Instruction Manual for ZOLL models including: AED Plus, AED Plus Automated External Defibrillator, Automated External Defibrillator, External Defibrillator, Defibrillator

Poster-Pedi

MHirshson

AED Plus

5 mag 2024 — △DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone. △Connect the electrode cable to the AED Plus unit after installing ...

[PDF] AED Plus - ZOLL Medicalwww.zollaeds-asia.com › aed-plus


File Info : application/pdf, 2 Pages, 733.54KB

PDF preview unavailable. Download the PDF instead.

9650-0300-05-SF G
Warning!
Use the AED Plus® unit only as described in this manual. Improper use of the
device can cause death or injury.
DO NOT use or place the AED Plus unit in service until you have read the AED
Plus Operator's and Administrator's Guides.
DO NOT use or place the AED Plus unit in service if the unit's status indicator
window (located on the left side of the handle) displays a red "X".
DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone. Connect the electrode cable to the AED Plus unit after installing batteries. Keep the electrode cable connected to the AED Plus unit at all times. This device is intended for use by personnel who have been trained in its
operation. Users should receive training in basic life support/AED, advanced life support or a physician-authorized emergency medical response training program.
Only use electrodes labeled "Infant/Child" on children less than 8 years old or
weighing less than 55 lbs (25 kg). Use CPR-D-padz® if patient is older than 8 years or weighs more than 55 lbs (25 kg).
Always stand clear of patient when delivering treatment. Defibrillation energy
delivered to the patient may be conducted through the patient's body and cause a lethal shock to those touching the patient.
DO NOT touch the electrode surfaces, the patient, or any conductive material
touching the patient during ECG analysis or defibrillation.
Move patient away from electrically conductive surfaces prior to use of equipment. DO NOT use the unit near or within puddles of water. Keep the patient as motionless as possible during ECG analysis. DO NOT use the unit near flammable agents, such as gasoline, oxygen-rich
atmospheres, or flammable anesthetics.
Avoid radio frequency interference from high-power sources that might cause the defibril-
lator to interpret cardiac rhythms incorrectly by turning off cell phones and 2-way radios.
Disconnect non-defibrillation protected electronic devices or equipment from
patient before defibrillation.
Dry victim's chest, if wet, before attaching electrodes. Apply freshly opened and undamaged electrodes, within the electrode expiration
date, to clean and dry skin to minimize burning.
DO NOT place electrodes directly over the patient's implanted pacemaker.
Pacemaker stimuli may degrade the accuracy of ECG rhythm analyses or the pacemaker may be damaged by defibrillator discharges.
Check labeling inside the ZOLL® AED Plus cover before using the cover as a Passive
Airway Support System (PASS) device to ensure it is intended for this use.
DO NOT use the Passive Airway Support System (PASS) if there is a suspected
head or neck injury. Place the patient on a firm surface before performing CPR.
DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may
explode if mistreated.
DO NOT use or stack the AED Plus unit with other equipment. If the unit is used or
stacked with other equipment, verify proper operation prior to use.
Caution!
DO NOT disassemble the unit. A shock hazard exists. Refer all servicing to
qualified personnel.
Use only 10 Duracell Coppertop CR123 batteries, made in the USA. Do not use any
batteries from other manufacturers. See the AED Plus Administrator's Guide (REF 9650-0301-05) for more information. Discard batteries properly after removal
from the unit.
If the device is stored outside the recommended environmental conditions, the
electrode pads and/or batteries may be damaged or their useful life reduced.
The CPR-D-padz Electrode can be connected to other ZOLL defibrillators with
Multifunction Cables. Defibrillation can be administered when connected to other ZOLL defibrillators.The CPR function does not operate with any device other than the AED Plus defibrillator.
Keep the unit away from magnetic resonance imaging equipment.
Important!
This symbol indicates that an AED Plus unit is equipped for treating adult and pediatric
patients. An AED Plus unit without this symbol is not equipped to treat pediatric patients
and will NOT work with Pedi-padz® II pediatric electrodes. To upgrade an AED Plus unit for
use with ZOLL Pedi-padz II pediatric electrodes, contact ZOLL Medical Corporation or an
authorized ZOLL distributor for information on the ZOLL AED Plus Pediatric Upgrade Kit.

Set-up and Check-out Procedure:
1. Insert 10 new batteries into AED Plus unit, left and right sides first. 2. Connect electrode cable to AED Plus unit and pack sealed electrodes inside unit
cover. Close cover. 3. Turn unit on and wait for "Unit OK" audio message. Verify that unit issues
appropriate "Adult Pads" or "Pediatric Pads" audio message. 4. Turn unit off. 5. Wait 2 minutes. Verify that green check symbol ( ) appears in status indicator
window (located on left side of handle) and that unit does not emit a beeping tone. 6. Place AED Plus unit in service. 7. Check AED Plus unit periodically to ensure that green check symbol ( ) appears
in status indicator window.
Battery Replacement
For AED Plus units running software version 5.32 or higher, replace batteries every 5 years or if unit prompts. For earlier software versions, replace batteries every 3 years and place a dated reminder note next to the On/Off button. Use only Duracell Coppertop CR123 batteries, made in the USA.
 Remove all batteries from battery compartment and discard before installing any new
batteries.
 Insert 10 new batteries into battery well, left and right sides first. Do not use old
batteries.
 Press button in battery well only after installation of new batteries.
Cleaning
 Clean and disinfect unit with soft, damp cloth using 90% isopropyl alcohol or soap
and water, or chlorine bleach (30 ml/liter water).
 Do not immerse any part of the unit in water.  Do not use ketones (MEK, acetone, etc.).  Avoid using abrasives (e.g., paper towels) on the LCD display, if so equipped.  Do not sterilize the unit.

TROUBLESHOOTING

Problem
Self-test failed.
"Change batteries" prompt.
Red "X" in Status Indicator window OR
beeping noise when unit is OFF.
Red "X" in Status Indicator window when unit is ON.

Recommended Action
Manually test by pressing and holding the ON/OFF button for more than 5 seconds. If unit fails test again, remove from service.
Replace all batteries at the same time.
Perform manual test. Check to see if cable is attached properly to unit. Replace batteries. If unit still does not operate correctly, remove from service.
Power cycle the unit. If Red "X" is still present in Status Indicator window, remove unit from service.

Federal (U.S.A.) law restricts this device for sale to or on the order of a physician.

AED Plus®
Automated External Defibrillator
Operator's Guide
ERC 2015

For Technical support or repair:

ZOLL Medical Corporation 269 Mill Rd. Chelmsford, MA 01824-4105 978-421-9655 www.zoll.com

ZOLL International Holding B.V. Einsteinweg 8A 6662 PW Elst Netherlands +31 (0) 481 366410

ZOLL Medical Switzerland A.G. Baarerstrasse 8 6300 Zug Switzerland
International Service: Contact your local distributor.
0123
© 2024 ZOLL Medical Corporation REF 9650-0300-05 Rev. G

start here

What you see. What you hear.
What you do.
Prompts you may see in the LCD screen: Text messages Shock counter Elapsed time ECG waveform

Person collapses-- suspected cardiac arrest.

Unit OK.

Turn unit on.

Indicator light flashes. Check responsiveness. Shout, "Are you OK?" as you shake the person.
Stay calm. Unit equipped to treat adult and pediatric patients.
Self-test; unit OK.

Indicator light flashes. Call for help. Call or send someone for help.
LCD screen

Indicator light flashes. Open airway.
Turn the victim onto his back and then open the airway using head tilt and chin lift; Place your hand on his forehead and gently tilt his head back; With your fingertips under the point of the victim's chin, lift the chin to open the airway.
Option: Place Passive Airway Support System (PASS) under shoulders to tilt head. DO NOT use PASS if suspected head or neck injury.

If not responsive, Open Airway.

Indicator light flashes. Check breathing.

Check Breathing. Look for chest movement; Listen at the victim's mouth for breath sounds; Feel for air on your cheek; Decide if breathing is normal, not normal or absent.

Indicator light flashes.
Attach defib pads to patient's bare chest.
Attach adult or infant/child electrode pads to patient's bare chest. Follow electrode pads package instructions.

Place electrode pads as shown on patients under 8 years of age or < 55 lbs (25 kg).

Other prompts you may hear:

What you should do:

Plug in cable. Unit failed.
Release shock button. Change batteries. (n) shocks delivered.

Attach electrode cable. Turn unit off and then on. If unit still fails, perform CPR.
Let go of shock button. Replace all 10 batteries.

Indicator light flashes.
Start CPR.
Administer CPR: 2 breaths for every 30 compressions.

After one shock...

Indicator light flashes. Shock advised. Don't touch patient. Press flashing shock button.
Give STAND CLEAR Warning. DO NOT touch patient. Press button. Shock delivered.
After 2 minutes of CPR...
Stop CPR.

Indicator light flashes. Don't touch patient. Analyzing.
No shock advised.

INDICATIONS FOR USE
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION indicated by:
 Unconsciousness and  Absence of normal breathing and  Absence of a pulse or signs of circulation.
When a victim is less than 8 years of age, or weighs less than 55 lbs (25 kg), the ZOLL AED Plus should be used with ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.

CONTRAINDICATIONS FOR USE
Do NOT use when a patient:
 Is conscious; or  Is breathing; or  Has a detectable pulse or other signs of circulation.
INTENDED PURPOSE
The intended purpose of the ZOLL AED Plus is to help victims of sudden cardiac arrest. Using compatible ZOLL defibrillation electrodes which connect the patient to the ZOLL AED Plus, the AED automatically analyzes the victim's ECG rhythm to determine whether an electrical shock should be delivered. If the AED detects a shockable rhythm (coarse ventricular fibrillation or wide complex ventricular tachycardia), it alerts the rescuer that a shock is advised and then it either issues

instructions to press the flashing shock button (semi-automatic model) or it automatically delivers the shock after a 3 second countdown (fully automatic model). The AED then prompts the rescuer to perform CPR for a pre-configured period of time.
INTENDED USER
The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in the use of the AED Plus device and basic life support, advanced life support, or other physicianauthorized emergency medical response to defibrillate victims of cardiac arrest. The CPR monitoring function provides a metronome designed to encourage rescuers to perform chest compressions at the ERC recommended rate of 100-120 compressions per minute. Voice and visual prompts encourage a compression depth of 5 - 6 cm for adult patients. The CPR monitoring function is not intended for use on patients under 8 years of age.

ZOLL Medical Corporation
269 Mill Road Chelmsford, MA 01824-4105
1-978-421-9655 www.zoll.com



References

Acrobat Elements 15.0 (Windows)