

Keeping the device surface clean and tidy contributes to prolonging the life of the sphygmomanometer.

Warning: Do not allow water or other liquids to penetrate the device. If this happens, the device and cuff will be damaged, and the sphygmomanometer will no longer be usable.

16. Specifications

Product Name Sphygmomanometer
Model ARM-30E+
Display LCD display
Measurement Method Oscillometric measurement
Measurement Point Arm
Measurement Range Blood pressure: SYS: 57 ~ 255 mmHg; DIA: 25 ~ 195 mmHg Pulse rate: 40~199 bpm
Cuff Pressure Accuracy ±3 mmHg (±0.4kPa)
Pulse Rate Accuracy ±5%
Low Battery 4.2V±0.1V: low battery; <4.0V±0.1V: device shuts down
Automatic Shutdown 1 minute of inactivity
Power Source 4 AAA DC batteries. 6V or DC 5V 1A with USB-C cable
Applied Part Type BF
Operating Mode Continuous operation
IP Classification IP21
Weight Approx. 220 g (without batteries)
Dimensions 118 mm (length)*98 mm (width)*62.5 mm (height)
Screen Dimensions 44.5 mm (L) x 58.5 mm (W) 2.9 inches
Cuff Dimensions 22~42 cm (8.6~16.5 inches)
Device Lifespan 5 years
Cuff Lifespan 10000 uses
Type of protection against electric shock ME equipment powered internally (when using batteries only) Class II ME equipment (if equipped with AC adapter)
Environmental operating conditions Temperature range: 5°C~40°C Humidity range: 15%~95% relative humidity Pressure range: 70kPa~106kPa
Transport and storage environment During transport, avoid subjecting the device to strong shocks, direct impacts, and exposure to direct sunlight or rain. Store the device and other components in a clean and safe place. Remove the cuff from the monitoring device. Gently fold the air tube into the cuff. The Sphygmomanometer should be stored in its packaging in a closed environment, free from corrosive gases and with good ventilation, where the temperature is between -20°C and 55°C, with a relative humidity of 10%~93%, and atmospheric pressure between 70kPa~106kPa.

The product has been clinically tested according to the requirements of the ISO 81060-2 standard.

Note: The specified power supply must meet the following conditions: Output voltage: 5V DC, Output current: 1000mA, Compliance with IEC 60601-1 standard, Class II. At least two MOPP isolations between the AC input and DC output. Compliance with US and Canadian deviation requirements.

Essential Functionality

  1. Measurement Range (Blood Pressure): SYS: 57 - 255 mmHg, DIA: 25 - 195 mmHg Pulse Rate: 40 - 199 bpm
  2. Cuff Pressure Accuracy: ±3 mmHg (±0.4 kPa)
  3. Pulse Rate Accuracy: ±5%

17. Appendix 1 EMC Information

Manufacturer's guidelines and declaration - Electromagnetic Emissions

The Sphygmomanometer is intended for use in the electromagnetic environment specified below. The customer or user of the Sphygmomanometer must ensure that it is used in such an environment.

Test on emissions Conformity Electromagnetic environment - guidance
RF EMISSIONS CISPR 11 Group 1 The Sphygmomanometer uses RF energy only for its internal function. As a result, RF emissions are very low and unlikely to cause interference with nearby electronic equipment.
RF EMISSIONS CISPR 11 Class B The Sphygmomanometer is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power network that supplies buildings for residential use.
Harmonic Emissions IEC61000-3-2 N.A.
Voltage Fluctuations/Flicker Emissions IEC61000-3-3 N.A.

Manufacturer's guidelines and declaration - Electromagnetic Immunity

The Sphygmomanometer is intended for use in the electromagnetic environment specified below. The customer or user of the Sphygmomanometer must ensure that it is used in such an environment.

Immunity Test IEC 60601 Test Level Conformity Level
Electrostatic Discharge (ESD) IEC 61000-4-2 ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transients/burst IEC 61000-4-4 Input/output of the signal ±1 kV repetition frequency 100 kHz Input/output of the signal ±1 kV repetition frequency 100 kHz
Surge IEC 61000-4-5 Not applicable Not applicable
Voltage dips, short interruptions and voltage variations on power supply lines IEC 61000-4-11 Not applicable Not applicable
Magnetic field at power frequency IEC 61000-4-8 30A/m, 50/60Hz 30A/m, 50/60Hz
RF conducted IEC61000-4-6 Input/output of the signal 3 V; 0.15MHz-80MHz 6 V in the ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 2Hz Input/output of the signal 3 V; 0.15MHz-80MHz 6 V in the ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 2Hz
RF radiated IEC61000-4-3 10 V/m 80 MHz - 2.7 GHz 80% AM at 2 Hz 10 V/m 80 MHz- 2.7 GHz 80% AM at 2 Hz

NOTE: UT is equivalent to the AC mains voltage before the test level is applied.

Manufacturer's guidelines and declaration - Electromagnetic Immunity

The Sphygmomanometer is intended for use in the electromagnetic environment specified below. The customer or user of the Sphygmomanometer must ensure that it is used in such an environment.

RF radiated IEC61000-4-39 (Test specifications for IMMUNITY OF THE ENCLOSURE to proximity magnetic fields) Frequency range (MHz) Modulation Test Level (V/m) Conformity Level (V/m)
385 380-390 TETRA 400, 460, FRS 460 Modulation pulse 18 Hz 1.8 0.3 27 27
450 430-470 GMRS FM de-2 modulation of ±5 kHz sinusoidal at 1 kHz 2 0.3 28 28
710-787, 745-780 Banda LTE 13, 17 Modulation pulse 217 Hz 0.2 0.3 9 9
810-960, 870-930 GSM 800/900, TETRA 800, DEN 820, CDMA 850, Banda LTE 5 Modulation pulse 18 Hz 2 0.3 28 28
1720-1990, 1970 GSM 1900; DECT; Banda LTE 1, 3, 4, 25; UMTS Modulation pulse 217 Hz 2 0.3 28 28
2450-2570 Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, Banda LTE 7 Modulation pulse 217 Hz 2 0.3 28 28
5240-5500, 5800-5785 WLAN 802.11 a/n Modulation pulse 217 Hz 0.2 0.3 9 9

Manufacturer's guidelines and declaration - Electromagnetic Immunity

RF radiated IEC61000-4-39 (Test specifications for IMMUNITY OF THE ENCLOSURE to proximity magnetic fields) Frequency (MHz) Modulation Test Level (A/m) Conformity Level (A/m)
30 kHz CW 8 8
134.2 kHz Pulse modulation 2.1 kHz 65 65
13.56 MHz Pulse modulation 50 kHz 7.5 7.5

Warning:

Note: Any serious incident that occurs in relation to the use of the device must be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient are located.

GIMA WARRANTY CONDITIONS

The standard Gima B2B warranty of 12 months applies.

Disposal: The product must not be disposed of with other household waste. Users must ensure the disposal of obsolete equipment by taking it to the designated collection point for the recycling of electrical and electronic equipment.

PDF preview unavailable. Download the PDF instead.

M49875IT Adobe PDF Library 25.1.250

Related Documents

Preview GIMA KD-595 Talking Automatic Blood Pressure Monitor User Manual
Comprehensive user manual for the GIMA KD-595 Talking Automatic Blood Pressure Monitor. Learn about setup, operation, specifications, troubleshooting, maintenance, and warranty.
Preview GIMA ARM-30E+ Blood Pressure Monitor User Manual
Comprehensive user manual for the GIMA ARM-30E+ upper arm blood pressure monitor, detailing setup, operation, safety precautions, troubleshooting, and technical specifications.
Preview GIMA Madrid Aneroid Sphygmomanometer - Professional Medical Product Information
Comprehensive guide to the GIMA Madrid Aneroid Sphygmomanometer (Model HS-201Q3), detailing its features, usage instructions, specifications, and maintenance. This document provides essential information for healthcare professionals and users.
Preview GIMA ABPM50 Ambulatory Blood Pressure Monitor Technical Specifications
Detailed technical specifications for the GIMA ABPM50 Ambulatory Blood Pressure Monitor, including manufacturer information, power requirements, and safety class.
Preview GIMA Pressure Activated Safety Lancets (PA) - Medical Device for Capillary Blood Sampling
Information on GIMA Pressure Activated Safety Lancets (Model PA), a sterile, single-use device for capillary blood sampling. Includes intended use, instructions, warnings, storage, and disposal.
Preview GIMA International Catalogue 2024 - Professional Medical Products
The GIMA International Catalogue 2024 presents over 9,500 professional medical products, including diagnostic tests, medical instruments, patient care equipment, and protective gear. GIMA emphasizes fast delivery, quality, innovation, and comprehensive customer service across 145 countries.
Preview PM10 Palm ECG Monitor User Manual
This user manual provides comprehensive information on the Gima PM10 Palm ECG Monitor, a personal electrocardiograph designed for home health care. It covers device operation, technical specifications, safety precautions, troubleshooting, and software usage for monitoring heart rate and ECG signals.
Preview GIMACARE Multi-Functional Monitoring System Owner's Manual
Comprehensive owner's manual for the GIMACARE Multi-Functional Monitoring System by GIMA, detailing setup, operation, testing procedures for blood glucose, beta-ketone, cholesterol, triglycerides, uric acid, hemoglobin, and lactate, maintenance, and troubleshooting.