Intended Use

This oscillometric blood pressure monitor is intended for measuring non-invasive blood pressure in people aged 12 years or older. It is clinically validated in patients with hypertension, hypotension, diabetes, pregnancy, pre-eclampsia, atherosclerosis, end-stage renal disease, obesity, and the elderly. The device can detect an irregular pulse suggestive of Atrial Fibrillation (AF). Please note that the device is not intended to diagnose AF. A diagnosis of AF can only be confirmed by ECG. The patient is advised to see a physician.

This device uses the same measuring technology as the award-winning «BP 3BTO-A» model tested according to the British and Irish Hypertension Society (BIHS) protocol.

Preparation for Measurement

  1. Sit on a back-supported chair and keep your legs uncrossed.
  2. Avoid thick or close-fitting garments on the upper arm.
  3. Place the artery mark on the cuff over your artery.
  4. Fit the cuff closely, but not too tight.
  5. Position the cuff 2-3 cm above your elbow.
  6. Keep your arm still and do not speak during the measurement.

Before Each Measurement

Device Components and Display

Device Components:

Display Indicators:

Using the Device for the First Time

Inserting the Batteries

Open the battery compartment 7, insert 4 new 1.5 V AA alkaline batteries, observing correct polarity.

Setting the Date and Time

  1. After battery insertion, the year flashes. Use the M-button 3 to set the year and the MAM button 4 to confirm.
  2. Set the month using the M-button, confirm with MAM button, then set the day.
  3. Repeat for hour and minutes.
  4. Once minutes are set and confirmed, date and time are displayed.
  5. To change date/time later, press and hold the MAM button for approx. 3 seconds until the year flashes.

Selecting the Correct Cuff

Microlife offers various cuff sizes. Select a size that matches your upper arm circumference for a close fit.

Cuff SizeUpper Arm Circumference
S17 - 22 cm
M22 - 32 cm
M - L22 - 42 cm
L32 - 42 cm
L - XL32 - 52 cm

Pre-shaped cuffs are optionally available. Only use Microlife cuffs. Contact Microlife Service if the enclosed cuff 8 does not fit.

Connect the cuff connector 9 into the cuff socket 5.

Selecting Standard or MAM Mode

Before each measurement, choose between standard (single measurement) or MAM (automatic triple measurement) mode.

Checklist for Taking a Reliable Measurement

Taking a Blood Pressure Measurement

  1. Select standard or MAM mode.
  2. Press the ON/OFF button 1 to start.
  3. The cuff inflates automatically. Relax, do not move or tense your arm muscles. Breathe normally and do not talk.
  4. The cuff fit check 18 on the display indicates placement. 18-A means suboptimal fit but measurement can proceed.
  5. Pumping stops when correct pressure is reached; pressure falls gradually.
  6. The pulse indicator 23 flashes during measurement.
  7. Results (systolic 12, diastolic 13, pulse 14) are displayed.
  8. Remove the cuff after measurement.
  9. Switch off the device (it switches off automatically after approx. 1 min).

AF detection is only activated in MAM mode.

You can stop measurement anytime by pressing the ON/OFF button.

This monitor is tested for use in pregnancy and pre-eclampsia. If high readings occur in pregnancy, re-measure after an hour. If still high, consult a doctor. The AFIB symbol can be ignored in pregnancy.

Manual Inflation

For high systolic pressure (e.g., >135 mmHg), you can set pressure individually. Press ON/OFF after the cuff inflates to ~30 mmHg, keep pressed until ~40 mmHg above expected systolic, then release.

How Not to Store a Reading

Press and hold the ON/OFF button 1 until «M» 16 flashes. Confirm deletion by pressing the MAM button 4. «CL» indicates deletion.

How to Evaluate Your Blood Pressure

The traffic light indicator 22 shows the range:

RangeSystolic (mmHg)Diastolic (mmHg)Recommendation
1. Blood pressure too high≥135≥85Seek medical advice
2. Blood pressure elevated130-13480-84Self-check
3. Blood pressure normal<130<80Self-check

The higher value determines the evaluation. Example: 140/80 mmHg or 130/90 mmHg indicates «blood pressure too high».

Appearance of the Pulse Arrhythmia (PAD)

The PAD symbol 20 indicates pulse irregularities. Results may deviate; repeat measurement. If the symbol appears regularly, consult your doctor.

Information for the Doctor on Frequent Appearance of the Arrhythmia Indicator

This device detects pulse irregularities. It does not replace a cardiac examination. In MAM mode, AF is also checked. If the symbol appears, select MAM mode and measure again.

Data Memory

The device stores the last 99 measurement values.

Viewing Stored Values

When the device is off, briefly press the M-button 3. It shows «M» 16 and «A» (average). Pressing M again shows previous values.

Readings with suboptimal cuff fit (18-A) are not included in the average.

The memory overwrites the oldest value when full (100th value). Consult a doctor before the memory is full.

Clearing All Values

To permanently remove all stored values, press and hold the M-button (device off) until «CL ALL» appears, then release. Press the MAM button 4 while «CL ALL» flashes to confirm. Individual values cannot be cleared.

Cancel deletion: Press ON/OFF button 1 while «CL ALL» flashes.

Battery Indicator and Battery Change

Low Battery

The battery symbol 17 flashes when batteries are low. The device can still measure reliably, but replacement batteries should be obtained.

Flat Battery - Replacement

The battery symbol 17 flashes when batteries are flat. No further measurements can be taken. Replace batteries.

  1. Open battery compartment 7.
  2. Replace batteries, observing polarity.
  3. Set date and time as described in Section 2. The year number will flash automatically.

Which Batteries and Which Procedure?

Use 4 new, long-life 1.5 V, size AA alkaline batteries. Do not use expired batteries. Remove batteries if the device is not used for a prolonged period.

Using Rechargeable Batteries

Use only «NiMH» type reusable batteries. Batteries must be removed and recharged when the flat battery symbol appears. Do not leave them in the device. Always remove rechargeable batteries if not using the device for a week or more. Batteries cannot be charged in the monitor; use an external charger.

Using a Mains Adapter

The device can be operated with a Microlife mains adapter (DC 6V, 600 mA). Use only the original Microlife adapter. Ensure the adapter and cable are undamaged.

  1. Plug the adapter cable into the mains adapter socket 6.
  2. Plug the adapter into the wall socket.

No battery current is consumed when the mains adapter is connected.

Error Messages

If an error occurs, the measurement is interrupted and an error message is displayed.

ErrorDescriptionPotential Cause and Remedy
«Err 1» [Signal too weak] (19)The pulse signals on the cuff are too weak.Re-position the cuff and repeat the measurement.*
«Err 2» [Error signal] (18-B)During measurement, error signals were detected by the cuff, caused by movement or muscle tension.Repeat the measurement, keeping your arm still.
«Err 3» [Abnormal cuff pressure] (18-C)An adequate pressure cannot be generated in the cuff. A leak may have occurred.Check cuff connection and fit. Replace batteries if necessary. Repeat measurement.
«Err 5» [Abnormal result]Measuring signals are inaccurate; no result can be displayed.Read checklist for reliable measurements and repeat.
«Err 6» [MAM Mode error]Too many errors in MAM mode, impossible to get a final result.Read checklist for reliable measurements and repeat.
«HI» [High pressure/pulse]Pressure in cuff too high (>299 mmHg) OR pulse too high (>200 bpm).Relax for 5 minutes and repeat measurement.*
«LO» [Low pulse]Pulse too low (<40 bpm).Repeat measurement.*

* Please immediately consult your doctor if this or any other problem occurs repeatedly.

Safety, Care, Accuracy Test and Disposal

Safety and Protection

Device Care

Clean the device with a soft, dry cloth.

Cleaning the Cuff

The cuff is washable.

  1. Remove cuff connector 9 from cuff tube 10. Pull bladder through cuff cover opening.
  2. Hand wash cuff cover in soapy water (not hotter than 30 °C).
  3. Dry cuff cover by air drying.
  4. Loop cuff tube back and place bladder flat in cuff cover.
  5. Reattach cuff connector.

The bladder must lie straight, not folded. Do not use fabric softener.

[Warning] Do not wash the cuff in a washing machine or dishwasher!

[Warning] Do not dry the cuff cover in a tumble dryer!

[Warning] Under no circumstances may you wash the inner bladder!

Accuracy Test

Recommend testing accuracy every 2 years or after mechanical impact. Contact Microlife-Service.

Disposal

[Disposal symbol] Batteries and electronic devices must be disposed of according to local regulations, not with domestic waste.

Guarantee

This device is covered by a 5-year guarantee from the date of purchase, valid with the completed guarantee card or receipt.

Technical Specifications

ParameterSpecification
Operating conditions10 - 40 °C / 50 - 104 °F; 15 - 95 % relative maximum humidity
Storage conditions-20 - +55 °C / -4 - +131 °F; 15 - 95 % relative maximum humidity
Weight402 g (including batteries)
Dimensions138 x 94.5 x 62.5 mm
Measuring procedureOscillometric, corresponding to Korotkoff method: Phase I systolic, Phase V diastolic
Measurement range20 - 280 mmHg – blood pressure; 40 - 200 beats per minute – pulse
Cuff pressure display range0 - 299 mmHg
Resolution1 mmHg
Static accuracyPressure within ± 3 mmHg
Pulse accuracy± 5% of the readout value
Voltage source4 x 1.5 V alkaline batteries; size AA; Mains adapter DC 6V, 600 mA (optional)
Battery lifetimeApprox. 920 measurements (using new batteries)
IP ClassIP20
Reference to standardsEN 1060-1 /-3 /-4; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service lifeDevice: 5 years or 10000 measurements; Accessories: 2 years

This device complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved.

PDF preview unavailable. Download the PDF instead.

ib-bp-b3-afib-bp3kt1-3n- en-0619 Acrobat Elements 9.0.0 (Windows)

Related Documents

Preview Microlife BP B3 AFIB Blood Pressure Monitor User Manual
Comprehensive user manual for the Microlife BP B3 AFIB Blood Pressure Monitor, detailing setup, usage, AFIB detection, safety precautions, and technical specifications for accurate home blood pressure monitoring.
Preview Microlife BP A200 AFIB: Accurate Blood Pressure & AFIB Detection
Explore the Microlife BP A200 AFIB, a state-of-the-art digital blood pressure monitor engineered for precise home health monitoring. It accurately measures systolic and diastolic blood pressure, as well as pulse rate, and features advanced Atrial Fibrillation (AFIB) detection technology.
Preview Microlife BP A200 AFib Blood Pressure Monitor User Manual and Guide
Comprehensive user manual for the Microlife BP A200 AFib digital blood pressure monitor. Learn about accurate home blood pressure measurement, AFib detection, device operation, safety, and maintenance.
Preview Microlife BP A200 AFIB Blood Pressure Monitor User Manual
Comprehensive user manual for the Microlife BP A200 AFIB blood pressure monitor, detailing features, usage instructions, AFIB detection, and health monitoring.
Preview Microlife BP A6 BT Blood Pressure Monitor - Instruction Manual
This is the official instruction manual for the Microlife BP A6 BT Blood Pressure Monitor. It guides users through preparation, accurate blood pressure and pulse measurement, and highlights key features such as Atrial Fibrillation (AFIB) detection and Bluetooth connectivity for health app integration.
Preview Microlife BP B3 AFIB Blood Pressure Monitor: User Manual & Features
Comprehensive user manual for the Microlife BP B3 AFIB Blood Pressure Monitor, detailing setup, usage, AF detection, safety precautions, and technical specifications. Learn how to accurately measure your blood pressure and monitor heart health.
Preview Microlife BP B6 Connect Bluetooth Blood Pressure Monitor User Manual
This user manual provides comprehensive instructions for the Microlife BP B6 Connect Bluetooth Blood Pressure Monitor. It covers setup, operation, features such as Atrial Fibrillation (AFIB) detection and Bluetooth connectivity for data transfer to the Microlife Connected Health+ app, as well as safety guidelines and technical specifications.
Preview Microlife BP A7 Touch BT Bluetooth Blood Pressure Monitor User Manual
Comprehensive user manual for the Microlife BP A7 Touch BT Bluetooth Blood Pressure Monitor, detailing setup, usage, features like AFIB detection, and technical specifications. Learn how to accurately measure blood pressure and monitor heart health.