HORIBA

Application Note

Measuring Size of Light Absorbing Materials

AN272

Introduction

Samples that absorb certain light wavelengths can pose challenges for dynamic light scattering (DLS) systems. DLS relies on measuring scattered light, not absorbed light, from nanoparticles. Nanogold, for instance, exhibits strong absorbance due to surface plasmon resonance, which significantly reduces the scattering signal, making it difficult to measure.

This application note demonstrates the measurement of a ~26 nm nanogold sample using the HORIBA SZ-100V2 nanoparticle analyzer. Two different experimental setups were tested: one with a 90° detector and cell center position, and another with a 173° detector and cell wall position. The absorption spectrum of the nanogold was also verified using a HORIBA Duetta Fluorescence and Absorbance Spectrometer.

Analytical Test Method

Test Procedure

  1. Add sample as-is to a quartz cuvette.
  2. Insert cuvette into the instrument and allow the temperature to equalize for 2-3 minutes at 25°C.
  3. Perform 3 measurements.

Experimental Results

Autocorrelation Functions

Figure 1a: Autocorrelation function with 90° detector and cell center position, showing poor signal. This setup yielded low signal and unstable data.

Figure 1b: Autocorrelation function with 173° detector and cell wall position, showing strong signal. This setup provided strong, stable data.

Z-average and Polydispersity Index Data

Table 1a. Z-average and Polydispersity Index with 90° detector and cell center position. Data shows poor matching with expected size and poor repeatability.
Measurement Z-average Polydispersity index
1 1923.1 0.351
2 --- ---
3 1067.0 0.223
Table 1b. Z-average and Polydispersity index with 173° detector and cell wall position. Data shows good matching with expected size and good repeatability.
Measurement Z-average Polydispersity index
1 26.2 0.463
2 26.2 0.468
3 25.9 0.483

Absorption Spectrum

Figure 2. Absorption Spectrum for nanogold taken with HORIBA Duetta, showing a 530 nm absorption peak. This peak aligns with the green laser wavelength of the SZ-100V2 (532 nm).

Discussion

The data from the 90° detector at the cell center position exhibited very low and unstable signals, as shown in Figure 1a and Table 1a. In contrast, the 173° detector at the cell wall position yielded strong, stable autocorrelation functions (Figure 1b) and Z-average values that closely matched the expected size of the nanogold (Table 1b).

The improved data quality when using the 173° detector and cell wall position can be attributed to minimizing the laser pathlength through the sample. This reduces the opportunity for the laser light to be absorbed, especially due to the surface plasmon resonance effect in nanogold. Conversely, the 90° detector at the cell center position involves a longer pathlength, leading to greater absorption of the laser signal and less scattered light reaching the detector.

This optimized technique, using the 173° detector with the cell wall position, effectively maximizes the scattering signal while minimizing laser absorption. It can also be applied to fluorescent samples, provided a filter is used to remove incoherent fluorescent signals. The absorption spectrum confirms the nanogold's peak absorption at 530 nm, which is consistent with the SZ-100V2's 532 nm laser, explaining the significant difference observed between the two measurement configurations.

PDF preview unavailable. Download the PDF instead.

AN272 Measuring Size of Light Absorbing Materials with DLS Adobe PDF Library 17.0 Adobe InDesign 20.2 (Windows)

Related Documents

Preview HORIBA ViewSizer 3000: Investigating the Lower Size Limit of 10 nm Silver Nanoparticles
An application note detailing how the HORIBA ViewSizer 3000 measures 10 nm silver nanoparticles using multispectral nanoparticle tracking analysis, comparing results with TEM data.
Preview HORIBA Scientific Instruments: Particle, Elemental, pH, and Spectroscopy Analysis Solutions
Explore HORIBA's comprehensive range of advanced scientific analytical instruments, including particle size analyzers, elemental analyzers, pH meters, and spectroscopy systems for research and quality control.
Preview Measuring 1 µm PSL Standard with the HORIBA LA-960V2 HL Cell
Application note AN255 detailing the procedure and results for measuring a 1 µm Polystyrene Latex (PSL) standard using the HORIBA LA-960V2 particle size analyzer with the High-concentration Low-viscosity (HL) Cell accessory.
Preview HORIBA Report 2024-2025: Driving Growth and Sustainability
Explore HORIBA's 2024-2025 report detailing their commitment to innovation, sustainability, and creating value through diverse people and technology. Discover their strategies in Energy & Environment, Bio & Healthcare, and Materials & Semiconductor.
Preview HORIBA Report 2024-2025: Driving Innovation for a Sustainable Future
HORIBA's 2024-2025 report detailing its mid-long term management plan (MLMAP2028), corporate strategy, financial performance, sustainability initiatives, and commitment to "Joy and Fun" and "HONMAMON" philosophy across its three focus fields: Energy & Environment, Bio & Healthcare, and Materials & Semiconductor.
Preview Monitoring Sub-surface Stress in 4H-SiC Using Raman Spectroscopy
This application note demonstrates how HORIBA's LabRAM Odyssey Raman spectrometer can be used to monitor and evaluate stress in 4H-SiC substrates during surface processing. It details the materials and methods used, presents data on peak shifts related to stress, and discusses the implications for process optimization and material behavior in electronic applications.
Preview HORIBA Report 2024-2025: Driving Growth and Contributing to a Sustainable Society
Explore HORIBA's strategic vision, value creation model, and commitment to addressing global challenges through innovation in energy, environment, bio-healthcare, and advanced materials. Discover HORIBA's 'Honest-to-Goodness' solutions and its dedication to fostering a culture of 'Joy and Fun for All'.
Preview HORIBA GD-Profiler 2: Advanced Surface Analysis for Rapid Depth Profiling
Discover the HORIBA GD-Profiler 2, a high-frequency glow discharge optical emission spectrometer for rapid and simple depth elemental analysis. Ideal for thin films, plating, heat treatment, surface modification, and coatings in R&D and quality control.