LSI LASTEM Instrument firmware upgrade guide

Doc. AN 01350 en 2 31/10/2024 Pag. 2/2 1 Purpose This document contains notes required to update the firmware of any LSI LASTEM device having boot

PDF preview unavailable. Download the PDF instead.

AN 01350 en
LSI LASTEM Instrument firmware upgrade guide

Doc. AN_01350_en_2

31/10/2024

Pag. 1/2

1 Purpose
This document contains notes required to update the firmware of any LSI LASTEM device having bootloader feature. Following equipment are supported:
· E-Log: version >= 2.32.00 · R/M-Log: version >= 2.12.00 except those equipped with an Ethernet port · Heat Shield Master unit: version >= 1.08.00. · DEA420 (SignalTransducerBox): version >= 1.00.01 · DEA485 (ModbusSensorBox): version >= 1.04.00
2 Upgrade procedure
1) Download any important data stored in the memory if any (e.g. configuration, measurements). 2) Unzip the zip file into a folder on your PC. 3) Make sure that you received a compatible firmware version from LSI LASTEM for your device
model/version. The name of the received file contains both the model of the addressed device and the new firmware version after the upgrade. The received file must be copied into the folder containing the update procedure and renamed with the name FW.hex. 4) Connect the PC (RS232 port or USB port using USB adapter) to the serial port used by the device for its setup upload (R/M-Log: RS232-1, DEA485: RS232-2). 5) Start the batch program FWupgService: a. If the PC serial port connected to the instrument is different than com1, indicate which port is
used (e.g. "FWupgService com3"). b. After starting the procedure, reboot the device using On/Off button, or Reset button if
available. On R/M-Log instruments the power off operated from keyboard is NOT sufficient, use instead the Reset button. Having E-Log and R/M-Log instrument respectively from versions 2.40.02 and 2.19.02 or greater, the power off/on cycle must be done keeping any of the instrument keyboard button pressed. c. When the device has been reset, press CTRL ­C; when procedure asks for stop, answer No (N) d. Check the result (step "Verifying"): if not corrected, re-start a new procedure, possibly by decreasing the communication speed (editing the batch program with a text editor, change the value on the line indicated by the set ComSpeed=115200, enter 9600). e. At the end of the operations, the procedure will automatically restart the device; verify if the device functionality is the one as expected. Heat Shield Master unit instruments require to reset the survey-mode using the specific command of the local user interface (see the instrument user's manual).

Doc. AN_01350_en_2

31/10/2024

Pag. 2/2

