AtriCure Isolator™ Transpolar™ Pen: Instructions for Use

Product Models: MAX1-A000163, MAX5-A000625

Description

The AtriCure Isolator™ Transpolar™ pen System includes the AtriCure® Ablation and Sensing Unit (ASU) or Multifunctional Ablation Generator (MAG), the Isolator™ Transpolar™ Pen (Pen) or Isolator™ Long Pen TT (Pen), a Footswitch, and the ASU Source Switch. The Pen is a single-patient-use electrosurgical instrument designed exclusively for use with the ASU and ASU Source Switch (ASB) or MAG. It is utilized for ablating cardiac tissues and serves as a surgical pacing and mapping tool. When connected to the ASU or MAG, the generator delivers bipolar radiofrequency (RF) energy between the Pen's electrodes, controlled by the operator via the Footswitch. The Pen is also designed for temporary pacing or monitoring when connected to an auxiliary pace, sense, or stimulation device. The ASB is an accessory that facilitates switching between the ASU and an external pacing/monitoring device, such as the Oscor PACE 203 H temporary pacemaker.

Nomenclature

Isolator™ Transpolar™ Pen (See Figure 1)

  • [1] Distal Tip
  • [2] Electrodes
  • [3] Malleable Shaft
  • [4] Handle
  • [5] Cable
  • [6] Connector
  • [7] Alignment Arrow

Figure 1 Description: A diagram showing the Isolator™ Transpolar™ pen with its components labeled numerically.

Isolator™ Long Pen TT Nomenclature (See Figure 2)

  • [1] Distal Tip
  • [2] Electrodes
  • [3] Malleable Shaft
  • [4] Rigid Shaft
  • [5] Handle
  • [6] Cable
  • [7] Connector
  • [8] Alignment Arrow

Figure 2 Description: A diagram showing the Isolator™ Long Pen TT with its components labeled numerically.

Indications for Use

  • The Isolator™ Transpolar™ Pen is a sterile, single-use electrosurgery device intended for ablating cardiac tissue during cardiac surgery using radiofrequency (RF) energy.
  • It may also be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.

Contraindications

  • The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
  • Do not ablate directly on cardiac valves.

Potential Complications

Possible complications related to the creation of spot or linear lesions in cardiac and soft tissues include:

  • Tissue perforation
  • Postoperative embolic complications
  • Extension of extracorporeal bypass
  • Perioperative heart rhythm disturbance (atrial and/or ventricular)
  • Pericardial effusion or tamponade
  • Injury to the great vessels
  • Valve leaflet damage
  • Conduction disturbances (SA/AV node)
  • Acute ischemic myocardial event

⚠️ WARNINGS

  • Do not touch the electrodes of the Pen while activating the ASU or MAG. Touching the Pen electrodes during activation could result in electrical shock or burn to the operator.
  • Do not touch the electrodes of the Pen to metal staples or clips, or to sutures while activating the ASU or MAG. This may damage the Pen or tissue, or result in an incomplete ablation.
  • Do not use abrasive cleaners or electrosurgical tip cleaners to clean debris from the Distal Tip. Do not turn on the ASU or MAG while cleaning the device. Use of abrasive cleaners or electrosurgical tip cleaners can damage the electrodes and result in device failure. Use saline-soaked gauze for cleaning debris.
  • Do not immerse the Pen in liquids, as this may damage the device.
  • To avoid shock/burn hazards, always wear appropriate surgical gloves when using the Pen and ASU or MAG.
  • To avoid shock/burn hazards, always remove the Pen from the patient during defibrillation.
  • No representation or warranty is made that device failure or cessation of function will not result in an adverse event, or that medical complications (including tissue perforation) will not follow the procedure, or that the use of the device will in all cases restore adequate cardiac function.
  • To avoid the risk of patient infection, inspect the product packaging prior to opening to ensure the sterility barrier is not breached. If the sterility barrier is breached, do not use the Pen.
  • The pen device is intended for single use only. Do not RESTERILIZE. Resterilization may cause loss of function or injury to the patient.
  • Use only connecting cables and auxiliary devices identified in these Instructions for Use to avoid risk of patient injury, operator injury, or equipment damage.
  • Do not allow the connectors of the Pen to get wet. Wet connectors may affect device performance.
  • Read auxiliary device manuals and observe warnings. The Pen can be used with a temporary pacemaker.
  • Do not connect the ASB or MAG auxiliary device cable to mains-voltage operated equipment without verifying isolation of the connected equipment to EN60601-1-1. Mains-voltage operated equipment may introduce dangerous leakage currents into the heart.
  • An auxiliary device may have an adverse effect on nearby radio or TV transmissions. Conversely, nearby electrical appliances may adversely influence the auxiliary device, causing data errors or malfunction.

? CAUTION:

  • Read all instructions carefully for the AtriCure® ASU and ASB Switch Matrix or MAG, Isolator™ Transpolar™ Pen, and any auxiliary device being used prior to using the devices. Failure to properly follow instructions may lead to electrical or thermal injury and improper functioning of the device.
  • Use of the Pen should be limited to properly trained and qualified medical personnel. Proper surgical procedures and techniques are the responsibility of the medical professional. Understanding the proper use of the Oscor PACE 203 H temporary pacemaker equipment is also the responsibility of the medical professional. Each surgeon must evaluate the appropriateness of any procedure based on their own medical training and experience, and the type of surgical procedure.
  • Patient and procedure selection is solely a medical responsibility, and the outcome depends on many variables, including patient pathology, and surgical and perfusion procedures.
  • Variations in specific procedures may occur due to individual physician techniques and patient anatomy.
  • To avoid damage to the device, do not drop or toss the Pen. If the Pen is dropped, do not use it; replace it with a new Pen.
  • Do not use the Pen in the presence of flammable materials.
  • Do not re-sterilize or reuse the Pen.
  • The distal tip of the Pen must be kept clean of debris during surgery to avoid loss of power. Before activating the ASU or MAG, inspect the distal tip for foreign matter. Foreign matter captured on the tip will adversely affect ablation.
  • The Pen is only compatible with the AtriCure® ASU/ASB or MAG. Use of the Pen with another manufacturer's generator may damage the device and result in patient injury.
  • The ASB or MAG should only be used with FDA-approved cardiac pacing and sensing devices.
  • Do not use the Pen for coagulation or ablation of veins or arteries.
  • Use caution to avoid trauma to tissues not within the target area of ablation. Tissues and/or structures behind the targeted tissue should be protected from potential thermal spread.
  • The Pen has an eight-hour useful life, tracked by the ASU or MAG. If a device that has reached its time limit expiration is plugged in, the Pen will no longer function, and the ASU or MAG will display a message indicating that the Pen must be replaced.
  • It is the responsibility of the user to dispose of this device in accordance with local regulations.
  • Excessive bending of the malleable stainless steel shaft will cause the shaft to harden and may increase the potential for breakage.

Instructions for Use

Cardiac Ablation Mode

  1. Using sterile technique, remove the Pen from its packaging. With the Connector Alignment Arrow symbol at the 12 o'clock position, push the Connector into the Pen receptacle on the front of the ASU/ASB or MAG. Verify that the connections between the Pen and the ASU or MAG are secure. If the connections are loose, do not use the Pen. Inspect the Cable and do not use the Pen if the cable is frayed or the insulation is damaged.
  2. Under direct visualization, place the distal tip against the targeted cardiac tissue. Ensure both electrodes are in contact with the targeted tissue.
  3. Painting Ablation Technique:
    1. Maintaining visualization, move the distal tip gently across the targeted cardiac tissue.
    2. While maintaining continuous contact between the tissue and the electrodes, move the device continuously in an oscillating manner at a rate of approximately 1 cm/sec.

Painting Lesion Depth*

TimeOscillationDepth
20 seconds2 cm oscillation2.0 - 4.0 mm

*Data was obtained from ablations performed on excised bovine myocardium and represent 95% confidence intervals. Results may vary based on live tissue properties.

  1. Stamping Ablation Technique:
    1. Apply constant firm pressure to the tissue without movement. Maintain full contact of the electrode surface with the tissue. A stamping lesion is approximately 8 mm x 6 mm.
    2. If creating longer linear lesions with the Stamp technique, overlap contiguous ablations by 50% to ensure a continuous and complete lesion.

Stamp Lesion Depth*

TimeDepth
10 Seconds3.3 - 3.8 mm
15 Seconds3.8 - 4.4 mm

*Data was obtained from ablations performed on excised bovine myocardium and represent 95% confidence intervals. Results may vary based on live tissue properties.

  1. Press the Footswitch to activate the ASU or MAG.
  2. When the Footswitch is pressed, the ASU or MAG will emit an audible tone indicating that current is flowing between the electrodes located at the distal tip of the Pen and through the tissue.
  3. Inspect the surgical area to ensure adequate ablation.
  4. Between ablations, wipe the distal tip clean with a saline-soaked gauze pad. Important: For optimal performance, keep the Pen electrodes clear of coagulum. To ensure the electrodes are clear of coagulum:
    1. Use saline-soaked gauze to clean the electrodes after each ablation. Coagulum is much easier to remove within the first several seconds after ablation. In a brief period, coagulum may dry, making removal more difficult.
    2. Check both electrodes before each ablation to ensure that the gold of the electrode is visible and coagulum is removed.
    3. If the Pen is idle between ablations, place the Pen distal tip onto saline-soaked gauze to prevent any coagulum not cleaned off the electrodes from drying.
  5. Repeat ablation if necessary.
  6. Upon completion of the surgical procedure, disconnect the Pen from the generator and discard the Pen after use. Follow local governing ordinances and recycling plans regarding disposal or recycling of device component.

Pacing and Mapping Mode

  1. Inspect the package and product to ensure the expiration date has not passed and no damage occurred to the product during shipping and handling.
  2. Using sterile technique, remove the Pen from its packaging.
  3. With the Connector Alignment Arrow symbol in the 12 o'clock position, push the Connector into the Pen receptacle on the front of the ASB or MAG. Verify that the connections between the Pen and the ASB or MAG are secure. If the connections are loose, do not use the Pen. Inspect the Cable and do not use the Pen if the cable is frayed or the insulation is damaged.
  4. A cable with standard 0.080 inch pin tip plugs is provided with the ASB or MAG to connect to the auxiliary device. The end of the cable with the single pin tip plug connects to the ASU Source Switch or MAG. The end of the cable with dual pin tip plugs connects to the auxiliary device.
  5. Turn on the temporary pacing or sensing equipment and ensure proper connections to validate electrical continuity. For detailed instructions, refer to the temporary pacemaker manual.
  6. Settings and procedures for the auxiliary device are determined according to the instructions for use provided with the auxiliary device. Set the auxiliary device to atrial asynchronous mode (sensing disabled or increased to maximum value).
  7. Note: The Pen will pace when the auxiliary device is in the ON position.
  8. Identify and expose the sites for pacing and sensing using standard surgical techniques. Under direct visualization, place the electrodes against the targeted tissue. Ensure both electrodes are in contact with targeted tissue.
  9. Ensure pacing is only active when the end effector is in contact with the targeted tissue. The device will pace when the auxiliary device pacing is on.
  10. For sensing (mapping), place the distal tip onto the targeted tissue to display the electrogram (EGM).
  11. Upon completion of the surgical procedure, disconnect the Pen from the ASU/ASB or MAG and discard the Pen after use. Follow local governing ordinances and recycling plans regarding disposal or recycling of device component.

How Supplied

The Isolator™ Transpolar™ pen is supplied as a STERILE instrument and is for single-patient use only. Sterility is guaranteed unless the package is opened or damaged. Do not resterilize.

Return of Used Product

If for any reason this product must be returned to AtriCure®, Inc., a return goods authorization (RGA) number is required from AtriCure®, Inc., prior to shipping.

If the product has been in contact with blood or body fluids, it must be thoroughly cleaned and disinfected before packing. It should be shipped in either the original carton or an equivalent carton to prevent damage during shipment, and it must be properly labeled with an RGA number and an indication of the biohazardous nature of the contents of shipment.

Instructions for cleaning and materials, including appropriate shipping containers, proper labeling, and an RGA number, may be obtained from AtriCure®, Inc.

? CAUTION: It is the responsibility of the health care institution to adequately prepare and identify the product for shipment.

Disclaimer Statements

Users assume responsibility for approving the acceptable condition of this product before it is used, and for ensuring that the product is only used in the manner described in these instructions for use, including, but not limited to, ensuring that the product is not re-used.

Under no circumstances will AtriCure, Inc. be responsible for any incidental, special, or consequential loss, damage, or expense, which is the result of the deliberate misuse or re-use of this product, including any loss, damage, or expense related to personal injury or damage to property.

Explanation of Symbols on Package Labeling

Refer to the outer package label to see which symbols apply to this product.

  • ⚕️ Caution
  • Rx ONLY: Caution: Federal Law (US) restricts this device to sale by or on the order of a physician.
  • ✔️ Non-Pyrogenic
  • LOT Lot Number
  • Do Not Re-Sterilize
  • ⚠️ Do Not Use if Package is Damaged
  • ⚕️ STERILE EO: Sterilized by Ethylene Oxide
  • ? Not made with Natural Rubber Latex
  • Do Not Re-Use
  • ? Follow Instructions for Use
  • Manufacturer: AtriCure Inc.
  • Address: 7555 Innovation Way, Mason, Ohio 45040 USA
  • Phone: +1 866 349 2342, +1 513 755 4100
  • Expiration Date
Models: MAX1 - A000163 Isolator Transpolar Pen, MAX1 - A000163, Isolator Transpolar Pen, Transpolar Pen, Pen

File Info : application/pdf, 2 Pages, 575.73KB

PDF preview unavailable. Download the PDF instead.

P000154.N Adobe PDF Library 17.0 Adobe InDesign 18.5 (Windows)

Related Documents

Preview AtriCure Isolator Synergy EnCompass Clamp and Guide System: Instructions for Use
Detailed instructions for the AtriCure Isolator Synergy EnCompass Clamp and Guide System, covering indications, contraindications, system description, setup, use, potential complications, and disposal for cardiac tissue ablation.
Preview AtriCure Isolator Synergy EnCompass Clamp and Guide System Instructions for Use
Detailed instructions for using the AtriCure Isolator Synergy EnCompass Clamp and Guide System (OSH, OLH, GPM100) for cardiac tissue ablation. Includes indications, contraindications, system description, setup, operation, potential complications, and disposal.
Preview AtriCure AtriClip LAA Exclusion System: Instructions for Use
Comprehensive instructions for the AtriCure AtriClip LAA Exclusion System, including the Gillinov-Cosgrove Clip. Covers device description, indications, contraindications, warnings, cautions, surgical procedure steps, handling, MRI safety, and symbol explanations.
Preview AtriCure cryoICE® System Cryoablation Probe: Instructions for Use
This document provides instructions for the safe and effective use of the AtriCure cryoICE® system cryoablation probe, a single-use medical device for cardiac arrhythmia treatment.
Preview AtriCure AtriClip® Flex-V® LAA Exclusion System Instructions for Use
Comprehensive instructions for the AtriCure AtriClip Flex-V LAA Exclusion System, detailing indications, contraindications, system description, warnings, complications, clip selection, deployment, and handling for medical professionals.
Preview AtriCure Cannula with Guide (CSK-6130, CSK-6131) Instructions for Use
Comprehensive instructions for the safe and effective use of the AtriCure Cannula with Guide (CSK-6130, CSK-6131) in laparoscopic and general surgery, including device description, warnings, indications, and operational procedures.
Preview AtriCure cryoICE BOX Instructions for Use
Comprehensive instructions for the AtriCure cryoICE BOX, model ACM1, detailing system overview, technical specifications, setup, device use, troubleshooting, and maintenance.
Preview EPi-Ease Epicardial Access Device Set-Up Guide
A comprehensive set-up guide for the AtriCure EPi-Ease Epicardial Access Device, detailing equipment, room setup, device description, and procedural steps for safe and effective use.