F3000 EN - 7 4.2. DESCRIPTION OF FUNCTIONS KEYS 4.2.1. /HOLD key The multifunction /HOLD key allows the herefollowed operations: A long pressure (2s) allows to switch on/off the instrument.


File Info : application/pdf, 13 Pages, 302.40KB

PDF preview unavailable. Download the PDF instead.

elektro-lijn-f3000-hl-en
ENGLISH User manual

 Copyright HT ITALIA 2015

Release EN 1.02 - 23/06/2015

F3000

Table of contents: 1. PRECAUTIONS AND SAFETY MEASURES ...............................................................2
1.1. Preliminary instructions ..................................................................................................... 2 1.2. During use ......................................................................................................................... 3 1.3. After use ............................................................................................................................ 3 1.4. Definition of measurement (overvoltage) category............................................................ 3
2. GENERAL DESCRIPTION ...........................................................................................4 3. PREPARATION FOR USE ...........................................................................................4
3.1. Initial checks...................................................................................................................... 4 3.2. Instrument power supply ................................................................................................... 4 3.3. Calibration ......................................................................................................................... 4 3.4. Storage.............................................................................................................................. 4
4. OPERATING INSTRUCTIONS.....................................................................................5
4.1. Instrument description ....................................................................................................... 5 4.1.1. Description of the controls .......................................................................................................... 5 4.1.2. Opening/closing flexible clamp jaw ............................................................................................ 6
4.2. Description of functions keys............................................................................................. 7

4.2.1.

/HOLD key.............................................................................................................................. 7

4.2.2. Backlight key ........................................................................................................................ 7 4.2.3. Disabling the Auto Power OFF function ..................................................................................... 7 4.3. Operating instruction ......................................................................................................... 8 4.3.1. AC Current measurement........................................................................................................... 8
5. MAINTENANCE ...........................................................................................................9
5.1. General information........................................................................................................... 9 5.2. Batteries replacement ....................................................................................................... 9 5.3. Cleaning the instrument .................................................................................................... 9 5.4. End of life .......................................................................................................................... 9
6. TECHNICAL SPECIFICATIONS ................................................................................10
6.1. Technical characteristics ................................................................................................. 10 6.1.1. Electrical specifications ............................................................................................................ 10 6.1.2. Reference guideline.................................................................................................................. 10 6.1.3. General characteristics ............................................................................................................. 10
6.2. ENVIRONMENT.............................................................................................................. 11 6.2.1. Environmental conditions for use ............................................................................................. 11
6.3. Accessories ..................................................................................................................... 11 6.3.1. Standard accessories ............................................................................................................... 11
7. SERVICE....................................................................................................................12
7.1. Warranty conditions......................................................................................................... 12 7.2. Service ............................................................................................................................ 12

EN - 1

F3000
1. PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1 relevant to electronic measuring instruments. For your safety and in order to prevent damaging the instrument, please carefully follow the procedures described in this manual and read all notes preceded by the symbol with the utmost attention.
CAUTION
Failure to comply with the caution notes and/or instructions may damage the instrument and/or its components and be a source of danger for the operator.
Before and after carrying out the measurements, carefully observe the following instructions:
 Do not carry out any current measurement in humid environments.  Do not carry out any measurements in case gas, explosive materials or flammables are
present, or in dusty environments.  Avoid contact with the circuit being measured if no measurements are being carried out.  Avoid contact with exposed metal parts, with unused measuring probes, circuits, etc.  Do not carry out any measurement in case you find anomalies in the instrument such
as deformation, breaks, substance leaks, absence of display on the screen, etc.
In this manual, and on the instrument, the following symbols are used:
CAUTION: observe the instructions given in this manual; an improper use could damage the instrument or its components.
Double insulated meter.
AC current
DC voltage
Ground reference
Use appropriate security measures to apply/remove the instrument on/from live conductors
1.1. PRELIMINARY INSTRUCTIONS  This clamp has been designed for use in environments of pollution degree 2.  It can be used for AC CURRENT measurements, on installations with measurement
category CAT IV 600V or CAT III 1000V refer to ground. For a definition of the abovementioned categories, see § 1.4  This instrument is not suitable for direct current measurements  Please take the standard safety precautions aimed at protect you against dangerous electric currents and protect the instrument against incorrect use  Do not test circuits exceeding the specified current and voltage limits  Check that the battery is correctly inserted
EN - 2

F3000
1.2. DURING USE Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the Caution notes and/or Instructions may damage the instrument and/or its components or be a source of danger for the operator.
 Before switching on the clamp, remove the conductor from the clamp jaw.  During current measurement, any other current near the clamp may affect
measurement precision.  When measuring current, always put the conductor as near as possible to the middle of
the clamp jaw, as described in § 4.1.2 to obtain the most accurate reading.  While measuring, if the value or the sign of the quantity being measured remain
unchanged, check if the HOLD function is enabled.
1.3. AFTER USE
 When measurement is complete, switch off the clamp.  If the instrument is not to be used for a long time, remove the battery.
1.4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY Standard IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements" defines what measurement category, commonly called overvoltage category, is. In § 6.7.4: Measured circuits, circuits are divided into the following measurement categories:
(OMISSIS)
 Measurement category IV is for measurements performed at the source of the lowvoltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
 Measurement category III is for measurements performed on installations inside buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.
 Measurement category II is for measurements performed on circuits directly connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
 Measurement category I is for measurements performed on circuits not directly connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user.
EN - 3

F3000
2. GENERAL DESCRIPTION
The instrument carries out the following measurements:
 AC TRMS current measurement up to 3000A with autorange  Use of high diameter flexible clamp  Data HOLD feature  Backlight feature
3. PREPARATION FOR USE
3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an electric as well as mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged. However, we recommend generally checking the instrument in order to detect possible damage suffered during transport. In case anomalies are found, immediately contact the forwarding agent. We also recommend checking that the packaging contains all components indicated in § 6.3.1. In case of discrepancy, please contact the Dealer. In case the instrument should be replaced, please carefully follow the instructions given in § 7.
3.2. INSTRUMENT POWER SUPPLY The instrument is supplied with 2x1.5V AAA LR03 batteries, included in the package. In order to prevent compromising its charge, the battery is not inserted in the instrument. For batteries installation, follow the instructions given in § 5.2. The " " symbol appears when the battery is nearly discharged. Replace the battery, following the instructions given in § 5.2.
3.3. CALIBRATION The instrument has the technical specifications described in this manual. The instrument's performance is guaranteed for 12 months.
3.4. STORAGE In order to guarantee precise measurement, after a long storage time under extreme environmental conditions, wait for the instrument to come back to normal condition (see § 6.2.1).
EN - 4

4. OPERATING INSTRUCTIONS
4.1. INSTRUMENT DESCRIPTION 4.1.1. Description of the controls

F3000
CAPTION: 1. Flexible clamp jaw 2. Opening/closing jaw connector 3. LCD display 4. /HOLD key 5. Backlight key 6. Battery cover

Fig. 1: Instrument description EN - 5

F3000 4.1.2. Opening/closing flexible clamp jaw
CAUTION
 Put the conductor as close as possible to the middle of the jaws in order to meet the meter accuracy specifications (see Fig. 2)
 When apply/remove the instrument on/from live conductors, if is not possible to de-energize the circuit on test, use appropriate security measures (e.g. protective gloves) to avoid dangerous electric shock for the operator and a possible damage of the instrument
 Disconnect the main voltage or wear protective gloves before performing measurement
 Put the hand below the LCD display during the measurement as indicated by the arrow in Fig. 2
For opening/closing the flexible clamp jaw following the below steps: 1. Rotate of a quarter in the counterclockwise direction the fixing jaw connector to unlock
the jaw (see Fig. 2) 2. Extract the flexible jaw and wrap around the conductor on test 3. Replace the jaw in the seat and rotate of a quarter in the clockwise direction the fixed
jaw connector up to align the two arrows on the plastic case to lock the jaw (see Fig. 2)
Fig. 2: Opening/closing flexible clamp jaw EN - 6

F3000 4.2. DESCRIPTION OF FUNCTIONS KEYS 4.2.1. /HOLD key The multifunction /HOLD key allows the herefollowed operations:  A long pressure (2s) allows to switch on/off the instrument. After switch on, a quick
autotest relative to the battery level charging is performed and the "FULL" message is displayed with completely charged batteries. With symbol " " replace the batteries (see § 5.2)  With instrument switched on, the key allows to activate/deactivate the HOLD feature in way to fix the result of the measured value at display. The "HOLD" symbol is shown 4.2.2. Backlight key Press the key for the backlight activation/deactivation. This light could help the operator reading the display while he's measuring in dark sites 4.2.3. Disabling the Auto Power OFF function In order to preserve internal batteries, the instrument switches automatically off approximately 20 minutes after it was last used. No symbol appears on the display when this function is active. When the instrument must be used for long periods of time, it may be useful to deactivate the Auto Power Off function as follows:  Switch off the instrument with the /HOLD key  Switch on the instrument by press and hold the /HOLD key for at least 2s  With "FULL" message at display press the key. The "AoFF" message is shown at display to indicate the disabling of the function  Switch off and on the instrument to automatically enable the feature
EN - 7

F3000
4.3. OPERATING INSTRUCTION
4.3.1. AC Current measurement
CAUTION
 Put the conductor as close as possible to the middle of the jaws in order to meet the meter accuracy specifications (see Fig. 2)
 When apply/remove the instrument on/from live conductors, if is not possible to de-energize the circuit on test, use appropriate security measures (e.g. protective gloves) to avoid dangerous electric shock for the operator and a possible damage of the instrument
 Disconnect the main voltage or wear protective gloves before performing measurement
 Put the hand below the LCD display during the measurement as indicated by the arrow in Fig. 2

Correct

Incorrect

Fig. 3: Use of the clamp for AC current measurement 1. Switch on the clamp by pressing the /HOLD key 2. Open the flexible clamp jaw and insert the cable in the middle of it (see Fig. 3). The
current value will be shown on the display 3. If "OL" message is shown at display the current under test is over the maximum value
that the instrument is able to measure 4. For HOLD feature refer to § 4.2
EN - 8

F3000
5. MAINTENANCE
5.1. GENERAL INFORMATION 1. The instrument you purchased is a precision instrument. While using and storing the
instrument, carefully observe the recommendations listed in this manual in order to prevent possible damage or danger during use. 2. Do not use the instrument in environments with high humidity levels or high temperatures. Do not expose to direct sunlight. 3. Always switch off the instrument after use. In case the instrument is not to be used for a long time, remove the battery to avoid liquid leaks that could damage the instrument's internal circuits. 5.2. BATTERIES REPLACEMENT When the LCD displays the " " symbol, replace the batteries
CAUTION
Only expert technicians should perform this operation. Before carrying out this operation, make sure you have removed the cable being tested from inside the clamp jaw.
Fig. 4: Batteries replacement 1. Switch off the clamp and remove the cable being tested from the clamp jaw. 2. Rotate the fixing screw of the battery cover from the " " to the " " position, remove the
cover and extract the batteries (see Fig. 4) 3. Fit two new batteries (see § 6.1.3) and pay attention to correct polarity (see Fig. 4) 4. Rotate the fixing screw of the battery cover from the " " to the " " position 5. Do not scatter old batteries into the environment. Use the relevant containers for
disposal 5.3. CLEANING THE INSTRUMENT Use a soft and dry cloth to clean the instrument. Never use wet cloths, solvents, water, etc. 5.4. END OF LIFE
CAUTION: the symbol on the instrument indicates that the appliance and its accessories must be collected separately and correctly disposed of.
EN - 9

F3000

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL CHARACTERISTICS Accuracy indicated as [%reading + (digit number*resolution)] at 23°C±5°C, < 80%RH

AC TRMS current (Autorange)
Range Resolution Accuracy (*), (**)

Bandwidth

Max overload reading

30.00A

0.01A

33.00A

300.0A

0.1A

(3.0%rdg + 5dgt)

45Hz  500Hz

330.0A

3000A

1A

3300A

NOTE: the display reading is zeroed for values < 10dgt (*) Accuracy relative to sinusoidal waveforms. For non-sinusoidal signals consider as follows:

Additional error due to the Crest Factor (CF) of non-sinusoidal signals: C.F: 1.0  2.0  Add 3.0%rdg C.F: 2.0  2.5  Add 5.0%rdg C.F: 2.5  3.0  Add 7.0%rdg

(**) Accuracy relative to conductor in the center of jaw, no magnetic/electric fields and reference temperature. The errors due to the conductor not in the center of the jaw are the followed:

 Error 2.0%rdg (distance 15mm from center of jaws)  Error 2.5%rdg (distance 25mm from center of jaws)  Error 3.0%rdg (distance 35mm from center of jaws)

6.1.1. Electrical specifications Conversion type: Sample rate: Temperature coefficient:

TRMS 4 times / second 0.2x(accuracy) /°C, <18°C o >28°C

6.1.2. Reference guideline Safety: EMC : Insulation: Pollution level: Max height of use: Measurement category:

IEC/EN61010-1, IEC/EN61010-2-032 IEC/EN61326-1 double insulation 2 2000m (6562 ft) CAT IV 600V, CAT III 1000V to ground

6.1.3. General characteristics Mechanical characteristics Dimensions (L x W x H): Weight (battery not included): Flexible jaw length: Max. cable diameter:

280 x 120 x 25mm (11 x 5 x 1in) 170g (6 ounces) 254mm (10in) 110mm (4in)

Power supply Battery type: Low battery indication: Battery duration: Auto Power OFF:

2x1.5V batteries type AAA LR03 symbol " " at display approx. 200 hours after 20 min of idleness (disabled)

Display Characteristics:

4 LCD, 3000 counts, decimal point and backlight
EN - 10

F3000

6.2. ENVIRONMENT

6.2.1. Environmental conditions for use

Reference temperature:

23° ± 5°C (73° ± 41°F)

Operating temperature:

0°C ÷ 30°C (32°F ÷ 86°F) (RH  80%)

30°C ÷ 40°C (86°F ÷ 104°F) (RH  75%)

40°C ÷ 50°C (104°F ÷ 122°F) (RH  45%)

Allowable relative humidity:

<80%RH

Storage temperature:

-20 ÷ 60 °C (-4°F ÷ 140°F)

Storage humidity:

<80%RH

This instrument satisfies the requirements of Low Voltage Directive 2006/95/EEC (LVD) and of EMC Directive 2004/108/EEC

6.3. ACCESSORIES
6.3.1. Standard accessories  User manual  Carrying bag  Batteries (not fitted)

EN - 11

F3000
7. SERVICE
7.1. WARRANTY CONDITIONS This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product.
The manufacturer declines any responsibility for injury to people or damage to property.
The warranty shall not apply in the following cases:
 Repair and/or replacement of accessories and battery (not covered by warranty).  Repairs that may become necessary as a consequence of an incorrect use of the
instrument or due to its use together with non-compatible appliances.  Repairs that may become necessary as a consequence of improper packaging.  Repairs which may become necessary as a consequence of interventions performed
by unauthorized personnel.  Modifications to the instrument performed without the manufacturer's explicit
authorization.  Use not provided for in the instrument's specifications or in the instruction manual.
The content of this manual cannot be reproduced in any form without the manufacturer's authorization.
Our products are patented and our trademarks are registered. The manufacturer reserves the right to make changes in the specifications and prices if this is due to improvements in technology
7.2. SERVICE If the instrument does not operate properly, before contacting the After-sales Service, please check the conditions of batteries and replace them, if necessary. Should the instrument still operate improperly, check that the product is operated according to the instructions given in this manual.
Should the instrument be returned to the After-sales Service or to a Dealer, transport will be at the Customer's charge. However, shipment will be agreed in advance. A report will always be enclosed to a shipment, stating the reasons for the product's return. Only use original packaging for shipment; any damage due to the use of non-original packaging material will be charged to the Customer.