Thank you very much for choosing our product. Please read and keep this manual carefully before use.
Table of Contents
Chapter 1 Getting Started............................................................................................................................ 3 1.1 Company Profile .................................................................................................................................. 3
Chapter 2 Product Introduction................................................................................................................. 4 2.1 Product Features................................................................................................................................... 4 2.2 Naming Rules.........................................................................................................................................5 2.3 Product Specification............................................................................................................................ 5 2.4 Dimensions (unit: mm)....................................................................................................................... 6 2.5 Technical Parameters........................................................................................................................... 7 2.6 Flow configuration..................................................................................................................................8 2.7 Reset Status........................................................................................................................................... 8 2.8 Port Definition.........................................................................................................................................8
Chapter 3 Description of Control Code....................................................................................................9 3.1 Overview................................................................................................................................................. 9 3.2 Installation and Debugging...................................................................................................................9 3.3 Code Instructions...................................................................................................................................9 3.3.1 Control Command Format............................................................................................................ 9 3.3.2 Command Format Instructions...................................................................................................10
Chapter 4 Common Problems & Solutions...........................................................................................16 Chapter 5 Technical Service..................................................................................... 

http://www.runzeliuti.com

2

Chapter 1 Getting Started
1.1 Company Profile
Nanjing Runze Fluid Control Equipment Co., Ltd. was established in 2014, We are a national high-tech enterprise focusing on R&D and production of fluid accessories for numerous analytical instruments. We engineer, manufacture and market differentiated standard products such as syringe pumps, multiport valves, peristaltic pumps, gastight syringes, plastic fittings, etc.. We persevere in providing our customers with best quality and service in the fields of environmental monitoring, biopharmaceuticals, medical equipment, industrial automation and laboratory instruments, etc.
RUNZE attracted experienced talents with strong capability on integrating software and hard-ware at electronic research, mechanical design, mold manufacturing. In past years, we have accumulated rich technical and practical experience that bring us honors of ISO9001, National High-tech Enterprise, Jiangsu Province Private Science and Technology Enterprise, 5A Bank Credit Assessment, 38 technical patents and multiple software copyright including 8 invention patents, 17 utility model patents, 13 design patents.
RUNZE Technology Center has a reliable EMC laboratory and a series of advanced processing equipment and testing instruments, such as imported white light interferometer, Keens plane rapid detector, five-axis machining center, high-precision nano-grinding machine, etc., Continuous and substantial investment in research and development, which enabled the company to obtain a huge advantage in the manufacture and sales of analytical instrument accessories.

http://www.runzeliuti.com

3

Chapter 2 Product Introduction
2.1 Product Features
(1) Name: The injector valve is an electric injector valve independently developed by RUNZE Fluid Control Equipment Co., Ltd., also known as electric rotary valve;
(2) Control: By receiving instructions from the host computer to control the operation of the stepper motor to make the rotor turn to the specified port position to achieve the switching of the flow path. Serial communication protocol supports RS232/RS485/CAN bus.
(3) Corrosion Resistance: This type of injector valve adopts sapphire valve core, which can be applied to various corrosive liquids. The valve head are made of PCTFE (polychlorotrifluoroethylene) and PPS, maintenance-free;
(4) Valve Core Structure: The valve core adopts a multi-directional self-adaptive plane fitting method, which can extend the service life of the product effectively. (This structure has been applied for utility patent, patent number: CN204852471U)
(5) Motive power: The rotation of the valve body adopts the imported NMB planetary gearbox motor as the power plant, which is extremely reliable.
(6) Orientation: The valve port is oriented by optical encoder, which can effectively solve the problem of inaccurate positioning after the gearbox is worn down.
(7) Drive: The two-phase bipolar stepper drive module is a constant current drive chip, which can control the motor current with high efficiency. It has a built-in error detection circuit and error detection (TSD/ISD) signal output function;
(8) Control Unit Interface: XH terminal with 2.54mm pitch (9) Mechanical Interface: 1/4-28UNF female thread (10)Usage: Widely used in the collection and distribution of fluid samples (11)Application: Environmental testing devices, experimental analysis instruments, medical analyzers, chromatographic instruments, etc. Before using the injector valve, please read the "Instruction Manual" carefully and follow the provisions in it.

http://www.runzeliuti.com

4

2.2 Naming Rules

Figure 2-2 Example: Slope type 10-port sapphire injector valve, naming: QHF-SV07B-X-T10-K1.2-S

2.3 Product Specification

SV-07B injector valves can be divided into 6 ports, 8 ports, and 10 ports according to the number of channels, as shown in the figure below:

6 ports

8 ports

10 ports

Figure 2-3-1
The 6/8/10 ports can be installed in place of other injector valves of our company. The size of the mounting plate is as shown in Figure 2-3-2 (unit: mm).

http://www.runzeliuti.com

5

Figure 2-3-2
2.4 Dimensions (unit: mm)

4-M3 Mounting Holes

Figure 2-4-1 6/8/10 channels dimensions

http://www.runzeliuti.com

6

2.5 Technical Parameters

Description

Parameter

Configuration

6 port

8 port

10 port

Orifice

1.2mm

1.2mm

1.2mm

Wetted material

UFE, Sapphire rotor/stator

Dead volume (Port to Port)

28.121L

28.67L

28.129L

Dead volume (Sealed rotor)

5.207L

5.756L

5.215L

Pressure rating Origin detection Liquid temperature

0-0.3Mpa (air) / 0-1.6Mpa (water) Auto reset to initial position when powered on (this function can be
opened or closed) 0-150

Connection

1/4-28UNF

Replaceable parts

Stator replaceable, sealed rotor irreplaceable

Transposition

Random start to any specific port

Driver

Non-optional

Valve switching time

2s/circle

2s/circle

3.3s/circle

Communication Baud rate
Address & Parameter setting

RS232/RS485/CAN RS232/RS485: 9600bps, 19200bps, 38400bps, 57600bps, 115200bps
CAN: 100Kbps, 200Kbps, 500Kbps, 1Mbps Via communication

Power supply

DC24V/3A

Max. power

60W

Operating temperature

-10-50

Operating humidity

80% relative humidity, non-condensing

Dimension (L*W*H)

60*51*150mm

Net weight

0.73kg

Liquid Figure 2-5-1 Dead volume (Port to Port)

Liquid Figure 2-5-2 Dead volume (Sealed rotor)

http://www.runzeliuti.com

7

2.6 Flow configuration
Center port is the public port and valve can switch to any specific port by programmable control via RS232/RS485/CAN communication. Flow configuration is shown as below:

Figure 2-6
2.7 Reset Status
Reset direction: CCW (unchangeable) Reset position: State 1 (as is shown in Figure 2-6)
2.8 Port Definition
GND

Power light

GND

Figure 2-8 Port Diagram

http://www.runzeliuti.com

Port Definition of Driver Board
8

Chapter 3 Description of Control Code

3.1 Overview
The data transmission between the injector valve and the host device (computer, single-chip computer, PLC, etc.) adopts serial communication (eg: RS-232/RS-485/CAN bus), the following describes the communication format: the communication adopts asynchronous serial communication, Commands and data frames use sum check, and the sum check is two bytes (2Byte). Commands and data in communication are in hexadecimal number, parameters are stored in little-endian mode.
Other instructions: Communication interface: RS-232, RS-485, CAN bus; Communication mode: two-way asynchronous, master-slave mode; Baud rate:RS-232/RS-485: 9600bps, 19200bps, 38400bps, 57600bps, 115200bps;
CAN: 100Kbps, 200Kbps, 500Kbps, 1Mbps; Data bits: 8 bits; Parity: none check. Response time <1 second after receiving the command.

3.2 Installation and Debugging
(1) Installation and debugging tools, see Instructions for Debugging Toolsfor details (2) Instructions for use, see SV-07B Quick Use Guidefor details

3.3 Code Instructions

3.3.1 Control Command Format a: Valve parameter setting command (factory command) b: Valve parameter query command (common command) c: Valve action command (common command) Table 3-3-1 Send Command (common command) Interpretation of 0xXX: 0x means hexadecimal, XX is a two-digit hexadecimal number. The value inputted into the software all should be XX.

http://www.runzeliuti.com

9

The message frame of "Send Command" is 8 bytes, and the complete format is as follows:

Start code Address code Function code

Parameters

End code

Sum check

BO

B1

B2

B3

B4

B5

B6

B7

STX

ADDR

FUNC

1st byte STX: Start code (0XCC)

1-8 bit 9-16 bit

ETX

Low byte

High byte

2nd byte ADDR: Address of slave device (0x00~0x7F)

Multicast Address (0x80~0xFE) Broadcast Address (0xFF)

3rd byte FUNC: Function code

4th and 5th byte: Parameters of function code

6th byte ETX: End code (0xDD)

7th and 8th byte: Sum check code from byte 1 to 6

Note: The above command format refers to the common command. In particular, if a password

bit is added to the factory command and the parameter bit has also changed, from the original 2 bytes

to 4 bytes, the command format is as follows:

The message frame of "Factory Command" is 14 bytes, and the complete format is as follows:

Table 3-3-2 Send Command (Factory Command)

Start Address Function

code code

code

Password

Parameters

B0

B1

B2

B3.B4.B5. B6

B7

B8

B9

STX ADDR

FUNC

PWD

1-8 9-16 17-2 bit bit 4 bit

Table 3-3-3 Response Command

B10 25-3 2 bit

End code B11
ETX

Sum check

B12

B13

Low High byte byte

Start code Address code Status code

Parameters

End code

Sum check

B0

B1

B2

B3

B4

B5

B6

B7

STX

ADDR

STATUS

1-8 bit

9-16 bit

ETX

Low byte

High byte

Note: The send command & response command format for the common commands are the

same, and all the response command message frames are 8 bytes.

3.3.2 Command Format Instructions

Definition of Start Code & End Code B0B5(B11)

Item Start code B0 End code B5 (B11)

Code 0XCC 0xDD

Remark

http://www.runzeliuti.com

10

Note: The send command & response command of the common command are the same. The

start and the end code are B0 and B5 respectively. The factory command is B11.

Definition of Address Bit B1

Item

Instruction

Code B1

Address Bit

Address

0xXX

Note: 1. The send command is the same as the response command

Remark

2. The XX in "0xXX" means that it can be set, the factory default is 0x00, and the parameter value

range is 0x000x7F.

Table 3-3-4 Control Command Instructions (B2~B10)

A: Valve parameter setting command (factory command) (B2~B10)

Command Code B2
0x00 0x01 0x02
0x03
0x0E 0x10

Command Instruction
Set address
Set RS232 baud rate
Set RS485 baud rate
Set CAN baud rate
Set automatic reset when power-on
Set CAN destination
address

Code (B3, B4, B5, B6) B3=0xFF B4=0xEE B5=0xBB B6=0xAA
B3=0xFF B4=0xEE B5=0xBB B6=0xAA
B3=0xFF B4=0xEE B5=0xBB B6=0xAA
B3=0xFF B4=0xEE B5=0xBB B6=0xAA
B3=0xFF B4=0xEE B5=0xBB B6=0xAA
B3=0xFF B4=0xEE B5=0xBB B6=0xAA

Parameters Instructions (B7, B8, B9, B10)
B7=0xXX (B8=0x00 B9=0x00 B10=0x00) The value range of XX is 007F in V1.9 & above version, 00FF in version below V1.9,
default value is 00 Totally 5 baud rates: the factory default is
9600bps (B8=0x00 B9=0x00 B10=0x00) B7=0x00 corresponding baud rate 9600bps B7=0x01 corresponding baud rate 19200bps B7=0x02 corresponding baud rate 38400bps B7=0x03 corresponding baud rate 57600bps B7=0x04 corresponding baud rate 115200bps Totally 4 baud rates: the factory default is 100K B8=0x00 B9=0x00 B10=0x00 B7=0x00 corresponding baud rate 100Kbps B7=0x01 corresponding baud rate 200Kbps B7=0x02 corresponding baud rate 500Kbps B7=0x03 corresponding baud rate 1Mbps B7=0x00 means non-automatic reset B7=0x01 means automatic reset Automatically reset to state 1 after power-on (the factory default of injector valve is
automatic reset) B7=0xXX
(B8=0x00 B9=0x00 B10=0x00) The value range of XX is 00FF, and the
default is 00

http://www.runzeliuti.com

11

0x50 0x51 0x52 0x53 0xFC 0xFF

Set address of multicast channel
1

B3=0xFF B4=0xEE B5=0xBB B6=0xAA

Set address of multicast channel
2

B3=0xFF B4=0xEE B5=0xBB B6=0xAA

Set address of multicast channel
3

B3=0xFF B4=0xEE B5=0xBB B6=0xAA

Set address of multicast channel
4

B3=0xFF B4=0xEE B5=0xBB B6=0xAA

Parameter lock
Restore factory settings

B3=0xFF B4=0xEE B5=0xBB B6=0xAA B3=0xFF B4=0xEE B5=0xBB B6=0xAA

B7=0xXX (B8=0x00 B9=0x00 B10=0x00) The value range of XX is 80FE, and the
default is 00 B7=0xXX
(B8=0x00 B9=0x00 B10=0x00) The value range of XX is 80FE, and the
default is 00 B7=0xXX
(B8=0x00 B9=0x00 B10=0x00) The value range of XX is 80FE, and the
default is 00 B7=0xXX
(B8=0x00 B9=0x00 B10=0x00 The value range of XX is 80FE, and the
default is 00
All the parameters are 0x00
All the parameters are 0x00

Example: Use the 0x50/51/52/53 command to set the multicast address (only uses 0x50/51/52

three commands in this example)

Use three sets of SV07B injector valves with the same software version. In the RS485

communication mode, set their addresses to 00, 01, 02 and make a mark. Firstly, Set parameter 0x81 of

SV07B multicast channel 1 address marked as 00 to 81 by the command 0x50, set parameter 0x81 of

multicast channel 2 address to 81 by command 0x51; Set parameter 0x82 of the multicast channel 1

address of SV07B marked as 01 to 82 by command 0x51, and the parameter 0x82 of multicast channel

2 address is set to 82 by command 0x52. Set parameter 0x83 of the multicast channel 1 address of

SV07B marked as 00 to 83 by command 0x50, and the parameter 0x83 of multicast channel 3 address

to 83 by command 0x52 (See the table below for details)

Item

Device

Multicast address

Device1 (address 0)
81(2)

Device2 (address 1)
81(2)

Broadcast address

FF(1)

FF(1)

Multicast address

82(2)

Broadcast address

FF(1)

FF(1)

Multicast address

83(2)

Device3 (address 2)
FF(1) 82(2) FF(1) 83(2)

http://www.runzeliuti.com

12

After the setting is completed, connect the three devices in parallel to the serial debugging tool,

and use our debugging tool MotorTest V0.8 for debugging. In MotorTest V0.8, set the address to 0x81,

the command to 0x44 and the parameter to 0x02, click to send and then the device 1 and device 2 will

have port switching action; Set the address to 0xFF ,the command to 0x44 and the parameter to 0x01,

click to send and then the device1.2 3 will all have port switching action; Set the address to 0x82, the

command to 0x44, and the parameter to 0x02, click send and then the device 2 and device 3 will have

port switching action; Set the address to 0xFF, the command to 0x44, and the parameter to 0x01 ,

click to send and then the device1.2 3 will all have port switching action; Set the address to 0x83, the

command to 0x44, and the parameter to 0x02 , click to send and then the device1 and device 3 will

have port switching action;

The newly added command to set the multicast address greatly meets the needs of

customer groups, making it easier and more convenient for customers. You can choose the

device you want to control so that you can complete your work more efficiently and quickly

during usage.

B: Valve parameter query command (common command) (B2~B4)

Command Code B2

Command Instruction

Response Parameter (B3, B4)

0x20

Query Address

B3=0x00 B4=0x00

0x21

Query RS232 baud rate

B3=0x00 B4=0x00

0x22 0x23

Query RS485 baud rate Query CAN baud rate

B3=0x00 B4=0x00 B3=0x00 B4=0x00

0x2E

Query Automatic reset when power on

B3=0x00 B4=0x00

0x30

Query CAN destination address

B3=0x00 B4=0x00

0x70

Query multicast channel 1 address

B3=0x00 B4=0x00

0x71

Query multicast channel 2 address

B3=0x00 B4=0x00

0x72

Query multicast channel 3 address

B3=0x00 B4=0x00

0x73

Query multicast channel 4 address

B3=0x00 B4=0x00

0x3E 0x3F 0x4A

Query current channel address Query current version Query motor status

B3=0x00 B4=0x00
B3=0x01 B4=0x09, above is just an example, if the response parameter is the same as above parameter, it means the current version is V1.9, see the version number on the label for details
B3=0x00 B4=0x00

http://www.runzeliuti.com

13

Note: After version V1.8, the software sub-version and the main version are merged into one

command

C: Valve action command (common command) (B2~B4)

Command Code B2

Command Instruction

Parameter Instructions (B3, B4)

0x44
0x45 0x4F

The motor rotates through the code disc, and select the best
path automatically
Reset
Reset on origin position

According to the actual number of ports of the injector valve, such as a 10-port injector valve, B3=0xXX B4=0x00,the value range of XX is 010A Note: 6-port State 1 1-6 connected
8-port State1 1-2 connected 10-port State1 1-2 connected
B3=0x00 B4=0x00 Injector valve turns to state 2
B3=0x00 B4=0x00 Injector valve runs to the encoder origin position, overlaps the position reset by the 0x45 command

0x49

Forced stop

B3=0x00 B4=0x00

Example 1, 0x44: Multi-port position switching

(1) Current position: Reset Position

Target: Position A, Port 1 is connected with Port 6

Command: 0x44 Parameter: 0x01

After send commands, valve will switch port position, as shown in below Figure 3-1

(2) Current position: Reset Position

Target: Position B, Port 1 is connected with Port 2

Command: 0x44 Parameter: 0x02

After send commands, valve will switch port position, as shown in below Figure 3-2

Figure 3-1

Figure 3-2

http://www.runzeliuti.com

14

Sum check (B6, B7)

Item

Instruction

Code (B6 B7)

Remark

Sum check

Sum check

0xXX 0xXX

The sum of the header to the end of the package

Note: The sum check bit of the factory command is B12, B13

Response parameters are B2 B3 B4.

Table 3-2-5 Response Parameters

Command Code B2

Command Instruction

0x00

Normal state

0x01

Frame error

Parameter Description= B3, B4
B3=0x00 B4=0x00 Example: When using the query command
"0x3e", the response command Of B3 B4 =0x01 0x00 ~ 0x0a 0x00
B3=0x00 B4=0x00

0x02

Parameter error

B3=0x00 B4=0x00

0x03

Optocoupler error

B3=0x00 B4=0x00

0x04

Motor busy

B3=0x00 B4=0x00

0x05

Motor stalled

B3=0x00 B4=0x00

0x06

Unknown position

B3=0x00 B4=0x00

0xFE

Task is being executed

B3=0x00 B4=0x00

0xFF

unknown error

B3=0x00 B4=0x00

Note: In RS485 communication, when sending an action command, when code B2 received FE

byte means the command is received and is being executed

Note:  The code B2 in the response command indicates the current running state of the

motor. Only when B2=0x00, the motor is in normal operation, other parameters are as shown in the

above table, corresponding to different abnormal states respectively. In principle, the motor should be

sent the 0X4A command to query the motor status at the end of the motor operation. Only when the

B2 parameter in the response command is 00, other commands can be executed correctly. Note: The code parameters of all the above commands are set in little-endian mode. Low data is

stored in the low bit of the address and high data is stored in the high bit of the address.

http://www.runzeliuti.com

15

Chapter 4 Common Problems & Solutions

Phenomenon
Not working when powered on
Not aspirating after switching Liquid has bubbles
No communication
The communications sent and received are consistent in RS232 Application Notice:

Possible problem Working voltage is not within the
specified range The power connection is loose or
disconnected Working current is not within the
specified range
The channel is blocked by particles
The connector interface is not tight
The TX and RX lines of RS232 are connected reversely or RS485 phase A & phase B are connected reversely
TX & RX are in short circuit

Solution
Check whether the voltage is within the specified range
Check the power connection
Check whether the current is within the specified range
Take out the pump tube and check for blocked particles Replace suitable connector Exchange the TX and RX line sequence of RS232, exchange the A, B phase sequence of
RS485
Check if there is a short circuit, if yes, replace the cable

 Please ensure that the voltage matches the standard voltage of the instrument.

 Please use original serial port wires

 Communication RS232, RS485, CAN are under Non-isolation mode, hot swapping

unsupported.

 Please cover the unused ports with suitable coned plugs when laid aside to avoid impurity

substance and air

 Do not disassemble the product parts at will. The tamper-evident label is not guaranteed.

 Please read above operation instructions and communication protocols carefully. Do not

input data randomly.

 Discard the instrument should be in line with the regulations on the disposal. Dispose of the

waste in accordance with national environmental protection requirements. Users should not throw

away at will.

 When using CAN bus protocols to connect multiple devices, please refer to the connection

method shown in Figure 4-1 below.

http://www.runzeliuti.com

16

Device 1 Device 2 Figure 4-1

Device N

 When using RS485 protocol to connect multiple devices, please refer to the connection

method shown in Figure 4-2 below (CAN resistance is removed)

Device 1 Device 2 Figure 4-2

Device N

http://www.runzeliuti.com

17

http://www.runzeliuti.com

18

