Owner's Manual for MEAN WELL models including: RPS-45-3.3, RPS-45-5, RPS-45-7.5, RPS-45-12, RPS-45-15, RPS-45-24, RPS-45-48, RPS-45 series 45W Reliable Green Medical Power Supply, RPS-45 series, 45W Reliable Green Medical Power Supply, Green Medical Power Supply, Power Supply, Supply

RPS-45-20240223

wangj

MEAN-WELL USA, INC. - РКС Компоненти - РАДІОМАГ


File Info : application/pdf, 4 Pages, 149.28KB

PDF preview unavailable. Download the PDF instead.

rps-45-spec
45W Reliable Green Medical Power Supply

RPS-45 series
User's Manual

EN 60601-1 IEC 60601-1
8
ANSI/AAMI ES60601-1 BS EN/EN60601-1 IEC60601-1 TPTC004
 Features
3"×2" compact size Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1 Suitable for BF application with appropriate system consideration Cooling by free air convection EMI class B for class  configuration No load power consumption<0.1W Extremely low leakage current Protections: Short circuit / Overload / Over voltage Lifetime > 50K hours Operating altitude up to 4000 meters 3 years warranty

++
 Applications
Oral irrigator Hemodialysis machine Medical computer monitors Sleep apnea devices
 GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description
RPS-45 is a 45W highly reliable green PCB type medical power supply with a high power density on the 3" by 2" footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-45 is able to be used for Class(no FG) system design. The extremely low leakage current is
less than 100A. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011,
perfectly fitting all kinds of BF rated "patient contact" medical system equipment.

Model Encoding RPS- 45 - 3.3
Output voltage Rated wattage Series name
Downloaded from Arrow.com.

File Name:RPS-45-SPEC 2024-02-23

45W Reliable Green Medical Power Supply

RPS-45 series

SPECIFICATION

ORDER NO.

RPS-45-3.3

RPS-45-5

RPS-45-7.5 RPS-45-12 RPS-45-15 RPS-45-24 RPS-45-48

DC VOLTAGE

3.3V

5V

7.5V

12V

15V

24V

48V

RATED CURRENT CURRENT RANGE

8A
0 ~ 8.8A

8A
0 ~ 8.8A

5.4A 0 ~ 5.95A

3.8A 0 ~ 4.18A

3A
0 ~ 3.3A

1.9A 0 ~ 2.1A

0.94A 0 ~ 1.03A

OUTPUT

RATED POWER

26.4W

PEAK LOAD(10sec.) Note.2 29W

RIPPLE & NOISE (max.) Note.3 60mVp-p

VOLTAGE ADJ.RANGE

3.1~3.6V

VOLTAGE TOLERANCE Note.4 ±2.0%

LINE REGULATION

±0.5%

LOAD REGULATION

±2.0%

40W
44W 60mVp-p
4.7~5.5V ±2.0% ±0.5% ±2.0%

40.5W
44.6W 80mVp-p
7.12~8.3V ±2.0% ±0.5% ±2.0%

45.6W
50.2W 100mVp-p
11.4~13.2V ±2.0% ±0.5% ±2.0%

45W
49.5W 100mVp-p
13.5~16.5V ±1.0% ±0.5% ±1.0%

45.6W
50.2W 120mVp-p
22.8~27.6V ±1.0% ±0.5% ±1.0%

45.1W
49.4W 120mVp-p
45.6~52.8V ±1.0% ±0.5% ±1.0%

SETUP, RISE TIME

500ms, 30ms / 230VAC 500ms, 30ms / 115VAC at full load

HOLD UP TIME (Typ.)

30ms / 230VAC 16ms / 115VAC at full load

VOLTAGE RANGE Note.5 80 ~ 264VAC

FREQUENCY RANGE

47 ~ 63Hz

INPUT EFFICIENCY (Typ.)

80.5%

83%

85%

88%

89%

90%

91%

AC CURRENT (Typ.)

1.2A / 115VAC 1A / 230VAC

INRUSH CURRENT (Typ.) OLD STAR 30A/115VAC 60A/230VAC

LEAKAGE CURRENT(max.) Note.6 Touch current< 100A/264VAC

OVERLOAD

115 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed

PROTECTION OVER VOLTAGE

3.8~5V

5.7~6.8V

8.6~11.3V

13.8~16.2V

Protection type : Shut down o/p voltage, re-power on to recover

17.2~20.3V 28.4~32.4V

55.2~64.8V

WORKING TEMP.

-30 ~ +70 (Refer to "Derating Curve")

WORKING HUMIDITY

20% ~ 90% RH non-condensing

ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing

TEMP. COEFFICIENT

±0.03% /  (0 ~ 50)

VIBRATION

10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

OPERATING ALTITUDE Note.7 4000 meters

SAFETY STANDARDS

IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;Design refer to BS EN/EN60335-1

ISOLATION LEVEL

Primary-Secondary: 2xMOPP

WITHSTAND VOLTAGE

I/P-O/P: 4KVAC

ISOLATION RESISTANCE I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH

Parameter

Standard

Test Level / Note

Conducted emission

BS EN/EN55011 (CISPR11)

Class B

SAFETY & EMC EMISSION

Radiated emission

BS EN/EN55011 (CISPR11)

Class B

EMC (Note. 8)

Harmonic current Voltage flicker

BS EN/EN61000-3-2 BS EN/EN61000-3-3

Class A -----

BS EN/EN55035, BS EN/EN60601-1-2

Parameter

Standard

Test Level / Note

ESD

BS EN/EN61000-4-2

Level 4, 15KV air ; Level 4, 8KV contact

RF field susceptibility

BS EN/EN61000-4-3

Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )

EMC IMMUNITY

EFT bursts Surge susceptibility

BS EN/EN61000-4-4 BS EN/EN61000-4-5

Level 3, 2KV Level 4, 2KV/Line-Line

Conducted susceptibility

BS EN/EN61000-4-6

Level 3, 10V

Magnetic field immunity

BS EN/EN61000-4-8

Level 4, 30A/m

Voltage dip, interruption

BS EN/EN61000-4-11

100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods

MTBF

3334.3K hrs min. Telcordia SR-332 (Bellcore) ; 726.2K hrs min. MIL-HDBK-217F (25)

OTHERS DIMENSION (L*W*H)

76.2*50.8*24mm or 3" * 2" *0.945" inch

PACKING

0.11Kg; 120pcs/14.2Kg/0.94CUFT

1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.

2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.

3. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1f & 47f parallel capacitor.

4. Tolerance : includes set up tolerance, line regulation and load regulation.

5. Derating may be needed under low input voltages. Please check the derating curve for more details.

NOTE

6. Touch current was measured from primary input to DC output. 7. The ambient temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan models for operating altitude higher than 2000m(6500ft).

8. The power supply is considered a component which will be installed into a final equipment. "All the EMC tests are been executed by mounting

the unit on a 360mm*360mm metal plate with 1mm of thickness." The final equipment must be re-confirmed that it still meets EMC directives.

For guidance on how to perform these EMC tests, please refer to "EMI testing of component power supplies."

(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

 Product Liability DisclaimerFor detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Downloaded from Arrow.com.

File Name:RPS-45-SPEC 2024-02-23

45W Reliable Green Medical Power Supply

RPS-45 series

Block Diagram

EMI FILTER

RECTIFIERS

POWER

I/P

&

&

SWITCHING

RECTIFIER

FILTER

O.L.P.

CONTROL

Derating Curve

100
80
60 50 40
20

-30

0

10

20

30

40

50

60

70 (HORIZONTAL)

AMBIENT TEMPERATURE ()

Static Characteristics
100 90 80 70 60 50 40
80 100 115 120 140 160 180 200 220 240 264
INPUT VOLTAGE (V) 60Hz

fosc : 65KHz

RECTIFIERS

+V

& FILTER

-V

DETECTION CIRCUIT
O.V.P.

LOAD (%)

LOAD (%)

Downloaded from Arrow.com.

File Name:RPS-45-SPEC 2024-02-23

45W Reliable Green Medical Power Supply

RPS-45 series

Mechanical Specification
3.175

Top View 76.2
69.85

3.175

50.8 44.45

FS1
3 2 1
CN1

AC FUSE T2A/250V

HS1

HS100

CN2 1 2
3
4 LED

SVR1

4-3.2

Case No. Unit:mm

3 max. 24

Side View

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing

Terminal

1

AC/N

2

No Pin

JST VHR or equivalent

JST SVH-21T-P1.1 or equivalent

3

AC/L

DC Output Connector (CN2) : JST B4P-VH or equivalent

Pin No. Assignment Mating Housing

Terminal

1

+V

2

+V

3

-V

JST VHR JST SVH-21T-P1.1

or equivalent

or equivalent

4

-V

Installation Manual
Please refer to : http://www.meanwell.com/manual.html

Downloaded from Arrow.com.

File Name:RPS-45-SPEC 2024-02-23



References

Acrobat Distiller 9.0.0 (Windows); modified using iText 5.5.6 ©2000-2015 iText Group NV (AGPL-version) (AGPL-version)