AVEIR™ VR Leadless Pacemaker

Setting the Pace for What's to Come

Long-Term Retrieval. Long Lasting. Mapping Prior to Fixation.

Advancing Today's Technology with a View to Tomorrow

Leadless pacemakers (LP) have revolutionized care for heart patients. With no visible or physical reminder of a pacemaker under the skin and fewer post-implant activity restrictions, patients can continue living their lives to the fullest.

The AVEIR™ VR Leadless Pacemaker (LP) is the next evolution in leadless technology, designed for:

At just 38.0 mm, the AVEIR VR LP has three times less volume than a standard AAA battery.

[Diagram: Comparison of AVEIR VR LP to a AAA battery, showing dimensions]

Long-Term Retrieval

The AVEIR™ VR Leadless Pacemaker is designed for long-term retrieval. Limited data is available for AVEIR VR LP. The AVEIR VR LP's predicate device has an overall long-term retrieval success rate of more than 88% with helix fixation with up to 9 years of retrieval experience.

[Diagram: Illustration of the AVEIR VR Leadless Pacemaker with its active fixation helix]

The AVEIR™ VR Leadless Pacemaker's active fixation helix uses a screw-in mechanism to enable both implantation and long-term retrieval of the LP.

Overall Long-Term Retrieval Success Rate:

>88% with helix fixation (Range 0 to 9 years, mean 3.1 ± 1.8 years, n=241)

[Diagram: Pie chart showing >88% success rate]

Tri-Loop Snare Retrieval Mechanism:

[Diagram: Sequence showing the Tri-Loop Snare Retrieval Mechanism engaging the AVEIR VR LP]

Long Lasting

Increased projected battery longevity over current available leadless pacemakers opens the door to more patients.

Mean ± SD: 17.6±6.6 years

95% CI: 16.6 to 18.6 years

[Bar chart: Estimated Total Longevity distribution of patients, showing mean and ranges]

The average battery longevity among Leadless II phase 2 IDE patients at 1 year follow-up is estimated to be 17.6 years. 48% of the study patients have an estimated battery longevity of over 20 years.

Battery Capacity:

Up to TWICE THE BATTERY CAPACITY of current VR leadless pacemakers.

Mapping Prior to Fixation

AVEIR™ VR Leadless Pacemaker mapping capability is designed to help reduce the number of repositioning attempts.

83.2% of patients had successful implants with no repositioning attempts.

96.4% of patients had successful implants in 1 or less repositioning attempts.

[Diagram: Comparison of AVEIR VR LP and Micra VR mapping capabilities. AVEIR VR LP can measure R-waves, impedance, and capture threshold before fixation by touching the tip electrode to the endocardial tissue. AVEIR VR LP is engaged with a rotational motion. Micra VR requires deployment before initial electrical readings and is inserted with forward tip pressure.]

Life-Changing Innovation

At Abbott, we adapt quickly to changes in the world around us, harnessing leading-edge science and technology to deliver the best possible solutions for some of the world's most important health challenges.

Our medical devices, like the AVEIR™ VR Leadless Pacemaker, use the most advanced technologies to keep hearts beating more regularly.

AVEIR DELIVERY AND RETRIEVAL CATHETERS

[Diagram: Illustration of the AVEIR Delivery Catheter and Retrieval Catheter (Tri-Loop Snare)]

Upgradable System

Patient therapy can be tailored by implanting an atrial or ventricular device alone, or both combined for dual chamber support. The option to upgrade over time allows you to meet your patient's needs today and adapt to common disease progression later.

  1. Option to Start with Ventricular Pacing: Treat patients for rare intermittent A-V block today.
  2. Add Atrial Pacing Later: Treat those same patients by adding an atrial device if sick sinus syndrome develops later to provide DDDR therapy.
  3. Achieve Dual Chamber Pacing: Now you have options to adapt to patient needs over time.

[Diagram: Sequence showing the progression from ventricular pacing to dual chamber pacing with AVEIR VR LP]

Enabling Safe MRI Scans

MR AVEIR is 1.5T and 3T MR Conditional

AVEIR VR LP is MR Conditional for full body scans using a 1.5T or 3T field strength MRI scanner.

AVEIR VR LP MR CONDITIONAL FEATURES*

Magnet (Tesla) 1.5T and 3T
Scan Type Full-Body
Scanner Mode First Level Controlled Operating Mode Or Normal Operating Mode

*For additional information about specific MR Conditional details, including warnings, precautions, adverse conditions to MRI scanning and potential adverse events, please refer to the MRI-Ready Leadless System Manual at medical.abbott/manuals or check our MRI Ready resources at cardiovascular.abbott/mriready

Workflow

Using the power of leadless technology, the AVEIR™ VR Leadless Pacemaker is implanted in the heart through a minimally invasive catheter procedure.

  1. CONNECT: Patient's skin electrodes to Merlin Programmer™ via AVEIR™ Link Module.
  2. INSERT: LP via ergonomic catheter for minimally invasive procedure, ensuring helix is fully covered during catheter navigation.
  3. MAP: Ventricular implanting location before fixation by touching endocardium with LP's tip electrode.
  4. FIXATE: The LP to the ventricular endocardium using the screw-in mechanism.
  5. RELEASE: LP and withdraw delivery catheter.

For a more detailed workflow, refer to the AVEIR™ VR LP Instructions For Use (IFU) or contact your Abbott Sales Representative for training.

About Abbott

A healthy heart is essential to good health. That's why Abbott is committed to advancing treatments for people with cardiovascular disease. Our breakthrough medical technologies help restore people's health so they can get back to living their best lives, faster.

Abbott focuses on innovative technologies that can improve the way doctors treat people with heart arrhythmias, or irregular heartbeats. Abbott's cardiac rhythm management devices keep the heart beating at a healthy pace with pacemakers, implantable cardiac defibrillators, and implantable cardiac monitors, all designed to get people's hearts working better, sooner.

Ordering Information

Contents: AVEIR™ VR Leadless Pacemaker & AVEIR™ Delivery Catheter

AVEIR™ VR Leadless Pacemaker

MODEL NUMBER DESCRIPTION LENGTH (MM/INCH) OUTER DIAMETER (MM/INCH) WEIGHT (G) VOLUME (CC)
LSP112V Aveir™ VR Leadless Pacemaker 38/1.5 6.5/0.26 2.4 1.1

AVEIR™ Delivery Catheter

MODEL NUMBER DESCRIPTION INTRODUCER SIZE (FR) DEFLECTION (DEGREES) OUTER DIAMETER OF GUIDE CATHETER BODY (MM/INCH) EFFECTIVE LENGTH (CM/INCH)
LSCD111 Aveir™ Delivery Catheter 25 ≥ 170 4.57/0.180 105/41.3

Contact your Abbott Sales Representative or visit cardiovascular.abbott/AVEIRVR

References

  1. Sattar et al. Complications of leadless vs conventional (lead) artificial pacemakers - a retrospective review. Journal of community hospital internal medicine perspectives vol. 10,4 328-333. 2 Aug. 2020, doi:10.1080/20009666.2020.1786901
  2. Reddy VY, Cantillon DJ, John IP. San Francisco, CA: 6 May 2016. A comparative study of acute and mid-term complications of leadless vs transvenous pacemakers. Late-Breaking Clinical Trials II. Presented at Heart Rhythm Society 2016; pp. 02-04. Abstract LBCT.
  3. Reddy, VY, et al. Worldwide Experience with Leadless Pacemaker Retrievals at 9 years. Presented at 15th Asia Pacific Heart Rhythm Society (APHRS) Scientific Session; Nov 18-20, 2022; Singapore.
  4. AVEIR™ DR Leadless Pacemaker and Delivery Catheter IFU. ARTEN600284235
  5. Micra VR IFU M991010A001 REV. B
  6. Reddy VY, Exner D, et al. 1-Year Outcomes of a Leadless Ventricular Pacemaker: The LEADLESS II (Phase 2) Trial. JACC: Clinical Electrophysiology 2023, DOI: 10.1016/j.jacep.2023.01.031
  7. AVEIR™ Retrieval Catheter IFU. ARTMT600174816
  8. AVEIR™ Introducer IFU. ARTEN600174817

Rx Only

Brief Summary: Prior to using these devices, please review the Instructions for Use for a complete listing of indications, contraindications, warnings, precautions, potential adverse events and directions for use.

Indications:

The AVEIR™ Leadless Pacemaker system is indicated for management of one or more of the following permanent conditions: Syncope, Pre-syncope, Fatigue, Disorientation. Rate-modulated pacing is indicated for patients with chronotropic incompetence, and for those who would benefit from increased stimulation rates concurrent with physical activity. Dual-chamber pacing is indicated for patients exhibiting: Sick sinus syndrome, Chronic, symptomatic second- and third-degree AV block, Recurrent Adams-Stokes syndrome, Symptomatic bilateral bundle-branch block when tachyarrhythmia and other causes have been ruled out. Atrial pacing is indicated for patients with: Sinus node dysfunction and normal AV and intraventricular conduction systems. Ventricular pacing is indicated for patients with: Significant bradycardia and normal sinus rhythm with only rare episodes of AV block or sinus arrest, Chronic atrial fibrillation, Severe physical disability. MR Conditional: The AVEIR Leadless Pacemaker is conditionally safe for use in the MRI environment and according to the instructions in the MRI-Ready Leadless System Manual.

Intended Use:

The AVEIR™ Leadless Pacemaker (LP) is designed to provide bradycardia pacing as a pulse generator with built-in battery and electrodes for implantation in the right ventricle and/or right atrium. The LP is intended to provide sensing of intrinsic cardiac signals and delivery of cardiac pacing therapy within the implanted chamber for the target treatment group. The LP is also intended to operate optionally with another co-implanted LP to provide dual-chamber pacing therapy.

The AVEIR™ Delivery Catheter is intended to be used in the peripheral vasculature and the cardiovascular system to deliver and manipulate an LP. Delivery and manipulation includes implanting an LP within the target chamber of the heart.

Contraindications:

Use of the AVEIRT™ Leadless Pacemaker is contraindicated in these cases:

Adverse Events:

Potential complications associated with the use of the AVEIR™ Leadless Pacemaker system are the same as with the use of single or dual chamber pacemakers with active fixation pacing leads including, but not limited to: Cardiac perforation, Cardiac tamponade, Pericardial effusion, Pericarditis, Endocarditis, Valve damage or regurgitation, Heart failure, Pneumothorax/hemothorax, Cardiac arrhythmias, Diaphragmatic/phrenic nerve stimulation / extra-cardiac stimulation, Palpitations, Hypotension, Syncope, Cerebrovascular accident, Infection, Hypersensitivity reaction to device materials, contrast media, medications, or direct toxic effect of contrast media on kidney function, Pacemaker syndrome, Inability to interrogate or program the LP due to programmer or LP malfunction, Intermittent or complete loss of capture, pacing or sensing (non-battery related), Oversensing, Increased capture threshold, Inappropriate sensor response, Corrupted, intermittent, or loss of i2i communications, Interruption of desired LP function due to electrical interference, either electromyogenic or electromagnetic, Battery malfunction/ premature battery depletion, Device-related complications (Premature deployment, Device dislodgement/ embolization of foreign material, Inability to release/re-dock of the LP from the catheter, Helix distortion), Additional surgery or intervention, Death. As with any percutaneous catheterization procedure, potential complications include, but are not limited to: Vascular access complications; such as perforation, dissection, puncture, groin pain, Bleeding or hematoma, Thrombus formation, Thromboembolism, Air embolism, Local and systemic infection, Peripheral nerve damage. General surgery risks and complications from comorbidities; such as dyspnea, respiratory failure, pneumonia, hypertension, cardiac failure, reaction to sedation, renal failure, anemia, and death.

Indicates a trademark of the Abbott group of companies.

† Indicates a third-party trademark, which is property of its respective owner.

[Image: Abbott logo]

Abbott

15900 Valley View Court, Sylmar, CA 91342

Tel: +1 818 362 6822

Abbott.com

© 2023 Abbott. All Rights Reserved.

MAT-2115556 v7.0 | Item approved for US use only.

PDF preview unavailable. Download the PDF instead.

BR.4807408206.AVEIR.VR.HCP.US.EXT.WR-Single Adobe PDF Library 17.0 Adobe InDesign 18.5 (Macintosh)

Related Documents

Preview Abbott AVEIR DR Dual Chamber Leadless Pacemaker System: Advanced Cardiac Pacing
Explore the Abbott AVEIR DR Dual Chamber Leadless Pacemaker System, the world's first dual chamber leadless pacing solution. Learn about its i2i™ technology, upgradeable design, long-term retrieval, and proven clinical outcomes for advanced cardiac care.
Preview AVEIR™ AR Right Atrial Leadless Pacemaker System Medicare Reimbursement Guide
A comprehensive guide for healthcare professionals on Medicare reimbursement for the AVEIR™ AR Right Atrial Leadless Pacemaker System, covering coverage, coding, claim submission, and relevant FAQs.
Preview AVEIR™ AR Atrial Leadless Pacemaker (LP) Prior Authorization Suggestions
Guidance for physicians on prior authorization for the AVEIR™ AR Atrial Leadless Pacemaker (LP) procedure, including CPT codes and suggested clinical information.
Preview AVEIR™ DR Dual Chamber Leadless Pacemaker Medicare Reimbursement Guide | Abbott
This guide provides comprehensive information on Medicare reimbursement for the AVEIR™ DR Dual Chamber Leadless Pacemaker System, covering coverage, CPT codes, billing procedures, and additional payment information.
Preview Medicare Advantage Policy Manual: Dual Chamber Leadless Pacemakers
This policy manual outlines the Medicare Advantage coverage criteria, indications, contraindications, and coding for Dual Chamber Leadless Pacemakers, including the Aveir DR system.
Preview St. Jude Medical MR Conditional System MRI Procedure Manual
Guide for healthcare professionals on performing MRI scans on patients with St. Jude Medical MR Conditional systems. Covers 3T and 1.5T MRI parameters, safety, device compatibility, and clinical procedures.
Preview Abbott Assurity MRI Dual Chamber Pacemaker - Technical Specifications and Product Information
Detailed technical specifications, product highlights, and ordering information for the Abbott Assurity MRI Dual Chamber Pacemaker. Includes information on MRI compatibility, pacing features, and AF management.
Preview Abbott MRI-Ready System MRI Scan Clinician Checklist
A checklist for clinicians to ensure proper procedures are followed when scanning patients with Abbott MRI-Ready Systems, including pre-scan, during-scan, and post-scan device settings.