S8-044-A1 Appendix 6 GA-map Dysbiosis Test Lx v2 Installation Guide v9
INSTALLATION GUIDE
for GA-map® Dysbiosis Test Lx v2

TABLE OF CONTENTS
Overview ..................................................................................................................................................................2 Installation check list................................................................................................................................................2 Recommended safety equipment............................................................................................................................2 Computer requirements ..........................................................................................................................................2 Laboratory requirements .........................................................................................................................................3 Equipment, materials, and reagents required .........................................................................................................4 Training Plan.............................................................................................................................................................9 Contact info..............................................................................................................................................................9
OVERVIEW
This guide provides the information required to prepare a laboratory for performing the GA-map® Dysbiosis Test Lx v2 assay. Requirements for instruments, materials, and reagents are described.
INSTALLATION CHECK LIST
The following installation checklist summarizes the elements that must be completed during the installation phase. Mark each item as it is completed. Once these prerequisites have been met, please contact Genetic Analysis AS to schedule the training.
 Instruments and consumable items for Step 1 Genomic DNA extraction (see tables on page 4-6) must be acquired
 Instruments and consumable items for Step 2 Amplification of the bacterial 16S rRNA gene ­ Step 6 Hybridization and Signal Detection (see tables on page 6-9) must be acquired
 All instruments, including the Luminex detection system, must be installed and ready for use  Operators must have been trained in use and maintenance of the Luminex detection system
RECOMMENDED SAFETY EQUIPMENT
Fecal samples should be treated as potentially infectious material until inactivation and require the use of BSL-2 grade laboratory equipment and precautions. This involves the use of appropriate PPE, biological safety cabinet, proper waste disposal and risk-minimizing routines for sample handling. Make sure to follow laboratory and local guidelines for EHS.
COMPUTER REQUIREMENTS
The GA-map® Analyzer is a cloud-based software compatible with most web browsers. See the GA-map® Analyzer Manual for details.

S8-044 A1 v9 Appendix 6

Page 2 of 9

LABORATORY REQUIREMENTS

The GA-map® Dysbiosis Test Lx v2 requires a typical molecular laboratory set-up. It is recommended that the laboratory contains, at a minimum, dedicated pre-PCR, template-free, and PCR/post-PCR zones. The zones should be separated to prevent contamination.
The table below contains an overview of the different laboratory procedures that are performed in the individual zones, in addition to instruments and equipment needed in the different workstations. Please refer to the Equipment, materials, and reagents required section for equipment specifications. To equip the lab with appropriate pipettes, please refer to the IFU for volumes to be pipetted.

Zone

Lab procedures

Workstation

Pre-PCR
Templatefree PCR/Post-PCR

Sample preparation and gDNA extraction. Addition of template for the amplification step.
Mastermix preparation for the amplification, clean-up, and End-Labeling steps.
Amplification reaction. The quantification step. Addition of template for the End-labeling step and the End-labeling reaction. The Hybridization and signal detection step.

Biological safety cabinet equipped with:
· Vortex mixer · Decapper for Lysing Matrix-E tubes · Appropriate single- and 8-channel pipettes w/ wide
orifice tips
Open bench equipped with:
· Bead beater · Plate centrifuge for Lysing Matrix-E tubes · Water bath · DNA extraction robot · Vortex mixer · Microcentrifuge · Dispenser pipette w/ tips · Single- and 8-channel pipettes with tips · Ice or cooling elements for other reagents and plate
PCR cabinet with available equipment:
· Vortex mixer · Microcentrifuge · Dispenser pipette w/ tips · Single channel pipettes w/ tips · Freezing block for enzymes · Ice or cooling elements for reagents and plate
Open bench equipped with:
· Plate centrifuge for quick spin · Thermal cycler · DNA quantification system · Ring magnet plate · Luminex detection platform · Vortex mixer · Microcentrifuge · Dispenser pipette w/ tips · Single- and 8-channel pipettes w/ tips · Ice or cooling elements for reagents and plate · Freezing block for plates

S8-044 A1 v9 Appendix 6

Page 3 of 9

EQUIPMENT, MATERIALS, AND REAGENTS REQUIRED
The lists below describe the required equipment, materials, and reagents for the GA-map® Dysbiosis Test Lx v2.
All items are required, unless otherwise specified. Genetic Analysis must be informed of any deviations from these requirements and a separate validation might be necessary.
The quantity for all equipment is based on the set-up described in the Laboratory Requirements section above. It should be noted that different laboratory set-ups (e.g. with additional zones) might affect the quantity.

FOR STEP 1 ­ GENOMIC DNA EXTRACTION

List of equipment when extracting DNA according to STEP 1 of the IFU ­ need of equipment may vary with choice of extraction kit and instruments.

Equipment

Specifications

Recommended options

type

 Decapper for Capacity: 8-12-channel

8-Channel Screw Cap Decapper

Lysing Matrix-E capper/decapper for matrix (4105MAT), Thermo Scientific

tubes

tubes

AlteCapTM Switch (404000) with Cassette 12/96 - Internal thread Matrix screw caps (404014), AltemisLab

 Bead beater

Capacity: 96-well rack Force: 1600-1800 RPM

FastPrep-96TM Homogenizer w/96well plate insert (116010500), MP Biomedicals

 Plate
centrifuge

Capacity: 96 well plate with Any min 6 cm height

Force: 1300 rcf

 Water bath

Capacity: 96 well plate

Any

Temperature: 65°C

 DNA extraction 96 well plate format

KingFisher Flex ­ 96 deep well head

robot

Magnetic particle processor (540 0630) w/magnetic micro-plate

Separator, installed with KF Flex 96

KF heating block (2407 5420),

Thermo Scientific

Bioer GenePure (NAP-32P)

Contact Genetic Analysis for approved instruments. Or request our standardized extraction validation procedure.

 Vortex mixer Speed: ~2800 rpm

Any

 Micro
Centrifuge

Capacity: 1.5ml/ 2ml tubes Any (spin down only)

S8-044 A1 v9 Appendix 6

Page 4 of 9

 Dispenser

Volumes to be dispensed: 20, Any

pipette w/tips 200, 250, 270 and 720 µl

Multipette®E3/E3x

(4987000371/4987000380),

Eppendorf

 Pipettes single Volumes: 10 ­ 1200 µl
channel w/tips

Wide orifice tips recommended for pipetting of fecal samples prior to gDNA extraction

 Pipettes 8-

Volumes: 100 ­ 1200 µl

channel w/tips

Wide orifice tips recommended for pipetting of fecal samples prior to gDNA extraction

 Ice or cooling For keeping reagents/sample Any

blocks

intermediates cold during

handling

List of materials

Material

Specifications

 Lysing Matrix-E Capacity: 2 ml

tubes

Bead type: 1.4 mm Ceramic

Spheres, 0.1 mm Silica

Spheres, and 4 mm Glass

Bead.

 Tube for Lysis Volume: 5, 15 or 50 ml

mix

 Deep well plate DeepWell 96 Plate, V-

for DNA

bottom

extraction robot

Recommended options
Lysing Matrix E, 96-tube rack, barcoded tubes, 1 Rack (116984001B), (note: individual Matrix tubes with screw cap), MP Biomedicals
Any
KingFisher Deepwell 96 Plate, Vbottom, (95040450), Thermo Scientific 96 Well Plate SW 2ml with V-Bottom PP (BAKR43001-0504), VWR



Deep well Tip

96 tip comb for DeepWell

KingFisher 96 tip comb for DW

Combs for DNA magnets

magnets, (97002534), Thermo

extraction robot

Scientific

96 Tip Comb PP (BAKR43001-0505),

VWR



Elution plate for 96 well microplate, 0.2 ml KingFisher 96 KF microplate (200 l),

DNA extraction

(97002540), Thermo Scientific

robot

Well Plate 96 SW 0.2ml with V-

Bottom PP (BAKR43001-0506), VWR



Adhesive PCR

Suitable for deep well plate Adhesive PCR Plate Seals (AB0558),

plate seal

for DNA extraction robot

Thermo Scientific

 Microtiter
sealing tape

Suitable for plates for DNA extraction robot

Adhesive Plate Seals (AB0580), Thermo Scientific



Microtiter plate Capacity: 96-well, 250 µl

Any

w/seal for

gDNA dilution

S8-044 A1 v9 Appendix 6

Page 5 of 9

List of reagents

Reagent

Specifications

 Extraction

Fecal sample of known

control positive quality

(optional)

 DNA extraction Validated for GA-map®

reagent kit

Dysbiosis Test Lx v2

 Ethanol (for

96-100% ethanol

extraction kit) 500 ml

 Acetone (for

99% acetone

extraction kit) 350 ml

 Water for
dilution of gDNA and as PCR ctrl neg.

Sterile, Nuclease-free water 500 ml

Recommended options Any
magTM maxi DNA purification kit, 288 tests (NAP40430), LGC Genomics Any
Any
Any

FOR STEP 2 ­ AMPLIFICATION OF THE BACTERIAL 16S RRNA GENE TO STEP 6 ­ HYBRIDIZATION AND SIGNAL DETECTION

List of equipment

Equipment type Specifications

Recommended options

Used in assay steps

 Plate centrifuge Capacity: 96-well plate

Any

(quick spin down only)

Step 2 Step 3 Step 4 Step 5 Step 6

 Thermal cycler Capacity: 96-well

VeritiTM 96-Well Thermal Cycler

Step 2

Lower maximum volume

(4375786), Applied Biosystems

Step 4

 75µl

VeritiProTM 96-Well Thermal Cycler

Temperature range: 4-105°C (A48141), Applied Biosystems

Step 5 Step 6

T100TM Thermal Cycler (186-1096), Bio-Rad

 DNA

System for quantifying

quantification dsDNA (0-100ng/µl)

system

Qubit® Fluorometer 3.0 (Q33216) or 4.0 (Q33238) or Flex (Q33327),
Invitrogen

Step 3

 Ring Magnet
Plate
 Luminex
detection platform

Capacity: 96-well Hybridization plate, 100µl
Luminex xMAP® technology Capacity: 96-well Multiplexing capacity:  50

FLUOstar OMEGA Microplate Reader with filter-based absorbance (415103) w/ appropriate filters for the quantification assay, BMG LabTech
96-Well Ring Magnet Plate (S380), Permagen
MAGPIX® system or NxTAG®Enabled MAGPIX® system with xPONENT® version 4.2 or higher (MAGPIXXPON4.1-CEIVD), Luminex*

Step 6 Step 6

S8-044 A1 v9 Appendix 6

Page 6 of 9

Laser type: Green and red

Luminex® 200TM system with xPONENT® version 4.2 or higher (LX200-XPON-IVD/RUO), Luminex*

 Vortex mixer

Speed: ~2800 rpm

Any

Step 2 Step 3

Step 4

Step 5

Step 6

 Micro

Capacity: 2ml tubes

Any

Centrifuge

(spin down only)

Step 2 Step 3

Step 4

Step 5

Step 6

 Dispenser

Volumes to be dispensed: Multipette®E3/E3x

pipette w/tips 20µl and 40µl

(4987000371/4987000380),

Eppendorf

Step 2 Step 5 Step 6

 Pipettes single Volumes: 0.5 ­ 1000µl

Any

channel w/tips

Step 2 Step 3

Step 4

Step 5

Step 6

 Pipettes 8-

Volumes: 0.5 ­ 300µl

Any

channel w/tips

Step 2 Step 3

Step 4

Step 5

Step 6



Ice or cooling

For keeping

Any

blocks

reagents/sample

intermediates cold during

handling

Step 2 Step 3 Step 4 Step 5

Step 6

 Freezing block Capacity: 1.5/ 2ml reagent Benchtop cooler (5115-0012),

for tubes

tubes

Thermo Scientific

For use during handling of

enzymes

Step 2 Step 4 Step 5

 Freezing block Capacity: 96-well microtiter PCR-Cooler 0.2ml (3881000031),

for plates

plate

Eppendorf

Step 5

For use during End-labeling

*Additional items for use and maintenance are required: kits for calibration and performance verification,

sheath fluid or drive fluid, and an ultrasonic bath (40-60 kHz)

S8-044 A1 v9 Appendix 6

Page 7 of 9

List of materials

Material

Specifications

 Microcentrifuge Volume: 1.5, 2 and 5ml
tubes for
mixing of reagents
 Microtiter plate Capacity: 96-well
for PCR/End- PCR grade labeling

Recommended options Safe-Lock tubes, Eppendorf
96-Well Semi-skirted Flat Deck PCR Plates (AB1400), Thermo Scientific

Used in assay steps
Step 2 Step 4 Step 5 Step 6
Step 2 Step 5

96-Well Semi-skirted, segmented

PCR Plate (AB0900), Thermo

Scientific



8-cap sealing

Suitable for microtiter plate Flat PCR Caps, strips of 8 (AB0784), Step 2

strips

Thermo Scientific

Step 5

 Microtiter

Suitable for microtiter plate Adhesive Plate Seals (AB0580),

Step 2

sealing tape

Thermo Scientific

Step 5



Tubes/plates

Suitable for the selected DNA For use with plate reader: NuncTM 96- Step 3

for DNA

quantification system

Well Microplate, Black (237108),

quantification

Thermo Scientific

For use with Qubit® 3.0/4.0 Fluorometer: QubitTM Assay Tubes (Q32856), Invitrogen

For use with Qubit® Flex

Fluorometer: QubitTM Flex Assay

Tube Strips (Q33252), Invitrogen



Hybridization

Capacity: 96-well, unskirted, 96-well Twin.tecTM PCR Plates,

Step 6

plate

0.2ml

Unskirted, Divisible (0030133358),

Suitable for Luminex

Eppendorf

detection platform

ThermowellTM 96-Well Polycarbonate

PCR Microplates, Model P (6509),

Corning

 Sealing film for Non-adhesive

Microseal® 'A' (MSA5001), Bio-Rad Step 6

Hybridization plate

For 96-well format

 Reagent

25ml

Any

Step 4

reservoir

Step 6

List of reagents

Reagent

Specifications

Recommended options

Used in assay steps



Water for PCR Sterile, Nuclease-free water, Any

control neg.

same as used for dilution of

gDNA

Step 2

 Assay kit for

For plate reader

DNA

quantification

Quant-iTTM PicoGreenTM dsDNA Assay Kit (P11496), ThermoFisher
Quant-iTTM 1X dsDNA HS Assay (Q33232), ThermoFisher

Step 3

S8-044 A1 v9 Appendix 6

Page 8 of 9

For Qubit® Fluorometer

QubitTM dsDNA HS assay kit (Q32854), ThermoFisher
QubitTM 1X dsDNA HS assay kit (Q33231), ThermoFisher
Quant-iTTM 1X dsDNA HS Assay (Q33232), ThermoFisher
TRAINING PLAN

Training will be scheduled following system set-up. Genetic Analysis will train users in performing the process of GA-map® Dysbiosis Test Lx v2 from fecal gDNA extraction (or from PCR, if the extraction method differs from that described in the IFU) to generation of the reports. Other sampling devices, extraction instruments and chemistry can be validated upon request using our standardized validation method. Basic instrument operation and use of the software will be included in the training. Those being trained will be required to have a basic knowledge of Microsoft Windows and use of general laboratory equipment and tools.
Operators must have been trained in use and maintenance of the Luminex system prior to the tech transfer.
CONTACT INFO

We are happy to help you with your inquiries. Technical Support: support@genetic-analysis.com

Genetic Analysis AS | Ulvenveien 80 | 0581 OSLO | Norway www.genetic-analysis.com | Phone: +47 48 32 16 10 | E-mail: info@genetic-analysis.com
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be considered unprotected by law. All trademarks and registered trademarks mentioned herein are the property of their respective owners. © 2021-2023 Genetic Analysis, all rights reserved.

S8-044 A1 v9 Appendix 6

Page 9 of 9

