UT3200plus PROGRAMMING MANUAL
Programming Manual
UT3200+ Series Multi-channel Temperature Tester

Programming Manual

UT3200+ Series

Warranty and Statement
Copyright
2023 Uni-Trend Technology (China) Co., Ltd.
Brand Information
UNI-T is the registered trademark of Uni-Trend Technology (China) Co., Ltd.
Statement
 UNI-T products are protected by patents (including obtained and pending) in China and other countries and regions.
 UNI-T reserves the right to change specifications and prices.  The information provided in this manual supersedes all previous publications.  The information provided in this manual is subject to change without notice.  UNI-T shall not be liable for any errors that may be contained in this manual. For any incidental or
consequential damages arising out of the use or the information and deductive functions provided in this manual.  No part of this manual shall be photocopied, reproduced or adapted without the prior written permission of UNI-T.
Product Certification
UNI-T has certified that the product conforms to China national product standard and industry product standard as well as ISO9001:2008 standard and ISO14001:2004 standard. UNI-T will go further to certificate product to meet the standard of other member of the international standards organization.

Instruments.uni-trend.com

2 / 13

Programming Manual

UT3200+ Series

SCPI
SCPIStandard Commands for Programmable Instrumentsis a standardized instrument programming language that builds on existing standards IEEE 488.1 and IEEE 488.2 and follows the floating point rules of IEEE 754 standard, ISO 646 message exchange 7-bit encoding notation (equivalent to ASCII programming) and many other standards.
This section introduces the format, symbols, parameters, and abbreviations of the SCPI command.
Command String Parse
The host computer can send a string of commands to the instrument and the command parser of the instrument starts to parsing after catching the terminator (\n) or an input buffer overflow. For example Valid command string: AAA:BBB CCC;DDD EEE;:FFF The instrument command parser is responsible for all command parsing and execution, and you must understand its parsing rules before writing a program.

Command Parse Rule

Command parser only parses and responds to ASCII data.

The command parser starts to command parsing when receive the end mark. The instrument only accept

three contents as the following as the end mark.

CR

CR+LF

LF

The command parser will terminate the parsing immediately after parsing an error, and the current command

will be invalidated.

The command parser is case-insensitive for parsing command strings.

he command parser supports abbreviated form of command and the detailed see the following section. In RS485 mode, add ADDRLocal address:: in front of SCPI, the local address can set to 1-32.

It's convenient to communicate with multiple devices via SCPI protocol.

For example: ADDR1::IDN?

 represents a blank

The end of data sent by the instrument defaults to 0x0A (LF).

Multiple instruction can be send via semicolon " ; ".

Instruments.uni-trend.com

3 / 13

Programming Manual

UT3200+ Series

Symbol Stipulation and Definition

This chapter uses some symbols that are not part of the command tree, but only for a better understanding of the command string.

Mark <......>
[......] {......}
Capital letter 

Description The text in angle brackets indicates the parameter of the command. For example: <float> represents floating point number <integer> represents integer parameter The text in square brackets indicates the optional command. When the curly brackets contain several parameter items, it means that only one item can be selected from them. Abbreviated form of the command.
Blank mark, it represents a blank and only for reading.

Command Tree Structure
SCPI commands have a tree-like structure with three level (note: the command parser of this instrument can parse any level), where the highest level is called the subsystem command. SCPI uses a colon (:) to separate high level commands from low level commands.

ROOT

AAA

BBB

CCC

DDD
Figure 1-1 Command Tree Structure

EEE

Instruments.uni-trend.com

4 / 13

Programming Manual

UT3200+ Series

For Example ROOT:CCC:DDD ppp
ROOT CCC DDD ppp

Subsystem command Second level Third level Parameter

Command and Parameter
A command tree is consist of command and [parameter], use a blank to separate (ASCII: 20H). For example AAA:BBB 1.234 Command [parameter]

Command

Command words can be in long command format or in abbreviated form. Long format facilitates engineers to better understand the meaning of the command string; abbreviated form is suitable for writing.

Parameter

Single character command, no parameter For Example AAA:BBB
Parameter can be string format and its abbreviated form is also follow the last section " command abbreviated rule" For example AAA:BBB1.23
Parameter can be numerical value format.

<integer> <float>

123, +123, -123
Floating point number of arbitrary form: <fixfloat>: fixed floating point number: 1.23, -1.23 <Sciloat>: floating point number represented by scientific notation: 1.23E+4, +1.23e-4 <mpfloat>: floating point number represented by multiplying power: 1.23k, 1.23M, 1.23G, 1.23u

Instruments.uni-trend.com

5 / 13

Programming Manual

UT3200+ Series

Table 0-1 Abbreviation of Multiplying Power
Numerical Value 1E18 (EXA) 1E15 (PETA) 1E12 (TERA) 1E9 (GIGA) 1E6 (MEGA) 1E3 (KILO) 1E-3 (MILLI) 1E-6 (MICRO) 1E-9 (NANO) 1E-12 (PICO) 1E-15 (PEMTO) 1E-18 (ATTO)

Multiplying Power EX PE T G MA K M U N P F A

SCPI is case-insensitive, so the written is different from standard name. For example :
"1M" represents 1 milli, not 1 mega.
"1MA" represents 1 mega.

Separator

The instrument command parser can only receive allowable separator. Other separator will cause error "Invalid separator".

;

Semicolon is for separating two commands.

For Example AAA:BBB 100.0 ; CCC:DDD

:

Colon is for separating command tree or restart the command tree.

For Example AAA : BBB : CCC 123.4; : DDD : EEE 567.8

?

Question mark is for querying.

For Example AAA ?



Blank is for separating the parameter.

For Example AAA:BBB1.234

Instruments.uni-trend.com

6 / 13

Programming Manual

Command Reference

All commands are explained by the subsystem command order.

MEAS

Measurement subsystem

SYST

System subsystem

FETCH

Fetch data subsystem

ERROR

ERROR subsystem

IDN?

Query subsystem

UT3200+ Series

MEAS Subsystem

MEAS subsystem is used to switch to different display page.

MEAS

:MODEL

{tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b}

:RATE

{fast,slow}

:START

{on,off}

:CMODEL

<para>,<level>

:CHANON

<para>,<on,off>

:LOW

<level>

:CLOW

<para>,<level>

:HIGH

<level>

:CHIGH

<para>,<level>

:SENSOR

{tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b}

MEAS:MODEL
MEAS:MODEL is used to set sensor type. Command Syntax MEAS:MODEL<tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b> Example SEND>MEAS:MODEL tc-k <NL> // Set the sensor type to Type K thermocouple. Query Syntax MEAS:MODEL? Query Return <tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b> Example SEND> MEAS:MODEL?<NL> RET> tc-t <NL>

Instruments.uni-trend.com

7 / 13

Programming Manual

UT3200+ Series

MEAS:RATE
MEAS:RATE is used to set sampling rate. Command Syntax MEAS:RATE<fast,slow> Example SEND>MEAS:RATE fast<NL> // Set sampling rate to fast. Query Syntax MEAS:RATE? Query Return <fast,slow> Example SEND> MEAS:RATE?<NL> RET> fast <NL>
MEAS:START
MEAS:START is used to enable the sampling. Command Syntax MEAS:START <on,off> Example SEND>MEAS:START off<NL> // Stop sampling. Query Syntax MEAS:START? Query Return <on,off> Example SEND> MEAS:START?<NL> RET> on <NL>
MEAS:CMODEL
MEAS:CMODEL is used to set the sensor type of each channel. Command Syntax MEAS:MODEL <para>,<tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b> For Example SEND>MEAS:CMODEL 1,TC-T<NL> // Set the sensor of CH001 to Type T. Query Syntax MEAS:CMODEL? // Acquire the sensor type of all channels. MEAS:CMODEL?<int> // Acquire the sensor type of single channel, the minimum of channel number is 1. Query Return < tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b > For Example SEND> MEAS:CMODEL?<NL> RET> < tc-t,tc-k,tc-j,tc-n,tc-e,tc-s,tc-r,tc-b ><NL> SEND> MEAS:CMODEL? 1<NL> // Acquire the sensor type of CH001. RET> < tc-t ><NL>

Instruments.uni-trend.com

8 / 13

Programming Manual

UT3200+ Series

MEAS:LOW
MEAS:LOW is used to set the lower limit of all channels. Command Syntax MEAS:LOW <float> For Example SEND>MEAS:LOW -200.0<NL> // Set the lower limit of all channels to -200.0. Query Syntax MEAS:LOW? Query Return <float,float> <NL> For Example SEND> MEAS:LOW? <NL> RET> <-2.00000e+02,-2.00000e+02> <NL>

MEAS:CLOW
MEAS:CLOW is used to set the lower limit of each channel. Command Syntax MEAS:CLOW <para>,<float> For Example SEND>MEAS:CLOW 1,-200.0<NL> // Set the lower limit of CH001 to -200.0.

MEAS:HIGH
MEAS:HIGH is used to set the upper limit of all channels. Command Syntax MEAS:HIGH <float> For Example SEND>MEAS:HIGH 1800.0<NL> // Set the upper limit of all channels to 1800.0. Query Syntax MEAS:HIGH? Query Return <float,float> <NL> For Example SEND> MEAS:HIGH? <NL> RET> <1.80000e+03, 1.80000e+03> <NL>

MEAS:CHIGH

MEAS:CHIGH is used to set the upper limit of each channel.

Command Syntax MEAS:CHIGH <para>,<float>

For Example SEND>MEAS:CHIGH 1,1800.0<NL> // Set the upper limit of CH001 to 1800.0.

Query Syntax

MEAS:CHIGH1

Query Response <float> <NL>

Example SEND> MEAS:CHIGH? 1<NL>

RET> <1.80000e+03> <NL>

Instruments.uni-trend.com

9 / 13

Programming Manual

MEAS:SENSOR

MEAS:SENSOR is used to acquire sensor type of each channel. Command Syntax MEAS:SENSOR Query Response <TC-T,TC-K,TC-J,TC-N,TC-E,TC-S,TC-R,TC-B> <NL> Example SEND> MEAS:SENSOR <NL> RET> <TC-T,TC-K,TC-J,TC-N,TC-E,TC-S,TC-R,TC-B> <NL>

SYST Subsystem

SYST subsystem is used to set SETUP page.
:COMP SYST :BEEP
:KEYTONE :UNIT

{on,off} {on,off} {on,off} {cel,kel,fah}

SYST:COMP

SYST:COMP is used to set the comparator state. Command Syntax SYST:COMP <on,off> For Example SEND>SYST:COMP on<NL> // Turn on the comparator. Query Syntax SYST:COMP? Query Return <on,off> <NL> For Example SEND> SYST:COMP? <NL> RET> <on> <NL>

SYST:BEEP
SYST:BEEP is used to set the comparator beep state. Command Syntax SYST:BEEP <on,off> For Example SEND>SYST:BEEP on<NL> // Turn on comparator beep. Query Syntax SYST:BEEP? Query Return <on,off> <NL> For Example SEND>SYST:BEEP? <NL> RET> <on> <NL>

Instruments.uni-trend.com

UT3200+ Series 10 / 13

Programming Manual

UT3200+ Series

SYST:KEYTONE
SYST:BEEP is used to set the state of key beep. Command Syntax SYST:KEYTONE <on,off> For Example SEND>SYST:KEYTONE on<NL> // Turn on key beep. Query Syntax SYST:KEYTONE? Query Return <on,off> <NL> For Example SEND>SYST:KEYTONE? <NL> RET> <on> <NL>

SYST:SYSINIT

Command Syntax Example

SYST:SYSINIT SEND> SYST:SYSINIT // Return to factory set.

SYST:UNIT

SYST:UNIT is used to set the temperature unit. Command Syntax SYST:UNIT <cel,kel,fah> Parameter <cel,kel,fah> cel: degree Celsius kel: Kelvin degree fah: Fahrenheit degree For Example SEND>SYST:UNIT cel<NL> // Set the temperature unit to degree Celsius. Query Syntax SYST:UNIT? Query Return <cel,kel,fah> <NL> For Example SEND> SYST:UNIT? <NL> RET> <cel> <NL>
FETCH Subsystem
FETCH subsystem is used to acquire the temperature data.
FETCH?

Instruments.uni-trend.com

11 / 13

Programming Manual

UT3200+ Series

FETCH?
FETCH? is used to fetch temperature data. Query Syntax FETCH? Query Return <float, float , float> <NL> For Example SEND>FETCH? <NL> RET> <+1.00000e-05, +1.00000e-05, +1.00000e-05> <NL>
ERROR Subsystem
ERROR subsystem is used to return error message. Query Syntax ERROR? Query Return Error string For Example SEND> ERR? <NL> RET>no error <NL>
*IDN? Subsystem
IDN? is used to query instrument ID. Query Syntax IDN?OR *IDN? Query Return <MODEL>,<Revision>,<SN>,<Manufacturer>

Modbus

Register Overview
All register addresses used by the instrument are listed below.
Notes: 1. Unless otherwise specified, the numeric value of instruction and response frame are hexadecimal. 2. The register only contains the instruction of acquiring the test result and starting/stopping the test. If user want to
customize other instructions, please contact UNI-T sake department. 3. Floating point number online conversion can refer to website
http://www.binaryconvert.com/convert_float.html

Register Address 0200 0202~0261

Name Start/Stop test Temperature value of channel 1~48

Numeric value 1 byte integer 4 bytes floating point number

Description Wirte-only register, data takes 1 register
Read-only register, data of each channel takes 2 registers.

Instruments.uni-trend.com

12 / 13

Programming Manual

UT3200+ Series

Start/Stop Test

Write 1 01 Station number

2 10 Write

3

4

02

00

Register

5

6

00

01

Register

quantity

7

8

9

02

00

01

Byte

Data

10

11

44

50

CRC16

0000Stop 0001Start

Written return

1

2

01

10

Slave

Write

station

3

4

02

00

Register

5

6

00

01

Register

quantity

7

8

00

71

CRC16

Acquire Test Result

Register 0202~0261 is used to acquire the test result of all channels.

For example: acquire the test result of CH1

Send

1

2

3

4

5

6

01

03

02

02

00

02

Slave

Read

Register

Register

station

quantity

7

8

64

73

CRC-16

Response

1

2

3

4

5

6

7

8

9

01

03

04

41

DC

44

5A

9C

CE

01

03

Byte

Float-point number with single

CRC-16

precision

B4~B7 is float-point number with single precision, byte order AA BB CC DD

Test data: 41 DC 44 5A converts to float-point number: 0x41DC445A = 27.5334; (If the channel is open circuit, then the test

result is 100000.)

Instruments.uni-trend.com

13 / 13

